An-Ju Hsueh1, Nurul Asyikeen Ab Mutalib2, Yusuke Shirato1, Hiroaki Suzuki3. 1. Graduate School of Science and Technology, University of Tsukuba, 1-1-1 Tennodai, Tsukuba, Ibaraki 305-8573, Japan. 2. Graduate School of Pure and Applied Sciences, University of Tsukuba, 1-1-1 Tennodai, Tsukuba, Ibaraki 305-8573, Japan. 3. Faculty of Pure and Applied Sciences, University of Tsukuba, 1-1-1 Tennodai, Tsukuba, Ibaraki 305-8573, Japan.
Abstract
Bipolar electrodes (BPEs) with arrays of cathodic and anodic poles were developed for use in closed bipolar systems. To increase the number of BPEs in the array, the anodic and cathodic poles were connected with each other using thin leads. A further increase in the number of BPEs was achieved by forming the cathodic and anodic poles of the BPEs and the leads in different layers. A device with 9 × 10 arrays of cathodes and anodes was thus realized. When using this device to sense hydrogen peroxide (H2O2), the sensitivity and linear range of calibration plots could be adjusted by changing the driving voltage and the area ratio between the cathodic and anodic poles. The devices were used to image H2O2 and obtain time-lapse images for the diffusion and dilution of H2O2. Furthermore, DNA detection was demonstrated using an electroactive intercalator. The sensitivity could be improved by making the anodic poles smaller with respect to the cathodic pole and concentrating the electrochemiluminescence (ECL) in a small area. The ECL intensity changed according to the target DNA concentration in the solution.
Bipolar electrodes (BPEs) with arrays of cathodic and anodic poles were developed for use in closed bipolar systems. To increase the number of BPEs in the array, the anodic and cathodic poles were connected with each other using thin leads. A further increase in the number of BPEs was achieved by forming the cathodic and anodic poles of the BPEs and the leads in different layers. A device with 9 × 10 arrays of cathodes and anodes was thus realized. When using this device to sense hydrogen peroxide (H2O2), the sensitivity and linear range of calibration plots could be adjusted by changing the driving voltage and the area ratio between the cathodic and anodic poles. The devices were used to image H2O2 and obtain time-lapse images for the diffusion and dilution of H2O2. Furthermore, DNA detection was demonstrated using an electroactive intercalator. The sensitivity could be improved by making the anodic poles smaller with respect to the cathodic pole and concentrating the electrochemiluminescence (ECL) in a small area. The ECL intensity changed according to the target DNA concentration in the solution.
There is an increasing demand for the imaging and multiplexed detection
of molecules related to cellular functions. For this purpose, devices
based on electrochemical principles have advantages with regard to
miniaturization, integration, and batch fabrication as well as high
sensitivity.[1] A straightforward approach
to miniaturizing and integrating a large number of sensing elements
is to simply use more individual components such as three-electrode
systems.[2,3] However, this results in a larger number
of contact pads to be connected to the external instrument. To solve
this problem, electrodes with the same roles can be connected by common
leads and arranged in the form of a matrix.[4−6] Although this
method significantly reduces the number of contact pads (from the
product of rows and columns to their sum), it still has limitations.
Eventually, a much larger number of sensing elements (e.g., hundreds,
thousands, or more) and the signal processing circuits may be all
integrated using the CMOS technology.[7−9] However, when the system
contains easily degraded parts (such as biomaterials) or when disposable
devices are desired, the high fabrication cost of the CMOS technology
is unfavorable.Bipolar electrochemistry is an attractive solution
to the aforementioned
problems.[10−13] Bipolar electrochemical systems contain one or more electrically
isolated strips of electrodes called the bipolar electrodes (BPEs).
