Mahado Ismail1, Catia Costa2, Katherine Longman1, Mark A Chambers3, Sarah Menzies4, Melanie J Bailey1,2. 1. Department of Chemistry, University of Surrey, Guildford, Surrey GU2 7XH, U.K. 2. Ion Beam Centre, University of Surrey, Guildford, Surrey GU2 7XH, U.K. 3. Faculty of Health and Medical Sciences, University of Surrey, Guildford, Surrey GU2 7XH, U.K. 4. Wexham Park Hospital, Frimley Health NHS Foundation Trust, Frimley, Surrey GU16 7UJ, U.K.
Abstract
A fingerprint offers a convenient, noninvasive sampling matrix for monitoring therapeutic drug use. However, a barrier to widespread adoption of fingerprint sampling is the fact that the sample volume is uncontrolled. Fingerprint samples (n = 140) were collected from patients receiving the antibiotic isoniazid as part of their treatment, as well as from a drug-naive control group (n = 50). The fingerprint samples were analyzed for isoniazid (INH) and acetylisoniazid (AcINH), using liquid chromatography high-resolution mass spectrometry. The data set was analyzed retrospectively for metabolites known to be present in eccrine sweat. INH or AcINH was detected in 89% of the fingerprints collected from patients and in 0% of the fingerprints collected from the control group. Metabolites lysine, ornithine, pyroglutamic acid, and taurine were concurrently detected alongside INH/AcINH and were used to determine whether the fingerprint sample was sufficient for testing. Given a sufficient sample volume, the fingerprint test for INH use has sensitivity, specificity, and accuracy of 100%. Normalization to taurine was found to reduce intradonor variability. Fingerprints are a novel and noninvasive approach to monitor INH therapy. Metabolites can be used as internal markers to demonstrate a sufficient sample volume for testing and reduce intradonor variability.
A fingerprint offers a convenient, noninvasive sampling matrix for monitoring therapeutic drug use. However, a barrier to widespread adoption of fingerprint sampling is the fact that the sample volume is uncontrolled. Fingerprint samples (n = 140) were collected from patients receiving the antibiotic isoniazid as part of their treatment, as well as from a drug-naive control group (n = 50). The fingerprint samples were analyzed for isoniazid (INH) and acetylisoniazid (AcINH), using liquid chromatography high-resolution mass spectrometry. The data set was analyzed retrospectively for metabolites known to be present in eccrine sweat. INH or AcINH was detected in 89% of the fingerprints collected from patients and in 0% of the fingerprints collected from the control group. Metabolites lysine, ornithine, pyroglutamic acid, and taurine were concurrently detected alongside INH/AcINH and were used to determine whether the fingerprint sample was sufficient for testing. Given a sufficient sample volume, the fingerprint test for INH use has sensitivity, specificity, and accuracy of 100%. Normalization to taurine was found to reduce intradonor variability. Fingerprints are a novel and noninvasive approach to monitor INH therapy. Metabolites can be used as internal markers to demonstrate a sufficient sample volume for testing and reduce intradonor variability.
