| Literature DB >> 35641642 |
Shaun T Lancaster1, Thomas Prohaska1,2, Johanna Irrgeher3,4.
Abstract
In inductively coupled plasma mass spectrometry, the most abundant Ca isotope (40Ca) suffers from isobaric interference with argon, hindering the potential for low detection limits of Ca. A powerful approach is to remove the interference by using a reaction gas in a reaction cell. Ammonia (NH3) has proven to be an effective reaction gas by process of a charge transfer reaction. However, NH3 is highly corrosive and toxic and cannot remove isobaric 40 K. Therefore, this work proposes the use of nitrous oxide (N2O) to mass shift the target analyte 40Ca to 40Ca16O+ as a non-corrosive and non-toxic alternative. Instrument performance testing demonstrated that N2O was capable of reaching equivalent detection limits (0.015 ng g-1) and background equivalence concentrations (0.041 ng g-1) to that of NH3 and limited by the blank only. Further investigation of matrix interferences with synthetic standards highlighted that the N2O approach supports the separation of potassium (K) and magnesium (Mg)-based interferences at tested concentrations of more than 600 times and almost 800 times higher than Ca respectively, whereas NH3 was found to only support the removal of Mg. This work highlights a clear advantage of N2O for low-level Ca determinations with high matrix loads, as well as compatibility with other instrumentation sensitive to corrosion that supports reaction cell technology.Entities:
Keywords: Calcium; ICP-MS/MS; Mass shift; N2O
Mesh:
Substances:
Year: 2022 PMID: 35641642 PMCID: PMC9482902 DOI: 10.1007/s00216-022-04146-9
Source DB: PubMed Journal: Anal Bioanal Chem ISSN: 1618-2642 Impact factor: 4.478
List of interferences for Ca on m/z 40. Mass resolution was calculated from IUPAC Periodic Table of the Elements and Isotopes [1]. Negative resolutions indicate that the atomic mass of the interfering species is greater than that of.40Ca
| Interference on 40Ca+ | Abundance (%) | Atomic mass (u) | Required mass resolution |
|---|---|---|---|
| 40Ca+ | 96.9 | 39.943 | – |
| 40Ar+ | 99.6 | 39.962 | 192,058 |
| 40 K+ | 0.0117 | 39.964 | − 28,394 |
| 24Mg16O+ | 78.8 | 39.980 | − 2302 |
| 80Se++ | 49.6 | 39.958 | 9229 |
| 80Kr++ | 2.29 | 39.958 | 9079 |
| 39K1H+ | 93.2 | 39.972 | − 4471 |
| 23Na17O+ | 0.038 | 39.989 | − 1520 |
Instrument and plasma conditions for measurements made using ICP-MS/MS
| Parameter | Q3 | MS/MS | NH3 DRC | N2O DRC |
|---|---|---|---|---|
| Measurement mode | Standard | Standard | DRC on mass | DRC mass shift |
| Cell gas | None | None | NH3 | N2O |
| Cell gas flow rate | None | None | 0.7 mL min−1 | 0.4 mL min−1 |
| RPa | 0 | 0 | 0 | 0 |
| RPq | 0.25 | 0.25 | 0.45 | 0.45 |
| Sample introduction | Peristaltic pump | Peristaltic pump | Peristaltic pump | Peristaltic pump |
| Nebulizer | PFA MicroFlow | PFA MicroFlow | PFA MicroFlow | PFA MicroFlow |
