BioFire FilmArray Pneumonia Panel enhances detection of pathogens and antimicrobial resistance in lower respiratory tract specimens.

Kosuke Kosai1, Norihiko Akamatsu2, Kenji Ota2, Fujiko Mitsumoto-Kaseida3, Kei Sakamoto3, Hiroo Hasegawa2, Koichi Izumikawa4, Hiroshi Mukae5, Katsunori Yanagihara2,3.   

Abstract

BACKGROUND: This study investigated the diagnostic utility of the BioFire FilmArray Pneumonia Panel (PN panel), an automated and multiplexed nucleic acid detection system that rapidly detects 26 pathogens (18 bacteria and eight viruses) and seven antimicrobial resistance markers in a single assay.
METHODS: We analyzed the targets in lower respiratory tract specimens using the PN panel and compared the detection results with those of bacterial culture methods and antimicrobial susceptibility testing.
RESULTS: Of the 57 samples analyzed, the PN panel detected 97 targets (84 bacteria, four viruses, and nine antimicrobial resistance markers). Detection of bacteria and antimicrobial resistance was three times greater than that of the bacterial culture (25 bacteria and two resistant isolates) against the targets available in the panel. The overall positive and negative percent agreements between the PN panel and culture methods for bacterial detection were 100.0% and 92.9%, respectively. Multiple pathogens were detected by the PN panel in 24 samples (42.1%), ranging from two pathogens in 11 samples (19.3%) to six pathogens in one sample (1.8%). The PN panel semiquantitatively detected higher copies (≥ 106 copies/mL) of bacterial targets if the bacteria were positive by the culture method. In contrast, the semiquantitative values obtained by the panel varied (104 to 107 ≤ copies/mL) among bacteria that were negative by the culture method.
CONCLUSIONS: The PN panel enhanced the detection of pathogens and antimicrobial resistance markers in lower respiratory tract specimens.
© 2022. The Author(s).

Entities:  

Keywords:  Automated system; Lower respiratory tract infections; Multiplex molecular assay; Rapid detection

Mesh:

Substances:

Year:  2022        PMID: 35659683      PMCID: PMC9166201          DOI: 10.1186/s12941-022-00512-8

Source DB:  PubMed          Journal:  Ann Clin Microbiol Antimicrob        ISSN: 1476-0711            Impact factor:   6.781


Background

Lower respiratory tract infections (LRIs) are a significant public health concern and a leading cause of death from infection worldwide [1]. Therapeutic strategies for LRIs, mainly antimicrobial selection, consist of empirical and definitive therapies [2-4]. Initial empirical therapy covers the major causative pathogens that are predicted, while antimicrobials for definitive therapy are selected against identified pathogens in consideration of antimicrobial susceptibility. Several guidelines and studies have reported the isolation frequency of pathogens and the risk factors for the acquisition of antibiotic-resistant bacteria [3-5]. However, it may be difficult to accurately predict etiological agents at the time of empiric therapy because of differences in regional epidemiology and medical settings. Additionally, culture-based bacterial identification and antimicrobial susceptibility testing (AST) are useful for definitive therapy but are time intensive. The FilmArray system is an automated and multiplexed nucleic acid detection system, and the BioFire FilmArray Pneumonia Panel (PN panel) simultaneously detects 33 targets (26 pathogens and seven antimicrobial resistance markers) in approximately 75 min in a simple assay procedure. In this study, we analyzed these targets in lower respiratory tract specimens using the PN panel and compared the results with those of bacterial culture and AST.