For the first time
in over a decade, tuberculosis (TB)-related
deaths have increased.[1] With increased
disruption to medical services and reduced access to treatment, the
ongoing pandemic of coronavirus disease (COVID-19) has reversed years
of progress in the fight to end TB. As we enter the final year to
achieve the historic targets set out in the first high-level United
Nations (UN) meeting on TB in 2018, the need for effective treatment
and disease management could not be more critical. It is estimated
that in 2019, 10.0 million fell ill with TB, and of those, 465,000
cases were multidrug- or rifampicin-resistant TB (MDR-RR TB).[2] Such cases arise when a patient is infected with Mycobacterium tuberculosis resistant to a particular
antibiotic or where M. tuberculosis acquires resistance during treatment.[3,4] Acquired resistance
can develop through improper use of antibiotics (examples including
poor adherence to a treatment regime or patients prematurely ceasing
treatment) or due to suboptimal in vivo drug concentration
in treatment regimens.[4,5] The global success rate for treatment
of drug-resistant TB is 57%, where treatment requires commitment to
a regime of second-line drugs over a period of 9–20 months
with continued counselling and support for adverse events.[2]Treatment adherence in patients with TB
is essential to limit transmission,
avoid relapse, and restrict the development of drug resistance. Various
strategies have been employed to increase adherence to treatment and
improve success rates.[6,7] Current methods to monitor adherence
include regular engagement with medical and nursing teams, including
provision of direct observation therapy (DOTS), tablet count, and
assessment of clinical improvement.[8,9] A urine dipstick
test for the frontline TB drug, isoniazid (INH; based on a colorimetric
test) can also be used on the spot if there are concerns or doubts
regarding treatment efficacy.[10−12] However, each of these approaches
have their limitations: tablet count is ineffective because it is
easily altered by the patient,[8,9] engagement with the
patient through DOTS is labor-intensive and expensive to administer,[9,13] and a urine dipstick test requires clinic access.Novel approaches
for monitoring TB treatment adherence include
the use of hair samples or saliva as a testing matrix.[14−17] Hair can provide a snapshot of historic drug use (therapeutic or
illicit), depending on the hair length and the stability of the drugs.
However, it presents challenges with external contamination and/or
effect of hair treatments (e.g., hair dye), which
can lead to false positive or negative results, respectively.[14] Additionally, the quantitative potential of
this matrix is poor, and publications have highlighted how failing
to take historic medication uptake into consideration can result in
false positive results or falsely high measurement of drug levels
in hair samples.[14,16] Saliva offers an alternative
approach, but a recent publication highlighted issues with highly
variable saliva–serum concentrations.[17]In addition to adherence monitoring, it is of interest to
clinicians
to be able to carry out therapeutic drug monitoring to ensure that
the drugs used to treat TB have been delivered at an appropriate therapeutic
dose. The risk of multi-drug-resistant tuberculosis (MDR-TB) increases
with suboptimal drug concentrations, which may arise due to nonadherence
as well as other factors (malabsorption for example).[18−20] This can be carried out by testing of venous blood samples using
liquid chromatography mass spectrometry (LC–MS),[21−24] or alternatively, the use of dried blood spots has been proposed.[25−27] Collection of blood requires specialist equipment utilized by trained
medical professionals, where additional complexities relating to necessary
laboratory preprocessing steps for storage and transport can cause
distress or discomfort in the donor. Such approaches are also prohibitively
expensive for the majority of countries with the highest burden of
infection.A fingerprint, unlike blood, offers a unique opportunity
for patients
to provide samples noninvasively, both saving clinical resources and
improving patient experience, as they are easily collected and transported.
Furthermore, in a post-pandemic climate, the simplicity of fingerprint
sampling could prove advantageous where access to clinics is limited,
especially for high-risk individuals such as those with TB.Prior research has shown the potential to use fingerprints for
illicit drug testing[28−34] and medical adherence monitoring.[35] Despite
a multitude of potential applications, uptake of fingerprint drug
testing has been stunted by the ability to distinguish between drug
use versus environmental exposure and the inability
to determine, or control for, sample volume. The issue of possible
contamination has been overcome through implementation of hand washing
procedures,[28,36] as well as application of appropriate
cutoff levels.[31] However, it is known that
the collection of reproducible fingerprints remains highly challenging,
where factors such as contact time, pressure, and time of collection
can influence the amount and composition of material collected.[37] Further, to control sample collection conditions,
previous attempts have utilized endogenous compounds (e.g., creatinine) to retrospectively adjust drug elimination profiles[29,30] and mathematical modeling using metabolic biomarkers to account
for variable sweat volume.[38]Here,
we describe for the first time the detection of INH and its
metabolite, acetylisoniazid (AcINH), in fingerprint samples collected
from patients undergoing treatment for TB. This provides proof of
principle that a fingerprint sample can be used to monitor INH use.
Additionally, we explore the variability in INH/AcINH and other endogenous
compounds simultaneously detected in the fingerprint depositions from
donors on INH treatment. We also demonstrate for the first time that
endogenous markers can be used to verify that a sufficient fingerprint
sample has been collected to eliminate false negative results.
