The Accuracy of Electrical Impedance Tomography for Breast Cancer Detection: A Systematic Review and Meta-Analysis.

Zahra Rezanejad Gatabi1, Mehri Mirhoseini2, Nasrin Khajeali3, Iman Rezanezhad Gatabi4, Maedeh Dabbaghianamiri5, Sara Dorri6.   

Abstract

Introduction: Incidence of breast cancer (BC) in 2020 is about 2.26 million new cases. It is the first common cancer accounting for 11.7% of all cancer worldwide. Disease complications and the mortality rate of breast cancer are highly dependent on the early diagnosis. Therefore, novel human breast-imaging techniques play an important role in minimizing the breast cancer morbidity and mortality rate. Electrical impedance tomography (EIT) is a noninvasive technique to image the breast using the electrical impedance behavior of the body tissues.
Objectives: The aims of this manuscript are as follows: (1) a comprehensive investigation of the accuracy of EIT for breast cancer diagnosis through searching pieces of evidence in the valid databases and (2) meta-analyses of the results.
Methods: The systematic search was performed in the electronic databases including PubMed, Web of Science, EMBASE, Science Direct, ProQuest, Scopus, and Google Scholar without time and language limitation until January 2021. Search terms were "EIT" and "Breast Cancer" with their synonyms. Relevant studies were included based on PRISMA and study objectives. Quality of studies and risk of bias were performed by QUADAS-2 tools. Then, relevant data were extracted in Excel form. The hierarchical/bivariate meta-analysis was performed with "metandi" package for the ROC plot of sensitivity and specificity. Forest plot of the Accuracy index and double arcsine transformations was applied to stabilize the variance. The heterogeneity of the studies was evaluated by the forest plots, χ2 test (assuming a significance at the a-level of 10%), and the I2 statistic for the Accuracy index.
Results: A total of 4027 articles were found. Finally, 12 of which met our criteria. Overall, these articles included studies of 5487 breast cancer patients. EIT had an overall pooled sensitivity and specificity of 75.88% (95% CI, 61.92% to 85.89%) and 82.04% (95% CI, 69.72% to 90.06%), respectively. The pooled diagnostic odds ratio was 14.37 (95% CI, 6.22% to 33.20%), and the pooled effect of accuracy was 0.79 with 95% CI (0.73, 0.83). Conclusions: This study showed that EIT can be used as a useful method alongside mammography. EIT sensitivity could not be compared with the sensitivity of MRI, but in terms of specificity, it can be considered as a new method that probably can get more attention. Furthermore, large-scale studies will be needed to support the evidence.
Copyright © 2022 Zahra Rezanejad Gatabi et al.

Entities:  

Mesh:

Year:  2022        PMID: 35711881      PMCID: PMC9186524          DOI: 10.1155/2022/8565490

Source DB:  PubMed          Journal:  Breast J        ISSN: 1075-122X            Impact factor:   2.269