Methods

Study design and samples

Sputum, tracheal aspirates, bronchoalveolar lavage fluid (BAL), and mini-BAL specimens, which were submitted to the clinical laboratory for microbiological evaluation, were collected at the Nagasaki University Hospital between October 2020 and January 2021. The hospital had 874 beds, and the average number of inpatients and outpatients were 669 and 1653 patients per day, respectively, during this period. The quality of sputum and tracheal aspirates was assessed according to the Miller and Jones classification [6]. Samples were randomly selected from specimens P1 (purulent, grade 1—pus amounting to less than one-third of the specimen), P2 (purulent, grade 2—pus amounting to one-third to two-thirds of the specimen), and P3 (purulent, grade 3—pus amounting to more than two-thirds of the specimen). De-duplication was performed if samples were repeatedly collected from a single patient, and samples from unique patients were included in this study. The specimens were stored at − 80 °C until assay using the PN panel.

Culture methods

Sputum and tracheal aspirate samples, which were prepared using Sputazyme (Kyokuto Pharmaceutical Industrial Co., Ltd.), and BAL and mini-BAL samples were streaked using 10 µL loop and cultured on Nissui Separated Plate Sheep Blood Agar/Chocolate Agar EXII (NISSUI PHARMACEUTICAL CO., LTD.), CHROMagar Candida/BTB agar (Prepared media) (KANTO CHEMICAL CO., INC.). Additionally, if the physician specifically requested testing for methicillin-resistant Staphylococcus aureus (MRSA), the samples were cultured on Pourmedia MRSA SELECTIVE AGAR II (Eiken Chemical Co., Ltd.). All bacteria that were considered pathogens, excluding normal flora, were picked up and identified even if they were cultured in rare or few quantities. After pure culture, isolates were identified using matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS), MALDI Biotyper system (Bruker Daltonik, GmbH). Streptococcus pneumoniae was identified using the optochin disc diffusion test. Antimicrobial susceptibility was determined using a BD Phoenix M50 (Becton Dickinson). MRSA and extended-spectrum β-lactamase (ESBL) producers were automatically determined using BD Phoenix M50. Antimicrobial susceptibility of Haemophilus influenzae was determined using minimal inhibitory concentration (MIC) plates customized by Eiken Chemical Co., Ltd.

Analysis using the BioFire FilmArray Pneumonia Panel

The BioFire FilmArray Pneumonia Panel (bioMérieux Japan Ltd.), which has been approved by Food and Drug Administration for use in clinical settings in the United States, simultaneously detects 18 bacterial targets (Acinetobacter calcoaceticus-baumannii complex, Enterobacter cloacae complex, Escherichia coli, H. influenzae, Klebsiella aerogenes, Klebsiella oxytoca, Klebsiella pneumoniae group, Moraxella catarrhalis, Proteus species, Pseudomonas aeruginosa, Serratia marcescens, S. aureus, Streptococcus agalactiae, S. pneumoniae, Streptococcus pyogenes, Chlamydia pneumoniae, Legionella pneumophila, and Mycoplasma pneumoniae), eight viral targets (adenovirus, coronavirus, human metapneumovirus, human rhinovirus/enterovirus, influenza A, influenza B, parainfluenza virus, and respiratory syncytial virus), and seven targets related to antimicrobial resistance (mecA/mecC and staphylococcal cassette chromosome mec [SCCmec] right-extremity junction [MREJ], KPC, NDM, OXA-48-like, VIM, IMP, and CTX-M) on the FilmArray system. The BioFire FilmArray Pneumonia Panel pouch, which is a closed and disposable system, contains all the reagents required for nucleic acid extraction and purification, reverse transcription, and PCR. The stored specimens were analyzed using the PN panel. Briefly, the pouch was hydrated by the hydration injection vial via the pouch hydration port. Approximately 200, 50, or 10 µL of the specimen was mixed with the sample buffer using a vortex mixer. After flashing in a centrifuge to remove foam, the mixture was transferred to the sample injection vial and injected into the pouch via the pouch sample port. The sample preparation process, including the sample volume used for the assay, was off-label. The prepared pouch was inserted into the instrument, and subsequent assay steps (nucleic acid extraction and purification, nested multiplex PCR, and detection of each target in the array) were automatically performed on the system. The system reports the detection results of pathogen and antimicrobial resistance markers and the semiquantification results of the detected bacterial targets excluding three atypical bacteria.

