Rajesh Madhuvilakku1,2, Yi-Kuang Yen1, Wei-Mon Yan2, Guang-Wei Huang1. 1. Department of Mechanical Engineering, National Taipei University of Technology, Taipei 106, Taiwan. 2. Department of Energy and Refrigeration Air-Conditioning Engineering, National Taipei University of Technology, Taipei 106, Taiwan.
Abstract
The analysis of pharmaceutical active ingredients plays an important role in quality control and clinical trials because they have a significant physiological effect on the human body even at low concentrations. Herein, a flexible three-electrode system using laser-scribed graphene (LSG) technology, which consists of Nafion/Fe3O4 nanohybrids immobilized on LSG as the working electrode and LSG counter and reference electrodes on a single polyimide film, is presented. A Nafion/Fe3O4/LSG electrode is constructed by drop coating a solution of Nafion/Fe3O4, which is electrostatically self-assembled between positively charged Fe3O4 and negatively charged Nafion on the LSG electrode and is used for the first time to determine a neurotoxicity drug (clioquinol; CQL) in biological samples. Owing to their porous 3D structure, an enriched surface area at the active edges and polar groups (OH, COOH, and -SO3H) in Nafion/Fe3O4/LSG electrodes resulted in excellent wettability to facilitate electrolyte diffusion, which gave ∼twofold enhancement in electrocatalytic activity over LSG electrodes. The experimental parameters affecting the analytical performance were investigated. The quantification of clioquinol on the Nafion/Fe3O4/LSG electrode surface was examined using differential pulse voltammetry and chronoamperometry techniques. The fabricated sensor displays preferable sensitivity (17.4 μA μM-1 cm-2), a wide linear range (1 nM to 100 μM), a very low detection limit (0.73 nM), and acceptable selectivity toward quantitative analysis of CQL. Furthermore, the reliability of the sensor was checked by CQL detection in spiked human blood serum and urine samples, and satisfactory recoveries were obtained.
The analysis of pharmaceutical active ingredients plays an important role in quality control and clinical trials because they have a significant physiological effect on the human body even at low concentrations. Herein, a flexible three-electrode system using laser-scribed graphene (LSG) technology, which consists of Nafion/Fe3O4 nanohybrids immobilized on LSG as the working electrode and LSG counter and reference electrodes on a single polyimide film, is presented. A Nafion/Fe3O4/LSG electrode is constructed by drop coating a solution of Nafion/Fe3O4, which is electrostatically self-assembled between positively charged Fe3O4 and negatively charged Nafion on the LSG electrode and is used for the first time to determine a neurotoxicity drug (clioquinol; CQL) in biological samples. Owing to their porous 3D structure, an enriched surface area at the active edges and polar groups (OH, COOH, and -SO3H) in Nafion/Fe3O4/LSG electrodes resulted in excellent wettability to facilitate electrolyte diffusion, which gave ∼twofold enhancement in electrocatalytic activity over LSG electrodes. The experimental parameters affecting the analytical performance were investigated. The quantification of clioquinol on the Nafion/Fe3O4/LSG electrode surface was examined using differential pulse voltammetry and chronoamperometry techniques. The fabricated sensor displays preferable sensitivity (17.4 μA μM-1 cm-2), a wide linear range (1 nM to 100 μM), a very low detection limit (0.73 nM), and acceptable selectivity toward quantitative analysis of CQL. Furthermore, the reliability of the sensor was checked by CQL detection in spiked human blood serum and urine samples, and satisfactory recoveries were obtained.
In
recent years, the analysis of drugs and their active ingredients
has become an integral part of drug quality control and clinical trials.
Clioquinol (5-chloro-7-iodo-8-hydroxyquinoline, CQL) is a halogenated
analogue of 8-hydroxyquinoline used to treat gastrointestinal amebiasis
and skin illness.[1] It can also serve as
a Zn2+ and Cu2+ chelator, and when combined
with cobalamin, it can treat Parkinson’s and Alzheimer’s
disorders.[2] An overdose or prolonged administration
of the CQL drug may cause subacute myelo-optic neuropathy (SMON),
which is a relatively uncommon neurological disorder. The SMON syndrome
is an iatrogenic disease in which nerve fibers in the spinal cord
and optic nerves degenerate. It was most prevalent in Japan. Several
symptoms are associated with SMON, including sensory losses, dysesthesia,
ataxic paralysis, impaired vision, and so forth.[3,4] Thus,
the use of CQL has been banned in Japan and some other countries.
