| Literature DB >> 35729182 |
Aléxia Thamara Gasparin1, Claudiane Isabel Franco Araujo2, Patricia Schmitt2, Mônica Ribas Cardoso2, Maiara Carolina Perussolo2, Thainá Caroline Schuartz de Jesus2, Erika Bergamo Santiago2, Ivan Lucas Reis Silva2, Ricardo Gurgel de Sousa2, Flavia Zhu Teng2, Evair Borges Severo2, Victor Henrique Alves Ribeiro2, Milena Andreuzo Cardoso2, Fernanda D'Amico Silva2, Carolina Rodrigues de Araujo Perazzoli2, João Samuel de Holanda Farias3, Bernardo Montesanti Machado de Almeida2, Sergio Renato Rogal Júnior2, Marcus Vinícius Mazega Figueredo2.
Abstract
The complete blood count (CBC) is one of the most requested tests by physicians. CBC tests, most realized in conventional hematological analyzers, are restricted to centralized laboratories due to frequent maintenance, large devices, and expensive costs required. On the other hand, most handheld CBC devices commercially available show high prices and are not liable to calibration or control procedures, which results in poor quality compared to standard hematology instruments. The Hilab system is a small-handed hematological platform that uses microscopy and chromatography techniques for blood cells and hematimetric parameters analysis through artificial intelligence, machine learning, and deep learning techniques. For clinical evaluation of the handheld CBC device, 450 blood samples were analyzed. The samples encompassed normal (82%) and pathological conditions (18%), such as thalassemias (2.2%), anemias (6.6%), and infections (9.2%). For all analytes, accuracy, precision, method comparison, and flagging capabilities of the Hilab System, were compared with the Sysmex XE-2100 (Sysmex, Japan) results. The sample source (venous and capillary) influences were also evaluated. Pearson correlation, Student t test, bias, and the Bland-Altman plot of each blood count analyte were calculated and shown. The significance level was set at p ≤ 0.05. For clinical evaluation, Hilab System and the Sysmex XE-2100 showed a strong correlation (r ≥ 0.9) for most evaluated parameters. In the precision study, analytes showed CV inside the limits established according to European Federation of Clinical Chemistry and Laboratory Medicine guidelines. The flagging capabilities of the Hilab system, compared to the manual microscopy technique, presented high sensibility, specificity, and accuracy. Venous and capillary samples (p > 0.05) do not differ statistically. Considering the need for point-of-care CBCs, the study indicated that the Hilab system provides fast, accurate, low cost, and robust analysis for reliable clinical use.Entities:
Mesh:
Year: 2022 PMID: 35729182 PMCID: PMC9213419 DOI: 10.1038/s41598-022-13913-8
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.996
Figure 1The Hilab system POCT hematology analyzer. Panel (A) shows Hilab lens device representation (Hilab, Brazil). (B) Components of the Hilab Lens test kit: (1) Hemocytometer, (2) Blood collection pipettes and Blood transfer pipettes, (3) Mixing-bottle of diluent 1; (4) Mixing-bottle of diluent 2; (5) Lancet. (C) Hilab Lens sample preparation workflow for capillary samples. Panel (D) shows the Hilab Flow device representation (Hilab, Brazil). (E) Components of the Hilab Flow test kit: (6) Capsule; (7) Blood collection pipette; (8) Lancet. (F) Hilab Flow sample preparation workflow for capillary samples. Both single-use test kits accompany isopropyl alcohol swab and curative. (G) Hilab’s software examples of evaluated blood cells.
Figure 2Bland–Altman plot of the method comparison study between the Hilab System and the Sysmex XE-2100. Pearson correlation, bias, Student t test p-value; upper limit of agreement (ULA), and lower limit of agreement (LLA) are demonstrated for each analyte.