Comparison between the PN panel and bacterial culture-based methods

The detection results of the PN panel were compared with those of bacterial culture methods and AST, and positive and negative percent agreements (PPA and NPA) were calculated as follows: PPA, the number of concordant positive results divided by the number of all positive results by culture-based methods; NPA, the number of concordant negative results divided by the number of all negative results by culture-based methods. The 95% confidence intervals were calculated using R (version 3.5.2) [7].

Results

A total of 57 specimens (35 sputum, nine tracheal aspirate, and 13 BAL or mini-BAL specimens) were included in this study. Of the 44 sputum and tracheal aspirate specimens, one, 13, and 30 were identified as P1, P2, and P3, respectively, according to the Miller and Jones classification. The average patient age was 66.7 years, and 36 patients (63.2%) were male. The frequencies of the detected bacteria and antimicrobial resistance are presented in Table 1. The PN panel detected 84 bacteria and nine antimicrobial resistance markers. The panel detected P. aeruginosa and S. aureus (17 targets each) most frequently, followed by the 10 K. pneumoniae group, six A. calcoaceticus-baumannii complex, six E. cloacae complex, and six H. influenzae. With respect to antimicrobial resistance markers, the PN panel detected five mecA/mecC and MREJ genes, as well as four CTX-M genes.
Table 1

Bacteria and antimicrobial resistance identified using the BioFire FilmArray Pneumonia Panel and culture-based methods

Bacterium and antimicrobial resistanceaBioFire FilmArray Pneumonia PanelbCulture-based methodsc
Bacterium
 Acinetobacter calcoaceticus-baumannii complex60
 Enterobacter cloacae complex61
 Escherichia coli41
 Haemophilus influenzae65
 Klebsiella aerogenes40
 Klebsiella oxytoca11
 Klebsiella pneumoniae group103d
 Moraxella catarrhalis20
 Pseudomonas aeruginosa178
 Serratia marcescens31
 Staphylococcus aureus173
 Streptococcus agalactiae30
 Streptococcus pneumoniae42
 Streptococcus pyogenes10
 Total8425
Antimicrobial resistance
 mecA/mecC and MREJ5e0f
 CTX-M4g2h
 Total92

aTargets available in the BioFire FilmArray Pneumonia Panel

bSample preparation process, including the sample volume used for the assay, was off-label

cCulture methods and antimicrobial susceptibility testing

dTwo and one were identified as K. pneumoniae and K. variicola, respectively

eDetected with S. aureus

fNo methicillin-resistant Staphylococcus aureus was detected using culture-based methods

gTwo were detected with K. pneumoniae group and two were detected with both K. pneumoniae group and E. coli

hOne K. pneumoniae and one E. coli were detected as ESBL producers

Bacteria and antimicrobial resistance identified using the BioFire FilmArray Pneumonia Panel and culture-based methods aTargets available in the BioFire FilmArray Pneumonia Panel bSample preparation process, including the sample volume used for the assay, was off-label cCulture methods and antimicrobial susceptibility testing dTwo and one were identified as K. pneumoniae and K. variicola, respectively eDetected with S. aureus fNo methicillin-resistant Staphylococcus aureus was detected using culture-based methods gTwo were detected with K. pneumoniae group and two were detected with both K. pneumoniae group and E. coli hOne K. pneumoniae and one E. coli were detected as ESBL producers Meanwhile, 25 bacteria were isolated using culture methods, with P. aeruginosa (eight isolates) isolated most frequently, followed by five H. influenzae, three strains in K. pneumoniae group, and three S. aureus. Two ESBL producers were isolated based on their AST results. The PN panel detected four rhinovirus/enterovirus targets, that were not tested using the reference method in this study. Table 2 shows the concordance of the detection results for bacteria between the PN panel and culture methods. The PN panel correctly detected all bacterial targets isolated by the culture methods if bacteria were included in the targets of the panel, indicating that the PPA was 100%. The NPAs ranged from 74.1 to 100% and the overall NPA for bacterial detection was 92.9%. Enterococcus faecalis, Enterococcus faecium, Stenotrophomonas maltophilia, Corynebacterium propinquum, and Candida albicans, which were not targeted by the PN panel, were isolated from one sample each using culture methods. The overall PPA and NPA for the detection of MRSA and ESBL producers were 100.0% and 93.8%, respectively.
Table 2

