| Literature DB >> 35684755 |
Nicolas Riesen1,2,3, Zane Q Peterkovic1, Bin Guan1,2, Alexandre François1, David G Lancaster1,2, Craig Priest1,2.
Abstract
The rapid development of optofluidic technologies in recent years has seen the need for sensing platforms with ease-of-use, simple sample manipulation, and high performance and sensitivity. Herein, an integrated optofluidic sensor consisting of a pillar array-based open microfluidic chip and caged dye-doped whispering gallery mode microspheres is demonstrated and shown to have potential for simple real-time monitoring of liquids. The open microfluidic chip allows for the wicking of a thin film of liquid across an open surface with subsequent evaporation-driven flow enabling continuous passive flow for sampling. The active dye-doped whispering gallery mode microspheres placed between pillars, avoid the use of cumbersome fibre tapers to couple light to the resonators as is required for passive microspheres. The performance of this integrated sensor is demonstrated using glucose solutions (0.05-0.3 g/mL) and the sensor response is shown to be dynamic and reversible. The sensor achieves a refractive index sensitivity of ~40 nm/RIU, with Q-factors of ~5 × 103 indicating a detection limit of ~3 × 10-3 RIU (~20 mg/mL glucose). Further enhancement of the detection limit is expected by increasing the microsphere Q-factor using high-index materials for the resonators, or alternatively, inducing lasing. The integrated sensors are expected to have significant potential for a host of downstream applications, particularly relating to point-of-care diagnostics.Entities:
Keywords: biological sensing and sensors; integrated optics devices; resonators
Mesh:
Substances:
Year: 2022 PMID: 35684755 PMCID: PMC9185560 DOI: 10.3390/s22114135
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.847
Figure 1(a) Schematic of the open microfluidic chip (not to scale) with caged whispering gallery microspheres incorporated. (b,c) SEM micrographs of the polystyrene microspheres caged within the open microfluidic chip.
Figure 2Experimental setup using inverted microscope for interrogation of whispering gallery microspheres integrated into the open microfluidic chip.
Figure 3(a) Typical normalised WGM-modulated fluorescence spectrum (blue: pure water, orange: 0.25 g/mL glucose in water) of 15 µm microspheres (FSR ~ 4.8 nm) lodged in an open microfluidic chip. (b) Wavelength shift of isolated TE and TM resonances from grey-dashed box in (a). (c) Wavelength shift of TE and TM modes within the microfluidic chip as a function of refractive index. The glucose concentration was converted to refractive index (RI) at 633 nm using [26], with the calibration shown in the inset. In these measurements, a different sphere and chip was used for each measurement. The water in the chip for the baseline measurement was also evaporated before the glucose solution was added.
Figure 4Wavelength shift of TE and TM WGMs within a longer microfluidic chip, as a function of glucose concentration. In this result progressively higher concentrations of glucose are added to the chip reservoir rather than allowing the liquid to first evaporate. In this case, a single chip and microsphere was used for all measurements.
Figure 5In-chip dynamics monitored for the microfluidic chip pre-loaded with water, as a 5 µL droplet of 0.25 g/mL glucose was added to the reservoir. The result showcases the potential of the microfluidic caged-sphere platform for practical real-time monitoring of liquids.
Figure 6Cycling tests between water and (a) 0.15 g/mL or (b) 0.25 g/mL glucose solutions in the chip. The TM resonance wavelength (at ~600 nm) response is seen to be reversible, opening up the opportunity for continuous monitoring of the refractive index of liquids using this platform.
Figure 7Visual demonstration of the behaviour of the microfluidic system (the 7 mm × 12 mm chip) using Ponceau-4R colour dye.