Despite this, CQL is widely used as a therapeutic agent in several
countries. Thus, developing simple, rapid, and trustworthy protocols
for monitoring and detecting CQL is imperative.Several approaches
have been developed to analyze CQL, including
gravimetry,[5] titrimetry,[6] calorimetry,[7] chromatography,[8] spectrophotometry,[9] and fluorescence detection.[1,10] Despite their advantages
of sensitivity and accuracy, these techniques are not easy in practical
analysis because they are very expensive to operate, complex, employ
costly instruments, and involve lengthy sample preparation procedures
and complicated manipulations. In this concern, electroanalytical
methods are highly efficient, straightforward, sensitive, and selective.
As far as we know, most of the research is devoted to develop high-performance
electrode materials or improve the properties of electrochemical sensors
and catalytic performances.[3,4,11−15] Nevertheless, to test at point-of-care (POC), portable, reusable,
and miniaturized CQL sensors are still indispensable.Graphene
and graphitic materials are ideal candidates for electrochemical
sensors because they possess a wide electrochemical potential window
and demonstrate electro-catalytic activity in a variety of redox reactions.[16−20] Fabrication of graphene-based materials often relies on chemical
vapor deposition, self-reduction of graphene oxide, or other approaches
that are usually pricey and labor-intensive.[21−23] In contrast,
graphene electrodes, however, can be imprinted using printing techniques
(e.g., inkjet printing and screen printing) using
graphene flakes made from chemically exfoliating bulk graphitic materials.[24−27] However, these printing techniques often require additional post-printing
processes (e.g., laser, thermal, or photonic annealing)
to make the graphene electrodes more electrically conductive, which
further complicates their fabrication.[19,28,29] Furthermore, solution processing of graphene could
result in restacking and reduced surface areas.[30] Therefore, implementation of a new approach may ultimately
provide a sustainable alternative for mass production of graphene
with low toxicity and a favorable environment for biomolecules. Laser-scribed
graphene (LSG) is an efficient and scalable technique for generating
conductive 3D porous graphene through laser etching of thermoplastic
resins such as polyimide (PI).[31−33] Currently, conventional substrate
electrodes like glassy carbon/metal electrodes are being replaced
with LSG electrodes because of their flexibility, large active surface
area, good conductivity, ease of manufacture, and low cost.[23,34,35] Considering all these features,
LSG electrodes show great potential as disposable and portable biosensor
strips.To date, much attention has been focused on using metal/metal
oxide
nanoparticles (NPs) on carbon-based materials as a signal transducing
and sensing material to improve biosensing capability.[36,37] The incorporation of NPs into LSGs has resulted in hybrid materials
that exhibit improved physical and chemical properties. A number of
metal oxides, such as CO3O4, MoO2, and Fe3O4, can be grown on the surface of
laser-induced graphene (LIG), demonstrating the versatility of this
method for synthesizing metal oxide–LIG hybrid nanostructures.[38] Among the metal oxide NPs decorating LIG, Fe3O4 composed of Fe2+ and Fe3+ has been widely recognized as an effective electrocatalyst due to
its excellent electronic conductivity, exceptional catalytic activity,
rapid electron transfer ability, low cost, simplicity in synthesis,
biocompatibility, chemical stability, and ease of functionalization.[39] Nanohybridization is considered to be a convenient
method of preparing hybrid networks with desirable physical properties
by incorporating inorganic polymers into carbon nanomaterials while
preserving the intrinsic characteristics of carbon.[40] The ionic polymer Nafion has been widely utilized for the
preparation and modification of different electrochemical sensors.[41,42] It is well known that Nafion comprises mainly hydrophobic backbone
chains (−CF2 groups) and hydrophilic side chains
(−SO3– groups), which can functionalize
graphene in alcoholic solutions after it is added to graphene.[43] Furthermore, sulfonic groups prevent graphene
layers from stacking.[44] A hybrid nanostructure
made by combining Fe3O4 NPs, Nafion, and graphene-based
nanomaterials accelerates electron transfer processes at the electrode/electrolyte
interface, making them suitable for use in electrochemical biosensors.
