Xiaojun Shen1, Pratik Shukla2, Sunita Nayak3, Vasanth Gopal4, Prabhakaran Subramanian4, Amy Sarah Benjamin3, Shivpuram Kalainathan5. 1. School of Electrical and Electronic Engineering, Nanyang Technological University, Singapore. 2. The Manufacturing Technology Centre (MTC), Coventry, UK. 3. Centre for Biomaterials, Cellular, and Molecular Theranostics, Vellore Institute of Technology, Vellore, Tamil Nadu, India. 4. Department of Metallurgy & Materials Engineering, Faculty of Engineering, University of Malta, Msida, Malta. 5. Department of Physics, School of Advanced Sciences, Vellore Institute of Technology, Vellore, Tamil Nadu, India.
Abstract
This paper focuses on the evaluation of mechanical and biological properties of laser shock peening (LSP) orthopaedic grade Ti-6Al-7Nb alloy. LSP surface treatment was conducted at laser energy of 3 to 7 J with overlaps of 33%-67%, and with a 3 mm laser spot size. Cell viability on laser shock peened surface was evaluated through in-vitro MTT assay, using osteoblast-like MG63 cells for the first-time. Residual stresses, microhardness, microstructure, sliding wear and wetting properties were investigated. Compressive residual stresses were found at various depths due to controlling the LSP parameters, compared to the as-received surface. The laser shock peened surfaces were hardened from 365HV0.05 to 405HV0.05, while the as-received surface was 320HV0.05. The average sub-grain size was refined from 14% to 36% after LSP. The wear resistance was also controllable by altering LSP parameters. The MTT results show that the cell viability on the laser shock peened surfaces was comparatively lower than that of the untreated surface after 24 h. However, after 72 h, the cell viability on modified surfaces were significantly improved. This work indicated that laser shock peened surfaces have a strong potential to decrease the pain from orthopaedic implant failures and promote the cytocompatibility between the bone and implant.
This paper focuses on the evaluation of mechanical and biological properties of laser shock peening (LSP) orthopaedic grade Ti-6Al-7Nb alloy. LSP surface treatment was conducted at laser energy of 3 to 7 J with overlaps of 33%-67%, and with a 3 mm laser spot size. Cell viability on laser shock peened surface was evaluated through in-vitro MTT assay, using osteoblast-like MG63 cells for the first-time. Residual stresses, microhardness, microstructure, sliding wear and wetting properties were investigated. Compressive residual stresses were found at various depths due to controlling the LSP parameters, compared to the as-received surface. The laser shock peened surfaces were hardened from 365HV0.05 to 405HV0.05, while the as-received surface was 320HV0.05. The average sub-grain size was refined from 14% to 36% after LSP. The wear resistance was also controllable by altering LSP parameters. The MTT results show that the cell viability on the laser shock peened surfaces was comparatively lower than that of the untreated surface after 24 h. However, after 72 h, the cell viability on modified surfaces were significantly improved. This work indicated that laser shock peened surfaces have a strong potential to decrease the pain from orthopaedic implant failures and promote the cytocompatibility between the bone and implant.
Long-term osseointegration of implants with bones have always been one of the most
crucial aims for considering the recovery of patients after implantations.
Once an implant is inserted into the human body, it would initially react
with host issue, involving a series of physiochemical changes, ranging from
molecular level to the cellular level.
The reaction determines the healing speed and long-term performance of
implants, affected by the surface physicochemical properties of biomedical materials
such as surface topography, wettability, electric properties, the pore size of
biomaterials and bioactive molecules. Therefore, surface modifications to improve
the biocompatibility of biomaterials are considered to be crucial to enhance the
integration of interface where cell/protein adheres.Although, surgical implant biomaterials such as metals, ceramics, and polymers are
improved by a variety of surface modification techniques, namely: etching; blasting;
shot peening/ultrasonic peening; and non-contact, laser-based modifications. In
particular, to acquire a certain surface roughness and enhance osseointegration,
Flamant et al.,
applied hydrofluoric acid (HF) to a dental ZrO2 to increase
roughness. It was found that a concentration of 40% HF acid gave a uniform surface
morphology with the shortest etching time. Blasting techniques including
sandblasting, alumina blasting and micro-abrasive blasting are also useful methods
to increases surface roughness to promote, biocompatibility. Günay-Bulutsuz et al.,
investigated the cell response of sandblasting on ultrafine and coarse
titanium surfaces. Ultrafine and blasted surfaces are more biocompatible for human
gingival fibroblast cells. Moreover, Granato et al.,
evaluated the effect of micro-abrasive blasting and
alumina-blast/acid-etching on pure Ti surfaces. In comparison to the as-machined
surface, both the surface treatments in terms of roughness lead to the improvement
of osseointegration in pure Ti substrates.Therefore, a variety of surface modification methodologies or surface fabrication
techniques, especially, the ones that are laser-based, were employed to improve the
osseointegration of implants. Compared to conventional chemical methods, laser-based
surface modifications enabled a wide variety of structures at the nano and the
micro-scale, and the process is fast, reproducible, controllable and contactless. It
is reported that laser introduces surface features such as grooves, dimples, lines
that are beneficial in strengthening the bonding of implants and human bones.
Raimbault et al.,
evaluated the cell behaviour on a femtosecond laser-fabricated a titanium
surface at nanoscale and microscale level. Cells showed more sensitivity to the
nanoscale structures. This indicated a potential for titanium modification in
orthopaedic or dental applications. Batal et al.,
employed a nanosecond pulse laser for surface texturing on CoCrMo alloy and
compared the cell viability of Saos-2 osteoblast like cells on an as-received and
various laser textured surfaces. A laser textured surface presented a better
performance in terms of cell viability. Moreover, a CO2 laser was used to
modify the surface of nylon 6,6 and the mesenchymal stem cell was employed for
evaluation. The viable cell counting increased to 60,000 cells/ml due to
CO2 laser surface treatment. Although, by applying a 3-D printing
technique, the Young’s modulus of orthopaedic scaffolds can be maintained close to
that of bones with interconnected porous network, which favours the osseointegration
and relieving the stress shielding effect.
It is also a new promising direction for achieving long-term osseointegration
of implants.Laser Shock Peening (LSP) is also a very useful method for surface modification
as it is more advanced in terms of penetration depth and quality of the part
being treated. It has been widely and originally used in aerospace and auto industry
for improving the fatigue properties of various metallic materials for decades.