1. Introduction

Incidence of breast cancer (BC) in 2020 is about 2.26 million new cases. It is the first common cancer accounting for 11.7% of all cancer worldwide [1]. It is the leading cause of cancer death in most countries [2]. The mortality rate of breast cancer varies from region to region. Compared to the USA and Europe, some Asian countries such as Japan and China have lower mortality rates due to breast cancer. Breast cancer mortality rate is higher than the mortality rate of colorectal cancer (8%) and lower than the mortality rate of lung cancer [3]. Similar to many other forms of cancers, breast cancer is the result of several environmental and hereditary factors. High-fat diets, alcohol intake, tobaccos, hormones, radiation, age, sex, and childbearing are identified as the primary risk factors [4, 5]. Statistical data show that many women with breast tumors refer to the hospital only after they feel a quite large lump. At that stage of the disease, a biopsy is required to determine whether the tumor is cancerous (malignant) or noncancerous (benign). For this purpose, clinicians often perform a fine-needle aspiration or a core biopsy [6]. The type of the tumor determines the subsequent treatments. While the core biopsy is a reliable technique to characterize the tumors, it is not hazard-free. The process requires taking out a considerable amount of tissue from the body at several locations. There are many reports on cancerous tumor adhering to the tip of the biopsy needle and introducing the disease to the upper cutaneous layers [7]. The early diagnosis of breast cancer plays a crucial role in minimizing the mortality rate of the disease. Magnetic resonance imaging (MRI), computed tomography (CT) scan, ultrasound imaging, and mammography are current techniques implemented for diagnostic imaging. Among these techniques, mammography became a standard technique for breast imaging and early-stage detection of carcinomas in the breast [8, 9]. Ultrasound imaging is also employed along with mammography to differentiate suspicious breast lesions [10]. While the mammography technique offers invaluable advantages to decrease mortality [11], it has some drawbacks as well. The disadvantage is that it applies ionizing radiation, which is associated with patient discomfort because of breast tissue compression [12]. In addition to the radiation exposure issues, the mammogram may result in false-positive results. It also has a sensitivity of lower than 100% [13]. Also, MRI has significant disadvantages as well, including its high cost, variability in performance, and moderate specificity [14], and application of CT scan has some limitations, such as portability, intermittent use, and X-ray exposure [15]. Hence, there is a need for the development of reliable noninvasive and nonhazardous imaging techniques for the detection and characterization of breast cancer tumors. Electrical impedance tomography (EIT) is a noninvasive technique to image the human breast using electrical impedance characteristics of the tissues [16]. In this imaging technique, an array of external electrodes is employed to extract and reconstruct an image of the inner part of a conductive object. The value of the electrical impedance in human tissues depends on the tissue's electrical storage potential. The electrical impedance is different in the normal tissues and pathologically changed tissues [17]. The variation of cellular water content, variation of extracellular fluids, packing density, and differences in the orientation of the cells are the main reasons for a significant increase in capacitance and conductance of malignant tumors, which results in lower impedance [18]. EIT is a rapid, compact, and inexpensive imaging solution [19-21]. The EIT technique can be employed with almost no age limitations. It can also be used on pregnant women because no ionizing radiation is emitted by the scanning device [16]. Several studies that have been performed previously focused on investigating the accuracy of EIT systems in the detection of breast cancer. These studies did not show comprehensive results in this field. However, to the best of the authors' knowledge, there are no comprehensive meta-analyses on this matter. On the other hand, from the clinical point of view, rare evidence is available for using this method for detecting breast cancer. So, there is still a gap between clinical use and research in this field. Therefore, the aim of this manuscript is as follows: (1) a comprehensive investigation of the accuracy of the electrical impedance tomography technique for breast cancer diagnosis through searching evidence in the valid databases and (2) meta-analyses of the results.

2. Materials and Methods

2.1. Search Strategy and Information Source

A comprehensive literature search according to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines was performed. The search was performed in PubMed, Web of Science, EMBASE, Science Direct, ProQuest, Scopus, and Google Scholar search engine (without time limitation until January 2021) with no language restrictions. The search syntax was developed based on Medical Subject Headings (MeSH) in PubMed. The keywords are ((Electrical Impedance Tomography) OR (EIT)) AND ((breast cancer) OR (breast neoplasm) OR (breast anomaly) OR (breast carcinoma) OR (breast lesions) OR (breast abnormality) OR (breast tumor) OR (breast palpable mass) OR (dense breast tissue) OR (Human Mammary Neoplasm) OR (Mammary cancer) OR (breast malignant neoplasm) OR (Breast malignant tumor)). The reference lists of relevant primary studies, reviews, and key journals for additional studies were also searched.

2.2. Study Selection Process

This systematic review and meta-analysis were conducted based on the PRISMA statement. EndNote software manager was used to manage the references. After removing duplications, two authors independently reviewed references based on the title and abstract. We excluded narrative reviews, opinion pieces, letters, and any other publications lacking primary data. Any disagreements between the investigators regarding the inclusion of articles were discussed and resolved with a third reviewer.

