Cinzia Peronace1, Rossana Tallerico1, Manuela Colosimo1, Vanessa Sacco1, Roberta Talarico1, Marco De Fazio1, Federica Pasceri1, Ilenia Talotta1, Giuseppina Panduri1, Jung-Hee Kim2, Erika Cione3, Pasquale Minchella1.
Abstract
Diagnostic laboratory tools are essential to keep everyone safe and track newly emerging variants; on the other hand, "filter" screening tests recognizing positivity are valuable tools to avoid hectic laboratory work that, besides COVID-19, are also part of the routine. Therefore, complementary assays, such as rapid antigen tests (RATs), are essential in controlling and monitoring virus spread within the community, especially in the asymptomatic population. A subset of nasopharyngeal swab specimens resulted in SARS-CoV-2 positive and investigated for genomic characterization were used for RAT validation. RATs were performed immediately after sampling, following the manufacturer's instructions (reading at 15 min). RT-PCRs were carried out within 24 h of specimens' collection. Out of 603 patients, 145 (24.05%) tested positive by RT-PCR and RAT and 451 (74.79%) tested negative by both methods; discordant results (RT-PCR+/RAT- or RT-PCR-/RAT+) were obtained in 7 patients (1.16%). RATs' overall specificity and sensitivity were 96.03% (95%CI: 91.55-98.53%) and 99.78% (95%CI: 98.77-99.99%), respectively, taking RT-PCR as the reference. Overall, RAT negative predictive value was 98.69% (95%CI 97.17-99.40%). The GeneFinder COVID-19 Ag Plus Rapid Test performed well as a screening test for early diagnosis of COVID-19, especially in asymptomatic subjects. The data suggested that patients with RT-PCR-proven COVID-19 testing negative by RAT are unlikely to be infectious. GeneFinder COVID-19 Ag Plus Rapid Test also works on variants of concern (VOC) delta and omicron BA.1 and BA.2.Entities:
Keywords: COVID-19; early diagnosis; rapid antigen detection test; variants of concern
Year: 2022 PMID: 35626282 PMCID: PMC9140403 DOI: 10.3390/diagnostics12051126
Source DB: PubMed Journal: Diagnostics (Basel) ISSN: 2075-4418
NAAT vs. GeneFinder COVID-19 Ag Plus Rapid Test.
| Test | RAT+ | RAT− | Total |
|---|---|---|---|
| NAAT+ | 145 | 6 * | 151 |
| NAAT− | 1 ** | 451 | 452 |
| Total | 146 | 457 | 603 |
* Six false-negative results were obtained from samples with RT-PCR cycle threshold Ct >28. ** In one case, we found a cross-reactivity result for RSV.
GeneFinder COVID-19 Ag Plus Rapid Test performance, compared to RT-PCR standard assay (NAAT).
| Sensitivity | Specificity | Overall Percent Agreement | PPV | NPV | Cohen’s | |
|---|---|---|---|---|---|---|
| 96.03% | 99.78% | 98.84% | 99.32% | 98.69% | 96.87 (SE: 0.01) | |
| Confidence Interval (CI 95%) | 91.55–98.53% | 98.77–99.99% | 97.62–99.53% | 95.43–99.90% | 97.17–99.40% | 0.9457–0.9918 |
The overall sensitivity of the GeneFinder COVID-19 Ag Plus Rapid Test according to the N gene cycle threshold in SH, NH and ANH patients.
| N Gene Cycle Threshold Value | SH ( | NH ( | ANH ( |
|---|---|---|---|
| <25 | 100% | 100% | 100% |
| 25–30 | 100% | 100% | 100% |
| >30 | 100% | 100% | - |
Figure 1(A) RAT test with different Ct values and (B) COI index.
Figure 2ROC curve analysis (A) ANH and SH; (B) ANH and NH.
Variant of Concern tested on GeneFinder COVID-19 Ag Plus Rapid Test.
| VOC | Clade | Lineage | N-Protein Mutations |
|---|---|---|---|
| Delta | 21k | B.1.617.2 | D63G; G215C; D377Y |
| Delta | 21k | AY.4 | D63G; R203M; G215C; D377Y |
| Delta | 21J | AY.39 | A35; D63G; R203M; G215C; D377Y |
| Delta | 21K | AY.42 | D63G; R203M; G215C; D377Y |
| Delta | 21J | AY.122 | A35; D63G; L139F; R203M; G215C; D377Y |
| Delta | 21J | AY.46.6 | D63G; R203M; G215C; D377Y |
| Delta | 21A | AY.53 | D63G; R203M; S327; D377Y; D399Y |
| Delta | 21I | AY.61 | D63G; T135I; R203M; D377Y |
| Delta | 21I | AY.9.2 | D63G; R203M; D377Y |
| Omicron | 21k | B.1.1.529 | P13L; GERS30G_del; R203K; G204R |
| Omicron | 21k | BA.1 | P13L; GERS30G_del; R203K; G204R |
| Omicron | 21k | BA.1.1 | P13L; GERS30G_del; R203K; G204R |
| Omicron | 21L | BA.2 | P13L; GERS30G_del; R203K; G204R; S413R |
| Omicron | 21L | BA.2 | P13L; GERS30G_del; R203K; G204R; S413R |