It is promising that LSP could also be employed in extending implant service
life by not only improving mechanical properties, but also enhancing the
biocompatibility. LSP has its own advantages in medical applications, compared to
conventionally employed surface modification techniques such as sandblasting, shot
peening, laser texturing and so on. For example, in comparison to sandblasting, no
contaminations will be left after LSP. In addition, the top-most benefit will be
induction of deeper compressive stress which is not available with sand, grit or
shot blasting techniques. LSP is also a cold working method and laser would not
contact metal surfaces directly rather than exploding the absorptive layer
generating plasma-driven shock- waves, which causes material deformations. With the
same capability like shot peening, LSP could also introduce a stable compressive
residual layer in the treated materials.
However, when it comes to the complex medical implants, components such as
acetabular, trabecular and proximal tibial, the focussed laser beam is able to
precisely process the corner of the complex components, where conventional shot
peening cannot reach, providing a better process control and benefit the material
strength exactly where it is needed. Besides that, due to the microstructural
deformation caused by high plasma pressure during peening, surface morphologies were
generated by applying laser energy and foot-print, overlap, which means LSP also
could benefit orthopaedic or dental implant osseointegration due to introducing
certain surface roughness that increasing the surface roughness is critical for
improving the biocompatibility. The features of topographies after LSP, namely:
grooves and dimples, contribute to the cells/proteins adhesion and proliferation on
the implant surfaces. This in turn strengthens the mechanical interlock between the
implant and the bones. More than that, the stable compressive residual stress layer
which benefits fatigue will be formed after LSP that is what other laser treatments
(such as laser texture) are not capable of delivering.When it comes to the mechanical properties of the implants, among the identified
failure mechanisms of the implant, fretting wear is destructive to implant materials
and occurs between two contacting solid bodies when exposed to relatively small
amplitude oscillatory motion under a certain load.
Such movements would generate fretting fatigue, which may cause material
surface damage, thereby, acting as crack initiation sites.
Furthermore, micro-cracks will extend from the material surface into the
depth direction under the fretting fatigue leading to the reduction of fatigue
strength. In a total hip implants system, the fretting wear normally occurs at the
junction of head-neck that normally consists of titanium stem and ceramics.
The metallic debris was considered to be the domain reason for causing
osteolysis, and polyethylene debris would even cause implant degradation. This
further results in aseptic loosening. Therefore, improving the wear resistance of a
prosthesis is one of the main purposes of a surface enhancing technique.As reported in our previous works and others,
LSP has presented its capability to not only improve the wear resistance, but
also alter the surface wettability of Ti-6Al-7Nb alloy by applying surface pattern
and surface microstructure refinement. Zhou et al.,
investigated the effects of LSP on friction and wear properties of medical
Ti-6Al-4V alloy in Hank’s solution. The results showed that LSP can reduce debris
peeling of Ti-6Al-4V implants. Zhang et al.,
demonstrated that LSP not only decreased the co-efficient of friction and
improved the wear resistance of AZ31B alloy, but also strengthened tensile strength
and fatigue performance as well as enhance the corrosion resistance of metallic
parts. Luo et al.,
examined the effects of multiple treatments on electrochemical corrosion
properties of Mg-Al-Mn alloy, and observed obvious improvement in electrochemical
corrosion.In this paper, a medical-grade Ti-6Al-7Nb alloy was treated with a Q-switch, Nd: YAG
nanosecond pulsed laser. X-ray Diffraction (XRD), Scanning Electron Microscope (SEM)
and Transmission Electron Microscope (TEM) were employed to observe both the
microstructures and dislocation activities before and after LSP surface treatment.
Mechanical property, namely, wear in simulated body fluid was investigated for the
first-time subject to laser peened metals. On account of this, laser energy and
overlap on the micro-hardness, tribological behaviour and worn surfaces are
investigated by means of Vickers microhardness indentation, SEM and 3-D surface
profiler. For a surface modification technique, the first and foremost aspect is
that it should not be any cytotoxic to any biomaterials. Based on this
pre-requisite, an in vitro biological evaluation was carried with determining the
cell viability of laser shock peened surfaces by using MTT cell proliferation assay
for 24 and 72 h. The cell morphology was characterised by fluorescence microscope.
In this work, the evaluation of the cell viability on the laser shock peened
surfaces was conducted for the first-time. No other previous publications, hitherto,
has shown investigations in relation to any parameters of laser shock peening
discussed herein at any depth. In addition, the work could directly offer a new
surface processing technique for improving both mechanical and biological properties
of implants, thereby, directly benefiting the end-users who will be suffering the
pain caused by implant failures such as corrosion, wear, fractures and
loosening.
Experimental design
Material characterisations
Ti-6Al-7Nb alloy and surface characterisations
A medial grade hot rolled Ti-6Al-7Nb alloy with standard ASTM F1295 was
deployed for the work herein. The chemical composition of the material is
listed in Table
1. The material was provided by Aircraftmaterials co., Ltd,
Stokenchurch, UK. The as-rolled Ti-6Al-7Nb rod was annealed at 870 4 h for
stress relaxation purposes. The microstructure of Ti-6Al-7Nb was
characterised by a SEM (Sigma 500VP, manufactured by Zeiss, Germany) and a
TEM (FEI, Tecnai-G
20, The Netherlands) operating at 200 kV. Prior to SEM, mechanical
grinding was carried out on a 3 mm-thickness disc by SiC abrasive paper from
320 grit to 2500 grit, followed by polishing from 9 to 1 µm and 2 h
vibration final polishing to mirror finishing surfaces. Koll’s reagent was
employed for etching and each samples was carried out for 15secs. The TEM
sample were cut from the top surface of the disc along the cross-section by
wire-cut electrical discharge machining (EDM). Then the cut pieces were
glued on a glass plate by a transparent thermoplastic glue and grinded to
100 µm. By heating the temperature, the TEM sample was carefully removed off
the glass plate. Finally, the sample was thinned by a low-angle ion mill
system until the material become electron transparent.
Table 1.
Showing the chemical composition of Ti-6Al-7Nb alloy (units in
wt%).
Element
Nb
Al
Ta
Fe
O
C
N
H
Ti
wt%
6.5–7.5
5.5–6.6
<0.5
<0.25
<0.2
<0.08
<0.05
<0.009
Bal
Showing the chemical composition of Ti-6Al-7Nb alloy (units in
wt%).