2.3. Inclusion Criteria

Articles were included based on the following inclusion criteria: Observational and interventional studies that assessed using the EIT to diagnose breast cancer Published articles in any language with full English abstracts were considered And explicitly calculated performance of the EII through performance metrics

2.4. Data Items

Two authors independently collected the data from articles and any conflicts were solved by the third author. The relevant data were extracted and collected in Excel form. Data items were author, year, study design, objectives, specificity, sensitivity, PPV, NPV, ACC, number of patients, age, TP, FP, FN, and TN.

2.5. Risk of Bias (Quality) Assessment

The quality of methodological of primary studies was assessed by the modified QUADAS-2 quality assessment tool which was developed by Whiting [22]. QUADAS is an 11-item generic tool that was developed specifically for implementation in diagnostic test accuracy reviews. QUADAS was developed using a formal consensus method informed by empirical evidence.

2.6. Effect Measures

In this study, the reference test was EIT and the gold standard was mammography. Specificity, sensitivity, ACC, and diagnostic odds ratio were the variables for the outcome measure. We conducted Cook's distance analysis as a measure of the influence of a study on the model.

2.7. Statistical Analysis

The statistical analyses were performed by using STATA V.13 statistical software (Stata Corp. 2013. Stata Statistical Software: Release 13. College Station, TX: Stata Corp LP). The hierarchical/bivariate meta-analysis was performed with the “metandi” package for the ROC plot of sensitivity and specificity. Patient frequencies within extracted 2 × 2 data tables were implemented to generate the forest plot of the Accuracy index that was applied using the “metaprop” command, and double arcsine transformations were applied to stabilize the variance. Every study that did not have enough outcome variables was eliminated. We evaluated the heterogeneity among the primary studies by the forest plots, χ2 test (assuming a significance at the a-level of 10%), and the I2 statistic for the Accuracy index.

3. Results

3.1. Study Characteristics

A total of 4027 articles were initially retrieved from the electronic databases. After removing the duplicates, the remaining 3544 studies were screened. A total of 3435 studies were excluded. Among 109 remaining papers, 88 were excluded because their methods were not available or the reference test was different or it did not provide relevant data. Finally, 12 articles were selected for inclusion in this study. Figure 1 illustrates the selection process of the articles based on the PRISMA flow diagram.
Figure 1

The flowchart illustrating the article search process and study inclusion.

The 12 selected studies included data for 5487 patients with breast cancer. These studies were conducted between 1999 and 2017. Most of these studies were conducted in Germany, China, and the USA. The main characteristics of each study and their sensitivity and specificity are summarized in Table 1.
Table 1

Summary of studies included in the meta-analysis.

IDAuthorYearSample sizeAge of participantsCountrySensitivity %Specificity %
1Melloul et al. [23]199912142 to 70Israel72.267
2Malich et al. [24]200058Germany75.972.4
3Malich et al. [25]2001100Germany8163
4Wersebe et al. [26]2002129Mean = 55Germany6269
5Fuchsjaeger et al. [27]200512821 to 89Austria94.674.7
6Szabo et al. [28]200514533 to 92Sweden8649
7Stojadinovic et al. [29]2006155030 to 45USA3895.1
8Stojadinovic et al. [30]2007215530 to 45USA26.494.7
9Wang et al. [31]201028625 to 45China86.772.9
10Amin et al. [7]20141917 to 55Bangladesh7587
11Daglar et al. [32]20162518 to 85Turkey7577.78
12Zhenggui et al. [33]201777118 to 83China92.198.9

3.2. Reporting Bias and Quality Assessment

According to QUADAS-2, the quality of the studies was evaluated. Each study was evaluated by two independent investigators. All disagreements were resolved by consensus. The quality of the included studies based on the QUADAS-2 tool showed a high risk of bias (Table 2). Seven articles had a high risk of bias, and 5 articles had a low risk of bias.
Table 2

QUADAS for the included studies.

