Olufunmilola Kolude1, Eyitayo E Emmanuel1, Ayomide O Aibinuomo1, Tope M Ipinnimo2,1, Mary O Ilesanmi1, John A Adu1.
Abstract
Background Antigen rapid diagnostic testing (Ag-RDT) for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) is an important testing technique used for the control and containment of COVID-19. This study aims to assess the quality of COVID-19 Ag-RDT in the testing sites in Ekiti State, Nigeria. Methods A validated World Health Organization (WHO) questionnaire was used to collect data from 138 testing sites located in 138 health facilities in Ekiti State. The assessment was based on the activities carried out in the last three months before the study. Results A total of 138 testing sites participated in the study out of which 121 (87.7%) were primary health facility testing sites. The mean number of samples tested in these sites was 26 samples, and 97 (70.3%) testing sites were tested below this. The average quality performance of the secondary/tertiary health facility testing sites (64.46 ± 10.47) was significantly higher than that of the primary health facility testing sites (53.13 ± 13.54) (p = 0.002). Additionally, the average quality performance of testing sites that tested 26 samples or more (61.61 ± 9.84) was significantly higher than that of the testing sites that tested below this mean cut-off (51.53 ± 13.97) (p < 0.001). Conclusion The majority of the testing sites that tested below the mean 26 samples, secondary/tertiary health facility testing sites, and sites that tested above the mean cut-off had higher average quality performance scores. Therefore, encouraging clinicians to refer for more testing of suspected cases and supportive supervision of COVID-19 Ag-RDT, especially in primary health facilities, is recommended.Entities:
Keywords: antigen-rdt; assessment of quality; covid-19; nigeria; testing sites
Year: 2022 PMID: 35586352 PMCID: PMC9108682 DOI: 10.7759/cureus.24173
Source DB: PubMed Journal: Cureus ISSN: 2168-8184
Type of testing sites and average number of samples tested
| Variables | Frequency, N = 138 | Percent (100%) |
| Types of testing sites | ||
| Primary health facilities | 121 | 87.7 |
| Secondary health facilities | 16 | 11.6 |
| Tertiary health facilities | 1 | 0.7 |
| Mean number of samples tested (26 samples) | ||
| Tested below 26 samples | 97 | 70.3 |
| Tested 26 samples and above | 41 | 29.7 |
Quality assessment of testing sites
| Sections | Average score ± standard deviation, N = 138 |
| Documents and records | 32.95 ± 18.11 |
| Personnel training and certification | 66.81 ± 32.71 |
| Physical infrastructure | 51.10 ± 23.56 |
| Safety | 48.08 ± 17.73 |
| Pre-testing phase | 66.05 ± 25.09 |
| Testing phase | 67.03 ± 21.01 |
| Post-testing phase | 74.15 ± 14.19 |
| External quality assessment | 30.02 ± 18.93 |
| Average quality performance | 54.52 ± 13.65 |
Relationship between the testing sites’ characteristics and quality assessment
SD: Standard deviation
| Characteristics of testing sites | ||||||
| Sections | Types of testing sites | Mean number of samples tested | ||||
| Primary health facilities (n = 121) | Secondary/tertiary health facilities (n = 17) | Tested below 26 samples (n = 97) | Tested 26 samples and above (n = 41) | |||
| Average score ± SD | Average score ± SD | T-test (p-value) | Average score ± SD | Average score ± SD | T-test (p-value) | |
| Documents and records | 31.71 ± 16.57 | 43.95 ± 24.70 | -2.603 (0.010) | 31.41 ± 16.38 | 36.59 ± 21.46 | -1.541 (0.126) |
| Personnel training and certification | 67.02 ± 33.31 | 47.78 ± 29.88 | 0.160 (0.873) | 66.70 ± 35.90 | 67.07 ± 23.90 | -0.061 (0.952) |
| Physical infrastructure | 47.78 ± 22.13 | 74.35 ± 20.67 | -4.545 (<0.001) | 43.94 ± 22.23 | 68.04 ± 17.22 | -6.195 (<0.001) |
| Safety | 46.07 ± 17.08 | 61.56 ± 16.20 | -3.428 (0.001) | 45.46 ± 18.06 | 54.27 ± 15.43 | -2.728 (0.007) |
| Pre-testing phase | 63.52 ± 25.20 | 83.04 ± 15.18 | -3.020 (0.003) | 61.86 ± 25.98 | 75.96 ± 19.81 | -3.112 (0.002) |
| Testing phase | 67.54 ± 20.66 | 65.28 ± 23.08 | 0.406 (0.686) | 65.92 ± 21.65 | 69.65 ± 21.65 | -0.952 (0.343) |
| Post-testing phase | 73.05 ± 14.25 | 81.25 ± 11.18 | -2.212 (0.029) | 71.25 ± 14.59 | 81.03 ± 10.50 | -3.885 (<0.001) |
| External quality assessment | 28.36 ± 17.86 | 40.62 ± 22.64 | -2.500 (0.014) | 25.70 ± 15.03 | 40.24 ± 23.08 | -4.393 (<0.001) |
| Average quality performance | 53.13 ± 13.54 | 64.46 ± 10.47 | -3.218 (0.002) | 51.53 ± 13.97 | 61.61 ± 9.84 | -4.195 (<0.001) |
Documentation and records
| 4.0 | Documentation and Records | Y | P | N | Comments |