Experimental methods
Laser shock peening parameters
The laser shock peening experiments were conducted by a Q-Switched Nd:YAG
laser with 3, 5 and 7 J at spot overlapping of 33%, 50%, and 67%. The
detailed LSP processing parameters are given in Table 2. The water layer was used
as the confinement layer (about 2 mm thickness), while the absorbtive layer
was a polyvinyl black tape. Figure 1(a) presents a schematic
diagram of LSP, (b) experimental and theoretical laser shock peened samples
and the schematic of LSP strategy.
Table 2.
Laser shock peening parameters employed for the surface treatment of
Ti-6Al-7Nb.
Parameters
Value
Pulse energy (J)
3, 5, 7
Laser wavelength (nm)
1064
Spot diameter (mm)
3
Number of laser impacts
1
Overlapping rate (%)
33%, 50%, 67%
Pulse duration (ns)
20
Radiance density19 (W.mm2.Sr.µm)
0.46 @ 3 J
0.77 @ 5 J
1.09 @ 7 J
Figure 1.
A schematic diagram of laser shock peening process in (a) and the
schematics of the laser peening strategy in (b).
Laser shock peening parameters employed for the surface treatment of
Ti-6Al-7Nb.A schematic diagram of laser shock peening process in (a) and the
schematics of the laser peening strategy in (b).
Microhardness
Vickers hardness tester manufactured by Struer, Germany conducted the
microhardness measurements. Each indentation was carried out with a load of
0.05 g along the longitudinal section of the laser shock peened specimen.
The average value of each point was calculated three times in the same
special area within the depth.
Wear testing
A ball-on-flat tribometer (manufactured by Ducom, India) with 6 mm diameter
bearing grade ZrO2 ball, was employed to conduct the
reciprocating wear experiments. The Ti-6Al-7Nb samples prior to and after
LSP were used as a flat material. Before the test, both counter moving ball
and Ti-6Al-7Nb alloy were ultrasonically cleaned with isopropanol and
distilled water. The wear test was conducted with a load of 5 N running for
18,000 cycles at frequency of 5 Hz and, the amplitude of the movement was
100 µm. The displacement was fixed at ±100 µm, which is the minimum
displacement that can be set with the tribometer. Faouvry energy ratio was
not measured because the tribometer used in the current study does not plot
fretting loop (frictional force vs displacement). The samples were placed in
the specimen rig, immersed in 20% foetal bovine serum (FBS) solution
(PH = 7.4), at the temperature of 37. The wear volume (V) of Ti-6Al-7Nb
samples before and after LSP were calculated using Klaffke’s equation
(1)
:Where
is the radius of the counterpart;
is the depth of wear scar,
is pie (3.14) and 4 is a constant.
Wetting properties and residual stress
The residual stress measurements were carried out by using an incremental
hole drilling method with a system designed by Stress Measurement, Shepshed,
UK. The measurement was conducted according to measurement good practice
guide no.3
and ASTM E837 standard.
Regarding the dynamic contact angle measurement, a needle-in method
was employed via an instrument OSCA1000 developed by Data
physics co., Ltd, Germany. The measurement of dynamic contact angles was
carried out automatically with filming the course from wetting (advance
contact angle) and de-wetting (receding contact angle). At the beginning,
the volume of the drop was continuously increased to 10 μl with motorised
piston. After keeping the syringe needle steady for 10 s, the volume of
liquid was withdrawn off the solid surfaces at the same the images are
recorded and evaluated.
Cell viability and fluoresce image
In order to check the biocompatibility against live cells the following assay
was carried out. The metal samples were sterilised by washing with sterile
water, ethanol and PBS and immersed in Dulbecco’s Modified Eagle Medium
(DMEM) w/o L-Glutamine incomplete media (HiMedia Laboratories, India) for
about 15 days. The spent media with immersed samples and leachate if any,
was used to carry out the study. Biocompatibility of the metal samples was
evaluated using osteosarcoma cells (MG-63). The cells were grown in DMEM
supplemented with 10% FBS, penicillin (100 I.U/ml) and streptomycin
(0.1 mg/ml) all obtained from HiMedia Laboratories, India. The cells were
maintained in a humidified incubator (ESCO Cellmate Biotech, Singapore) at
37 °C with 5% CO2 environment. The cells (MG-63) were seeded onto
the 24 well plate at 1 × 104 cells per well and allowed to attach
overnight with complete media. After 24 h the complete media was aspirated
and replaced by the immersed media of the metal samples and the study was
done for 24 and 72 h after the change.After the first day and third day the
(3-(4,5-Dimethylthiazol-2-yl)−2,5-Diphenyltetrazolium Bromide) MTT reagent
(HiMedia Laboratories, India) was added (1 mg/ml) and incubated for 1 h.
Thereafter, DMSO (Himedia) was added to dissolve the formed formazan
crystals. The dissolved solution was transferred to 96 well plate and the
absorbance was measured using a spectrophotometer (Read Well, Touch Robonik,
India) at a wavelength of 570 nm. Reported values are the mean of 3
replicates and expressed as percentages of the control values.For fluorescence study, the metal samples were sterilised with ethanol
followed by wash with PBS and kept in UV light overnight. Equal number of
osteosarcoma cells (MG-63) (1 × 104 cells) were seeded on the
surface of the metal samples. The cells were grown in DMEM supplemented with
10% FBS, penicillin (100 I.U/ml) and streptomycin (0.1 mg/ml). The cells
were maintained in a humidified incubator (ESCO Cellmate Biotech, Singapore)
at 37°C with 5% CO2 environment. Very minimal amount of media was
added initially to prevent the runaway of cells directly on to the surface
treated well plate. After 4 h more media was added to the cells and allowed
to culture. The imaging was done on the third day after 20 min of incubation
of cells with nuclear staining dye 4′,6-diamidino-2-phenylindole (DAPI from
Himedia).
Statistical analysis
The whole experiment was repeated three times as mean values with standard
deviation. The data were statistically analysed using analysis of variance
(ANOVA) for assessing the significance level of the differences between the
groups. The possibility p less than 0.05 was considered
significant differences.