As a result, Fe3O4 combined with LSG and Nafion
is expected to produce a remarkably effective electrocatalyst for
the sensitive determination of therapeutic agents. Recently, electrochemical
sensors based on LSG have been successfully applied for the detection
of small biomolecules, potential biomarkers, and neurotransmitters.[31,45−52] Nasraoui et al. described the fabrication of LIG
modified with functionalized multiwalled carbon nanotubes/AuNPs for
nitrite detection in water samples.[53] Zhang et al. developed LIG decorated with Cu NPs using electroless
deposition technique to fabricated glucose sensor strips to detect
low glucose levels in the serum samples.[54] Although, LIG electrode patterns with bimetals (Ni/Au) have been
leveraged for non-enzymatic glucose sensors.[55] Poly(3, 4-ethylenedioxythiophene) was electrochemically deposited
on LSG to fabricate flexible electrochemical sensors for the simultaneous
detection of DA, AA, and UA.[56] You’s
group synthesized Nobel metal NPs–LIG hybrid nanostructures
using laser induction, which were later used to fabricate a flexible
immunosensor to detect Escherichia coli O157:H7.[38] In addition, the operation
of supercapacitors on Fe3O4 NP-anchored LIG
(LIG/Fe3O4) electrodes was studied.[57−59] To our knowledge, no study has reported the detection of CQL using
LSGs modified with Nafion/Fe3O4.In line
with these perspectives, we proposed a Nafion/Fe3O4 nanostructured anchored LSG-based electrochemical sensor
strip for the detection of the neurotoxic drug clioquinol. To characterize
the physicochemical properties of these electrodes, various techniques
(analytical and spectroscopic) were used. The homogenous distribution
of Fe3O4 coupled with the high electrostatic
interaction of Nafion–LSG in a super-hydrophobic environment
provides a high surface-to-volume ratio that makes electrolyte access
to the electrode surfaces easier and allows for quick and sensitive
electrochemical detection.[60,61] Furthermore, the fabricated
sensor strip successfully quantitated CQL in urine and serum samples.
Experimental Section
Materials and Methods
The detailed
information about the materials, reagents, and apparatus is provided
in the Supporting Information.
Design and Fabrication of LSG Electrodes
To prepare
the LSG sensor, the electrode patterns for the LSG sensor
were first drawn by using computer software, and the PI films were
treated with alcohol and distilled water and dried at 60 °C before
they were laser-irradiated. A sensor strip was designed with three
electrodes (working, reference, and counter). The pre-designed patterns
were directly laser-engraved on PI films using a portable laser scribing
system (Universal Laser Systems PLS 6.75, laser peak power 75 W) with
a wavelength of 10.6 μm, and a pulse duration of ≈14
μs was used to irradiate a PI sheet to produce graphene electrodes
as the photothermal process leads to molecular rearrangements in air
to form a black carbonized layer, namely LSG.[62,63] Typical power values were between 2.4 and 5.4 W, with pulse frequency
(PPI) ranging from 1 to 1000. The PPI feature allows for controlling
the number of laser pulses per inch, emitted from the laser gun. Consequently,
setting PPI to 100% corresponds to a pulse rate of 1000 per inch.
Various experimental settings were tuned, including peak power (15%),
speed (10%), PPI (1000 mm), and Z-distance (5 mm).
Following the fabrication of the LSGEs on PI sheets, an Ag paste was
applied to the reference electrode and cured for 30 min at 60 °C.
A PDMS passivation layer was used to precisely define the working
electrode (3 mm in diameter), and the layer was cured for 30 min at
85 °C. Afterward, copper tape glued with silver paste was used
to connect the LSG electrodes externally for better conductivity.
Functionalization of LSG Electrodes
Scheme depicts the
fabrication and functionalization of Nafion/Fe3O4/LSG. A laser scribing process transforms orange-colored PI material
into black 3D carbonaceous materials shown in the schematic.[31,56,58] Three-electrode LSG sensors (working,
reference, and counter) were fabricated on a single PI substrate with
measurements of 2.5 cm (length) × 1.5 cm (width). PDMS passivation
was used to separate the working electrode (d = 3
mm) from the rest of the pattern.
