Background: Therapeutic drug monitoring (TDM) is useful to assess clozapine adherence and optimize treatment. However, analysis of venous blood levels by liquid chromatography tandem mass spectrometry (LC-MS/MS) is often logistically complicated and process time is prolonged. Objective: To assess the feasibility and reliability of a new point-of-care device, (MyCare™ Insite), using capillary blood for clozapine therapeutic monitoring. Methods: Matched venous and capillary blood samples were collected from patients treated with clozapine on a stable dose. Samples were analyzed by LC-MS/MS and MyCare Insite Clozapine Test. Clozapine plasma levels were compared between methods using linear regression model. Both patients and treatment team completed questionnaires about the feasibility of blood sampling. Results: Of the total sample (44 patients, 61% males, mean age 43 ± 12 years), mean daily clozapine dose was 293 ± 134 mg/day. Linear regression model demonstrated high correlation with R 2 = 0.83 (p < 0.0001) and mean difference of 26 ± 162 ng/ml. More than 60% of the patients found the clozapine TDM to be important. Most of the participants (58%) favored the capillary sampling and 11% claimed that testing method would affect their adherence to TDM. Moreover, a larger portion (72%) strongly preferred to be tested at the office rather than at the lab. Conclusions: The point-of-care device offers an accessible and satisfactory measurement of clozapine blood levels. Both patients and healthcare providers reported preference for capillary sampling as well as for the in-office TDM procedure. The immediate results provided by the device can facilitate rapid and informed clinical decisions and therefore improve clozapine treatment outcomes.
Background: Therapeutic drug monitoring (TDM) is useful to assess clozapine adherence and optimize treatment. However, analysis of venous blood levels by liquid chromatography tandem mass spectrometry (LC-MS/MS) is often logistically complicated and process time is prolonged. Objective: To assess the feasibility and reliability of a new point-of-care device, (MyCare™ Insite), using capillary blood for clozapine therapeutic monitoring. Methods: Matched venous and capillary blood samples were collected from patients treated with clozapine on a stable dose. Samples were analyzed by LC-MS/MS and MyCare Insite Clozapine Test. Clozapine plasma levels were compared between methods using linear regression model. Both patients and treatment team completed questionnaires about the feasibility of blood sampling. Results: Of the total sample (44 patients, 61% males, mean age 43 ± 12 years), mean daily clozapine dose was 293 ± 134 mg/day. Linear regression model demonstrated high correlation with R 2 = 0.83 (p < 0.0001) and mean difference of 26 ± 162 ng/ml. More than 60% of the patients found the clozapine TDM to be important. Most of the participants (58%) favored the capillary sampling and 11% claimed that testing method would affect their adherence to TDM. Moreover, a larger portion (72%) strongly preferred to be tested at the office rather than at the lab. Conclusions: The point-of-care device offers an accessible and satisfactory measurement of clozapine blood levels. Both patients and healthcare providers reported preference for capillary sampling as well as for the in-office TDM procedure. The immediate results provided by the device can facilitate rapid and informed clinical decisions and therefore improve clozapine treatment outcomes.
Clozapine is a unique and effective antipsychotic compound and is known as the most
efficient agent for treatment resistant schizophrenia (TRS). It is estimated that as
many as 30% of individuals with schizophrenia meet the criteria for TRS and
clozapine is considered valuable in 30–75% of this subgroup.[1,2] Clozapine is not only highly
effective; it is also associated with reduced mortality in comparison with other
antipsychotic treatments, as demonstrated repeatedly.[3,4] However, clozapine use is
associated with barriers and the compound is underused in spite of its notable
advantages.[2,5]
Complex pharmacodynamic profile and fear of severe side effects are among those barriers.Measuring serum drug levels is available for many psychopharmacologic agents,
including mood stabilizers, tricyclic antidepressants, and several antipsychotic
drugs, including clozapine.
Therapeutic drug monitoring (TDM) offers many advantages such as reflecting
adherence to treatment, a crucial issue in schizophrenia treatment where adherence
is estimated to be less than 50%,
assisting customized dosing decisions, and avoiding drug toxicity. Serum drug
levels are affected by age, medical conditions, genetics, drug-drug interactions,
and pharmacokinetic variability; all can differ immensely between patients and
during the patient’s life course.
Clozapine blood levels are correlated with clinical outcome: response is
associated with blood levels above 350 ng/ml,[9,10] whereas levels higher than
600 ng/ml might be associated with increased risk to develop side effects.