Results
Microstructure characterisation
As-received surface and cross-sectional microstructural
characterisations
Both the as-received surface and the laser shock peened Ti-6Al-7Nb were
characterised to find that laser shock peening affected grains, leading to
grain refinement. This has a significant relationship with the mechanical
properties of metal alloys.
The SEM images (as shown in Figure 2(a)) showed that the surface
microstructure of as-received Ti-6Al-7Nb alloy consists of globular and
acicular α and retained β grains. The acicular and globular α-grains
distribute non-uniformly. The smaller α-grain surround around the large
coarse one, and between them, retained β-grains, that were distributed
randomly. What is more, the size of α-grain varies from several micrometres
to tens of micrometres due to the hot rolling process. Observing at high
magnifications, sub-grains are inside the coarse grains and the sub-grain
boundaries are quite sharp. Cross-sectional microstructure is given in Figure 2(b), as
mentioned above (hot rolling), the grains are elongated along rolling
direction. Like surface, the grains also distribute randomly and
non-uniformly. At higher magnification, we also can see sub-grains and
sub-grains boundaries. With the grain intercept method (GIM), the
as-received sub-grain size is 266.5 nm.
Figure 2.
SEM images showing surface in (a) and longitudinal section in(b) of
the microstructure of Ti-6Al-7Nb.
SEM images showing surface in (a) and longitudinal section in(b) of
the microstructure of Ti-6Al-7Nb.
The longitudinal section microstructure of laser shock peened
Ti-6Al-7Nb
Figure 3 shows the
longitudinal section microstructure of Ti-6Al-7Nb alloy subject to LSP
parameter of 7 J, 67% (at the magnification of 10 K). Compared to the
as-received microstructure, amongst all LSP parameters, there is no obvious
sign showing that any refinement occurred by the grain size.
Figure 3.
SEM images showing the longitudinal microstructure of Ti-6Al-7Nb
subject to LSP parameter of 7 J at 67%.
SEM images showing the longitudinal microstructure of Ti-6Al-7Nb
subject to LSP parameter of 7 J at 67%.Therefore, increasing the magnifications to 40 K, inside coarse grains, the
sub-grains were clearly observed as shown in Figure 4. With GIM, the average
sub-grain size is presented in Table 3.
Figure 4.
SEM images showing the sub-grains before and after laser shock
peening at parameters of (a) 3 J, 33%, (b) 3 J, 50%, (c) 3 J, 67%,
(d) 5 J, 33%, (e)5 J, 50%, (f) 5 J, 67%, (g) 7 J, 33%, (h) 7 J, 50%,
(i) 7 J, 67%, and (j) untreated.
Table 3.
The average sub-grain size (nm) after laser shock peening calculated
by grain intercept method.
33%
50%
67%
3 J
251.08
216.9
185
5 J
175.75
159.77
152.83
7 J
152.83
146.46
140.4
The average sub-grain size (nm) after laser shock peening calculated
by grain intercept method.SEM images showing the sub-grains before and after laser shock
peening at parameters of (a) 3 J, 33%, (b) 3 J, 50%, (c) 3 J, 67%,
(d) 5 J, 33%, (e)5 J, 50%, (f) 5 J, 67%, (g) 7 J, 33%, (h) 7 J, 50%,
(i) 7 J, 67%, and (j) untreated.
Phase determination by X-ray diffraction
The LSP induced high pressure shock waves could result in plastic
deformation, the formation of residual stress and new microstructural phases
in K403 nickel alloy.
In order to investigate the effects of LSP on the microstructure of
Ti-6Al-7Nb, X-ray diffraction (XRD) method was conducted to measure the
different diffraction patterns. Therefore, the XRD patterns of laser shock
peened samples (2θ from 10° to 90°) are showed in Figure 5. It can be seen that the
main phase of Ti-6Al-7Nb alloy are still peaks of α(100), α(002), α(101),
α(102), α(110), α(103), α(200), α(201) and β(110). There are not any new
peaks formed after LSP. This indicates that there is no phase transformation
and no new crystalline phases generated which is normally the case with
laser shock peening of metals and alloys.
Figure 5.
X-ray diffraction patterns and magnified α (100) peak of Ti-6Al-7Nb
alloy in (a) 33%, (b) 50% and (c) 67%.
X-ray diffraction patterns and magnified α (100) peak of Ti-6Al-7Nb
alloy in (a) 33%, (b) 50% and (c) 67%.The peak broadening and shifting is shown in the magnified peaks images in
Figure 5. This
is due to the changes in lattice strain and crystallite size caused by
high-strain plastic deformation. In the peak of α(100) (Figure 5(a-1), (b-1) and (c-1)), the peak was broadened with
increasing the laser energy (if the overlap was kept constant). It is
reported that if peaks shift towards higher 2 theta angles, then the induced
residual stress is compressive. However, we did not see such trend about the
peak shifting relating the residual stress. Therefore, the broadening is the
domain factor for examining the lattice strain and crystallite size.
TEM observation of dislocations at grain boundaries after laser shock
peening
The microstructure in the region of top surface of laser shock peened
Ti-6Al-7Nb alloy was characterised by using TEM shown in Figures 6 and 7. Published
literatures has reported that LSP could lead to surface and near-surface
microstructure deformation movements, especially high-density
dislocations.[25,26] At the top surface,
there are many dislocations features including dislocation tangles,
dislocation walls, dislocation cells, and dislocation pile-ups. In Figure 6(a) to (d), dislocation
lines, dislocations walls are accumulated near the grain interfaces as the
impeding of the grain boundary. With increasing the density of dislocations,
the dislocation tangles will transform into dislocation walls (Figure 6(e)).
Furthermore, these will further form new sub-grain boundaries, and the
sub-grains were even refined into nano-size which can be classified as
‘Laser Shock Peening Nano-crystallisation’ phenomenon.
It has been shown that the higher the density of the sub-grains, the
higher the yield stress of material due to the increased sub-grain boundary.
In Figure 6(f) and
(g), more
dislocations move towards the sub-grain or grain boundaries, and unable to
pass the boundary, thereby, a process of ‘pile-up’ occurs at the grain
boundary. The grain boundary strengthening theory has a great influence on
the yield strength, as described by Hall-Petch relationship as shown in
equation (2):
Figure 6.
TEM image showing (a) -(b) dislocation pile-ups; (c)-(e) Slip band
dislocation tangles; and (e)-(f) dislocations near the grain
/sub-grain boundary.
