Habtu Adane Aytolign1, Desalegn Muche Wudineh2, Yophtahe Woldegerima Berhe1, Wubie Birlie Checkol1, Misganaw Mengie Workie1, Shimelis Seid Tegegne3, Awoke Alemneh Ayalew4.
Abstract
Background: Anesthesia Equipment malfunction is one of the most common factors contributing to intraoperative surgical patient morbidity and mortality. It is impossible to give anesthesia without proper anesthesia machine checks and airway equipment preparation. Therefore, all anesthesia professionals should make sure that the anesthetic machine and equipment are working correctly. Method: An institutional-based prospective observational study was conducted at the University Comprehensive Specialized Hospital, Operation rooms, from April 10, 2020 to May 10, 2020. About 90 anesthetists were working regularly in the operation theater both emergency and elective patients. Those include; 26 Msc holders, 17 MSc students, 7 BSc anesthetists, and 40 graduating BSc students. These descriptive data were presented with frequency, percentage, and table. Result: The overall compliance rate was 87%. Whereas; 12.46% of clinicians have not met the standard. Out of standards that were not performed, 25.81% were not available from the setup.Entities:
Keywords: Airway management; Equipment safety; Protocol compliance; Quality of health care
Year: 2022 PMID: 35734739 PMCID: PMC9207033 DOI: 10.1016/j.amsu.2022.103775
Source DB: PubMed Journal: Ann Med Surg (Lond) ISSN: 2049-0801
Anesthesia machine checklist at university of Gondar comprehensive specialized hospital Operation Theatre, (202).
| Variables | yes | Not | Percent (%) |
|---|---|---|---|
| 1. Is the electrical supply to the machine checked? | 192 | 10 | 95 |
| 2. Is the oxygen supply (cylinder) to the machine is checked with | |||
| 2a. Cylinder pressure | 197 | 5 | 97 |
| 2b. Pressure regulators setting | 174 | 28 | 86 |
| 2 c. Oxygen failure alarm | 190 | 12 | 94 |
| 3. Rotameter checked with | |||
| 3a. Bobbins move freely | 191 | 11 | 95.55 |
| 3b. Anti-hypoxic device checked | 198 | 4 | 98.02 |
| 3 c. Oxygen flush checked | 178 | 24 | 88.12 |
| 4 c, Leak checked | 200 | 2 | 99.01 |
| 4. Vaporizer checked for | |||
| 4 a. Correctly seated | 186 | 16 | 92.08 |
| 4 b. Filled and checked for leak | 185 | 17 | 91.58 |
| 5 a. Configuration and connection | 190 | 12 | 94.06 |
| 5 b. Soda-lime inspected | 183 | 19 | 90.59 |
| 5 c. Pressure leak test performed | 194 | 8 | 96.04 |
| 5 d. Function of APL valve and unidirectional valve | 190 | 12 | 94.06 |
| 5 e. Gases delivery through the face mask | 179 | 23 | 88.61 |
| 6 a. Ventilator tubing configuration | 170 | 32 | 84.16 |
| 6 b. Appropriate control setting | |||
| 6 c. Ventilator function checked and gases delivery through face mask | 175 | 27 | 86.63 |
| 7. Scavenging Is connected and functioning checked | 160 | 42 | 79.21 |
| 8. Alternative means of ventilation | 130 | 72 | 64.37 |
| 9. Monitoring checked | |||
| 9 a. Pulse oximeter checked | 200 | 2 | 99.01 |
| 9 b. Capnography checked and functional | 20 | 182 | 9.90 |
| 9 c. NIBP measurement functional | 195 | 7 | 96.53 |
| Unanticipated difficult airway preparation | |||
Airway equipment preparation at University of Gondar comprehensive specialized hospital Operation Theatre, (202).
| 1 | Different size ETT | 185 | 17 | 91.58 |
|---|---|---|---|---|
| 2 | Laryngeal mask (3 sets per O.R) | 194 | 8 | 96.04 |
| 3 | Different size laryngoscope blade | 192 | 10 | 95.05 |
| 4 | Suction apparatus (functional) | 167 | 35 | 82.67 |
| 5 | ETT introducer bougie (1 set per O.R) and stylet | 166 | 36 | 82.18 |
| 6 | Airways (oral or nasal) | 140 | 62 | 69.31 |