Yet, measuring clozapine levels is complicated and often avoided. The current
available customary analytic method of liquid chromatography tandem mass
spectrometry (LC-MS/MS) requires transportation to a special lab and a multistep
analysis process that may take several days.New and advanced technologies can enable rapid and easier ways to measure clozapine
blood levels. MyCareTM Insite by Saladax is a new point-of-care (POC)
immunoassay method using capillary blood sampling. This method offers three main
advantages: (1) The analysis requires capillary blood instead of venous blood, which
is preferred by both caregivers and patients.
(2) The test is conducted as an in-office procedure, with no need for
laboratory outsourcing. (3) The results of the test are received immediately on the
spot in less than 7 minutes. Preliminary results of this novel POC device were
recently published,
demonstrating the validity of clozapine blood level measurements.In the current study, we aimed to evaluate the feasibility of the POC analysis among
both patients and healthcare practitioners. In addition, we aimed to corroborate and
extend the initial data on the validity and accuracy of this innovative methodology,
by comparison with the standard LC-MS/MS analysis.
Methods
Sample
Study population for this observational study included 44 subjects treated at
Geha Mental Health Center during the years 2019–2020. The sample included both
inpatients and patients treated in the day care unit, in a public mental health
center in Israel that belongs to Clalit Health Services (CHS) health care
organization. According to the inclusion criteria all participants were (1)
adults aged over 18 years. (2) Diagnosed with schizophrenia, schizoaffective
disorder, or other psychotic conditions. (3) Prescribed clozapine for at least 4
weeks at a stable dose at study entry. (4) Capable to provide a written informed
consent. The study was approved by the institutional review board (approval
number: 0007-19-GEH). Due to the exploratory nature of the study, the results
obtained from the POC device were not available during medical
decision-making.
Study procedures
Following provision of written informed consent to take part in the study,
participants were asked to provide a finger-prick capillary blood sample for
testing in addition to the regular venous blood sample that was collected in the
same session. Samples were collected in the morning (08:00), before breakfast
and the morning dose administration. The intrapatient results of the 2 adjacent
assays, the LC-MS/MS and Insite POC test, were later analyzed and compared.
Testing protocol
Venous sample: Each venous sample, containing 5 ml of blood in an EDTA test
tube, was centrifuged to separate the plasma. The frozen plasma (at −20°)
was later sent to the toxicology lab at ‘Carmel’ medical center (Haifa,
Israel) for analysis by the standard LC-MS/MS technology.Capillary sample: An additional capillary sample, containing 0.01 ml of
blood, was retrieved from the fourth finger of the nondominant hand of each
participant. Each collected whole- blood sample was tested via the MyCare™
Insite POC device using the MyCare Clozapine Test. The MyCare Clozapine
immunoassay is based on an antigen- antibody reaction causing nanoparticle
aggregation that is measured photometrically. The entire procedure was
performed by a trained health care professional (physician or a nurse) at
the clinic room within minutes.
Feasibility evaluation
All participants were requested to complete a questionnaire regarding the
test usability and their personal TDM preference. The questionnaire also
included a series of questions regarding adherence to treatment, habits
(tobacco use and caffeine consumption, etc.) and adverse effects under
clozapine treatment. A second questionnaire, designated for the treating
team (six treating physicians), was composed of clinical global assessment
of each patient. In addition, the physicians were requested to evaluate
their own experience with the POC device.
Statistical analysis
Clozapine blood levels were compared between methods using linear regression.
Significance was set at p < 0.05. Computer software used for
data analysis was SPSS for Windows, version 20.0 (IBM Corp. Armonk, NY,
USA).
Results
Sample characteristics
The sample included 44 patients, 61% males, mean age 43 ± 12 years, mean body
weight was 80 ± 16 kg and 41% of the participants reported cigarette smoking
(Table 1). Mean
daily clozapine dose was 293 ± 134 mg/day (ranging between 50 and 600 mg/day)
(Table 1) with
one-fourth of the patients prescribed a daily clozapine dose above 350 mg/day.
According to clinical global assessment, most of the patients (79%) presented
moderate to severe psychotic symptoms as well as negative symptoms.
Moderate-to-severe side-effects were reported by 44% of them, more than
evaluated by the treatment team.
Table 1.
Characteristics of the patient sample (n = 44).