Comparison of detection results against bacteria and antimicrobial resistance between the BioFire FilmArray Pneumonia Panela and culture-based methods

Bacterium and antimicrobial resistancePN+, CM+PN+, CM−PN−, CM+PN−, CM−PPA (95% CI)NPA (95% CI)
Bacterium
 A. calcoaceticus-baumannii complex0605189.5 (78.5–96.0)
 E. cloacae complex15051100.0 (2.5–100.0)91.1 (80.4–97.0)
 E. coli13053100.0 (2.5–100.0)94.6 (85.1–98.9)
 H. influenzae51051100.0 (47.8–100.0)98.1 (89.7–100.0)
 K. aerogenes0405393.0 (83.0–98.1)
 K. oxytoca10056100.0 (2.5–100.0)100.0 (93.6–100.0)
 K. pneumoniae group37047100.0 (29.2–100.0)87.0 (75.1–94.6)
 M. catarrhalis0205596.5 (87.9–99.6)
 Proteus species00057100.0 (93.7–100.0)
 P. aeruginosa89040100.0 (63.1–100.0)81.6 (68.0–91.2)
 S. marcescens12054100.0 (2.5–100.0)96.4 (87.7–99.6)
 S. aureus314040100.0 (29.2–100.0)74.1 (60.3–85.0)
 S. agalactiae0305494.7 (85.4–98.9)
 S. pneumoniae22053100.0 (15.8–100.0)96.4 (87.5–99.6)
 S. pyogenes0105698.2 (90.6–100.0)
 Total25590771100.0 (86.3–100.0)92.9 (90.9–94.5)
Antimicrobial resistance
 MRSAb0505291.2 (80.7–97.1)
 ESBL producerc22053100.0 (15.8–100.0)96.4 (87.5–99.6)
 Total270105100.0 (15.8–100.0)93.8 (87.5–97.5)

PN BioFire FilmArray Pneumonia Panel, CM culture-based methods (culture methods and antimicrobial susceptibility testing), MRSA methicillin-resistant Staphylococcus aureus, ESBL extended-spectrum β-lactamase, PPA positive percent agreement, NPA negative percent agreement, CI confidence interval

aSample preparation process, including the sample volume used for the assay, was off-label

bMRSA was detected as mecA/mecC and MREJ genes using the BioFire FilmArray Pneumonia Panel

cESBL producers were detected as CTX-M genes using the BioFire FilmArray Pneumonia Panel

Comparison of detection results against bacteria and antimicrobial resistance between the BioFire FilmArray Pneumonia Panela and culture-based methods PN BioFire FilmArray Pneumonia Panel, CM culture-based methods (culture methods and antimicrobial susceptibility testing), MRSA methicillin-resistant Staphylococcus aureus, ESBL extended-spectrum β-lactamase, PPA positive percent agreement, NPA negative percent agreement, CI confidence interval aSample preparation process, including the sample volume used for the assay, was off-label bMRSA was detected as mecA/mecC and MREJ genes using the BioFire FilmArray Pneumonia Panel cESBL producers were detected as CTX-M genes using the BioFire FilmArray Pneumonia Panel Table 3 presents the number of samples in which multiple pathogens were detected by the PN panel and culture methods. Of the 57 samples included in the study, multiple pathogens were detected in 24 samples (42.1%) using the PN panel, ranging from two pathogens in 11 samples (19.3%) to six pathogens in one sample (1.8%). Conversely, the culture method detected two pathogens in four samples (7.0%).
Table 3