In the initial development stage, open bipolar systems with the BPEs
immersed in a single solution were mainly used (Figure A). When a voltage is applied between two
driving electrodes immersed in the solution, a potential gradient
is generated. The potential differences generated at the interface
between the solution and the BPE polarize the two ends of the BPE
to work as the cathode and the anode. Because the BPEs are not connected
to external instruments, the redox reaction on one pole related to
detection is reported via electrochemiluminescence (ECL) on the other
pole. Open bipolar systems are advantageous in integrating a large
number of BPEs.[14] However, electroactive
materials that may react on the anodic and/or cathodic poles of the
BPE coexist in the same solution, which could influence the output
signal. To address this issue, closed bipolar systems with the cathodic
and anodic poles immersed in different solutions are widely used now
(Figure B). Although
the open and closed bipolar systems only differ structurally by a
wall that separates the two solutions, they have quite dissimilar
working principles. In a closed bipolar system, the relationship between
the driving voltage, E, and the potentials of the
driving electrodes and the poles of the BPE with respect to the solution
they contact is expressed by Kirchhoff’s second law as follows,
when the ohmic drop in the solutions can be neglectedhere, Δφ(D1), Δφ(D2),
Δφ(C), and Δφ(A) are the potentials of the
driving electrodes 1 and 2, the cathodic pole, and the anodic pole
with respect to the solution, respectively.
Figure 1
Schematic diagram and
working principles of (A) open and (B) closed
bipolar electrochemical systems. The figure on the right side of panel
A illustrates the relation between the potential gradient in the solution
and the polarization in different parts of the BPE.
Schematic diagram and
working principles of (A) open and (B) closed
bipolar electrochemical systems. The figure on the right side of panel
A illustrates the relation between the potential gradient in the solution
and the polarization in different parts of the BPE.Compared with the open bipolar system, there is a limitation
in
the number of integrated BPEs in the closed bipolar system because
of the presence of the wall that separates two solutions. Indeed,
previous closed bipolar systems for multiplexed detection integrated
only a limited number of BPEs.[15−22] Nevertheless, to realize high-density BPE arrays, bundles of insulated
fibers or pores in a membrane were used to form a large number of
tiny BPEs, and the resultant device was used to demonstrate chemical
imaging.[23−26] However, this type of system has not been used in applications other
than imaging, and batch fabrication and simplification of the entire
measurement setup are also the remaining issues. A reason for the
limitations in multiplexed detection using closed bipolar systems
with a small number of BPEs lies in the use of simple strips of rectangular
BPEs. The cathodic and anodic poles are the critical parts for redox
reactions, while the part between them merely functions as a conductor
to pass electrons. Therefore, as long as the connecting parts can
pass electrons without any problems, their layout may be optimized
to further increase the number of integrated BPEs.In this study,
we fabricated a closed bipolar system with BPE arrays
consisting of a matrix of cathodic and anodic poles connected with
leads for chemical imaging and multiplexed sensing. The system was
successfully used for imaging hydrogen peroxide (H2O2) and multiplexed detection of DNA.
Experimental
Section
Reagents and Materials
All reagents
and materials used for fabricating and characterizing the devices
were obtained from commercial sources: glass wafers (TEMPAX Float;
diameter: 3 inch; thickness: 500 μm) from Schott Japan (Tokyo,
Japan); positive photoresist (S1818G) from Dow Chemical (Midland,
MI); polyimide precursor solution (SP-341) from Toray Industries (Tokyo,
Japan); poly(dimethylsiloxane) (PDMS; KE-1300T) and curing agent (CAT1300)
from Shin-Etsu Chemical (Tokyo, Japan); platinum wire (diameter: 0.5
mm) from Nilaco (Tokyo, Japan); tris (2,2′-bipyridyl) dichloro-ruthenium(II)
hexahydrate (Ru(bpy)3Cl2·6H2O), buffer of tris(hydroxyethyl) aminomethane (Tris), ethylenediaminetetraacetic
acid (EDTA) (TE buffer, pH 7.4), 6-mercapto-1-hexanol (MCH), and methylene
blue (MB) from Sigma-Aldrich Japan (Tokyo, Japan); tri-n-propylamine
(TPA) and other reagents from Wako Pure Chemical Industries (Osaka,
Japan). All chemicals were reagent grade. Deionized Milli-Q water
(18.2 MΩ cm; Millipore, Billerica, MA) was used to prepare all
solutions.Single-stranded probe DNA (pDNA) and target DNA (tDNA)
were obtained from Eurofins Genomics (Tokyo, Japan) with the following
sequences:pDNA: 5′-SH-C6-GCA TCG TAA GTC GTC AGT CAG
CTA-3′tDNA: 5′-TAG CTG ACT GAC GAC TTA CGA TGC-3′Solutions containing pDNA, MCH, or tDNA were prepared with TE buffer
(pH 7.4) containing 10 mM tris, 1 mM EDTA, and 100 mM NaCl. An MB
solution (100 μM) was prepared with 50 mM phosphate buffer solution
(PBS) at pH 7.4 containing 100 mM KNO3.