Figure 7.
The TEM images showing the microstructure of laser shock peened
Ti-6Al-7Nb at the top surface (a) and sub-surface(b and c)
corresponding selected area electron diffraction separately.
Where d is the grain size,
is the yield stress,
is the material constant for the starting stress and
is the material strengthening co-efficient. This H-P
relationship indicates that the smaller the grain size, the higher the
applied stress needed to propagate dislocations through the material.
Therefore, that is an important factor in the microstructural aspect that
LSP can benefit the fatigue resistance of the metallic materials.TEM image showing (a) -(b) dislocation pile-ups; (c)-(e) Slip band
dislocation tangles; and (e)-(f) dislocations near the grain
/sub-grain boundary.The TEM images showing the microstructure of laser shock peened
Ti-6Al-7Nb at the top surface (a) and sub-surface(b and c)
corresponding selected area electron diffraction separately.Figure 7(a) to
(c) with the
selected area electron diffraction (SEAD) was taken from top surface to the
inner substrates. It can be seen from Figure 7, the SEAD pattern in the
image was ring-shaped which indicated that the original sub-grains have been
refined and the orientation distributed randomly. In Figure 7(b), there are high
intensity dislocations forming dislocations tangles and dislocations wall.
However, the corresponding SEAD image presents the arc-shaped diffraction
patterns which means the sub-grain refinement has not been completed yet. A
similar trend can also be found in Figure 7(c) corresponding SEAD
image. There is some matrix apparent in the SEAD.
Mechanical properties
Residual stress
The residual stress distributions along the surface and sub-surface before
and after LSP are presented in Figure 8. In terms of the untreated
curves, the residual stress at the near surface is tensile due to the hot
rolling process which rendered to uneven cooling. Although, the such
residual stress is self-phase-balanced, it still affects the performance of
titanium alloy under external. For instance, force deformation, stability,
fatigue, and other aspects may have adverse effects. Therefore, surface
modification method is necessary to change the metallic surface condition.
In comparison, after LSP, the residual stresses are transferred into
compressive by LSP from the near surface to a certain depth. More detailed
info can be found in our previous published literature.
Figure 8.
A plot showing the cross-sectional distributions of axial directional
residual stress of Ti-6Al-7Nb before and after LSP.
A plot showing the cross-sectional distributions of axial directional
residual stress of Ti-6Al-7Nb before and after LSP.
Surface hardness subject to laser shock peening
Improving micro-hardness can benefit the wear resistance, which is one of the
main failure issues in many surface modification methods, as the main
failure mode of orthopaedic implants is wear. The Vickers micro-hardness
distributions in the cross-section of the Ti-6Al-7Nb alloy, before and after
LSP are shown in Figure
9. The longitudinal section microhardness of the untreated
scatters from 325HV0.05 to 335 HV0.05. After laser
shock peening, it was seen that the Ti-6Al-7Nb alloy are surface hardened
and the improvement of microhardness range from 365 HV0.05 to
425HV0.05. Yin et al.,
examined the effects of LSP on microhardness of Ti-6Al-4 V. The
microhardness was improved from 390 HV0.5 to 470HV0.5,
due to two factors which separately are a severe plastic layer was generated
and the grain refinement according to Hall-Petch formula.
Figure 9.
The microhardness map showing the through thickness microhardness
distribution of Ti-6Al-7Nb alloy before and after multiple laser
shock peening in surface and sub-surface.
The microhardness map showing the through thickness microhardness
distribution of Ti-6Al-7Nb alloy before and after multiple laser
shock peening in surface and sub-surface.If we keep the overlap constant, looking at the microhardness distribution
according to laser energy we can see that the depth of the surface hardened
layer and microhardness are proportional to the laser energy. For instance,
at 33%, 7 J laser-induced around 410 HV0.05 at the near-surface
followed by 375 HV0.05 (5 J) and 365HV0.05 (3 J). The
microhardness decreases gradient along the depth direction, eventually,
leading to a stable value as same as the untreated means the surface
hardened effect are disappeared beyond this point. Ge et al.,
laser peened the magnesium with 6.5, 8.5 and 10.5 J laser energy. The
plastic deformation layer depths at three different laser energies are 0.6,
0.7, and 0.8 mm, respectively. The surface hardened effect will reduce along
the cross-section direction and the affected depth is called plastic
deformation layer. A similar trend was also be found in 50% and 67% curves.
Amongst these three overlaps, we can see that the gap in the 67% curves are
much narrow which means the surface hardened effect is not so obviously
improved by 7 J laser.By keeping laser energy constant, it can be seen that microhardness and
surface hardness depths are also proportional to overlap. For example, the
micro-hardness of 3 J at the near-surface is 365HV0.05 (33%),
372HV0.05 (50%) and 377HV0.05 (67%). In 5 and 7 J,
the microhardness at the near-surface was improved to 374 HV0.05
(5 J, 33%), 389 HV0.05 (5 J, 50%), 402 HV0.05(5 J,
67%) and 409 HV0.05 (7 J, 33%), 416 HV0.05(7 J, 50%)
425 HV0.05 (7 J, 67%).
Wear behaviour in simulated body fluid
Wear behaviour
Figure 10 shows
the COF versus number fretting wear cycles of Ti-6Al-7Nb alloy subject
to LSP parameter of 7 J group in SBF. All laser shock peened curves are
similar which are characterised by an initial run-in stage with a sharp
increase, followed by a steady-state regime. As shown in Figure 10(b), it
was found that in the run-in step, the untreated curve increases faster
than that of laser shock peened curves. This can be attributed to the
surface hardness induced by LSP. A low COF indicated an easy sliding of
two contacting bodies. However, with the increasing number cycles of
wear tests, there is no obvious difference amongst the untreated and
laser shock peened samples in 3 and 5 J groups when they gradually
stabilised in the steady-state step. As present in the microhardness
section, the surface hardened depth after 3 and 5 J LSP is comparatively
shorter than that of 7 J. What is more, although surface hardness was
induced by LSP, the surface roughness was also increased which was
detrimental to the wear resistance. Once the cycles are more than
certain values, the surface hardened layer of 3 and 7 J samples have
already worn out, whereas, the worn surface was the matrix and that was
why COFs of the laser shock peened was very closed to that of the
untreated.
Figure 10.