| Spray chamber | Peltier cooled SilQ cyclonic spray chamber | Peltier cooled SilQ cyclonic spray chamber | Peltier cooled SilQ cyclonic spray chamber | Peltier cooled SilQ cyclonic spray chamber |
| Spray chamber temperature | 5 °C | 5 °C | 5 °C | 5 °C |
| Interface cones | Nickel | Nickel | Nickel | Nickel |
| RF power | 1600 W | 1600 W | 1600 W | 1600 W |
| Ar nebulizer gas flow | 0.98 L min−1 | 0.98 L min−1 | 0.98 L min−1 | 0.98 L min−1 |
| Ar auxiliary gas flow | 1.2 L min−1 | 1.2 L min−1 | 1.2 L min−1 | 1.2 L min−1 |
| Ar plasma gas flow | 16 L min−1 | 16 L min−1 | 16 L min−1 | 16 L min−1 |
| QID fixed voltage | − 12 V | − 12 V | − 12 V | − 12 V |
| Hyperskimmer park voltage | 5 V | 5 V | 5 V | 5 V |
| OmniRing park voltage | − 185 V | − 185 V | − 185 V | − 185 V |
| Inner target lens voltage | 2 V | 2 V | 2 V | 2 V |
| Outer target lens voltage | − 7 V | − 7 V | − 7 V | − 7 V |
| Deflector exit voltage | − 8 V | − 8 V | − 8 V | − 8 V |
| Differential aperture voltage | − 3.5 V | − 3.5 V | − 3.5 V | − 3.5 V |
| Q1 AC rod offset | − 7 V | − 6 V | − 5 V | − 7.5 V |
| Q1 rod offset | − 11 V | − 2 V | − 0 V | 0 V |
| Cell rod offset | − 33 V | − 33 V | − 5 V | − 2 V |
| Axial field voltage | 0 V | 0 V | 150 V | 250 V |
| Cell entrance voltage | − 10 V | − 5 V | − 6.5 V | − 7.5 V |
| Cell exit voltage | − 2 V | − 2 V | − 7 V | − 5 V |
| Q3 AC rod offset | − 2.5 V | − 2.5 V | − 8.5 V | − 8 V |
| Q3 rod offset | − 2 V | − 2 V | − 13 V | − 10 V |
| Dwell time | 50 ms | 50 ms | 50 ms | 50 ms |
| Scans | 6 | 6 | 6 | 6 |
| Replicates | 6 | 6 | 6 | 6 |
ICP-MS/MS measurement conditions for the tested internal standards using N2O and NH3 DRC
| Internal standard | Reaction gas | Q1 | Q3 |
|---|---|---|---|
| Sc | N2O | 45 | 61 |
| Y | N2O | 89 | 105 |
| In | N2O | 115 | 115 |
| Sc | NH3 | 45 | 45 |
| Y | NH3 | 89 | 89 |
| In | NH3 | 115 | 115 |
Fig. 1Optimization of sensitivity for 10 ng g−1 Ca (solid line) using A N2O and B NH3 cell gas flow rate for the removal of 40Ar interference on.40Ca. The background signal, measured using a blank solution, is also indicated (dotted line). Variation in the signal to background ratio (SBR) for NH3 DRC is shown in C. Error bars present represent one standard deviation of six replicates
Fig. 2Variation of 45Sc (solid line), 89Y (dashed line), and.115In (dotted line) internal standard signal with A N2O and B NH3 cell gas flow rate. Error bars present represent one standard deviation of six replicates
Comparison of calibration parameters for the measurement of 44Ca with no reaction gas in Q3 and MS/MS mode, and.40Ca in MS/MS mode with N2O and NH3 reaction gas
| Parameter | 44Ca | 44Ca | 40Ca | 40Ca | 40Ca | 40Ca |
|---|---|---|---|---|---|---|
| Internal standard | Indium | Indium | Indium | Scandium | Yttrium | Indium |
| Slope (cps/ng g−1) | 8260 | 3650 | 226,000 | 129,000 | 129,000 | 128,000 |
| Intercept (cps) | 150,000 | 56,500 | 82,500 | 52,600 | 52,700 | 53,300 |
| Correlation, | 0.99983 | 0.99968 | 0.99832 | 0.99941 | 0.99924 | 0.99930 |