| 4.1 | Are the following guidelines specific for SARS-CoV-2 antigen rapid diagnostic testing available at the testing facility? Instruction: The Evaluator/Auditor/Mentor/Supervisor must establish and determine the relevant latest International & National Guidelines before assessing the testing facility. For each guideline(s) listed below, award 1 point if the current version of the guideline is available, award 0 points if unavailable (and 0.5 points if available but not the current version or otherwise) | ||||
| a. | Africa CDC Interim Guidance on the Use of Rapid Antigen Tests for COVID-19 Response | ||||
| b. | Quality Assurance Framework for SARS-COV-2 antigen rapid testing for the diagnosis of COVID-19 (by the Africa CDC/ASLM) | ||||
| c. | WHO “Interim” Guidance antigen detection in the diagnosis of SARS-COV-2 infection using immunoassays (latest guidance) | ||||
| d. | Are there SOPs and/or job aides in place to implement safety practices? | ||||
| e. | Are there SOPs and/or job aides in place on how to dispose of infectious and non-infectious waste? | ||||
| f. | Are there SOPs and/or job aides in place to manage spills of COVID-19 samples, blood, and other body fluids? | ||||
| 4.2 | Is the national SARS-CoV-2 antigen rapid diagnostic testing algorithm available at the testing facility/site? | ||||
| 4.3 | Are there national guidelines describing how client identification should be recorded in the SARS-CoV-2 antigen rapid diagnostic testing register? | ||||
| 4.4 | For each of the SARS-CoV-2 Ag-RDT kits in use at the testing facility, are the manufacturer instructions/manuals/inserts available and accessible to testers? | ||||
| 4.5 | Are SOPs and/or job aides in place for each SARS-CoV-2 antigen rapid diagnostic test used in the testing algorithm available and posted at the testing point? | ||||
| 4.6 | Are national biosafety guidelines for COVID-19 for infection prevention and control available and accessible to the testers? | ||||
| 4.7 | Does the testing facility have national/regional guidelines for monitoring COVID-19 Ag-RDT key performance and quality indicators? | ||||
| 4.8 | Have the site performed and documented risk assessment according to the National Guidelines or WHO Interim Guidelines? | ||||
| 4.9 | Are there a national standardized SARS-CoV-2 Ag-RDT tests register/logbook available for recording test results? | ||||
| 4.10 | Are there records indicating all testers have demonstrated competency in SARS-CoV-2 antigen rapid diagnostic testing prior to client testing? | ||||
| 4.11* | Are there records in the testing register/logbook for swapped results received from the referral lab? | N/A | |||
| Documents & records scores attained | |||||
| Documents & records maximum possible score | 16 | ||||
| Documents & records % score attained instruction: % score computation = scores attained divided by maximum possible score x 100% | |||||
Personnel training and certification
| 5.0 Personnel Training and Certification | Y | P | N | Comments | |
| 5.1 | Have all testers received a comprehensive training on COVID-19 Ag rapid testing? | ||||
| 5.2 | Are the testers trained on the use of standardized SARS-CoV-2 antigen rapid diagnostic test registers/logbooks? | ||||
| 5.3 | Are the testers trained on quality control (QC) processes? | ||||
| 5.4 | Are the testers trained on safety, risk assessment, and waste management procedures and practices? | ||||
| 5.5 | Are only certified testers allowed to perform the SARS-CoV-2 antigen rapid diagnostic test? | ||||
| Personnel training & certification scores attained | |||||
| Personnel training & certification maximum possible score | 5 | ||||
| Personnel training & certification % score attained instruction: % score computation = scores attained divided by maximum possible score x 100% | |||||
Physical facility
| 6.0 | Physical Facility | Y | P | N | Comments |
| 6.1 | Is there a designated area for SARS-CoV-2 antigen rapid diagnostic testing? | ||||
| 6.2 | Is the testing area clean and organized for SARS-CoV-2 antigen rapid diagnostic testing? | ||||
| 6.3 | Is sufficient lighting available in the designated testing area? | ||||
| 6.4 | Are the SARS-CoV-2 Ag-RDT kits stored within the temperature range based on the manufacturers’ instructions? | ||||
| 6.5 | Is there sufficient and secure storage space for test kits and other consumables? | ||||
| 6.6 | Is the space allocated for SARS-CoV-2 antigen rapid diagnostic testing adequate to perform the work without compromising the quality and safety of patients and testing personnel/health workers? | ||||
