| Literature DB >> 35592423 |
Maria Koivisto1, Milad Mosallaei2, Tarja Toimela1, Sampo Tuukkanen2, Tuula Heinonen2.
Abstract
Contractility is one of the most crucial functions of the heart because it is directly related to the maintenance of blood perfusion throughout the body. Both increase and decrease in contractility may cause fatal consequences. Therefore, drug discovery would benefit greatly from reliable testing of candidate molecule effects on contractility capacity. In this study, we further developed a dual-axis piezoelectric force sensor together with our human cell-based vascularized cardiac tissue constructs for cardiac contraction force measurements. The capability to detect drug-induced inotropic effects was tested with a set of known positive and negative inotropic compounds of isoprenaline, milrinone, omecamtiv mecarbil, propranolol, or verapamil in different concentrations. Both positive and negative inotropic effects were measurable, showing that our cardiac contraction force measurement system including a piezoelectric cantilever sensor and a human cell-based cardiac tissue constructs has the potential to be used for testing of inotropic drug effects.Entities:
Keywords: cardiac tissue model; conformal coating; contraction force; force measurement; human-induced pluripotent stem cell–derived cardiomyocytes; in vitro model; inotropic drug
Year: 2022 PMID: 35592423 PMCID: PMC9110810 DOI: 10.3389/fphar.2022.871569
Source DB: PubMed Journal: Front Pharmacol ISSN: 1663-9812 Impact factor: 5.988
FIGURE 1Attachment of the sensor in the sample holder of parylene coating equipment and keeping them in a standing condition with a piece of wire for uniform coating of the polymer in all sides of the sensor during the deposition.
FIGURE 2(A) Measurement setup where the culture plate is placed on a heater plate under a microscope, and the measurement sensor is brought into contact with a cardiac tissue construct. (B) Microscope view of the sensor tip attached onto the cardiac tissue construct.
Mean beating rate and peak-to-peak force amplitudes from the baseline measurements and their percentage changes in the cumulative concentrations of the tested drugs normalized to the baseline measurements.
| Drug | Concentration | Mean beating rate | SD | Mean peak-to-peak force | SD | Number of wells |
|---|---|---|---|---|---|---|
| ISO | 0 nM | 25.87 BPM | 6.44 | 88.95 µN | 82.15 | 6 |
| 10 nM | 176.05% | 33.79 | 105.28% | 9.84 | 6 | |
| 30 nM | 183.86% | 32.59 | 109.56% | 10.62 | 6 | |
| 100 nM | 193.72% | 31.50 | 115.76% | 12.72 | 6 | |
| MIL | 0 µM | 20.76 BPM | 5.95 | 465.44 µN | 622.6 | 9 |
| 1 µM | 108.13% | 16.73 | 103.06% | 9.91 | 9 | |
| 10 µM | 118.06% | 19.11 | 112.15% | 12.38 | 9 | |
| 100 µM | 232.67% | 50.05 | 103.51% | 14.19 | 9 | |
| OME | 0 nM | 22.22 BPM | 6.77 | 37.45 µN | 16.90 | 10 |
| 30 nM | 96.31% | 12.49 | 98.45% | 5.819 | 10 | |
| 100 nM | 94.76% | 17.67 | 95.53% | 13.20 | 10 | |
| 300 nM | 94.51% | 19.36 | 95.81% | 12.23 | 10 | |
| PRO | 0 nM | 21.47 BPM | 4.856 | 673.27 µN | 822.83 | 9 |
| 30 nM | 98.62% | 14.43 | 97.18% | 9.39 | 9 | |
| 100 nM | 90.73% | 16.24 | 94.62% | 10.79 | 9 | |
| 300 nM | 63.16% | 26.93 | 93.63% | 9.03 | 9 | |
| VER | 0 nM | 15.10 BPM | 7.77 | 347.28 µN | 645.90 | 6 |
| 10 nM | 116.01% | 31.83 | 99.53% | 12.20 | 6 | |
| 30 nM | 115.62% | 27.07 | 93.90% | 15.86 | 6 | |
| 100 nM | 157.02% | 25.54 | 85.71% | 8.26 | 4 |
FIGURE 3Percentage change of the beating rate and force in the cumulative concentrations of (A) isoprenaline (ISO), (B) milrinone (MIL), (C) omecamtiv mecarbil (OME), (D) propranolol (PRO), and (E) verapamil (VER) exposures. Mean and SD of the beating rate and peak-to-peak amplitudes from each repeat R1–R3 (number of measurements per repeat). *N = 1 at 100 nM. (F) Beating from an example well during baseline and 100 nM isoprenaline force measurements.
FIGURE 4Cardiac troponin T (green)–positive cardiomyocytes and the von Willebrand factor (red)–positive vascular network. Scale bar: 200 µm.