Structure and Fabrication of Bipolar Electrode
Array
Structures of the fabricated BPE arrays and the experimental
setup are shown in Figure . The BPEs were fabricated using a thin-film process, and
the details are described in the Supporting information. Platinum BPEs were used for basic characterization and imaging,
whereas gold BPEs were used for immobilizing pDNA strands. Each BPE
consisted of square areas for the cathodic and anodic poles and a
lead that connects them. The width of the lead was 70 μm for
all BPEs.
Figure 2
Devices with an array of BPEs. Electrode layouts of (A) type I
device with arrays of 6 × 8 cathodic and anodic poles and (B)
type II device with arrays of 9 × 10 cathodic and anodic poles.
Cross section along the Y–Y′ line is shown on the bottom
right of panel B. (C) Experimental setup with the BPEs on a glass
substrate and PDMS chambers with driving electrodes placed atop the
glass substrate.
Devices with an array of BPEs. Electrode layouts of (A) type I
device with arrays of 6 × 8 cathodic and anodic poles and (B)
type II device with arrays of 9 × 10 cathodic and anodic poles.
Cross section along the Y–Y′ line is shown on the bottom
right of panel B. (C) Experimental setup with the BPEs on a glass
substrate and PDMS chambers with driving electrodes placed atop the
glass substrate.We fabricated two types
of devices named types I and II. In the
type I device (Figure A), the cathodic and anodic poles and leads were formed on the same
glass plane. Areas other than the anodic and cathodic poles were insulated
with a polyimide layer. Platinum and gold BPEs were formed using arrays
of 4 × 5 and 6 × 8 cathodic and anodic poles, respectively.
Dimensions of the cathodic and anodic poles were generally 500 μm
× 500 μm. BPEs with different cathode and anode areas were
also fabricated (Figure S1), which will
be discussed later. For the gold BPEs, dimensions of the cathodic
and anodic poles were 1000 μm × 1000 μm and 100 μm
× 100 μm, respectively.In the type II device (Figure B), the arrays of
the cathodic and anodic poles and
the leads were formed in different layers. First, only the leads were
formed on the glass substrate, and then the polyimide insulating layer
was formed. The insulating layer was opened at the ends of the leads
for the corresponding cathodic and anodic poles. Then, square patterns
that will function as the cathodic and anodic poles were formed to
cover the exposed lead areas and establish electrical connections
(Figure S2). Only platinum BPEs were used
for type II devices, and the dimensions of the cathodic and anodic
poles were 500 μm × 500 μm.
Fabrication
of the Chamber and Construction
of the Setup
A PDMS chamber (height: 10 mm) was formed by
replica molding, and the details are described in the Supporting information. The chamber was placed
on the glass substrate with BPEs and slightly pressed for fixing (Figure C). The arrays of
anodic and cathodic poles were exposed in the chamber. Platinum wires
(diameter: 0.5 mm) serving as the driving electrodes were fixed at
the longitudinal ends of the chamber using a polyimide tape. The ECL
chamber was then filled with 50 mM PBS (pH 7.4) containing 5 mM Ru(bpy)32+, 25 mM TPA, and 100 mM KNO3 as a
supporting electrolyte. The sensing chamber was filled with 50 mM
PBS (pH 7.4) containing the analyte to be detected and 100 mM KNO3.