IDAuthorQUADAS questions
Risk of biasApplicability concern
PatientselectionReferencestandardIndex testFlow andtimingPatientselectionIndex testReferencestandard
1Melloul et al. [23]YesYesYesYesNoYesYes
2Malich et al. [24]YesNoYesYes?YesYes
3Malich et al. [25]YesNoYesYesYesNoYes
4Wersebe et al. [26]YesNoYesYesYesNoYes
5Fuchsjaeger et al. [27]YesYesYesYes?YesYes
6Szabo et al. [28]YesYesYesYesYesYesYes
7Stojadinovic et al. [29]YesNoYesYesYesNoYes
8Stojadinovic et al. [30]YesNoYesYesYesNoYes
9Wang et al. [31]YesYesYesYesYesYesYes
10Amin et al. [7]YesYesYesYes?YesYes
11Daglar et al. [32]YesNoYesYesYesYesYes
12Zhenggui et al. [33]YesNoYesYesYesYesYes

3.3. Meta-Analysis

The summary ROC curve presents the summary estimates for EIT (Figure 2).
Figure 2

Summary ROC plot of sensitivity and specificity of EIT.

Table 3 illustrates the bivariate and the hierarchical summary receiver operating characteristic (HSROC) parameter estimates along with their standard errors and approximate 95% confidence intervals. Pooled sensitivity and specificity were 75.88 and 82.04 with the standard error of 6.16 and 5.14, respectively. The pooled diagnostic odds ratio (DOR) was 14.37 with a 6.14 standard deviation.
Table 3

Meta-analysis of diagnostic accuracy.

Parameter estimationSE95% CI
Sensitivity75.886.1661.9285.89
Specificity82.045.1469.7290.06
DOR14.376.146.2233.20

Diagnostic odds ratio.

Figure 2 shows the summary ROC plot of sensitivity and specificity of EIT. This diagram is a graphical representation of a fitted model for simultaneous meta-analysis of Se and Sp on a ROC diagram. Each small circle of information is specific to a study, and the size of which is related to the weight assigned to it in the meta-analysis. The confidence interval and the prediction intervals are also marked with dashes. The pooled sensitivity and feature are also marked with a red dot. This graph uses the hierarchical summary receiver operating characteristic (HSROC) method to estimate results. Cook's distance analysis was conducted as a measure of the influence of a study on the model parameters [34]. The results indicate that two studies (ID numbers 8 and 12) had the most influence on the parameters (Figure 3).
Figure 3

Left panel: Cook's distance. Right panel: standardized residuals (standardized predicted random effects).

Table 4 shows the results of the pooled effect of accuracy of 0.79 with 95% CI (0.73, 0.83) and P value = <0.01.
Table 4

Meta-analysis of accuracy.

StudyAccuracy95% CIWeight (%)
Melloul et al. [23]0.680.590.758.57
Malich et al. [24]0.740.620.847.18
Malich et al. [25]0.740.650.828.26
Wersebe et al. [26]0.650.570.738.67
Fuchsjaeger et al. [27]0.800.730.868.66
Szabo et al. [28]0.690.610.768.84
Stojadinovic et al. [29]0.850.830.8610.28
Stojadinovic et al. [30]0.820.800.8310.33
Wang et al. [31]0.790.740.849.57
Amin et al. [7]0.840.620.944.41
Daglar et al. [32]0.760.570.895.11
Zhenggui et al. [33]0.940.920.9510.11
Pooled effect0.790.730.83100
Significant heterogeneity was found between the studies (I2 = 93.26% and P value <0.01). The forest plot is shown in Figure 4. Due to the heterogeneity of the studies, the data were combined using a random-effect model. Different comparisons with different diagnostic methods may cause heterogeneity in our study.
Figure 4

Forest plot of accuracy.

Funnel plot was used to examine the publication bias, and the distribution of points in the two halves of the drawn triangle is discussed. Each point of the number is related to a study, which if these points are drawn relatively uniformly in all parts of the triangle, indicates the absence of publication bias. Finally, no significant publication bias was found through the funnel plot (P value = 0.108) (Figure 5).
Figure 5

Funnel plot of accuracy.