Scheme 1
Schematic Illustration of the Preparation
of Nafion/Fe3O4 Nanohybrids, Fabrication of
LSG Electrodes Using Laser
scribing, and Subsequent Modification with Nafion/Fe3O4via Drop Casting to Produce Nafion/Fe3O4/LSG Electrodes
Preparation of Fe3O4 NPs
According to the literature, highly water-dispersible
Fe3O4 NPs were synthesized using the co-precipitation
method.[64] In the first step, 0.32 M FeCl3 and 0.31 M FeCl2 were mixed vigorously for 30
min in 0.45 M HCl solution. Second, the ferrous/ferric salt mixed
solution was added to 250 mL of 0.3 M NaOH solution and vigorously
stirred for 3 h at ambient temperature in a N2 atmosphere.
After 2 h, the obtained black colloidal Fe3O4 NPs were separated by centrifugation and washed with water. The
Fe3O4 NPs were then redispersed in 10 mL of
deionized water to obtain Fe3O4 NP stock solution
(a total concentration of 0.2 mg/mL) and stored at room temperature
for later use.
Fabrication of the Flexible
Nafion/Fe3O4/LSG-Modified Electrodes
The LSG working
electrode was immersed in a 450 μL solution of 1:1 (v/v) EDC
(0.5 M): NHS (0.1 M) for 30 min to activate the carboxylic group on
the LGS electrode surface. To make a Nafion/Fe3O4 suspension, 10 μL of Fe3O4 NP solution
was ultrasonically mixed with 10 μL of Nafion alcoholic solution
(5%) for 45 min. Following this, 8 μL of the Nafion/Fe3O4 suspension was drop-coated on the activated LSG electrode
surface, followed by 30 min of incubation to ensure effective adsorption.
This ultrasonication treatment produced free radicals in the Nafion
polymer chain; later, these free radicals could react with unsaturated
C=C groups in the LSG, resulting in a chemically bonded Nafion/Fe3O4/LSG electrode strip.[65] Finally, the electrodes were rinsed in phosphate-buffered saline
(PBS, pH 7.0) to remove any residual Nafion molecules, yielding the
desired Nafion/Fe3O4/LSG flexible electrode.
Later on, the electrodes were stored at 4 °C until they were
required again.
Results and Discussion
Surface Morphological Characterization
Scanning electron
microscopy (SEM) and transmission electron microscopy
(TEM) were used to examine the surface morphology of the fabricated
LSG, Fe3O4/LSG, and Nafion/Fe3O4/LSG electrodes. The SEM and TEM micrographs of the various
fabricated LSG strip electrodes are shown in Figure a–f, respectively. The SEM image of
LSG (Figure a) shows
a three-dimensional porous graphene flake structure with a large accessible
surface area, which is consistent with previous reports of LSG obtained
by laser irradiation of the PI substrate.[59,66] Similarly, the SEM image of the Fe3O4/LSG
electrode (Figure b) exhibits that cubical Fe3O4 NPs are well-covered
over the LSG flake surface skeleton, particularly at the flake’s
edge. Figure c shows
the typical SEM image of the Nafion/Fe3O4/LSG
electrode, which indicates that aggregated Nafion/Fe3O4 hybrid structures have compactly coated most of the LSG electrode
surface, confirming the adsorption of Nafion/Fe3O4 on LSG. A plausible mechanism for an aggregated system within the
LSG seems to impregnate Nafion-formed interfacial interactions (hydrogen
bonding) between the −COOH group of LSG and −SO3H of Nafion.[67] Also, TEM images
of the LSG electrode (Figure d) reveal few-layered graphene flakes with rumpled edges.