Detection of ECL
The device with
the BPEs and PDMS chamber was placed in the dark, and a constant voltage
was applied between the two driving electrodes using a DC power supply
(PMX110-0.6A, Kikusui Electronics, Kanagawa, Japan). ECL was captured
with a CCD camera (VB-7010, Keyence, Osaka, Japan) attached to a fluorescence
stereomicroscope (VB-G25, Keyence, Osaka, Japan). The sensitivity
and exposure time of the camera were, respectively, set at ISO 200
and 60 s for multiplexed sensing and H2O2 imaging,
while they were set at ISO 1600 and 30 s for DNA sensing. ECL intensity
data were extracted with ImageJ (NIH) from a single pixel at the center
of the related anodic poles in the captured photograph.
Modification of BPE Arrays for DNA Detection
A type
I device with gold BPEs was used for DNA detection. To clean
the surface of the BPEs, chips with BPEs were immersed for 10 min
in a solution containing 50 mM KOH and 30% H2O2 (w/w). Then, the surface of the cathodic poles was cleaned by scanning
90 cycles at potentials between −1.0 and +1.2 V (vs Ag/AgCl)
at a scan rate of 1.0 V/s in a solution containing 100 mM KNO3 and 10 mM K3[Fe(CN)6] until the shape
of the voltammogram stabilized.To modify the surface of BPEs,
a piece of PDMS with five straight flow channels (height: 100 μm)
was attached to the cathode side of the BPEs (Figure S3). The flow channels were used to accommodate solutions
containing pDNA, MCH, or tDNA. The surface of cathodic poles was first
kept in a 20 μM solution of thiolated single-stranded pDNA overnight
and then in 1.0 mM MCH for 1 h, forming a self-assembled monolayer
containing pDNA and MCH on the surface. MCH was used to block the
surface of the gold electrode.[27,28] After each step, the
PDMS flow channels were removed, and the BPEs were rinsed with pure
water and dried with nitrogen gas.For the hybridization of
pDNA and tDNA, the PDMS flow channels
were again attached to the glass substrate with BPEs after pDNA modification.
Then, standard solutions containing tDNA were injected into the flow
channels and incubated for 2 h. Then, the flow channels were removed,
and the BPEs were rinsed with pure water to remove excess tDNA. Next,
the BPEs were immersed in a 100 μM MB solution and incubated
for 1 h. The BPEs were rinsed with pure water and then used in the
experiments. The whole procedure was performed at room temperature.
Characterization Using the Three-Electrode
System
Immobilization and hybridization of DNAs were first
examined using one of the cathodic poles as the working electrode
in a conventional three-electrode system (Figure S4). The PDMS chamber shown in Figure C was attached to the glass substrate with
BPEs, and only the cathodic chamber was filled with 50 mM PBS containing
100 mM KNO3 (pH 7.4). A commercial liquid-junction Ag/AgCl
reference electrode (2060A-10T, Horiba; Kyoto, Japan) was inserted
into the solution. The internal solution of the Ag/AgCl electrode
was 3.0 M KCl. The platinum wire fixed at the end of the PDMS chamber
was used as the auxiliary electrode.
Detection
of tDNA
ECL from the BPEs
with hybridized pDNA and tDNA on the cathodic poles was detected using
the setup mentioned earlier in Figure C. The anodic chamber was filled with the ECL solution,
whereas the cathodic chamber was filled with 50 mM PBS (pH 7.4) containing
100 mM KNO3.
Results and Discussion
Multiplexed Sensing and Imaging of H2O2
In the fabricated devices, the cathodic and
anodic poles were connected with 70 μm wide lead patterns of
different lengths. Table S1 summarizes
the measured resistances of the shortest and longest leads used in
type I and II devices. Even with the longest leads, the resistance
was approximately 300 Ω. In contrast, the current flowing through
a lead was anticipated to be on the order of a few μA, as per
our previous results obtained using BPEs with cathodic and anodic
poles of similar sizes.[29] Therefore, the
ohmic drop at the leads was expected to be negligible. To confirm
this, we filled the sensing chamber of a type I device with a 4 ×
5 array with 50 mM PBS (pH 7.4) containing 100 mM KNO3 and
examined the ECL generated following the reduction of dissolved oxygen
on the cathodic poles. The ECL images obtained at two different driving
voltages are shown in Figure S5. Uniform
ECL was observed from all anodic poles, although the length of leads
in the device ranged from 5.7 to 22.5 mm. This result demonstrates
that the conductivity of 70 μm wide leads is sufficient and
that the lead length only has negligible influence.Another
type I device with the same cathodic and anodic areas (1000 μm
× 1000 μm) was then used to detect H2O2 in standard solutions of different concentrations. Figure A shows ECL images obtained
with 1, 2.5, 5, and 10 mM H2O2. Under a fixed
driving voltage, the ECL became brighter with the increase in H2O2 concentration. Figure B shows the dependence of ECL intensity on
the H2O2 concentration at various applied voltages.