Characteristic
Result
%
SD
Range
Sex
Male
27
61
Female
17
39
Mean age (years)
43
12
18–71
Mean weight (kg)
79.7
15.5
Smoking tobacco
18
41
Mean clozapine dose (mg/day)
292.6
133.6
50–600
SD, standard deviation.
Characteristics of the patient sample (n = 44).SD, standard deviation.
TDM preferences and test feasibility
When asked about their point of view and preferences regarding TDM, more than 60%
of the patients declared TDM of clozapine to be of importance to them. Almost
half (49%) of the patients preferred frequent TDM testing (medium or higher
frequency of TDM). Most of the participants (58%) favored the capillary blood
sampling, with 11% claiming that testing method would affect their adherence to
TDM. Moreover, a larger portion (72%) strongly preferred to be tested in the
physician’s office rather than the lab. Treatment team (81%) also preferred the
capillary testing and predicted most patients would prefer it (Figure 1).
Figure 1.
Therapeutic drug measurement preferences. (a) Preferred sampling method
of treatment team (n = 37). (b) Preferred sampling
method of patients (n = 38). (c) Preferred sampling
setting of patients (n = 39).
Therapeutic drug measurement preferences. (a) Preferred sampling method
of treatment team (n = 37). (b) Preferred sampling
method of patients (n = 38). (c) Preferred sampling
setting of patients (n = 39).
Clozapine blood levels
Treating physicians predicted that most patients (62%) would present clozapine
blood levels detectable but below the therapeutic threshold, 35% of the
participants would be within therapeutic range and only few (3%) were predicted
to have clozapine level higher than the range. Analyzing blood samples found the
blood level of 57% and 64% of the sample to be below the therapeutic threshold
by the LC-MS/MS and the POC device, respectively. Therapeutic blood levels of
clozapine (range 350–650 ng/ml) were found in 18% and 21% of the patient’s
samples, with the two methods, respectively. Clozapine blood levels above the
upper limit of the therapeutic threshold (above 650 ng/ml) were found in 25% and
16% of the sample, respectively.
Test validity
Linear regression model of TDM measurements from the two methods demonstrated
high correlation with R2 = 0.83
(p < 0.0001) and mean difference (LC-MS/MS minus POC) of
26 ± 162 ng/ml (median of 4.5 ng/ml) (Figure 2). Concordance rate between
samples according to categories was found to be 96% for levels below 350 ng/dl,
62.5% for levels between 350 and 650 ng/ml, and 63.6% for levels above 650 ng/dl
(Table 2).
Figure 2.
Comparison of the two assays using linear regression.
POC, point of care.
Table 2.
The concordance rate between the two assays and clozapine blood levels
(ng/ml).
Lab
Insite
<350
351–650
>650
<350
96%
351–650
62.5%
>650
63.6%
Comparison of the two assays using linear regression.POC, point of care.The concordance rate between the two assays and clozapine blood levels
(ng/ml).
Discussion
In this study, we aimed to assess the feasibility and validity of a novel rapid TDM
immunoassay which is performed as an office-based procedure. Our main findings were
that most caregivers and patients prefer this testing method over the ‘gold
standard’ assay and that the new device provides valid clozapine blood level
results.Our results regarding test validity, with a satisfactory
R2 of 0.83 (p < 0.0001), are
consistent with a previous UK study regarding the same POC device.
The authors assessed clozapine blood levels of 309 patients by the two
methods, the standard laboratory LC-MS/MS assay and the MyCare™ Insite device.
Sample characteristics and inclusion criteria were similar to those applied in the
UK study. The latter revealed similarity between the results obtained by the two
methods, with correlation coefficient of 0.89 and a slope of 1.0 [95% confidence
interval (CI) 0.9–1.0]. Within-patient differences between measures were ⩽10%.In our study, agreement rate between the results obtained by the two methods was
better with lower clozapine blood levels (96%) in comparison with higher levels
(63%). This may imply that the results obtained by the POC tend to be lower compared
with the results from the LC-MS/MS. These findings are comparable with the previous
UK POC study.