The rumpled formation is considered to be the result of thermal expansion
caused by laser irradiation.[58] Furthermore,
the selected area electron diffraction (SAED) pattern for LSG (Figure d inset) reveals
distinct poly-crystalline lattice fringes. TEM micrographs of Fe3O4/LSG (Figure e) show that Fe3O4 NPs (∼45
nm average size) are densely anchored over the LSG, and the SAED pattern
confirms that the Fe3O4 has poly-crystalline
inverse cubic spinel composition (Figure e-insert). Figure f presents a typical TEM micrograph of Nafion/Fe3O4/LSG electrodes, in which Nafion/Fe3O4 interacts incoherently with the LSG matrix, resulting
in bigger particles with an average size of 80–120 nm. The
inset figure in Figure f displays that the SAED pattern exhibits well-defined diffraction
rings, and highly crystalline dots confirm the effective adsorption
of the Nafion/Fe3O4 hybrids on the LSG. Besides,
the EDX (energy-dispersive X-ray) spectra and elemental mapping reveal
that the nitrogen and oxygen elements are embedded uniformly in the
LSG electrode (Figure g,j). Similarly, the carbon, oxygen, and iron elements are evenly
distributed on the Fe3O4/LSG electrode (Figure h,k). Furthermore,
the EDX spectrum and elemental mapping images of Nafion/Fe3O4/LSG (Figure i,l) validate the presence of C, O, Fe, S, and F in its structure.
It also implies that the simple ultrasonication method used for the
preparation of hybrid nanocomposites is effective.
Figure 1
(a–f) SEM (scale
bar for 10 μm) and TEM (scale bar
of 200 nm; the inset shows the SEAD pattern) images of LSG, Fe3O4/LSG, and Nafion/Fe3O4/LSG
electrodes, (g–l) EDX spectra and elemental mapping images
of LSG, Fe3O4/LSG, and Nafion/Fe3O4/LSG electrodes.
(a–f) SEM (scale
bar for 10 μm) and TEM (scale bar
of 200 nm; the inset shows the SEAD pattern) images of LSG, Fe3O4/LSG, and Nafion/Fe3O4/LSG
electrodes, (g–l) EDX spectra and elemental mapping images
of LSG, Fe3O4/LSG, and Nafion/Fe3O4/LSG electrodes.In order to determine the accessibility of water-based electrolytes
into the electrode, measurements of contact angles (wetting angle)
were made on the electrode strip. Thus, we assessed the wettability
of LSG, Fe3O4/LSG, and Nafion/Fe3O4/LSG by measuring the fixed water contact angle. Figure S1 shows the image of a water drop on
the surface of LSG, Fe3O4/LSG, and Nafion/Fe3O4/LSG electrodes, and each electrode showed different
wetting characteristics. The pristine LSG showed a higher degree of
contact angles (90°) that could be attributed to its hydrophobic
property. Fe3O4/LSG exhibited a considerably
smaller contact angle of 68°, suggesting that micro–nanostructures
could be incorporated into 3D network frameworks to increase the hydrophobic
properties.[68] More importantly, the Nafion/Fe3O4/LSG strip electrode possesses an excellent wettability
behavior with substantially reduced contact angles to 12° due
to their polar groups (OH, COOH, and −SO3H), which
facilitates electrolyte diffusion.[67]
Structural and Composition Characterization
The X-ray diffraction (XRD) was used to investigate the crystalline
structure of LSG, Fe3O4/LSG, and Nafion/Fe3O4/LSG, and the resulting diffraction patterns
are shown in Figure a
Figure 2
(a) XRD patterns for LSG, Fe3O4/LSG and Nafion/Fe3O4/LSG. (b) Raman spectra for LSG, Fe3O4/LSG, and Nafion/Fe3O4/LSG electrodes.
