Ana Carolina Dos Santos1, Ludimila Labanca1, Larissa Resende Assumpção1, Patricia Cotta Mancini1, Sirley Alves da Silva Carvalho1, Luciana Macedo de Resende1.
Abstract
Introduction Distortion product otoacoustic emissions (DPOAE) and their suppression may be considered useful in monitoring cochlear function and the efferent auditory pathway inhibitory effect. Nonetheless, the establishment of reliable parameters of response variations is of great importance. Objectives To verify the replicability of test and retest in the research of the inhibitory effect of the efferent pathway using contralateral suppressing stimulus during DPOAE recording for clinical applicability. Methods Cross-sectional study with 48 volunteers, aged 18 to 30 years, with normal audiometric thresholds. The procedures included were audiometric and immittance measures to overrule any conductive or sensorineural conditions and DPOAE recordings without and with contralateral suppression with a 60 dBHL white noise. Distortion product otoacoustic emissions amplitudes were analyzed and compared in both conditions with Wilcoxon test, and the Spearman correlation test was used to assess test-retest reliability. Results The comparative analysis showed differences between amplitudes in test and retest conditions only in 1,500 Hz for DPOAE measures with all other tested frequencies showing no differences, and no difference was observed in all recorded frequencies in the test and retest comparison for DPOAE suppression. The degree of correlation between test and retest of DPOAE amplitude was good at 6,000 Hz and strong (r > 0.880) at the other frequencies. For DPOAE with suppression, all frequencies presented strong correlation between test and retest: 1,500 Hz (r = 0.880), 2,000 Hz (r = 0.882), 3,000 Hz (r = 0.940), and 6,000 Hz (r = 0.957). Conclusions The study found good replicability in contralateral suppression of DPOAE with potential clinical applicability, and we recommend conducting the test from 2000Hz to higher frequencies for more reliable results. Fundação Otorrinolaringologia. This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commecial purposes, or adapted, remixed, transformed or built upon. ( https://creativecommons.org/licenses/by-nc-nd/4.0/ ).Entities:
Keywords: auditory efferent pathway; cochlear hair cells; hearing loss; noise effects; reproducibility of results; young adult
Year: 2021 PMID: 35602270 PMCID: PMC9122768 DOI: 10.1055/s-0041-1733928
Source DB: PubMed Journal: Int Arch Otorhinolaryngol ISSN: 1809-4864
Pass/fail criteria for the minimum amplitudes per frequency of the distortion product otoacoustic emissions established by the manufacturer (Biologic Co)
| Pass/Fail Criteria | ||||
|---|---|---|---|---|
| f2 Frequency (Hz) | DP | And/Or | DP-NF | Noisy Message |
| 8,000 | −13 | And | 6 | −5 |
| 7,000 | −10 | And | 6 | −5 |
| 6,000 | −7 | And | 6 | −5 |
| 5,000 | −6 | And | 6 | −5 |
| 4,000 | −5 | And | 6 | −5 |
| 3,000 | −8 | And | 6 | −5 |
| 2,000 | −7 | And | 6 | −5 |
| 1,500 | −3 | And | 6 | −5 |
Abbreviations: DP, distortion product; IEEP, inhibitory effect of the efferent pathway; NF, noise floor.
Comparative analysis of the left and right ears' distortion product otoacoustic emissions
| Descriptive statistics | 1,500 Hz | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| DP (dB) | NF (dB) | DP-NF (dB) | |||||||
| RE | LE | RE | LE | RE | LE | ||||
|
| 11.45 | 11.13 | 0.590 | −8.45 | −8.75 | 0.587 | 19.98 | 19.89 | 0.743 |
|
| 14.00 | 13.10 | −8.15 | −8.25 | 21.30 | 20.20 | |||
|
| −17.30 | −21.10 | −19.00 | −18.00 | 0.10 | −3.10 | |||
|
| 23.80 | 24.20 | 9.70 | 1.00 | 33.90 | 37.20 | |||
|
| |||||||||
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
| ||||
|
| 7.43 | 6.94 | 0.249 | −12.40 | −12.55 | 1.000 | 19.83 | 19.49 | 0.357 |
|
| 8.00 | 8.20 | −13.80 | −12.90 | 19.35 | 20.00 | |||
|
| −13.60 | −16.60 | −17.50 | −18.60 | 0.60 | 1.00 | |||
|
| 22.20 | 20.80 | −3.50 | −0.80 | 34.20 | 34.00 | |||
|
| |||||||||
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
| ||||
|
| 3.28 | 2.76 | 0.498 | −16.35 | −16.12 | 0.775 | 19.27 | 18.88 | 0.685 |
|
| 4.35 | 3.70 | −16.85 | −17.20 | 20.40 | 19.65 | |||
|
| −16.80 | −17.30 | −23.90 | −22.50 | −17.60 | −5.40 | |||
|
| 16.80 | 15.30 | 4.00 | −5.50 | 35.50 | 34.30 | |||
|
| |||||||||
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
| ||||
|
| −2.90 | −3.73 | 0.800 | −20.44 | −20.30 | 0.293 | 17.54 | 16.36 | 0.767 |
|
| −3.30 | −4.40 | −21.85 | −21.50 | 18.85 | 17.85 | |||
|
| −30.20 | −27.10 | −29.10 | −28.00 | −5.90 | −4.60 | |||
|
| 24.90 | 13.10 | −11.60 | −10.70 | 38.50 | 38.20 | |||
Abbreviations: DP, Distortion product amplitude; LE, left ear; NF, Noise floor; RE, right ear; p-value (Wilcoxon test).
