Evaluation and assessment of new disposable strip for determination of plasma potassium concentration.

I Gibb.   

Abstract

Overall laboratory reproducibility for the Ames Seralyzer and plasma potassium test strips was less than 4% within the plasma reference range; and linearity extended from 2.1 to 10.2 mmol/l. Neither bilirubin nor lipaemia interfered in the analysis and selectivity for potassium over sodium was 450:1. Laboratory comparison with automated flame photometry and an indirect ion selective electrode showed a Seralyzer bias of +0.10 (+/- 25D 0.26) and +0.11 (0.27) mmol/l. Against flame photometry, comparison of analyses by six doctors from a coronary care unit and four nurses from a renal dialysis unit showed smaller positive biases (0.06 and 0.02 mmol/l) but substantially larger error ranges (0.43 and 0.55 mmol/l, respectively). Some 49% of the nurses' and 37% of the doctors' results differed from laboratory values by more than 0.2 mmol/l. Potassium test strip results proved acceptably comparable and reproducible when produced by an experienced analyst, but greater variability in clinical hands rendered the system unsuitable for precise monitoring of potassium concentration.

Entities:  

Mesh:

Substances:

Year:  1987        PMID: 3558863      PMCID: PMC1140903          DOI: 10.1136/jcp.40.3.298

Source DB:  PubMed          Journal:  J Clin Pathol        ISSN: 0021-9746            Impact factor:   3.411


  3 in total

1.  A study of the accuracy and precision of clinical chemistry determinations in 170 Canadian laboratories.

Authors:  D B TONKS
Journal:  Clin Chem       Date:  1963-04       Impact factor: 8.327

2.  Medical significance of laboratory results.

Authors:  R N Barnett
Journal:  Am J Clin Pathol       Date:  1968-12       Impact factor: 2.493

3.  Solid-phase colorimetric determination of potassium.

Authors:  S C Charlton; R L Fleming; A Zipp
Journal:  Clin Chem       Date:  1982-09       Impact factor: 8.327

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.