Assumpta Nantume1, Noah Kiwanuka2, Asad Muyinda3, Teresa Cauvel1, Sona Shah1.
Abstract
Objective: The purpose of this study was to evaluate the accuracy and reliability of neoGuard in comparison to a conventional bedside monitor on patients in a low-resource clinical setting. Design: This was a single-arm methods comparison study involving the use of a wearable vital signs monitor (neoGuardTM) versus a conventional bedside monitor (Edan iM8). Setting: The study was conducted at Jinja Regional Referral Hospital, a tertiary care hospital situated in Eastern Uganda. Participants: Thirty patients (10 male, 20 female) were enrolled from the adult recovery ward at JRRH. Participants were eligible for the study if they were at least 18 years of age, had 2 sets of normal vital sign measurements obtained 1 h apart, and were able and willing to provide informed consent. Main Outcome and Measures: The primary outcome measures were (i) bias (mean deviation) and (ii) limits of agreement [95% CI]. Bland-Altman plots were generated to illustrate the level of agreement between the neoGuardTM technology and the Edan iM8 monitor.Entities:
Keywords: digital health; newborn health; technology; vital signs; wearable; wireless
Year: 2022 PMID: 35656284 PMCID: PMC9152187 DOI: 10.1177/20552076221102262
Source DB: PubMed Journal: Digit Health ISSN: 2055-2076
Figure 1.The neoGuard vital signs monitoring system.
Summary statistics for Bland-Altman plots for measurement of SpO2, PR, RR and temperature.
| SpO2 | PR | RR | Temp | |
|---|---|---|---|---|
| Expected values | ||||
| Bias | 0.0 | 0.0 | 0.0 | 0.0 |
| Total SD | 2.0 | 3.0 | 3.0 | 0.25 |
| Upper LOA | 4.0 | 6.0 | 7.0 | 0.50 |
| Lower LOA | −4.0 | −6.0 | −7.0 | −0.50 |
| Actual values | ||||
| Bias | 1.73 | 0.56 | 0.90 | 0.06 |
| Total SD | 2.44 | 5.18 | 6.34 | 0.70 |
| Upper LOA [95% CI] | 6.51 [5.61, 7.85] | 10.71 [9.98, 11.57] | 13.33 [11.42, 16.04] | 1.44 [1.17, 1.84] |
| Lower LOA [95% CI] | −3.06 [−4.40, −2.15] | −9.58 [−10.45, −8.86] | −11.53 [−14.24, −9.62] | −1.32 [−1.72, −1.06] |
Figure 2.Bland-Altman plot for comparison of neoGuard (SpO2n) and Edan iM8 (SpO2e) oxygen saturation readings.
Figure 5.Bland-Altman plot for comparison of neoGuard (TEMPn) and Edan iM8 (TEMPe) temperature readings.
Figure 3.Bland-Altman plot for comparison of neoGuard (PRn) and Edan iM8 (PRe) pulse rate readings.
Figure 4.Bland-Altman plot for comparison of neoGuard (RRn) and Edan iM8 (RRe) respiratory rate readings.