| LOD (ng g−1) | 0.14 | 0.27 | 0.015 | 0.015 | 0.017 | 0.017 |
| LOQ (ng g−1) | 0.43 | 0.81 | 0.049 | 0.049 | 0.057 | 0.056 |
| BEC (ng g−1) | 18 | 15 | 0.37 | 0.41 | 0.41 | 0.42 |
| Stability test accuracy (%)* | – | – | 101 | 101 | 99.3 | 97.0 |
| Stability test RSD (%)* | – | – | 2.39 | 1.57 | 2.20 | 2.92 |
*Based on 20 replicate measurements of a 0.5-ng g.−1 Ca standard
Total least squares regression results for the comparison of determined background Ca concentrations (w = 0–7.5 ng g−1) using isotopes 44Ca and 40Ca in high-concentration Mg (w = 0–4.0 µg g−1) and K (w = 0–4.4 µg g.−1) single element standards
| DRC mode | Slope | Standard error of the slope | Significant? | Lower confidence interval | Upper confidence interval | |
|---|---|---|---|---|---|---|
| Potassium interference | ||||||
| N2O at 0.4 mL min−1 (Sc Int. Std.) | 0.9917 | 0.0062 | > 0.150 | No | 0.96518 | 1.01825 |
| N2O at 0.4 mL min−1 (Y Int. Std.) | 0.9916 | 0.0061 | > 0.150 | No | 0.96532 | 1.01797 |
| N2O at 0.4 mL min−1 (In Int. Std.) | 0.9915 | 0.0061 | > 0.150 | No | 0.96526 | 1.01777 |
| NH3 at 0.7 mL min−1 (In Int. Std.) | 1.1335 | 0.0022 | < 0.001 | Yes | 1.12386 | 1.14314 |
| Magnesium interference | ||||||
| N2O at 0.4 mL min−1 (Sc Int. Std.) | 0.9982 | 0.0107 | > 0.150 | No | 0.95212 | 1.04429 |
| N2O at 0.4 mL min−1 (Y Int. Std.) | 0.9986 | 0.0111 | > 0.150 | No | 0.95084 | 1.04626 |
| N2O at 0.4 mL min−1 (In Int. Std.) | 0.9987 | 0.0104 | > 0.150 | No | 0.95399 | 1.04336 |
| NH3 at 0.7 mL min−1 (In Int. Std.) | 1.0062 | 0.0128 | > 0.150 | No | 0.95135 | 1.06103 |
Fig. 3Contribution of 40 K interference on 40Ca measurements of single element K standards (w = 0–4.4 µg g.−1) using NH3 DRC. Error bars present represent the combined standard deviation of the difference between the observed w(Ca) 40 and w(Ca) 44 (calculated by the law of propagation of uncertainties)
Fig. 4Variation of the relative signal to background ratio (SBR) with NH3 flow rate (normalized to the SBR observed at 0.7 mL min−1 NH3) for 10 ng g−1 40Ca signal with a 25 ng g−1.39 K background (solid line) and blank background on m/z 40 (dotted line). Error bars represent one combined standard deviation of six replicates (calculated by the law of propagation of uncertainties)
Measured mass fractions and recoveries of river water CRMs (diluted to 1 ng mL.−1) using N2O and NH3 DRC
| CRM | Certified mass fraction (µg g−1) | N2O DRC (Sc internal standard) | N2O DRC (Y internal standard) | N2O DRC (In internal standard) | NH3 DRC (In internal standard) | ||||
|---|---|---|---|---|---|---|---|---|---|
| Measured (µg g−1) | Recovery (%) | Measured (µg g−1) | Recovery (%) | Measured (µg g−1) | Recovery (%) | Measured (µg g−1) | Recovery (%) | ||
| SLRS-3 | 6.0 ± 0.4 | 6.09 ± 0.09 | 101 | 5.95 ± 0.10 | 99.2 | 5.72 ± 0.09 | 95.4 | 5.87 ± 0.09 | 97.9 |
| SLRS-5 | 10.5 ± 0.4 | 10.8 ± 0.1 | 103 | 10.8 ± 0.1 | 103 | 10.2 ± 0.1 | 96.9 | 10.3 ± 0.1 | 98.5 |