4. Discussion

This study investigates the accuracy of EIT for detecting breast cancer. Most studies have reported several sensitivity and specificity values for the accuracy of anatomic EIT to diagnose breast cancer. The meta-analysis of 12 studies included in our research showed higher diagnostic accuracy for EIT (sensitivity and specificity of 75.88 ± 6.16 and 82.04 ± 5.14, respectively) and a DOR of 14.37 ± 6.14. The results of this meta-analysis revealed overall high sensitivity and specificity. Based on our knowledge, our study is the first systematic review and meta-analysis to investigate the accuracy of the EIT for breast cancer detection. In this field, only one systematic review was performed by Zain et al. in 2015 and their review did not reach the meta-analysis. Their review showed that the range sensitivity of the EIT system to the human breast was between 17% and 94.6%. The range of specificity was between 49% and 97.1%. The accuracy (ACC) of EIT was between 69% and 80.5% [35]. Zhang in the meta-analysis of diagnostic accuracy of magnetic resonance imaging and mammography for breast cancer reported that MRI sensitivity was 0.92 (95% CI, [0.89, 0.94]) and specificity of 0.70 (95% CI, [0.66, 0.73]) [36]. While Medeiros reported MRI pooled sensitivity of 90% (95% CI, 88–92%) and specificity of 75% (95% CI, 70–79%), respectively [37], and Xiang 0.97 (95% CI, 0.95–0.98) and 0.52 (95% CI, 0.46–0.58) as well [38]. So, we can say EIT sensitivity could not be compared with the sensitivity of MRI, but in terms of specificity, it can be considered as a new method that probably can get more attention. However, the stage of tumor and grade may affect these results. Although MRI is an accepted tool in detecting tumors, breast MRI is expensive and can only diagnose breast lesions in certain clinical situations [39]. As such, alternative devices could be favorable. EIT is a radiation-free and noninvasive method that can be employed as an inexpensive, portable, and radiation-free technique for imaging [20, 21]. Hence, it is a great candidate for bedside diagnosis and intraoperative imaging applications [40]. However, EIT was not acceptable for clinical diagnosis until now, and more research is needed for its development. Several studies were conducted using meta-analyses for mammography accuracy stand-alone or in comparison with other imaging techniques. The range of pooled sensitivity of 75–97% and the specificity of 66–96% with 95% CI [36, 38, 41] was reported. The sensitivity of 75% and specificity of 82% of our study showed that EIT can be used as a useful method alongside mammography. In comparison with ultrasound or mammography, EIT is less effective due to its low specificity [27]. It appears that age is an important determinative factor in this regard. As seen in studies [29, 30], EIT may have more efficiency for people who are 30–45 years old. Breast cancers in young women are usually diagnosed at a more advanced stage, requiring more serious and expensive treatments with a reduction in the survivorship rate and the quality of life [42]. This late diagnosis is due to the lack of mammography evaluation for people under 40 years old. In addition, small and nonpalpable probably benign lesions remain unnoticed with either clinical, self-breast examination, or screening surveillance until they have grown sufficiently large enough that the woman herself detects a palpable mass [42, 43]. Therefore, for people under 40, the use of low-cost modalities with acceptable sensitivity and specificity at the initial stages of the disease is required to reduce the rate of unnecessary biopsies [42]. Screening surveillance in the probably benign lesions is more cost-effective than surgical intervention and prevents unnecessary biopsies [44]. However, despite all efforts for follow-ups, approximately 30 to 47% of participants did not complete the screening surveillance [45, 46]. EIT has demonstrated good results in detecting nonpalpable breast cancer as a safe diagnostic tool in younger women [42]. Subgroup analysis can provide more specific information for diagnosis in special groups. In this study, we did not perform subgroup analysis because of limited initial information in selected studies. Respectively, discrimination between patients of different ages was not possible, for example, in patients under 40, the sensitivity of mammography to detect cancer is decreased [47]. Different quality of articles may affect the result of meta-analysis. However, it showed that the use of EIT can be useful because of its sensitivity in detecting cancer in cases that were not detected by mammography and ultrasound [48].