(a) XRD patterns for LSG, Fe3O4/LSG and Nafion/Fe3O4/LSG. (b) Raman spectra for LSG, Fe3O4/LSG, and Nafion/Fe3O4/LSG electrodes.The XRD pattern of LSG, Fe3O4/LSG, and Nafion/Fe3O4/LSG sensor strips exhibits
a broad, intense
diffraction peak centered at 2θ = 25.5° corresponding to
the (002) plane, giving an interlayer spacing of 0.38 nm between (002)
planes, indicating a high degree of graphitization. Additionally,
the broadening depicts a c-axis periodicity of the
complex 3D graphitic structure.[31] Furthermore,
2θ = 42.9 corresponds to a peak resulting from reflections (100)
associated with in-plane structures. For Fe3O4/LSG and Nafion/Fe3O4/LSG, the major peaks
observed at 2θ = 35.48, 37.11, 42.27, 43.12, 53.5, 57.0, 62.62,
and 71.05 are assigned to the (311), (222), (400), (311), (422), (511),
(440), and (620) facets, respectively, typically corresponding to
the simple cubic (sc) structure of Fe3O4 (JCPDS
no. 75-0033) along with (002) and (100) planes,[39] which ensures the successful anchoring of Fe3O4 on the surface of LSG.The Raman spectra of LSG,
Fe3O4/LSG, and
Nafion/Fe3O4/LSG are shown in Figure b. As shown in Figure b, there are three prominent
peaks observed. The peaks located at 1330 cm–1 (D
band) and 1550 cm–1 (G band) correspond to A1g vibrational modes of the sp3-hybridized disordered
carbon lattice and the E2g vibration mode in sp2-hybridized ordered carbon lattice, respectively.[69] A 2D peak of the LSG can be observed at 2700 cm–1. The 2D profile is evidence for the formation of a randomly stacked
layer of graphene.[70] Additionally, the
D band and G band positions and width of the Fe3O4/LSG and Nafion/Fe3O4/LSG are slightly different
from LSG as a result of larger structural defects and small graphene
domains. ID/IG values of LSG, Fe3O4/LSG, and Nafion/Fe3O4/LSG are 0.82, 0.89, and 0.96, respectively,
which indicate that the −SO3H organic moiety has
successfully functionalized LSG.[67] Additionally,
the characteristic Raman mode of Fe3O4 can be
detected at 616 cm–1, corresponding to the A1g mode of Fe–O vibration, confirming that the composite
structure of Nafion/Fe3O4 is anchored on the
LSG electrode surface.[58]We further
characterized the functionalization of LSG with Nafion/Fe3O4 using X-ray photoelectron spectroscopy (XPS)
spectra. Nafion/Fe3O4 bonded into LSG is evident
from signals for C, F, S, Fe, and O in the wide-scan spectra (Figure a). The C1s core-level
spectrum (Figure b)
shows the characteristic peaks of C=C (284.3 eV), C–O–C
(285.2 eV), C–F (288.7 eV), and C–F2 (291.3
eV), which correspond with Nafion bonded on LSG.61Figure c illustrates that
the peaks at 161.7 and 163.1 eV belong to the −SO3– side chain within Nafion, and the C–S=O
configuration verifies that the sulfonate groups of Nafion are in
strong contact with LSG.[71]Figure d shows two peaks at 530.4
and 531.9 eV attributed to the C–O and C=O surface oxygen
complex. Fe2p spectra in Figure e demonstrate two intense peaks at 710.2 and 722.3
eV corresponding to the binding energies of Fe2p3/2 and
Fe2p1/2 bonds, respectively, resulting in a slight chemical
shift difference between Fe2+ and Fe3+ in Fe3O4. In addition, the occurrence of a satellite
peak between these two peaks centered at about 712.7 and 717.5 eV
can be ascribed to the formation Fe3+ and Fe2+.[72] N2 adsorption/desorption
was used to determine the textural properties of the samples. The
results are illustrated in Figure S2. It
is noted that both LSG and Nafion/Fe3O4/LSG
samples have isothermal shapes that fall into both type I and type
IV (according to IUPAC classification), indicating a mesoporous and
microporous state, respectively.[73]Figure S2 (inset) shows pore size distribution
curves, demonstrating that the pores in the synthesized LSG materials
are less than 5 nm in size. BET surface areas of 10.65 and 15.36 m2/g with pore volumes of 0.910 and 0.266 cm3 g–1 were observed for LSG and Nafion/Fe3O4/LSG, respectively. Therefore, it can be concluded that after
the immobilization of Nafion/Fe3O4 in the LSG
matrix, the surface area of the material would increase due to superior
physical properties.
Figure 3
(a) XPS survey spectra of Nafion/Fe3O4/LSG
electrodes and their (b) C1s, (c) S2p, (d) O1s, and (e) Fe2p core
level spectra (inset: high-resolution spectra of Fe2p3/2).
(a) XPS survey spectra of Nafion/Fe3O4/LSG
electrodes and their (b) C1s, (c) S2p, (d) O1s, and (e) Fe2p core
level spectra (inset: high-resolution spectra of Fe2p3/2).