Materials
and Methods
Sample Collection
A favorable ethical opinion was obtained
from the National Research Ethics Service (NRES—REC reference
16//LO/1663) for the collection and analysis of fingerprint samples
taken from individuals receiving treatment for TB at Wexham Park Hospital
(Frimley Health NHS Foundation Trust). Fingerprint samples (all five
fingers of the right hand) were collected from (a) 28 patients (giving n = 140 samples) taking INH treatment for TB and (b) 10
individuals (giving n = 50 samples) who had never
taken INH (negative control group). Samples were collected as per
a previous work[31] to include one set of
samples “as presented” and another after handwashing.
Fingerprint samples were collected on 2 × 2 cm squares of Whatman
1-Chr grade chromatography paper. Samples were collected using kitchen
scales (Sainsbury’s Color) to measure the pressure applied
during sample collection (800–1200 g for 10 s).
Materials
INH, AcINH, and isoniazid-d4 (INH-d4; used as an internal
standard) were purchased from Toronto Research Chemicals (North York,
Canada). Optima-grade LC–MS solvents of methanol, acetonitrile
(ACN), and water were used to prepare solutions and solvent mixtures
(Fisher Scientific, Leicestershire, UK). Formic acid was added to
the mobile phase solvents at 0.1% (v/v) (Fischer Scientific, Leicestershire,
UK). Dichloromethane was used to prepare solvent extraction mixtures
and was of analytical or reagent grade (Sigma-Aldrich, Dorset, UK).
A solution of artificial eccrine perspiration was used to simulate
a fingerprint matrix in the experimental design of fingerprint experiments
(Pickering Laboratories, Inc., Obertaufkirchen, Germany).
Sample Preparation
and Analysis
Fingerprint samples
were extracted as previously described.[31] Extracted samples were reconstituted in 100 μL of 50:50 (v/v)
ACN and water (with 0.1% formic acid) containing 50 ng/mL INH-d4 as the internal standard.Chromatographic
separation was performed on a Thermo Scientific Ultimate 3000 ultrahigh-performance
LC (UHPLC) (Thermo Scientific, Bremen) using a Kinetex C18 column
(100 × 2.1 mm, 5 μm) operated at 30 °C at a flow rate
of 0.25 mL/min. Gradient analysis was performed with an initial mobile
phase comprising 95% water (0.1% formic acid) and 5% ACN (0.1% formic
acid), increased to 80% ACN (0.1% formic acid) and 20% water (0.1%
formic acid) over 2 min, and kept constant for 0.5 min before returning
to the initial mobile phase composition. The samples were introduced
to a Q Exactive Plus Hybrid Quadrupole Orbitrap mass spectrometer
(Thermo Scientific, Bremen) via the standard electrospray ionization
(ESI) interface. Table S1 describes the
ESI and MS parameters used for the analysis.Analysis of 0.5–10
ng of INH and AcINH standards extracted
from paper produced R2 values >0.999
and
within a day precision of ≤3% (n = 5). The
lower limit of detection was 10 pg for both analytes (see Figure S1). To evaluate matrix effects, 10 μL
of the analyte standard at 100, 500, and 1000 ng/mL concentrations
in methanol (equivalent to deposited masses of 1, 5, and 10 ng) was
pipetted onto fingerprints (n = 5 donors) from nonpatients
collected “as presented” and after handwashing, as well
as a blank paper to monitor the response of AcINH and INH with and
without fingerprints present. For all standards, the presence of a
fingerprint suppressed the INH signal by ∼20%. This matrix
effect varied by ≤9% for the five different donors. For AcINH,
no overall signal suppression was observed, and the corresponding
variability between donors was ≤14% (see Figure S2). A blank mobile phase injection was performed after
each calibration standard in triplicate to evaluate the potential
for carryover at each concentration level. No carryover was observed
at any calibration level for INH, AcINH, or INH-d4. TraceFinder software (Thermo, UK) was used for retrospective
interrogation of the data to search for compounds reported in previously
published studies on MS applied to fingerprints.[28,37] Peak assignment of sweat metabolites was carried out using the accurate
mass of the protonated molecular ion (with a 3 ppm acceptance criterion).