Comparative analysis of distortion product otoacoustic emissions with the right and left ear suppression effect
| Descriptive statistics | 1,500 Hz | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| DP (dB) | NF (dB) | DP-NF (dB) | IEEP (dB) | |||||||||
| RE | LE | RE | LE | RE | LE | RE | LE | |||||
|
| 11.18 | 10.41 | 0.266 | −6.08 | −6.34 | 0.626 | 17.26 | 16.75 | 0.448 | 2.72 | 3.13 | 0.770 |
|
| 13.05 | 12.85 | −5.75 | −6.45 | 18.35 | 16.90 | 2.70 | 2.75 | ||||
|
| −13.70 | −14.90 | −15.80 | −17.80 | −4.40 | 0.60 | −9.00 | −6.00 | ||||
|
| 23.80 | 23.40 | 4.30 | 9.20 | 39.60 | 34.90 | 14.30 | 16.40 | ||||
|
| ||||||||||||
|
|
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
| |||||
|
| 7.18 | 6.61 | 0.488 | −11.16 | −11.50 | 0.512 | 18.33 | 18.12 | 0.828 | 1.50 | 1.38 | 0.580 |
|
| 6.80 | 7.25 | −11.15 | −11.90 | 17.30 | 16.75 | 1.60 | 1.35 | ||||
|
| −8.40 | −10.40 | −19.00 | −18.10 | 6.10 | 2.50 | −8.10 | −6.90 | ||||
|
| 22.30 | 21.00 | 1.80 | 0.00 | 38.00 | 37.00 | 15.10 | 11.50 | ||||
|
| ||||||||||||
|
|
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
| |||||
|
| 4.44 | 2.39 | 0.069 | −16.28 | −16.23 | 0.985 | 19.94 | 18.83 | 0.859 | −0.68 | 0.05 | 0.254 |
|
| 5.20 | 3.05 | −17.40 | −17.05 | 19.70 | 19.80 | −0.25 | 0.40 | ||||
|
| −15.70 | −31.50 | −23.70 | −22.20 | −18.30 | −11.40 | −17.20 | −14.20 | ||||
|
| 21.60 | 15.50 | 3.00 | −3.60 | 37.20 | 31.10 | 8.60 | 11.60 | ||||
|
| ||||||||||||
|
|
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
| |||||
|
| −2.71 | −3.80 | 0.253 | −20.19 | −21.06 | 0.251 | 17.47 | 17.26 | 0.984 | 0.07 | −0.90 | 0.597 |
|
| −2.70 | −4.45 | −21.20 | −21.05 | 18.60 | 18.35 | −0.10 | −0.45 | ||||
|
| −24.20 | −25.30 | −30.70 | −30.60 | 2.40 | −5.10 | −9.50 | −11.90 | ||||
|
| 25.00 | 11.60 | −10.80 | −11.90 | 38.60 | 34.40 | 12.20 | 8.50 | ||||
Abbreviations: DP, distortion product amplitude; IEEP, inhibitory effect of the efferent pathway; LE, left ear; NF, noise floor; RE, right ear; S, suppression.
p -value (Wilcoxon test).
Fig. 1Test and retest distortion products otoacoustic emissions’ amplitudes and distortion product otoacoustic emissions' amplitudes in the test and retest with suppressing noise ( n = 96 ears)
Comparative analysis of distortion product otoacoustic emissions amplitude in the test and retest condition
| Descriptive statistics | 1,500 Hz | 2,000 Hz | 3,000 Hz | 6,000 Hz | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Test | Retest | Test | Retest | Test | Retest | Test | Retest | ||||||
|
|
| 11.29 | 8.99 | 0.001 | 7.19 | 6.89 | 0.335 | 3.02 | 3.12 | 0.953 | −3.31 | −3.19 | 0.567 |
|
|
| 13.25 | 10.80 | 8.00 | 7.65 | 3.85 | 4.15 | −3.45 | −2.75 | ||||
|
| −21.10 | −18.90 | −16.60 | −19.80 | −17.30 | −27.10 | −30.20 | −34.50 | |||||
|
| 24.20 | 23.80 | 22.20 | 22.40 | 16.80 | 17.30 | 24.90 | 25.00 | |||||
|
|
| −8.60 | −8.51 | 0.866 | −12.47 | −12.17 | 0.874 | −16.23 | −16.62 | 0.576 | −20.37 | −20.54 | 0.799 |
|
|
| −8.15 | −9.00 | −13.05 | −12.35 | −17.00 | −17.05 | −21.70 | −20.65 | ||||
|
| −19.00 | −16.90 | −18.60 | −18.20 | −23.90 | −22.70 | −29.10 | −31.60 | |||||
|
| 9.70 | 3.70 | −0.80 | 5.00 | 4.00 | −6.70 | −10.70 | −11.70 | |||||
|
|
| 19.93 | 19.67 | 0.908 | 19.66 | 19.04 | 0.186 | 19.07 | 19.73 | 0.805 | 16.95 | 17.35 | 0.626 |
|
|
| 20.75 | 20.05 | 19.60 | 19.30 | 19.90 | 21.10 | 18.60 | 19.55 | ||||
|
| −3.10 | −11.60 | 0.60 | −2.30 | −17.60 | −10.20 | −5.90 | −7.00 | |||||
|
| 37.20 | 35.40 | 34.20 | 35.30 | 35.50 | 36.20 | 38.50 | 39.30 | |||||
Abbreviations: DP, distortion product amplitude; NF, noise floor; S, suppression.