Number of samples in which multiple pathogens were detected using the BioFire FilmArray Pneumonia Panel and culture methods

BioFire FilmArray Pneumonia PanelaCulture methodsTotal
012
0161017
1106016
227211
31416
40314
51102
61001
Total3122457

Detected pathogens were included, even though they were not available in the BioFire FilmArray Pneumonia Panel

aSample preparation process, including the sample volume used for the assay, was off-label

Number of samples in which multiple pathogens were detected using the BioFire FilmArray Pneumonia Panel and culture methods Detected pathogens were included, even though they were not available in the BioFire FilmArray Pneumonia Panel aSample preparation process, including the sample volume used for the assay, was off-label Table 4 summarizes the semiquantitative values of the bacteria measured using the PN panel. The PN panel detected bacterial targets of 106 copies/mL or greater if the bacteria were positive by the culture method. Conversely, semiquantitative values determined by the panel ranged from ≥ 107 copies/mL (13 bacteria) to 104 copies/mL (24 bacteria) in culture-negative samples.
Table 4

Semiquantitative values of bacteria measured by the BioFire FilmArray Pneumonia Panel in culture-positive and culture-negative samples

BioFire FilmArray Pneumonia Panel (copies/mL)aCulture methodsTotal
PositiveNegative
≥ 107231336
10621012
10501212
10402424
Total255984

aSample preparation process, including the sample volume used for the assay, was off-label

Semiquantitative values of bacteria measured by the BioFire FilmArray Pneumonia Panel in culture-positive and culture-negative samples aSample preparation process, including the sample volume used for the assay, was off-label

Discussion

Our study demonstrated that the PN panel rapidly and effectively detected a variety of pathogens in the lower respiratory tract specimens. The PN panel detected 97 targets (84 bacteria, four viruses, and nine antimicrobial resistance markers), and the detection of bacteria and antimicrobial resistance was three times greater than that of the culture-based methods (25 bacteria and two resistant isolates) (Table 1). Several previous studies have shown similar results in that the panel detected more bacterial targets than the culture method [8, 9]. A recent study reported that the PN panel identified nearly twice as many total bacterial targets as standard-of-care culture in BAL specimens [8]. Another study reported that the PN panel detected one or more bacterial targets in an additional 20% of patients compared to culture methods [10]. In addition, our results also showed that the PN panel correctly detected all bacteria that were detected by the culture methods (PPA, 100%) if they were included in the targets of the panel (Table 2), and that the overall NPA for bacterial detection was 92.9%. Similarly, previous studies reported high performance of the PN panel for bacterial detection with PPAs of 90.0–98.4% and NPAs of 96.8–98.1% [8, 9, 11]. In this study, only four bacterial species (E. faecalis, E. faecium, S. maltophilia, C. propinquum) and one C. albicans, which were unavailable in the PN panel, were isolated using the culture method. A recent study from Taiwan reported that the PN panel may cover approximately 70–90% of the most prevalent pathogens causing moderate to severe community-acquired pneumonia in adults, as well as 70–80% of those causing healthcare-associated pneumonia [9]. The current diagnostic strategies require multiple tests to detect possible pathogens. Bacterial culture is the gold standard for bacterial detection, but it is time consuming. The PN panel, which can simultaneously detect major pathogens in a short time, can be useful for the diagnosis of LRIs. Although atypical bacteria were not detected by the PN panel in this study, recent studies have reported that the PN panel detected bacteria such as L. pneumophila and M. pneumoniae in a few samples [9, 12]. We found four samples in which rhinovirus/enterovirus was simultaneously detected with the bacteria. Previous studies have reported that co-infection with viruses and bacteria contributes to increased disease severity and mortality [13, 14]. Although we could not assess the relationship between co-infection and disease severity, comprehensive pathogen detection by the PN panel might contribute to elucidating the epidemiology and pathophysiology of co-infection. The PN panel can semiquantify the bacterial targets. In this study, of the 59 analytes with negative results by the culture methods, bacterial targets of 104–5 copies/mL were detected in 36 analytes (61.0%) (Table 4). Quantitative values below the defined thresholds, such as 103 CFU/mL for the protected specimen brush and 104 CFU/mL for BAL, might be used as indicators for discontinuing antibiotics against ventilator-associated pneumonia (VAP) [3]. Therefore, the semiquantitative values determined by the PN panel may be useful for the management of VAP. Meanwhile, we should consider the possibility of bacterial colonization, regardless of the detection method used. However, because P. aeruginosa and MRSA colonization are reported as the risk factors for subsequent infection [15, 16], detecting colonized bacteria would be helpful for prompt initiation of treatment when infection occurs. Detection using culture methods depends on pathogen viability. Therefore, it is possible that the PN panel detected nonviable bacteria in this study, as one of the causes of discrepant results in bacterial detection between the PN panel and culture methods. There are other possibilities that the PN panel detected bacteria that were low-abundance or not cultured because of fastidious growth characteristics [8]. Therefore, the results of molecular assays should be carefully interpreted when used for patient management, in consideration of the differences in characteristics between molecular assays and culture methods. Additionally, even if antimicrobial resistance markers are not detected, antimicrobial susceptibility can be decreased through other resistance mechanisms. Therefore, culture methods are necessary to test antimicrobial susceptibility. Further studies are required to establish practical strategies for complementary use of PN panel and traditional culture-based methods. This study had some limitations. First, we did not collect all the samples submitted to our laboratory during the study period. Therefore, the data did not represent the epidemiology in our hospital. Second, we did not examine viral targets detected by the panel using a reference method and performed culture for MRSA using selective media only when requested by physicians. Third, the sample preparation process, including the sample volume, for the assay by the PN panel was off-label.