Comparative plot showing the friction co-efficient (COF) versus
number of cycles for Ti-6Al-7Nb alloy before and after laser
shock peening in 7 J, 33%; 7 J, 50%; 7 J, 67% (a); the magnified
run-in stage in (b); the magnified steady-state step shown in
(c).
Comparative plot showing the friction co-efficient (COF) versus
number of cycles for Ti-6Al-7Nb alloy before and after laser
shock peening in 7 J, 33%; 7 J, 50%; 7 J, 67% (a); the magnified
run-in stage in (b); the magnified steady-state step shown in
(c).When it comes to 7 J groups (Figure 10), at the run-in stage
(Figure
10(b)), COFs of untreated also increased higher than that of
the laser shock peened samples. However, in the steady-state step, we
can see that the average of COFs of laser shock peened samples are all
lower than that of untreated. In addition, the gradient decreasing
sequence can be observed that the sequence is 7 J,
33% > 7 J, 50%
J, 67%.COFs of 7 J, 50% and 7 J, 67% are comparatively close to each other. What
is more, the wear volume loss of Ti-6Al-7Nb samples before and after LSP
are calculated with maximum worn depth and presented in Figure 11. The
wear volume loss of the untreated surface is
5.72 × 10−3 mm3. In comparison to the
untreated surface, 5 J, 33% lost the maximum volume amongst all the
laser shock peened samples with a value of
8.23 × 10−3 mm3, followed by 3 J, 50%,
(7.01 × 10−3 mm3), then 3 J, 67%
(7 × 10−3 mm3) and finally 3 J, 33%
(6.89 × 10−3 mm3) that the wear volume loss
are higher than that of the untreated. Once the laser energy and overlap
are increased to 5 J, 50%, the wear resistance effect of LSP start to
show. Marked by blue in the Figure 11 which values are
lower than that of the untreated surface, 7 J, 50% experienced the
minimum volume loss (3.2 × 10−3 mm3), onwards is
5 J, 67% (3.67 × 10−3 mm3), 7 J, 33%
(4.12 × 10−3 mm3),7 J, 67%
(4.34 × 10−3 mm3) and 5 J, 50%
(5.06 × 10−3 mm3). It is well-known that
enhanced hardness and microstructural refinement are both induced after
LSP surface treatment to ultimately, improve wear resistance, but the
surface roughness is also increased which is detrimental to wear
performance. Although, the surfaces are hardened and the average size of
sub-grains are minimised, the wear volume loss is still higher after the
LSP. This is because of the increase of surface roughness after LSP. In
addition, the wear volume loss of sample 5 J, 33% is comparative close
to the one of the untreated sample. Therefore, it can be considered as a
neutralised result among roughness (negative to the wear resistance),
surface hardness and microstructure refinement (positive to the wear
resistance).
Figure 11.
The plot showing wear volume loss of Ti-6Al-7Nb alloy before and
after laser shock peening.
The plot showing wear volume loss of Ti-6Al-7Nb alloy before and
after laser shock peening.
The worn surfaces
According to the worn surfaces, Ti-6Al-7Nb samples can be divided into 3
categories which are: (1) the untreated and treated at 5 J, 50%; (2)
3 J, 33%, 3 J, 50%, 3 J, 67% and 5 J, 33%; (3) 5 J, 67%, 7 J, 33%, 7 J,
50% and 7 J, 67%. Therefore, Figure 12 shows the typical
appearance of fretted scars on Ti-6Al-7Nb before and after LSP
(including untreated, 5 J, 33% and 7 J, 67%). This is consistent with
the wear volume loss. As shown in Figure 12(a), the ploughing
wear mechanism dominated in the wear process of the untreated and 5 J,
50%. In the high magnification images, the micro-ploughing and spalling
features on the abraded surfaces can be seen. With respect to the
samples 3 J, 33%, 3 J, 50%, 3 J, 67% and 5 J, 33%, as exhibited by Figure 12(b),
the worn surfaces exhibited much more severe wear among the others.
Extensive debris, grooves, material adhesions and delamination were
found after the adhesive and abrasive wear along the sliding wear
direction. Although the hardened surface layer in laser shock peened
samples benefited the wear resistance, the asperities on the surface
induced by LSP facilitated the removal of the top material off the
surface. Compared to that, as shown in Figure 12(c), 7 J, 50% exhibits
less wear in the SBF as the surface hardening at least is increased to
400 HV0.05. This enable them to get better performance in
wear. From the worn SEM images, adhesion and abrasive wear still
co-existed on the worn surfaces as the adhesion material, spalling, and
delamination was found. However, the delamination area ratio is
proportional to the wear volume loss. Especially in 7 J, 50%, from the
SEM image, the wear scar consists of most delamination and less adhesion
material. Therefore, it was seen that the delamination wear is the main
wear mechanism.
Figure 12.
(Continued)
(Continued)In general, the diameter of the wear scars are proportional to the wear
scar depth as shown in Figure 13. However, some
samples such as 3 J, 67% and 5 J, 33% are not consistent with that
trend. This may be due to the FBS environment. The wear is a process
that the removal of the passive oxide layer exposing the metallic
surfaces, thereby, undergoing anodic dissolution. As such, the wear scar
acts as a cathode. The body simulated solution aggravates the anodic
dissolution and material removal during the wear process. What is more,
in the present study, the wear results is a neutralisation result
amongst surface roughness, surface hardness and the microstructural
refinement. In terms of wear scar depth, limited surface hardening with
low surface roughness may not be worse than higher surface hardening
with higher surface roughness. That is why the two set of parameters in
the middle exhibit the maximum wear scar depth.
Figure 13.
The plot showing the average diameter of wear scars for
Ti-6Al-7Nb alloy before and after laser shock peening.
The plot showing the average diameter of wear scars for
Ti-6Al-7Nb alloy before and after laser shock peening.
Wettability
The advancing contact angle of Ti-6Al-7Nb titanium before and after LSP with
distilled water is presented in Figure 14. As a base line, the contact
angle of untreated samples with distilled water of 53.96°. Comparing this, all
laser shock peened contact angles are higher than the untreated which was also
concluded in the work of Prabhakaran et al.
and Caralapatti and Narayanswamy.