Samples extracted from a blank paper and analyzed using the same method
were inspected to ensure that any compounds of interest were detected
in the fingerprints at signal intensities above 3 times the levels
measured in the blank samples. The metabolites investigated and their
observed m/z values are given in Table S2.
Results
Detection
rates for INH analytes in participant fingerprints are
summarized in Table . INH analytes were not detected in the fingerprints from the negative
control group (Figure S3). AcINH was detected
more frequently than INH, and 89% of fingerprints were positive for
either INH or AcINH, reducing to 84% after handwashing. Based on these
data, a fingerprint test for INH use performs better when both the
parent drug and metabolite are used for detection, and handwashing
should be avoided immediately prior to testing.
Table 1
Detection Rates of INH and AcINH from
Individual Fingerprint Samples Collected from Patients (Taking INH
Treatment) and the Negative Control Group (Not Taking INH Treatment)
n (no. of fingerprint samples)
AcINH only (n, %)
INH only (n, %)
both analytes (n, %)
one analyte (n, %)
neither analyte (n, %)
unwashed hands
patients
140
29
1
95
124
15
21%
1%
68%
89%
11%
negative control
50
0
0
0
0
50
0%
0%
0%
0%
100%
washed hands
patients
140
40
4
40
117
23
29%
3%
29%
84%
16%
negative control
50
0
0
0
0
50
0%
0%
0%
0%
100%
The proportion of patient samples that were
negative for both analytes
was 11%, increasing to 16% after handwashing. There are several explanations
for these negative results: (i) the patients had not taken their medication
as directed, (ii) the sampling window exceeded the detection window
of the drug, or (iii) an insufficient fingerprint sample had been
deposited.Despite efforts to control the fingerprint deposition
parameters
(deposition pressure, collection area, and contact time), a considerable
intradonor range in the AcINH level was observed (Figure S4). This could arise from several factors: variation
in volume of the fingerprint sample or variation from finger to finger
in the concentration of the analyte.To decouple these possibilities,
a retrospective inspection of
the data was carried out to find biomarkers of fingerprint sweat.
A range of analytes previously reported in fingerprints and eccrine
sweat were considered (Table S2). Of these,
protonated molecular ions corresponding to the amino acid or amino
acid derivatives lysine, ornithine, pyroglutamic acid and taurine
were most frequently detected, and Figure shows the intradonor variation in these
compounds for a selection of donors. Figure clearly shows a relationship between the
detection of endogenous metabolites and the detection of AcINH. For
example, AcINH was detected in all fingerprints of donors FP033 and
FP038, as well as pyroglutamic acid and taurine. Nondetection of AcINH
from any fingerprint of donor FP037 coincided with the nondetection
of the candidate eccrine sweat markers, which suggests that an insufficient
fingerprint material was collected. Therefore, it cannot be determined
if the participant had taken INH as the sample volume is insufficient.
For donors FP030 and FP039, detection of AcINH in only some donated
fingerprints corresponded well with the detection of eccrine sweat
markers. NC004 was chosen at random from the negative control group
to demonstrate that eccrine sweat markers are not only associated
with INH use.
Figure 1
Average (n = 5, ±standard deviation)
peak
area of the extracted ion chromatogram for ions assigned to the protonated
molecular mass of (A) lysine (m/z 147.1134), (B) ornithine (m/z 133.0977),
(C) pyroglutamic acid (m/z 130.0504),
and (D) taurine (m/z 126.0225) for
a selection of fingerprint donors. FP037 and FP030 are examples of
donors with poor INH detection, FP033 and FP038 are examples of good
fingerprint donors, FP039 gives a mix of fingerprint depositions,
and NC004 is a patient selected at random from the negative control
group. Plot (E) shows the average (n = 5, ±standard
deviation) analyte to internal standard (A/IS) ratio measured for
INH and AcINH in the selected patient samples.