| 6.7 | Is the COVID-19 sample collection area separated from the patient examination/testing areas/room(s)? | ||||
| Physical facility scores attained | |||||
| Physical facility maximum possible score | 7 | ||||
| Physical facility % score attained instruction: % score computation = scores attained divided by maximum possible score x 100% | |||||
Safety
| 7.0 Safety | Y | P | N | Comments | |
| 7.1 | Is there a biohazard sign on the doors of the rooms/or at the designated area where COVID-19 RDT testing is done? | ||||
| 7.2 | Are the following personal protective equipment available in sufficient quantity at the testing facility? Instruction: For each of the listed PPE below, award 1 point if available in sufficient quantities, 0 points if unavailable and 0.5 points if available but in insufficient quantities | ||||
| a. | Gloves | ||||
| b. | Gowns/Laboratory coats | ||||
| c. | Eye protection or face shields | ||||
| d. | Respirator (FFP2 or N95) | ||||
| e. | Face masks | ||||
| 7.3 | Is PPE consistently and properly used by all testers through the testing process? Instructions: The evaluator/assessor/supervisor observes for consistent use of the PPE throughout the processes before scoring this requirement. | ||||
| 7.4 | Is there an installation/sink/facility dedicated to handwashing? | ||||
| 7.5 | Is there clean water and soap available for handwashing? | ||||
| 7.6 | Is an appropriate disinfectant to clean the work area available? | ||||
| 7.7 | Are leak-proof biohazard bags and appropriate waste bins available and properly used on the testing site? | ||||
| 7.8 | Are sharps, infectious, and non-infectious waste handled properly? | ||||
| 7.9 | Are infectious and non-infectious waste containers emptied regularly per the SOP and/or job aides? | ||||
| Safety scores attained | |||||
| Safety maximum possible score | 13 | ||||
| Safety % score attained instruction: % score computation = scores attained divided by maximum possible score x 100% | |||||
Pre-testing phase
| 8.0 Pre-testing Phase | Y | P | N | Comments | |
| 8.1 | Is the national SARS-CoV-2 antigen rapid diagnostic testing algorithm being used at the testing facility/point and adhered to? | ||||
| 8.2 | Is there a process in place for an alternative SARS-CoV-2 antigen rapid diagnostic testing algorithm in case of an expired or shortage of test kit(s)? | ||||
| 8.3 | Are only national or WHO Emergency Use Listing Procedure (EUL) approval SARS-CoV-2 Ag-RDT approved test kits available for use currently? | ||||
| 8.4 | Are all the test kits currently in use within the expiration date? | ||||
| 8.5 | Are the test kits labeled with the date received and initials? | ||||
| 8.6 | Are there sufficient supplies available for client sample collection? | ||||
| 8.7 | Are client identifiers recorded in the SARS-CoV-2 antigen rapid diagnostic testing register per national guidelines and on test devices? | ||||
| *8.8 | If samples are shipped to a reference lab, does the laboratory have appropriate packaging for referring specimens (triple package if air transport, or any package in conformity with local regulations or recommendations)? | N/A | |||
| 8.9 | Is a transportation system for sample referral (bus, ambulance, national postal service, etc.) already set up? | ||||
| 8.10 | Is/are the person(s) in charge of shipments trained for the transport of infectious substances? | ||||
| Pre-testing phase scores attained | |||||
| Total of “Not Applicable” | |||||
| Pre-testing phase maximum possible score | 10 | Guidance: In case, checklist item 5.8 is not applicable, then subtract 1 point from the total 10 points. | |||
| Pre-testing phase % score attained instruction: % score computation = scores attained divided by maximum possible score x 100% | |||||
Testing phase
| 9.0 Testing Phase | Y | P | N | Comment | |
| 9.1 | Are timers available and used routinely for SARS-CoV-2 antigen rapid diagnostic testing? | ||||
| 9.2 | Did the tester put on the appropriate PPE for testing? | ||||
| 9.3 | Are sample collection devices used correctly? | ||||
| 9.4 | Are testing procedures adequately followed? | ||||
| 9.5 | Is IQC performed and documented according to National/International Guidelines or kit manufacturer instructions? | ||||
| 9.6 | Did the tester put on the appropriate PPE for testing? | ||||
| 9.7 | Did the tester adhere to the manufacturer's instructions for using the SARS-CoV-2 antigen RDT? | ||||
| 9.8 | Did the tester have all the necessary supplies to perform the SARS-CoV-2 antigen RDT procedure before starting the sample testing process? | ||||