An enhanced sensitivity was clearly observed upon increasing the driving
voltage, and the ECL intensity increased linearly at lower concentrations.
The ECL intensity tends to become saturated because the ECL reaction
on the anodic pole becomes rate-limiting at higher H2O2 concentrations.
Figure 3
Device response to H2O2. (A) ECL images obtained
with 1, 2.5, 5, and 10 mM H2O2. Driving voltage:
2.4 V. (B) Dependence of ECL intensity on H2O2 concentration obtained at different driving voltages (1.6, 2.0,
and 2.4 V). (C) Dependence of ECL intensity on H2O2 concentration obtained with different area ratios of cathodic
and anodic poles (Ac/Aa = 0.25, 0.5, 1, and 2.25). Driving voltage: 2.0 V. The
symbols and error bars in B (n = 9) and C (n = 3) indicate, respectively, the means and standard deviations
of ECL intensities, which were measured experimentally from separate
anodic poles in the same device under the designated conditions.
Device response to H2O2. (A) ECL images obtained
with 1, 2.5, 5, and 10 mM H2O2. Driving voltage:
2.4 V. (B) Dependence of ECL intensity on H2O2 concentration obtained at different driving voltages (1.6, 2.0,
and 2.4 V). (C) Dependence of ECL intensity on H2O2 concentration obtained with different area ratios of cathodic
and anodic poles (Ac/Aa = 0.25, 0.5, 1, and 2.25). Driving voltage: 2.0 V. The
symbols and error bars in B (n = 9) and C (n = 3) indicate, respectively, the means and standard deviations
of ECL intensities, which were measured experimentally from separate
anodic poles in the same device under the designated conditions.The above experiments used an array of BPEs with
cathodic and anodic
poles having the same area. When the device is used to not only detect
the target analyte but also measure its concentration, it is important
to know which side of the poles is rate-limiting. To check this influence,
BPEs with cathodic poles of different areas were used (Figure S1), while the area of the anodic poles
was fixed (1000 μm × 1000 μm), and the results are
shown in Figure C.
At an area ratio of Ac/Aa = 1 (Ac and Aa are the areas of cathodic and anodic poles, respectively),
a linear relation between the ECL intensity and the H2O2 concentration was observed below 7.5 mM. Upon increasing
the Ac/Aa ratio,
the sensitivity increased; however, the linear range of the calibration
plot decreased at the same time. The opposite tendency was observed
when decreasing Ac/Aa. As shown in Figure B,C, the sensitivity and the linear range of the calibration
plot can be adjusted by changing both the applied driving voltage
and the Ac/Aa ratio. When the analyte concentration is very low (such as the case
of DNA sensing, which will be discussed later), Ac/Aa should be as large as
possible subject to the limitation of observing ECL in small spots,
and the driving voltage should be as small as possible to minimize
the influence of redox-active interferents.Figure A,B shows
the diffusion and dilution of H2O2 after dropping
20 μL of 100 mM H2O2 solution at different
locations in the sensing chamber of a type I device. The array of
cathodic poles is shown on the right side of the figure, with the
location of dropped H2O2 solution indicated
by orange arrows. Note that the ECL image of the spreading H2O2 on the array of cathodic poles in the sensing chamber
is a mirror image of that of the ECL image in the anodic chamber.
In both cases shown in Figure , ECL was initially observed in a localized area near the
dropped solution. As time elapsed, however, the area became larger
and the ECL intensity decreased overall, suggesting that the H2O2 concentration decreased as a result of diffusion
in the horizontal direction and a change in the diffusion layer due
to the consumption of H2O2 by the cathodic poles.
