Abstract
This paper presents a new microwave antenna design with improved heating at the tip for interstitial hyperthermia and thermocoagulation. Temperature distribution patterns surrounding a conventional insulated antenna and the new sleeved coaxial slot radiator are measured in a saline phantom under transient and steady-state conditions. The new design provides an improved steady-state temperature distribution over the active region of the radiator. In contrast to the cold zone at the tip of a conventional insulated antenna the new sleeved coaxial slot antenna provides maximal (sixfold greater) power deposition at its tip.Entities:
Mesh:
Year: 1987 PMID: 3559297 DOI: 10.3109/02656738709140371
Source DB: PubMed Journal: Int J Hyperthermia ISSN: 0265-6736 Impact factor: 3.914