| 9.9 | Did the tester set up the workstation correctly? | ||||
| 9.10 | Did the tester check the expiry date of the SARS-CoV-2 antigen RDT? | ||||
| 9.11 | Did the tester check that the test device and the desiccant pack in the foil pouch were not damaged or invalid? | ||||
| 9.12 | Did the tester insert the swab into an extraction buffer tube and, while squeezing the buffer tube, stir the swab? | ||||
| 9.13 | Did the tester remove the swab while squeezing the sides of the tube to extract the liquid from the swab? | ||||
| 9.14 | Did the tester press the nozzle cap tightly onto the tube? | ||||
| 9.15 | Did the tester apply the required number of drops of the extracted specimen to the specimen well of the test device? | ||||
| 9.16 | Are incorrect/invalid QC results properly recorded? | ||||
| 9.17 | Are appropriate steps taken and documented when QC results are incorrect and/or invalid? | ||||
| 9.18 | Are QC records reviewed by the person in charge routinely? | ||||
| Testing phase scores attained | |||||
| Testing phase maximum possible score | 18 | ||||
| Testing phase % score attained instruction: % score computation = scores attained divided by maximum possible score x 100% | |||||
Post-testing
| 10.0 Post-testing Phase | Y | P | N | Comments | |
| 10.1 | Is a national standardized register/logbook being used correctly and consistently to record SARS-CoV-2 antigen RDT test results? | ||||
| 10.2 | Does the SARS-CoV-2 antigen RDT tests testing register/logbook include all of the key quality elements? | ||||
| 10.3 | Are all the elements in the register/logbook recorded/captured correctly? (e.g., client demographics, kit names, lot numbers, expiration dates, tester name, individual and final SARS-CoV-2 antigen RDT test results, etc.)? | ||||
| 10.4 | Is there documented evidence that clinicians are immediately notified of SARS-CoV-2 test positive results to inform timely patient isolation and management/treatment? | ||||
| 10.5 | Are invalid test results recorded in the register/logbook as well? | ||||
| 10.6 | Are invalid tests repeated and results properly recorded in the register/logbook? | ||||
| 10.7 | Are all client documents and records securely kept throughout all phases of the testing process? | ||||
| 10.8 | Are all registers/logbooks and other documents kept in a secure location when not in use? | ||||
| 10.9 | Are registers/logbooks properly labeled and archived when full? | ||||
| Testing phase scores attained | |||||
| Testing phase maximum possible score | 9 | ||||
| Testing phase % score attained instruction: % score computation = scores attained divided by maximum possible score x 100% | |||||
External quality audit
| 11.0 External Quality Audit (PT, Supervision, and Retesting) | Y | P | N | Comments | ||
| 11.1 | Is the testing facility enrolled in an EQA/PT program? | |||||
| 11.2 | Do all testers at the testing facility test the EQA/PT samples? | |||||
| 11.3 | Does the person in charge at the testing facility review the EQA/PT results before submission to NRL or the designee? | |||||
| 11.4 | Is an EQA/PT report received from National Reference Laboratory and reviewed by testers and/or the person in charge at the testing facility? | |||||
| 11.5 | Does the testing facility conduct sample retesting using other laboratories as an alternative to SARS-CoV-2 Ag-RDT EQA? | |||||
| 11.6 | Does the testing facility implement corrective action in case of unsatisfactory results? | |||||
| 11.7 | Does the testing facility receive periodic supervisory visits? | |||||
| 11.8 | Is feedback provided during the supervisory visit and documented? | |||||
| 11.9 | If testers need to be retrained, are they being retrained during the supervisory visit? | |||||
| N/A | Guidance: In case checklist items 8.10-8.13, whichever is not applicable, subtract 1 point from the total 13 points for each item. | |||||
| 11.10* | Does the site collect swap samples for retesting according to country guidelines (e.g., collection of every 20th client sample)? | |||||
| 11.11* | Are the swap samples collected for retesting properly documented? | |||||
| 11.12* | Are swap samples stored properly? | |||||
| 11.13* | Are the identifier’s samples sent for retesting properly recorded? | |||||
| External quality audit scores attained | ||||||
| External quality audit maximum possible score | 13 | |||||
| External quality audit % score attained instruction: % score computation = scores attained divided by maximum possible score x 100% | ||||||