Imaging was also conducted using a type II device with an array of
9 × 10 cathodes and anodes (Figure ). The same tendency was observed with higher
spatial resolution.
Figure 4
Change in ECL intensity on the array of anodic poles when
a H2O2 solution (100 mM, 20 μL) was dropped
onto
1 mL of 50 mM PBS (pH 7.4) containing 100 mM KNO3 at (A)
the center or (B) the upper left part of the array of cathodic poles.
The type I device contained an array of 6 × 8 cathodes and anodes.
Left: ECL images taken at 60 and 120 s after dropping the solution.
The rightmost figure is the layout of the array of cathodic poles,
with the orange arrow indicating the location of the dropped H2O2 solution. Applied voltage: 2.4 V.
Figure 5
Diffusion of H2O2 in a type II device with
an array of 9 × 10 cathodic and anodic poles. From left to right:
ECL images captured at 60 and 120 s after dropping 10 μL of
100 mM H2O2 solution onto 1 mL of 50 mM PBS
(pH 7.4) containing 100 mM KNO3, and a layout of the array
of cathodic poles in the sensing chamber. The orange arrow indicates
the location where the H2O2 solution was dropped.
Applied voltage: 2.4 V.
Change in ECL intensity on the array of anodic poles when
a H2O2 solution (100 mM, 20 μL) was dropped
onto
1 mL of 50 mM PBS (pH 7.4) containing 100 mM KNO3 at (A)
the center or (B) the upper left part of the array of cathodic poles.
The type I device contained an array of 6 × 8 cathodes and anodes.
Left: ECL images taken at 60 and 120 s after dropping the solution.
The rightmost figure is the layout of the array of cathodic poles,
with the orange arrow indicating the location of the dropped H2O2 solution. Applied voltage: 2.4 V.Diffusion of H2O2 in a type II device with
an array of 9 × 10 cathodic and anodic poles. From left to right:
ECL images captured at 60 and 120 s after dropping 10 μL of
100 mM H2O2 solution onto 1 mL of 50 mM PBS
(pH 7.4) containing 100 mM KNO3, and a layout of the array
of cathodic poles in the sensing chamber. The orange arrow indicates
the location where the H2O2 solution was dropped.
Applied voltage: 2.4 V.Resolution of the images
can be improved by simply decreasing the
size of cathodic and anodic poles and the distance between neighboring
poles, as well as increasing the density of cathodic and anodic poles.
The current thin-film process is capable of fabricating the cathodic
and anodic poles of submicrometer dimensions. Here, a trivial but
challenging problem is where and how to place the lead patterns. In
the type II device, the lead patterns were separated from the patterns
of cathodic and anodic poles and formed in different layers. A straightforward
method will be further separating the lead patterns into more layers
and making them thinner while maintaining sufficient conductivity.
As we stressed earlier, an advantage of our devices is using conventional
thin-film technology to batch-fabricate arrays of BPEs, which will
help lower the production cost of the devices.
Application
to DNA Sensing
Figure shows the principle
of DNA detection. The electroactive MB works as an intercalator for
insertion into double-stranded DNAs, and this can be used to detect
hybridization between tDNA and pDNA on the cathodic poles of the gold
BPEs. MB intercalated in the double strand is reduced by applying
an appropriate potential to the electrode, and therefore the hybridization
can be detected from the increase in current.[30,31] First, we checked the modification of gold BPEs and detection using
MB, using one of the cathodes of the BPEs as the working electrode
in a conventional three-electrode system (Figure S4). Figure A shows changes in the cyclic voltammograms after each modification
step of the electrode surface. With a bare gold electrode, a wave
corresponding to the reduction of dissolved oxygen was observed. After
immobilizing pDNA, there was a distinct decrease in current, suggesting
that the immobilized pDNA blocked the diffusion of oxygen. Further
modification of the electrode surface with MCH decreased the current
significantly and the voltammogram became more featureless, indicating
the successful modification with MCH. Cyclic voltammograms of free
MB in solutions without DNAs were also measured (Figure B) as compared to those of
MB intercalated in double-stranded DNAs. Redox peaks of MB were observed,
and the peak currents increased with the increase in MB concentration.