Conclusions

The PN panel effectively detected a variety of pathogens in lower respiratory tract specimens. Further studies are needed to clarify its effects on patient management and cost-effectiveness.
  14 in total

1.  A STUDY OF TECHNIQUES FOR THE EXAMINATION OF SPUTUM IN A FIELD SURVEY OF CHRONIC BRONCHITIS.

Authors:  D L MILLER
Journal:  Am Rev Respir Dis       Date:  1963-10

2.  Evaluation of the BioFire FilmArray Pneumonia Panel for Detection of Viral and Bacterial Pathogens in Lower Respiratory Tract Specimens in the Setting of a Tertiary Care Academic Medical Center.

Authors:  Daniel M Webber; Meghan A Wallace; C A Burnham; Neil W Anderson
Journal:  J Clin Microbiol       Date:  2020-06-24       Impact factor: 5.948

3.  Increased incidence of co-infection in critically ill patients with influenza.

Authors:  Ignacio Martin-Loeches; Marcus J Schultz; Jean-Louis Vincent; Francisco Alvarez-Lerma; Lieuwe D Bos; Jordi Solé-Violán; Antoni Torres; Alejandro Rodriguez
Journal:  Intensive Care Med       Date:  2016-10-05       Impact factor: 17.440

4.  Management of Adults With Hospital-acquired and Ventilator-associated Pneumonia: 2016 Clinical Practice Guidelines by the Infectious Diseases Society of America and the American Thoracic Society.

Authors:  Andre C Kalil; Mark L Metersky; Michael Klompas; John Muscedere; Daniel A Sweeney; Lucy B Palmer; Lena M Napolitano; Naomi P O'Grady; John G Bartlett; Jordi Carratalà; Ali A El Solh; Santiago Ewig; Paul D Fey; Thomas M File; Marcos I Restrepo; Jason A Roberts; Grant W Waterer; Peggy Cruse; Shandra L Knight; Jan L Brozek
Journal:  Clin Infect Dis       Date:  2016-07-14       Impact factor: 9.079

5.  Estimates of the global, regional, and national morbidity, mortality, and aetiologies of lower respiratory tract infections in 195 countries: a systematic analysis for the Global Burden of Disease Study 2015.