Average (n = 5, ±standard deviation)
peak
area of the extracted ion chromatogram for ions assigned to the protonated
molecular mass of (A) lysine (m/z 147.1134), (B) ornithine (m/z 133.0977),
(C) pyroglutamic acid (m/z 130.0504),
and (D) taurine (m/z 126.0225) for
a selection of fingerprint donors. FP037 and FP030 are examples of
donors with poor INH detection, FP033 and FP038 are examples of good
fingerprint donors, FP039 gives a mix of fingerprint depositions,
and NC004 is a patient selected at random from the negative control
group. Plot (E) shows the average (n = 5, ±standard
deviation) analyte to internal standard (A/IS) ratio measured for
INH and AcINH in the selected patient samples.The results in Figure provide evidence that detection of INH/AcINH typically coincides
with the detection of eccrine sweat metabolite markers. We propose
that these compounds can therefore be used as internal markers to
determine whether the fingerprint sample volume was sufficient for
INH testing to take place. The detection rate for INH use was recalculated
based on a requirement for the simultaneous detection of various sweat
metabolites and combinations thereof. Figure shows the performance of six candidate testing
protocols, under which a fingerprint sample should contain lysine
(A); ornithine (B); pyroglutamic acid (C); taurine (D); ≥2
metabolites (E); and ≥3 metabolites (F) to be considered a
valid sample, benchmarked against a test based on the detection of
INH/AcINH only (G). Test protocols (C) and (F) resulted in the elimination
of all false negatives, with <11% of samples being classed as invalid.
Figure 2
Number
of patient samples testing positive for INH use (blue, based
on INH or AcINH); negative for INH use (green, based on INH or AcINH);
and invalid test (yellow) based on the criteria of detection of one
or a combination of the selected four sweat metabolites.
Number
of patient samples testing positive for INH use (blue, based
on INH or AcINH); negative for INH use (green, based on INH or AcINH);
and invalid test (yellow) based on the criteria of detection of one
or a combination of the selected four sweat metabolites.The performance (sensitivity, precision, specificity, and
accuracy)
of fingerprint testing protocols C, F, and G is shown in Table . Detection based
on pyroglutamic acid (protocol C) or three amino acids (protocol F)
showed improved performance when compared to detection of INH/AcINH
alone (protocol G).
Table 2
Calculated Accuracy,
Precision, Sensitivity,
and Specificity for INH Use Using the Criteria of Detection of Pyroglutamic
Acid (Test Protocol C), Three Amino Acids (from Lysine, Ornithine,
Pyroglutamic Acid, and Taurine; Test Protocol F), or None (Test Protocol
G)
detection
based on pyroglutamic acid (C) (%)
detection
based on 3 metabolites (F) (%)
detection
based on INH and/or AcINH (G) (%)
accuracy
100
100
92
precision
0
0
8
sensitivity
100
100
100
specificity
100
100
77
In addition
to identifying samples for which insufficient fingerprint
material was collected, this work shows that the metabolite signals
can also be used for data normalization to control for the volume
of the material deposited. Figure S4 shows
the considerable intradonor variability in AcINH per fingerprint.
In Figure A, samples
in which the metabolite taurine was not detected are rejected to exclude
poor depositions. The AcINH signal was then normalized to taurine.
In Figure A, the coefficient
of variation (CV) of fingerprint samples deposited by the same donor
are plotted after internal standard normalization (blue) and taurine
normalization (green). Normalizing to taurine reduces the CV for every
donor.
Figure 3
(A) Comparison per donor of the CV of AcINH measurements from fingerprints
using only the internal standard for normalization (blue) and using
normalization to taurine (orange). (B) Box plots showing the distribution
(range, upper and lower quartiles, and median) of CVs obtained for
the different fingerprint sets using different normalization strategies.