p -value (Wilcoxon test).
Comparative analysis of distortion product otoacoustic emissions with suppression in the test and retest conditions
| Descriptive statistics | 1,500 Hz | 2,000 Hz | 3,000 Hz | 6,000 Hz | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Test | Retest | Test | Retest | Test | Retest | Test | Retest | ||||||
|
|
| 10.79 | 11.04 | 0.965 | 6.90 | 6.69 | 0.202 | 3.41 | 3.10 | 0.209 | −3.26 | −2.93 | 0.157 |
|
| 13.00 | 12.35 | 7.10 | 7.15 | 4.40 | 4.25 | −3.35 | −2.70 | |||||
|
| −14.90 | −16.70 | −10.40 | −29.80 | −31.50 | −25.90 | −25.30 | −24.50 | |||||
|
| 23.80 | 23.30 | 22.30 | 22.10 | 21.60 | 16.60 | 25.00 | 25.30 | |||||
|
|
| −6.21 | −6.04 | 0.684 | −11.33 | −11.15 | 0.605 | −16.25 | −16.05 | 0.289 | −20.62 | −20.89 | 0.128 |
|
| −5.95 | −5.70 | −11.70 | −11.20 | −17.25 | −16.30 | −21.05 | −21.95 | |||||
|
| −17.80 | −15.70 | −19.00 | −21.50 | −23.70 | −22.40 | −30.70 | −31.10 | |||||
|
| 9.20 | 7.20 | 1.80 | 3.00 | 3.00 | −5.00 | −10.80 | −9.10 | |||||
|
|
| 17.01 | 17.08 | 0.709 | 18.23 | 17.62 | 0.660 | 19.39 | 19.17 | 0.242 | 17.37 | 17.96 | 0.128 |
|
| 17.00 | 17.50 | 17.10 | 17.35 | 19.70 | 19.70 | 18.60 | 18.70 | |||||
|
| −4.40 | −5.00 | 2.50 | −18.90 | −18.30 | −7.50 | −5.10 | −4.70 | |||||
|
| 39.60 | 33.40 | 38.00 | 37.60 | 37.20 | 33.50 | 38.60 | 40.60 | |||||
|
|
| 2.93 | 2.59 | 0.560 | 1.44 | 1.42 | 0.933 | −0.31 | 0.56 | 0.181 | −0.42 | −0.60 | 0.708 |
|
| 2.70 | 1.70 | 1.50 | 1.00 | 0.30 | 0.85 | −0.25 | −0.30 | |||||
|
| −9.00 | −7.70 | −8.10 | −13.40 | −17.20 | −11.90 | −11.90 | −25.00 | |||||
|
| 16.40 | 26.20 | 15.10 | 17.60 | 11.60 | 8.00 | 12.20 | 19.60 | |||||
Abbreviations: DP, distortion product amplitude; IEEP, inhibitory effect of the efferent pathway; NF, noise floor.
p -value (Wilcoxon test).
Fig. 2Test and retest values of the inhibitory effect of the efferent pathway ( n = 96 ears)
Correlation coefficient between test and retest conditions of the distortion product otoacoustic emissions amplitudes with and without suppression, and the inhibitory effect of the efferent pathway
| 1,500 Hz | 2,000 Hz | 3,000 Hz | 6,000 Hz | ||
|---|---|---|---|---|---|
|
|
| 0.761 | 0.973 | 0.957 | 0.960 |
|
| < 0.001 | < 0.001 | < 0.001 | < 0.001 | |
|
|
| 0.930 | 0.929 | 0.974 | 0.979 |
|
| < 0.001 | < 0.001 | < 0.001 | < 0.001 | |
|
|
| 0.563 | 0.464 | 0.289 | 0.281 |
|
| 0.010 | 0.033 | 0.128 | 0.153 |
Abbreviations: DP, distortion product; IEEP, inhibitory effect of the efferent pathway.
p-value = significance probability (intraclass correlation coefficient).