Electron transfer between the electrode and MB intercalated in double-stranded
DNAs (made of pDNA and tDNA) was also confirmed from the cyclic voltammogram
(Figure C). The peaks
were observed at the same potentials as in Figure B, suggesting that electron transfer occurred
between the electrode and MB. The reduction peak current increased
with the increase in tDNA concentration (Figure D), demonstrating that tDNA can be detected
using pDNA immobilized on the gold electrode.
Figure 6
Principle of DNA detection
on BPE, with illustrations of the hybridization
of pDNA and tDNA and the electron transfer to MB.
Figure 7
DNA sensing
using MB as a DNA intercalator and a three-electrode
system. (A) Changes in cyclic voltammograms in 50 mM PBS (pH 7.4)
containing 100 mM KNO3 after sequential modification with
pDNA and MCH on a bare gold electrode. (B) Cyclic voltammograms obtained
with 20, 100, 500, and 1000 μM MB and a bare gold electrode
as the working electrode. (C) Cyclic voltammograms after incubating
1, 5, or 10 μM tDNA and 100 μM MB on the gold electrode
with immobilized pDNA and MCH. (D) Dependence of the reduction peak
current in panel C on tDNA concentration (CtDNA) (n = 3). The red circles and error bars indicate
the averages and standard deviations. All cyclic voltammograms were
obtained at a scan rate of 50 mV/s.
Principle of DNA detection
on BPE, with illustrations of the hybridization
of pDNA and tDNA and the electron transfer to MB.DNA sensing
using MB as a DNA intercalator and a three-electrode
system. (A) Changes in cyclic voltammograms in 50 mM PBS (pH 7.4)
containing 100 mM KNO3 after sequential modification with
pDNA and MCH on a bare gold electrode. (B) Cyclic voltammograms obtained
with 20, 100, 500, and 1000 μM MB and a bare gold electrode
as the working electrode. (C) Cyclic voltammograms after incubating
1, 5, or 10 μM tDNA and 100 μM MB on the gold electrode
with immobilized pDNA and MCH. (D) Dependence of the reduction peak
current in panel C on tDNA concentration (CtDNA) (n = 3). The red circles and error bars indicate
the averages and standard deviations. All cyclic voltammograms were
obtained at a scan rate of 50 mV/s.Next, DNA sensing was performed in a closed bipolar system with
an array of 4 × 5 gold BPEs (Figure A). Five rows of BPEs were placed in five
separate PDMS flow channels, and five solutions containing tDNA at
0, 0.5, 1, 5, and 10 μM were introduced into the flow channels
to hybridize the tDNA with the immobilized pDNA. After removing the
flow channels and rinsing the BPEs with pure water, the array of BPEs
was immersed in 100 μM MB solution to intercalate MB in the
double-stranded DNAs, followed by ECL detection. We first used BPEs
with Ac/Aa = 1. However, under the rate-limiting condition of the cathodic
pole, the ECL was too weak to be detected. Therefore, the area of
the anodic poles was decreased to 1/100 with respect to that of the
cathodic poles (Ac/Aa = 100) to concentrate the emitted ECL within smaller areas
(Figure A). At this
area ratio, ECL could be detected, as shown in Figure B (and magnified images in Figure S6), and the ECL intensity increased with increasing
tDNA concentration (Figure C,D).
Figure 8
DNA sensing using a BPE array. (A) BPE array used for
this purpose,
with the area ratio of the cathodic to the anodic pole (Ac/Aa) of 100. Areas surrounded
by the red lines correspond to the white dotted line in panel B. (B)
ECL images obtained after incubating the BPEs with 0, 0.5, 1, 5, and
10 μM tDNA solutions and 100 μM MB solution in each flow
channel. Driving voltage: 2.1 V. A magnified image of panel B is provided
as Figure S6 in the Supporting information.
(C) Dependence of ECL intensity on the concentration of tDNA obtained
at driving voltages of 2.0 and 2.1 V. “BG” indicates
the background intensity obtained from areas without ECL emission.