The highest water contact angle is 79.94°, which was 3 J, 33%. Meanwhile,
the lowest for water was 64.09°, at 7 J, 67%. What is more, observing the figure
form longitudinal direction, contact angle values rise with the increase of
laser energy at the same overlap. However, in the transverse way, the contact
angles decreased with the increase of overlap at the same laser energy
level.
Figure 14.
Advancing contact angle on both untreated and laser shock peened
specimens for distilled water.
Advancing contact angle on both untreated and laser shock peened
specimens for distilled water.
Cell morphology and viability
Figure 15 presents the
fluorescence images of the attached cells on the titanium surfaces prior to and
after LSP after 72 h. It was seen that the morphology of cells was well
distributed on both surfaces prior to and after LSP. In the Figure 15(b) to (d) (3 J group from 33% to 67% overlap),
the cell numbers are increased with the respective laser overlap. However,
compare to the untreated, the cell numbers are quite lower after 3 J, laser
energy processing. In 5 J groups (Figure 15(e)–(g)), the cell numbers are
not increased with overlap as cells on the 50% (Figure 15(f)) and 67% (Figure 15(g)) are
comparatively lower than that of 5 J, 33% (Figure 15(e)) overlap. In comparison to
the untreated surfaces, the numbers in 5 J, 33% and 5 J, 67% are higher than
that of the untreated surface. In terms of 7 J groups (7 J, 33%, 7 J, 50% and
7 J, 67%), only the cell numbers in 7 J, 33% (Figure 15(h)) is lower than that of the
untreated surface and the lowest among the three variables. 7 J, 50% (Figure 15(i)) and 7 J,
67% (Figure 15(j)) are
comparatively close to each other.
Fluorescence image showing cell morphology prior to and after Ti-6Al-7Nb
alloy incubated for 72 h: (a) Untreated, (b) 3 J, 33% overlapping, (c)
3 J, 50%, (d)3 J, 67%, (e) 5 J, 33%, (f) 5 J, 50%, (g) 5 J, 67%, (h)
7 J, 33%, (i) 7 J, 50% and (j) 7 J, 67%.Figure 16 presents the
cell viability of MG-63 cells after 24 and 72 h on the as-control and the laser
shock peened surface with MTT assay. Overall, it was observed that the cells can
attach onto both surfaces of untreated and the laser shock peened surfaces. This
meant that LSP was not cytotoxic to Ti-6Al-7Nb alloys. Then, in the data for
24 h group, except 5 J, 33% and 7 J, 67%; the optical density value of the rest
of the laser shock peened are all lower than untreated. This means the cell
viability in the above samples were not higher than that of the untreated. There
it indicates that laser overlap is correlated with the cell viability after
laser peening surface treatment with these particular parameters.
Figure 16.
shows the absorbance of MG-63 cells on the surfaces of positive control,
untreated and the laser shock peened for 24 and 72 h (Data = mean ± St.
Dev; N = 3, *p < 0.05 compared to
the positive control at the same time point).
shows the absorbance of MG-63 cells on the surfaces of positive control,
untreated and the laser shock peened for 24 and 72 h (Data = mean ± St.
Dev; N = 3, *p < 0.05 compared to
the positive control at the same time point).After 72 h, it was observed that all optical density values of the laser shock
peened surfaces were equal to or higher than that of untreated. 5 J, 50% reached
the highest cell growth with a mean value of 0.82, followed by 5 J, 33% (0.79),
3 J, 67% (0.73) and 7 J, 50% (0.712).
Discussion
The fluorescence images suggest that not all LSP parameters are beneficial to the
osseointegration of implant materials, despite having a high level of compressive
residual stress. It was observed that cell numbers after some LSP were much lower
than that of the untreated in the confocal images. These LSP parameters were: 3 J,
33%, 3 J, 50%, 3 J, 67%, 5 J, 50%, and 7 J 33%. Furthermore, in the 1 day MTT
result, cell viability of 3 J groups, 5 J, 50%, 7 J, 33% and 7 J, 50% are
comparatively close to or lower than that of the untreated. This was postulated due
to the differences amongst the surface morphologies after LSP. From another point of
view, LSP surface treatment can works as a filter function that LSP selects high
viability cell and eliminates the unhealthy cells and in turn, could render greater
benefits. This meant in the early stages, certain LSP parameters are detrimental to
the cell number, and somehow, the cell on the laser shock peened surfaces will
decrease. These results are also in agreement with the work of Bagherifard et al.,
who seeded osteoblast cells on shot peened and untreated surfaces for 1, 3
and 7 days. It was found that after 1 and 3 days, the viability of the cells of shot
peened were lower than that of the untreated. Even after 7 days culture, there were
still no significant improvements after shot peening on the surfaces. However, in
our study, compared to 24 h result, the cell viabilities of all laser shock peened
surfaces were significantly increased after 72 h, while the untreated surfaces of
Ti-6Al-7Nb, were slightly decreased. This might be due to the rough surface that
must be beneficial to the bone-implant interface ossoeintegration. It is the micro
features such as dimples, grooves and nano features (reliefs) induced by LSP that
improved the cell viability on the treated surfaces. Moreover, the increased value
was found to be comparatively large in 3 J group, and sample treated at 5 J,
50%.Overall, amongst nine LSP parameters, three of them (5 J, 33%, 5 J, 67% and 7 J, 6%),
exhibited consistent performance in both 24 and 72 h experiments. Compared to the
untreated; the increase of 5 J, 67% and 7 J, 67%, are not obvious. In addition,
after considering the performance in 24 h experiments, although, 5 J, 50%, was the
highest (0.82) in 72 h experiment, 5 J, 33% was the optimal LSP parameters for cell
viability as it rendered high cell viability from evaluating all the
experiments.It has been reported in the literatures that the biocompatibility of the implant was
affected by surface finishing, and the corresponding surface wettability. Surface
topographies were characterised by parameters such as root mean squared, maximum
valley depth, maximum peak height and skewness etc. With that said, the parameters
of 3-D mean average roughness Sa were the most used to present implant surface
finishing. However, such parameters are not suitable to stand for the periodic
surface structures after surface modification as different surface morphologies may
have the same surface roughness with totally different wetting properties.