4.1. Limitations of the Study

Unfortunately, it is not possible to analyze the histopathology of patients to determine the efficiency of EIT in detecting the type of different mass in young women Different comparisons with different diagnostic methods may cause heterogeneity in our study Insufficient information and unclear mean age in different groups in articles

5. Conclusion

This paper demonstrated an acceptable accuracy of EIT for breast cancer detection. However, it is concluded that ETL is more considered in the research and clinical application was not very clear yet. This study showed that EIT can be used as a useful method alongside mammography. EIT sensitivity could not be compared with MRI, but in terms of specificity, it can be considered as a new method that probably can get more attention. Furthermore, large-scale studies will be needed to support the evidence.
  39 in total

1.  Diagnostic potential of targeted electrical impedance scanning in classifying suspicious breast lesions.

Authors:  Annika Wersebe; Katja Siegmann; Ute Krainick; Nikos Fersis; Ulrich Vogel; Claus D Claussen; Markus Müller-Schimpfle
Journal:  Invest Radiol       Date:  2002-02       Impact factor: 6.016

2.  Adjunctive diagnostic value of targeted electrical impedance imaging to conventional methods in the evaluation of breast lesions.

Authors:  B K Szabó; A Saracco; B Wilczek; B Boné; P Aspelin
Journal:  Acta Radiol       Date:  2005-12       Impact factor: 1.990

Review 3.  Magnetic resonance imaging of the breast: opportunities to improve breast cancer management.

Authors:  Nola Hylton
Journal:  J Clin Oncol       Date:  2005-03-10       Impact factor: 44.544

4.  Breast cancer screening with tomosynthesis--initial experiences.

Authors:  A Tingberg; D Förnvik; S Mattsson; T Svahn; P Timberg; S Zackrisson
Journal:  Radiat Prot Dosimetry       Date:  2011-07-05       Impact factor: 0.972

5.  Double-phase 99mTc-sestamibi scintimammography and trans-scan in diagnosing breast cancer.

Authors:  M Melloul; A Paz; G Ohana; O Laver; D Michalevich; R Koren; Y Wolloch; R Gal
Journal:  J Nucl Med       Date:  1999-03       Impact factor: 10.057

6.  The negative predictive value of electrical impedance scanning in BI-RADS category IV breast lesions.

Authors:  Michael H Fuchsjaeger; Daniel Flöry; Cäcilia S Reiner; Margarete Rudas; Christopher C Riedl; Thomas H Helbich
Journal:  Invest Radiol       Date:  2005-07       Impact factor: 6.016

7.  Global Cancer Statistics 2020: GLOBOCAN Estimates of Incidence and Mortality Worldwide for 36 Cancers in 185 Countries.

Authors:  Hyuna Sung; Jacques Ferlay; Rebecca L Siegel; Mathieu Laversanne; Isabelle Soerjomataram; Ahmedin Jemal; Freddie Bray
Journal:  CA Cancer J Clin       Date:  2021-02-04       Impact factor: 508.702

8.  Sandblasting improves the performance of electrodes of miniature electrical impedance tomography via double layer capacitance.

Authors:  Zahra Rezanejad Gatabi; Raheleh Mohammadpour; Javad Rezanejad Gatabi; Mehri Mirhoseini; Mohsen Ahmadi; Pezhman Sasanpour
Journal:  Heliyon       Date:  2020-04-01

9.  Evaluation of QUADAS, a tool for the quality assessment of diagnostic accuracy studies.

Authors:  Penny F Whiting; Marie E Weswood; Anne W S Rutjes; Johannes B Reitsma; Patrick N M Bossuyt; Jos Kleijnen
Journal:  BMC Med Res Methodol       Date:  2006-03-06       Impact factor: 4.615

Review 10.  Epidemiology of breast cancer, a paradigm of the "common soil" hypothesis.

Authors:  Licia Iacoviello; Marialaura Bonaccio; Giovanni de Gaetano; Maria Benedetta Donati
Journal:  Semin Cancer Biol       Date:  2020-02-20       Impact factor: 15.707

View more
  1 in total

1.  HSPA5 Could Be a Prognostic Biomarker Correlated with Immune Infiltration in Breast Cancer.

Authors:  Chao Zhang; Qing Liu; Yubo Zhou; Jianfen Hua; Ruijun Su; Jun Ai
Journal:  Dis Markers       Date:  2022-09-20       Impact factor: 3.464

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.