2-year outcomes of phased radiofrequency ablation for atrial fibrillation with the second-generation PVAC Gold ablation catheter.

M N Klaver1,2, L I S Wintgens3,4, M C E F Wijffels3, V F van Dijk3, A Alipour5, S M Chaldoupi6, R Derksen7, J Peper3, J C Balt3, L V A Boersma3,4.   

Abstract

PURPOSE: The second-generation multi-electrode catheter, PVAC Gold, was designed to improve the safe delivery of phased radiofrequency energy using a "single shot" approach for pulmonary vein isolation (PVI), while retaining efficacy. This large registry presents long-term performance in a daily practice setting.
METHODS: A total of 1011 patients undergoing first time ablation for atrial fibrillation (AF) using PVAC Gold were included, 639 patients with PVI for paroxysmal AF (PAF PVI) and 372 patients with persistent or long-standing persistent AF, divided into 175 patients receiving PVI only (PersAF PVI) and 197 patients receiving PVI with additional substrate ablation (PersAF PVI +).
RESULTS: At 24-month follow-up, single procedure freedom from atrial tachyarrhythmia (ATA) was 58% (368/639) in the PAF PVI group, 44% (77/175) in the PersAF PVI group, and 29% (57/197) in the PersAF PVI + group. Allowing one repeat procedure in 33% of patients, 76%, 65%, and 54% were free from ATA at 24 months, respectively. Pulmonary vein reconnection was observed in 98% of patients with recurrent arrhythmia after PVI.
CONCLUSIONS: Although phased RF ablation with PVAC Gold is quick and safe, the efficacy outcomes are modest compared to current mainstream ablation strategies.
© 2022. The Author(s).

Entities:  

Keywords:  Atrial fibrillation; Catheter ablation; Long-term performance; PVAC Gold; Phased radiofrequency; Pulmonary vein isolation

Year:  2022        PMID: 35604577     DOI: 10.1007/s10840-022-01249-2

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  3 in total

1.  Pulmonary vein isolation by duty-cycled bipolar and unipolar radiofrequency energy with a multielectrode ablation catheter.

Authors:  Lucas V A Boersma; Maurits C E F Wijffels; Hakan Oral; Eric F D Wever; Fred Morady
Journal:  Heart Rhythm       Date:  2008-09-04       Impact factor: 6.343

2.  Predictors of durable pulmonary vein isolation after second-generation cryoballoon ablation with a single short freeze strategy - Different criteria for the best freeze of the 4 individual PVs.

Authors:  Shinsuke Miyazaki; Takatsugu Kajiyama; Tomonori Watanabe; Hiroaki Nakamura; Hitoshi Hachiya; Hiroshi Tada; Yoshito Iesaka
Journal:  Int J Cardiol       Date:  2019-11-09       Impact factor: 4.164