(A) Comparison per donor of the CV of AcINH measurements from fingerprints
using only the internal standard for normalization (blue) and using
normalization to taurine (orange). (B) Box plots showing the distribution
(range, upper and lower quartiles, and median) of CVs obtained for
the different fingerprint sets using different normalization strategies.In Figure B, box
plots depicting the CVs of fingerprint sample sets are presented with
different normalization strategies. The box plot on the left depicts
the CVs of the fingerprint sample sets produced by normalizing only
to the internal standard and without any consideration for the quality
of the fingerprint. The middle box plot shows the same data, but now
rejecting samples for which taurine was not detected. The far-right
box plot shows the effect of additionally normalizing to taurine.
The median CV reduces from 65 to 55 and then to 26% using these normalization
strategies. Normalizing to the other endogenous compounds was found
to improve the CV with inconsistent results (Figure S5).
Discussion
This work has demonstrated how markers of
eccrine sweat can significantly
improve fingerprint drug testing by highlighting and rejecting samples
for which insufficient material has been deposited. This approach
resulted in sensitivity, specificity, and accuracy of 100% but resulted
in 11% of fingerprint samples (n = 15) being inadequate
for testing purposes. Of the 28 patients (n = 140
fingerprint samples), only two patients returned invalid tests for
all five collected fingerprints (FP037 and FP041). The right index
and right middle fingers were the only ones that did not return any
invalid results for any patient. Future work should explore increasing
the number of viable samples by exploring methods to enhance the deposition
of fingerprint residues, for example, by increasing the deposition
time and/or contact pressure or application of more sensitive analytical
methods.Finally, the ability to relate the level of the drug
deposited
in a fingerprint to the blood plasma level would provide much broader
opportunities for fingerprint-based diagnostics. Variation in the
mass of the drug or metabolite per fingerprint across a set of fingerprints
currently poses a barrier to any meaningful comparison of fingerprint
samples with any other biofluid. This work also showed that normalizing
AcINH to amino acid taurine can be valuable in reducing the intradonor
variability across a set of deposited fingerprints. This is a first
step toward providing a quantitative test from a fingerprint, and
future studies should explore whether the intradonor variability can
be reduced further through targeted analysis of taurine and the other
fingerprint biomarkers identified here. A method to standardize or
account for the mass of fingerprint deposition could enable quantitative
analysis with future applications, including the potential for reliable
monitoring of antibiotics in the fingerprints of patients undergoing
treatment for TB and other infectious diseases.
Conclusions
INH
and AcINH can be detected in natural fingerprints from patients
receiving TB treatment. Our findings show that a fingerprint test
for INH use should test for both INH and its metabolite AcINH, and
handwashing should be avoided immediately prior to testing. Additionally,
markers of eccrine sweat can be used to monitor whether sufficient
fingerprint has been collected and reduce intradonor variability.
The test protocol based on eccrine sweat markers developed to assure
the sample quality gave an accuracy of 100%.
Authors: Katharine E Schmitz; Melbourne F Hovell; Charlene A Wong; Norma J Kelley; Donata Nilsen; Elaine J Blumberg; Linda L Hill; Carol L Sipan; Bo Kolody; Dale A Chatfield Journal: Clin Nurs Res Date: 2010-05 Impact factor: 2.075
Authors: Simone H J van den Elsen; Onno W Akkerman; Mireille Wessels; Erwin M Jongedijk; Samiksha Ghimire; Tjip S van der Werf; Mathieu S Bolhuis; Daan J Touw; Jan-Willem C Alffenaar Journal: Eur Respir J Date: 2020-10-22 Impact factor: 16.671
Authors: Roy Gerona; Anita Wen; Aaron T Chin; Catherine A Koss; Peter Bacchetti; John Metcalfe; Monica Gandhi Journal: PLoS One Date: 2016-05-18 Impact factor: 3.240
Authors: Morgan Panitchpakdi; Kelly C Weldon; Alan K Jarmusch; Emily C Gentry; Arianna Choi; Yadira Sepulveda; Shaden Aguirre; Kunyang Sun; Jeremiah D Momper; Pieter C Dorrestein; Shirley M Tsunoda Journal: PLoS One Date: 2022-07-26 Impact factor: 3.752