Accuracy, specificity, sensibility, kappa coefficient, and balanced accuracy of the method comparison study, comparing the Hilab system to Sysmex XE-2100.
| Accuracy (%) | Specificity (%) | Sensibility (%) | Kappa | Balanced accuracy (%) | |
|---|---|---|---|---|---|
| RBC | 99.3 | 93.0 | 99.7 | 0.94 | 96.2 |
| HB | 99.1 | 99.7 | 100.0 | 0.96 | 99.3 |
| HT | 98.7 | 96.3 | 98.9 | 0.94 | 97.6 |
| MCV | 97.6 | 88.5 | 98.9 | 0.89 | 93.7 |
| MCH | 97.0 | 93.6 | 97.6 | 0.89 | 95.6 |
| PLT | 99.8 | 99.9 | 99.8 | 0.95 | 99.0 |
| WBC | 98.0 | 93.5 | 98.6 | 0.89 | 96.0 |
| NEU | 99.4 | 96.5 | 99.6 | 0.95 | 98.07 |
| MON | 99.8 | 99.9 | 99.9 | 0.95 | 99.9 |
| LINF | 99.6 | 99.9 | 99.6 | 0.86 | 99.8 |
| EOS/BAS | 80.0 | 99.0 | 98.0 | 0.81 | 89.1 |
Hilab system precision study. Coefficient of variation (CV) and standard deviation (SD) of each range are demonstrated.
| Target range | Repeatability study | Reproducibility study | |||
|---|---|---|---|---|---|
| CV (%) | SD | CV (%) | SD | ||
| RBC | 3.0–4.0 × 106/mm3 | 4.15 | 0.152 × 106 | 4.10 | 0.148 × 106 |
| 4.0–4.5 × 106/mm3 | 0.32 | 0.020 × 106 | 2.77 | 0.131 × 106 | |
| 4.0–5.0 × 106/mm3 | 3.61 | 0.245 × 106 | 2.99 | 0.192 × 106 | |
| 5.0–6.0 × 106/mm3 | 3.04 | 0.156 × 106 | 3.70 | 0.390 × 106 | |
| WBC | 1.0–2.5 × 103/mm3 | 10.97 | 0.306 × 103 | 9.33 | 0.307 × 103 |
| 2.5–4.5 × 103/mm3 | 5.64 | 0.412 × 103 | 5.83 | 0.200 × 103 | |
| 4.5–6.0 × 103/mm3 | 5.38 | 0.225 × 103 | 6.55 | 0.449 × 103 | |
| 6.0–7.5 × 103/mm3 | 6.90 | 0.278 × 103 | 5.17 | 0.244 × 103 | |
| PLT | 50–90 × 103/mm3 | 7.24 | 7.0 × 103 | 7.70 | 10.5 × 103 |
| 90–150 × 103/mm3 | 5.78 | 20.5 × 103 | 8.93 | 20.3 × 103 | |
| 150–250 × 103/mm3 | 6.50 | 21.9 × 103 | 8.80 | 24.3 × 103 | |
| 250–400 × 103/mm3 | 8.66 | 19.15 × 103 | 11.70 | 20.2 × 103 | |
| HB | 4.0–8.0 g/dL | 1.3 | 0.206 | 2.0 | 0.215 |
| 8.1–12.9 g/dL | 2.1 | 0.395 | 1.7 | 0.301 | |
| 13.0–15.9 g/dL | 1.6 | 0.289 | 1.0 | 0.190 | |
| 16.0–19.0 g/dL | 1.9 | 0.277 | 1.5 | 0.310 | |
Figure 3Anticoagulant influence and the effect of sample type. Graphs indicate the Hilab system results for venous (plus K3EDTA) × fingerstick blood samples. Mean (), standard deviation (SD), and p-values of Paired Student t test are demonstrated for each analyte.
Accuracy, specificity, sensibility, kappa coefficient, and balanced accuracy of the flagging study, comparing the Hilab system to the microscopy technique.
| Disturbance | Accuracy (%) | Specificity (%) | Sensitivity (%) | Kappa | Balanced accuracy (%) |
|---|---|---|---|---|---|
| RBC morphology | 97.06 | 100.0 | 97.06 | 0.98 | 98.1 |
| PLT | 100.0 | 100.0 | 100.0 | 1.00 | 100.0 |
| Immature cells | 95.73 | 100.0 | 95.73 | 0.89 | 89.9 |