As suggested by the authors, some variation between methods is expected due
to use of different types of blood samples (venous plasma versus
capillary whole blood)
and calibration calculations applied by the assay manufacturer to report
plasma results from whole blood samples. In addition, it is important to note that
the gold standard LS-MS/MS is not infallible and may be subjected to errors that may
affect agreement rate.The current study, beyond confirming the test validity, evaluated user experience and
testing method preferences. Patients and their treating physicians were queried
regarding various aspects of TDM, testing procedure and personal viewpoints. The
participants’ responses reflect a clear trend in favor of Insite capillary testing,
indicating that the POC device may contribute to increased TDM adherence. Another
important clinical consideration is the immediacy and proximity of the assay, which
makes the POC device useful to detect and prevent clozapine overdose.Such preference is consistent with clozapine-treated patients and their
practitioners’ attitude toward other POC testing, as reported in a study of a
portable capillary white blood count (WBC) monitoring device.
This is also in line with data on patients satisfaction with POC testing
methods in general practice, like INR capillary testing for anticoagulation monitoring
and sugar blood level monitoring.
Capillary POC methods are gaining more interest and a new portable WBC device
was recently compared well with the gold standard laboratory assay.
The possibility to combine POC WBC monitoring and TDM can be clinically-
beneficial for clozapine- treated patients.In spite of the clear advantages, such as convenient application and instant results,
the POC immunoassay is associated with several built-in drawbacks, which must be
addressed. As mentioned, the results obtained are less accurate at higher drug
levels. Furthermore, the test has a detection upper limit of 1390 ng/ml, which makes
it less suitable to precisely measure clozapine overdose. Therefore, it is advisable
to apply the same testing method when repeated measures are required per-patient and
interpret the obtained results within the clinical context. Another technical
disadvantage is lacking the ability to measure clozapine metabolite levels and
specifically, norclozapine. The ratio of clozapine to norclozapine concentrations,
which can be provided by the LC-MS/MS method, has been suggested to have clinical
value in assessing recent compliance and pharmacokinetic changes,
however, evidence on this subject is conflicting.
Study limitations
Study limitations are inherent to the small sample size of the current
investigation. On one hand, due to the limited scope of the study, all samples
were undertaken by a small team of trained device operators, hence contributing
to its consistency. On the other hand, the small sample size may reflect on
higher effect of the outlier results.In addition, most of the sample measures (capillary as well as venous samples)
detected clozapine blood levels on the lower range. Test accuracy in the
presence on higher clozapine blood levels should be studied more thoroughly in
the future.Further research, including larger samples of patients from diverse care settings
and with various clinical characteristics, is essential to further evaluate both
POC device feasibility and user experience.
Conclusions
In conclusion, clozapine TDM is a valuable tool to ascertain both efficacy and safety
of treatment. The POC device offers a rapid, accessible, and satisfactory measure of
clozapine blood levels. Both patients and healthcare providers reported preference
of capillary sampling as well as the in-office TDM procedure. Using POC immunoassay
may contribute to increase TDM adherence and therefore improving rate and outcome of
clozapine treatment among this difficult-to-treat population.
Authors: C Hiemke; N Bergemann; H W Clement; A Conca; J Deckert; K Domschke; G Eckermann; K Egberts; M Gerlach; C Greiner; G Gründer; E Haen; U Havemann-Reinecke; G Hefner; R Helmer; G Janssen; E Jaquenoud; G Laux; T Messer; R Mössner; M J Müller; M Paulzen; B Pfuhlmann; P Riederer; A Saria; B Schoppek; G Schoretsanitis; M Schwarz; M Silva Gracia; B Stegmann; W Steimer; J C Stingl; M Uhr; S Ulrich; S Unterecker; R Waschgler; G Zernig; G Zurek; P Baumann Journal: Pharmacopsychiatry Date: 2017-09-14 Impact factor: 5.788
Authors: M H Kronig; R A Munne; S Szymanski; A Z Safferman; S Pollack; T Cooper; J M Kane; J A Lieberman Journal: Am J Psychiatry Date: 1995-02 Impact factor: 18.112
Authors: Georgios Schoretsanitis; John M Kane; Can-Jun Ruan; Edoardo Spina; Christoph Hiemke; Jose de Leon Journal: Expert Rev Clin Pharmacol Date: 2019-05-31 Impact factor: 5.045
Authors: G Remington; J Lee; O Agid; H Takeuchi; G Foussias; M Hahn; G Fervaha; L Burton; V Powell Journal: Expert Opin Drug Saf Date: 2016-06-01 Impact factor: 4.250
Authors: Karen Woods; James D Douketis; Terri Schnurr; Krystyna Kinnon; Peter Powers; Mark A Crowther Journal: Thromb Res Date: 2004 Impact factor: 3.944