(D) Dependence of ECL intensity normalized by the background intensity
on tDNA concentration. Panels (C) and (D) display the average values
and standard deviations of ECL intensities obtained from the four
anodic poles at the same CtDNA (n = 4).
DNA sensing using a BPE array. (A) BPE array used for
this purpose,
with the area ratio of the cathodic to the anodic pole (Ac/Aa) of 100. Areas surrounded
by the red lines correspond to the white dotted line in panel B. (B)
ECL images obtained after incubating the BPEs with 0, 0.5, 1, 5, and
10 μM tDNA solutions and 100 μM MB solution in each flow
channel. Driving voltage: 2.1 V. A magnified image of panel B is provided
as Figure S6 in the Supporting information.
(C) Dependence of ECL intensity on the concentration of tDNA obtained
at driving voltages of 2.0 and 2.1 V. “BG” indicates
the background intensity obtained from areas without ECL emission.
(D) Dependence of ECL intensity normalized by the background intensity
on tDNA concentration. Panels (C) and (D) display the average values
and standard deviations of ECL intensities obtained from the four
anodic poles at the same CtDNA (n = 4).In this experiment, we
checked the response down to 500 nM. If
the detection limit is defined as three times the standard deviation
of the background, then it is around this concentration, which is
relatively high compared with previous works using MB as an intercalator
but based on different detection principles.[32−35] The relatively high detection
limit is not due to the detection chemistry but mainly to the method
of measuring the ECL intensity using image analysis. Here, the ECL
was captured by digital photography and its intensity was calculated
from the images. The detection limit could be lowered by measuring
the ECL intensity using photomultipliers or photodiodes instead. This
approach is applicable in the case of a single sensor. However, it
will be extremely difficult for a large number of ECL spots because
the same number of photomultipliers or photodiodes will then have
to be collected or integrated, and an additional structure is required
to shield the ECL from neighboring sensing sites. Therefore, the application
of these techniques will be unrealistic. In comparison, it is more
realistic to extract ECL intensity information using the appropriate
software from captured digital images. Recently, there is a growing
interest in using smartphones for processing the signals from sensors,[36−41] which may be suitable for using our devices in practical applications.
To improve the detection sensitivity in this case, one option would
be changing the DNA detection chemistry. On the other hand, there
is also room for improving various electrochemical aspects of the
device. We have previously proposed coulometric devices coupled with
silver deposition to realize highly sensitive detection.[42,43] In this technique, the analyte is first converted into silver, whose
amount is then measured by coulometry. The amount of deposited silver
is directly related to the charge generated in the redox reactions,
which is the integral of current. The integral of the ECL intensity
over the exposure time is directly related to the charge generated
during this period. Therefore, if the intercalator is first converted
into silver and then ECL is generated on the cathodic pole (instead
of the anodic pole) following the oxidation of silver, much stronger
ECL could be recorded within a short period of time to reduce the
influence of the background. This could effectively lower the detection
limit.
Conclusions
Unlike
open bipolar systems in which all parts of the BPEs are
immersed in a single solution, it was not straightforward to batch-fabricate
devices with arrays of many cathodic and anodic poles for closed bipolar
systems. Here, we increased the number of cathodic and anodic poles
of BPEs by simply connecting them with thin leads and arranging the
cathodic and anodic poles in the form of matrices. The number of integrated
BPEs was further increased by separating the cathodic/anodic poles
and the leads into different layers.The fabricated device was
used to image H2O2 as an electroactive analyte,
as well as its diffusion and dilution.
The sensitivity and linear range of calibration plots can be adjusted
by changing the area ratio Ac/Aa between the cathodic and anodic poles. The
devices were also used for multiplexed DNA detection. To facilitate
measurement and improve the detection sensitivity, it is effective
to reduce the Ac/Aa ratio. The ECL intensity can be correlated with the concentration
of DNA to be detected.Although it is challenging to record
many ECL signals and achieve
high sensitivity, our devices fabricated using a simple method realized
a large number of integrated sensing sites. These devices may be useful
basic tools for clinical analysis and cell engineering.