| Literature DB >> 35627347 |
Petr P S Otahal1, Eliska Fialova1,2, Josef Vosahlik1, Hannah Wiedner3, Claudia Grossi4, Arturo Vargas4, Nathalie Michielsen5, Tuukka Turtiainen6, Aurelian Luca7, Katarzyna Wołoszczuk8, Thomas R Beck9.
Abstract
An international comparison of continuous monitors measuring radon activity concentration was performed to validate the traceability of the European radon calibration facilities. It was carried out by comparing the secondary standards used by these previous facilities, ranging from 100 Bq·m-3 to 300 Bq·m-3. Secondary standards were individually compared to a secondary reference device previously calibrated in a reference radon atmosphere traceable to a primary standard. The intercomparison was organized by the National Institute for Nuclear, Chemical, and Biological Protection (SUJCHBO) in the period from October 2019 to April 2020 within the European Metrology Program for Innovation and Research (EMPIR), JRP-Contract 16ENV10 MetroRADON. Eight European laboratories participated in this study. The results of the experiment are presented and discussed.Entities:
Keywords: calibration; interlaboratory comparison; radon; radon activity concentration; traceability
Mesh:
Substances:
Year: 2022 PMID: 35627347 PMCID: PMC9141648 DOI: 10.3390/ijerph19105810
Source DB: PubMed Journal: Int J Environ Res Public Health ISSN: 1660-4601 Impact factor: 4.614
Radon calibration facilities participating in the interlaboratory comparison (sorted alphabetically by acronym) [7].
| Acronym | Institute | Country | Measuring Device |
|---|---|---|---|
| BEV-PTP | Physikalisch-technischer Prüfdienst, Bundesamt für Eich- und Vermessungswesen, Wien | Austria | AG * PQ2000 Pro |
| BfS | Federal Office for Radiation Protection, Berlin | Germany | AG DF2000 |
| CLOR | Central Laboratory for Radiological Protection, Warsaw | Poland | AG DF2000 |
| IFIN-HH | Horia Hulubei National Institute for R&D in Physics and Nuclear Engineering, Bucharest | Romania | RadonScout |
| IRSN | Institut de Radioprotection et de Sûreté Nucléaire, Fontenay-aux-Roses | France | AG PQ2000 |
| STUK | Radiation and Nuclear Safety Authority, Helsinki | Finland | AG PQ2000 |
| SUJCHBO | National Institute for Nuclear, Chemical and Biological Protection, Milin | Czech Republic | AG DF2000 |
| INTE-UPC | Laboratory of 222Rn studies of the Institut de Tecniques Energetiques of the Universitat Politecnica de Catalunya, Barcelona | Spain | AG PQ2000 Pro |
* AG: AlphaGuard.
Technical parameters of the intercomparison.
| Parameter | Value |
|---|---|
| Integration time | 60 min |
| Exposure time | at least 24 h |
| Temperature | 22.3 ± 1.1 °C |
| Relative humidity | 51.2 ± 3.8% |
| Air pressure | 955.8 ± 8.6 hPa |
Figure 1Results of participants’ devices in a zero-radon-activity concentration atmosphere.
Results for a level of radon activity concentration of 200 Bq·m−3.
| Participants’ Code | av,Rn | s(av,Rn) | z-Score | D |
|---|---|---|---|---|
| 1 | 201 | 9 | 0.1 | 0.5 |
| 2 | 203 | 6 | 0.3 | 1.5 |
| 3 | 196 | 6 | −0.4 | −2.0 |
| 4 | 208 | 18 | 0.8 | 4.0 |
| 5 | 194 | 9 | −0.6 | −3.0 |
| 6 | 201 | 15 | 0.1 | 0.5 |
| 7 | 202 | 8 | 0.2 | 1.0 |
| 8 | 202 | 4 | 0.2 | 1.0 |
av,Rn: the average participant´s value of radon activity concentration in the exposure period (each participant sent their results to SUJCHBO after they received the monitor back and downloaded the data). s(av,Rn): participant´s measurement uncertainties (each participant sent his results to SUJCHBO after they received the monitor back and downloaded the data)—it is an expanded uncertainty—the product of the standard measurement uncertainty and the coverage factor k = 2.
Figure 2Participants’ devices results for the level of 200 Bq·m−3.
Results for a level of radon activity concentration of 300 Bq·m−3.
| Participants’ Code | av,Rn | s(av,Rn) | z-Score | D |
|---|---|---|---|---|
| 1 | 306 | 12 | 0.6 | 2.0 |
| 2 | 290 | 6 | −1.0 | −3.3 |
| 3 | 288 | 8 | −1.2 | −4.0 |
| 4 | 292 | 18 | −0.8 | −2.7 |
| 5 | 290 | 10 | −1.0 | −3.3 |
| 6 | 291 | 21 | −0.9 | −3.0 |
| 7 | 300 | 10 | 0 | 0 |
| 8 | 302 | 3 | 0.2 | 0.7 |
av,Rn: the average participant´s value of radon activity concentration in the exposure period (each participant sent their results to SUJCHBO after they received the monitor back and downloaded the data). s(av,Rn): participant´s measurement uncertainties (each participant sent their results to SUJCHBO after they received the monitor back and downloaded the data)—it is an expanded uncertainty—the product of the standard measurement uncertainty and the coverage factor k = 2.
Figure 3Participants’ results for a level of 300 Bq·m−3.