Therefore, in this study, considering the relationship between the cell viability
and the overlap is not appropriate as the surface morphologies induced by overlap
vary from each other at the same laser energy. Up on keeping the overlap consistent
and laser energy variable was because the surface morphologies are nearly with
varying in the depth of the valley and the height of the peak. It is more reasonable
to analyse the relationship by keeping the overlap consistent. Figure 17 showed the scatter distribution
of the contact angle (water) versus the optic density. It can be seen that the
highest cell viability exhibits in the range between contact angles of 70°–75°. In
33%, 50% and 67% laser overlapping groups, 5 J, 33%, 5 J, 50% and 3 J, 67%, present
the highest cell viability in the range. If the contact angle is lower than 70° or
higher than 75°, the cell viability will comparatively be lower than that in this
range. The fitted third-degree polynomial equation indicated that the optimal
contact angle should to be located at 71.56°. Tamada and Ikada,
found that the optimal water contact angle for cell adhesion was
approximately 70°, which was in good agreement from the results of our research
herein. From the above discussion, surface wetting properties is the directly
influential factor while the surface roughness was also determining the cell
viability via surface wetting properties. This is because, surface roughness is also
an important factor that determines the surface wettability.
Figure 17.
Plot of contact angle versus optical density with the fitting equation.
Plot of contact angle versus optical density with the fitting equation.Combining with the mechanical, interfacial results, properties such as
biocompatibility, mechanics, and wettability cannot all be benefited by only one
simple LSP parameter. For example, with respect to residual stress and
microhardness, 7 J, 67% are optimal LSP parameters leading to maximum compressive
residual stress and surface hardening. However, such benefits may not necessarily
produce the best surface morphology that creates a beneficial biological response.
The morphology may also affect the wettability by increasing the contract angle.
Therefore, the evaluations of the performance post different LSP conditions need to
be considered comprehensively, with a particular requirement and or a specification.
It has to be compromised between mechanical and biological properties.Regarding the LSP parameters of 3 J, although, the cell viability after 72 h was
higher than that of 7 J treated surfaces, the wear resistance was inadequate after
LSP. This is not acceptable for the medical industry applications as wear debris may
be worn out inside the human body, leading to early failure of the implant. Thus,
low laser energy is not applicable for improving the wear resistance.In the 5 J LSP group, regarding cell viability performance, 5 J, 33% and 5 J, 50%
were comparatively better than 5 J, 67%. With that said, the wear loss volumes are
higher due to the lower surface hardening. Especially, the wear volume loss of 5 J,
33% was the highest amongst all LSP samples. In other words, compared to the
untreated, 5 J, 67% improved the wear resistance, but did not enhance the cell
viability. The surfaces treated with 5 J, 50% compromised of wear volume loss and
rendered cell viability. As for 72 h culture, the cell viability is the highest
amongst samples post all LSP parameters with the comparative low wear loss volume.
For further improvement, multiple LSP impacts can be applied to the 5 J, 50% for the
deeper surface hardening layer, thereby, enhancing the wear resistance.As mentioned above, 7 J samples received the deepest surface hardening and exhibited
the best performance in wear resistance. However, the improvements of the cell
viability are comparatively lower than that of other LSP parameters. This is due to
the decrease of contact angle leading to contact angle value lower than 70°. The
pros and cons of LSP on cell viability and wear resistance are summarised in Table 4. It can be seen
that compromising the results of the cell viability and wear resistance, 7 J, 50%
was the optimal LSP parameter for improving the comprehensive properties of
Ti-6Al-7Nb alloy.
Table 4.
The pros and cons of laser shock peening on cell viability and wear
resistance.
Wear resistance
Cell viability
3 J, 33%
×
√√
3 J, 50%
×
√
3 J, 67%
×
√√
5 J, 33%
××
√√
5 J, 50%
-
√√
5 J, 67%
√√
√
7 J, 33%
√√
√
7 J, 50%
√√√
√√
7 J, 67%
√√
-
×: Worse than the untreated; -: Comparatively closed to the untreated; √:
Better than the untreated.
The pros and cons of laser shock peening on cell viability and wear
resistance.×: Worse than the untreated; -: Comparatively closed to the untreated; √:
Better than the untreated.
Conclusions
This paper focuses on laser shock peening, a medical-grade Ti-6Al-7Nb alloy with a
range of laser energies (3, 5 and 7 J) and overlaps (33%, 50% and 67%) with a
spot-size of 3 mm, 20 ns pulse duration and a frequency of 5 Hz. The evaluation was
carried out from microstructural, mechanical, interfacial and biological aspects.
The corresponding conclusions were drawn as follows:By LSP, the average sub grain size can be refined to 36.8% at maximum. The
evaluation of phase transformation revealed that there was no new phase
formed after LSP. The peak was broadened with increasing the energy, whilst
keeping the overlap constant. Dislocation tangles, dislocation walls,
dislocation cells and dislocation pile-ups were observed using TEM.
According to Hall-Petch theory, these deformation features benefit the
mechanical properties of Ti-6Al-7Nb alloy after performing laser shock
peening.The surface was hardened after LSP. Laser energy and overlap were
proportional to microhardness at the near-surface. LSP parameters of 5 J,
67%, 7 J, 33%, 7 J, 50% and 7 J, 67% all improved the wear resistance in
SBF, while parameters of 3 J, 33%, 3 J, 50%, 3 J, 67% and 5 J, 33% could
aggravate the implant failure due to wear. LSP induced surface roughness is
detrimental to the wear resistance of Ti-6Al-7Nb alloy, if the surface
hardening effect is not enough.From fluoresce images after 72 h culture experiment, it was seen that LSP
surface treatment was not toxic to MG63 cell adhesion. There is not a
dramatic increase in the cell numbers on both surfaces before and after LSP.
By contrast, the cell numbers of non-optimal LSP parameters are even quite
lower than that of untreated surfaces. The MTT assay results showed that,
after 72 h the cell viability of laser shock peened samples are all higher
than that of untreated. The cell viability is directly related to the
surface wetting properties as more contact angle is close to 71.5°, the
higher the cell viability it presents.Considering with both mechanical and biological performance, 7 J, 50% is the
optimal LSP processing parameters for improving the implant properties.This work has shown that laser shock peening has a potential to be deployed to
improve the implant osseointegration, as it not only improves the mechanical
properties, but also, the cell viability. However, such cell viability, for example
the culture days are not enough to “characterize the biocompatibility of laser shock
peened titanium samples. Further work is underway to verify/confirm this with
additional biological experiments, namely: mesenchymal stem cell proliferation,
adhesion, and differentiation.