Authors: 
Journal:  Lancet Infect Dis       Date:  2017-08-23       Impact factor: 25.071

6.  Accuracy of American Thoracic Society/Infectious Diseases Society of America criteria in predicting infection or colonization with multidrug-resistant bacteria at intensive-care unit admission.

Authors:  S Nseir; G Grailles; A Soury-Lavergne; F Minacori; I Alves; A Durocher
Journal:  Clin Microbiol Infect       Date:  2009-08-20       Impact factor: 8.067

7.  Viral-bacterial coinfection affects the presentation and alters the prognosis of severe community-acquired pneumonia.

Authors:  Guillaume Voiriot; Benoit Visseaux; Johana Cohen; Liem Binh Luong Nguyen; Mathilde Neuville; Caroline Morbieu; Charles Burdet; Aguila Radjou; François-Xavier Lescure; Roland Smonig; Laurence Armand-Lefèvre; Bruno Mourvillier; Yazdan Yazdanpanah; Jean-Francois Soubirou; Stephane Ruckly; Nadhira Houhou-Fidouh; Jean-François Timsit
Journal:  Crit Care       Date:  2016-10-25       Impact factor: 9.097

8.  P. aeruginosa colonization at ICU admission as a risk factor for developing P. aeruginosa ICU pneumonia.

Authors:  Fleur P Paling; Martin Wolkewitz; Pieter Depuydt; Liesbet de Bus; Frangiscos Sifakis; Marc J M Bonten; Jan A J W Kluytmans
Journal:  Antimicrob Resist Infect Control       Date:  2017-04-20       Impact factor: 4.887

9.  Practical Comparison of the BioFire FilmArray Pneumonia Panel to Routine Diagnostic Methods and Potential Impact on Antimicrobial Stewardship in Adult Hospitalized Patients with Lower Respiratory Tract Infections.

Authors:  Blake W Buchan; Sam Windham; Joan-Miquel Balada-Llasat; Amy Leber; Amanda Harrington; Ryan Relich; Caitlin Murphy; Jennifer Dien Bard; Samia Naccache; Shira Ronen; Amanda Hopp; Derya Mahmutoglu; Matthew L Faron; Nathan A Ledeboer; Amanda Carroll; Hannah Stone; Oluseun Akerele; Kathy Everhart; Andrew Bonwit; Christina Kwong; Rebecca Buckner; Del Warren; Randal Fowler; Sukantha Chandrasekaran; Holly Huse; Shelley Campeau; Romney Humphries; Corrin Graue; Angela Huang
Journal:  J Clin Microbiol       Date:  2020-06-24       Impact factor: 5.948

10.  Multicenter Evaluation of the BioFire FilmArray Pneumonia/Pneumonia Plus Panel for Detection and Quantification of Agents of Lower Respiratory Tract Infection.

Authors:  Caitlin N Murphy; Randal Fowler; Joan Miquel Balada-Llasat; Amanda Carroll; Hanna Stone; Oluseun Akerele; Blake Buchan; Sam Windham; Amanda Hopp; Shira Ronen; Ryan F Relich; Rebecca Buckner; Del A Warren; Romney Humphries; Shelly Campeau; Holly Huse; Suki Chandrasekaran; Amy Leber; Kathy Everhart; Amanda Harrington; Christina Kwong; Andrew Bonwit; Jennifer Dien Bard; Samia Naccache; Cynthia Zimmerman; Barbara Jones; Cory Rindlisbacher; Maggie Buccambuso; Angela Clark; Margarita Rogatcheva; Corrin Graue; Kevin M Bourzac
Journal:  J Clin Microbiol       Date:  2020-06-24       Impact factor: 5.948

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.