| Literature DB >> 35632806 |
Florence Abravanel1,2, Delphine Parraud1, Sabine Chapuy-Regaud1,2, Marcel Miedouge1, Estelle Bonnin1, Margaux Larrieu1, Alexandre Aversenq1, Sébastien Lhomme1,2, Jacques Izopet1,2.
Abstract
To evaluate the diagnostic performance of the Liaison® Murex anti-HEV IgM and IgG assays running on the Liaison® instrument and compare the results with those obtained with Wantai HEV assays. We tested samples collected in immunocompetent and immunocompromised patients during the acute (HEV RNA positive, anti-HEV IgM positive) and the post-viremic phase (HEV RNA negative, anti-HEV IgM positive) of infections. The specificity was assessed by testing HEV RNA negative/anti-HEV IgG-IgM negative samples. The clinical sensitivity of the Liaison® IgM assay was 100% for acute-phase samples (56/56) and 57.4% (27/47) for post-viremic samples from immunocompetent patients. It was 93.8% (30/32) for acute-phase (viremic) samples and 71%% (22/31) for post-viremic samples from immunocompromised patients. The clinical sensitivity of the Liaison® IgG assay was 100% for viremic samples (56/56) and 94.6% (43/47) for post-viremic samples from immunocompetent patients. It was 84.3% (27/32) for viremic samples and 93.5% (29/31) for post-viremic samples from immunocompromised patients. Specificity was very high (>99%) in both populations. We checked the limit of detection stated for the Liaison® IgG assay (0.3 U/mL). The clinical performance of the Liaison® ANTI-HEV assays was good. These rapid, automated assays for detecting anti-HEV antibodies will greatly enhance the arsenal for diagnosing HEV infections.Entities:
Keywords: hepatitis E virus; immunocompetent patients; immunocompromised patients; serology
Mesh:
Substances:
Year: 2022 PMID: 35632806 PMCID: PMC9145211 DOI: 10.3390/v14051065
Source DB: PubMed Journal: Viruses ISSN: 1999-4915 Impact factor: 5.818
Figure 1Clinical sensitivities of the Liaison® Murex assays in immunocompetent and immunocompromised patients.
Evaluation of the limit of detection of the Liaison® Murex IgG assay: Results of the serial 2-fold dilutions of an HEV IgG-positive sample.
| Expected Value (IU/mL) | Murex IgG HEV Quantification(IU/mL) | Murex IgG HEV Interpretation |
|---|---|---|
| 2.63 | 3.16 | positive |
| 1.32 | 1.61 | positive |
| 0.66 | 0.7 | positive |
| 0.33 | 0.41 | positive |
| 0.16 | 0.23 | negative |
| 0.08 | 0.11 | negative |
Figure 2Correlation between the expected IgG HEV titers and the Liaison® IgG HEV results. The dashed line represents the limit of detection of the Liaison® Murex IgG HEV assay.