Electrochemical
Characterization of the Electrode
Strip
To prove the electrochemical characteristics of the
fabricated LSG, Fe3O4/LSG, Nafion/LSG, and Nafion/Fe3O4/LSG electrode strips, cyclic voltammetry (CV)
performance acquired in 0.1 M KCl containing 1.0 mM Fe(CN)63–/4– as a redox couple at a scan rate of
100 mV/s is shown in Figure a. A pair of well-defined redox peaks with a peak-to-peak
separation (ΔEp) of 80 mV (ΔEp = Epa – Epc, where Epa and Epc are the anodic and cathodic peak potentials,
respectively) was observed for a pristine LSG electrode, which incorporated
a quasi-reversible electron transfer of the Fe3+/Fe2+ redox couple. Despite Fe3O4 NPs being
embedded in the LSG surface, the redox peak current intensity was
slightly increased Ipa = 18.37 μA
(0.203 V) and Ipc = −21.72 μA
(0.062 V) than that of pristine LSG. As compared to the LSG electrode,
Fe3O4 NP-embedded LSG would provide an enhanced
activity surface area and a higher electrocatalytic effect, which
would lead to a higher current response. In contrast, when the LSG
electrode was modified with a Nafion membrane, the redox peaks were
severely suppressed (Ipa = 3.28 μA
and Ipc = −6.05 μA) with
a peak separation of 141 mV. This is because Nafion can act as a mass
transfer blocking layer for the diffusion of [Fe(CN)6]3–/4– into the electrode surface and hinder the
electron transfer. After modification with Nafion/Fe3O4/LSG, ion exchange occurred between Fe3O4/LSG and the Nafion membrane, resulting in an increase in the redox
peak current [∼twofold higher than that observed with LSG in
the potassium ferricyanide solution (Ipa = 27.17 μA and Ipc = −27.73
μA)], indicating that Nafion/Fe3O4/LSG
can transfer electrons effectively.[74,75] These results
suggest that the Fe3O4 NPs play a critical role
in facilitating electron transfer. On one side, the micro–nanostructure
confers an accessible active site. Additionally, surface-anchored
Nafion/Fe3O4 can enhance electron transfer across
electrodes and electrolytes due to their excellent wetting characteristics
and superior catalytic activity.
Figure 4
(a) CVs of LSG, Fe3O4/LSG, Nafion/LSG, and
Nafion/Fe3O4/LSG electrodes in 1.0 mM Fe(CN)63–/4– and 0.1 M KCl as supporting
electrolyte solutions at the 100 mV/s scan rate. (b) CV behavior of
Nafion/Fe3O4/LSG in 1.0 mM Fe(CN)63–/4– in 0.1 M KCl at different scan rates
(10–100 mV/s). (c) Peak current vs square
root of scan rate calibration plots. (d) EIS spectra of LSG, Fe3O4/LSG, Nafion/LSG, and Nafion/Fe3O4/LSG electrodes measured in the frequency range from 100 kHz
to 1 Hz in 1.0 mM Fe(CN)63–/4– and 0.1 M
KCl electrolyte (inset: Randles-equivalent circuit model).
(a) CVs of LSG, Fe3O4/LSG, Nafion/LSG, and
Nafion/Fe3O4/LSG electrodes in 1.0 mM Fe(CN)63–/4– and 0.1 M KCl as supporting
electrolyte solutions at the 100 mV/s scan rate. (b) CV behavior of
Nafion/Fe3O4/LSG in 1.0 mM Fe(CN)63–/4– in 0.1 M KCl at different scan rates
(10–100 mV/s). (c) Peak current vs square
root of scan rate calibration plots. (d) EIS spectra of LSG, Fe3O4/LSG, Nafion/LSG, and Nafion/Fe3O4/LSG electrodes measured in the frequency range from 100 kHz
to 1 Hz in 1.0 mM Fe(CN)63–/4– and 0.1 M
KCl electrolyte (inset: Randles-equivalent circuit model).The electrochemical activity of Nafion/Fe3O4/LSG was investigated further using CV in 0.1 M KCl containing
1.0
mM ferro/ferri redox mediators at various scan rates within the range
10–100 mV/s, as shown in Figure b. The results showed that as the scan rate increased,
redox peak currents and peak-to-peak separation potential increased,
implying an interfacial behavior with quasi-reversible electron transfer
kinetics. Figure c
shows that Ipa and Ipc were linearly related to the square root of the scan rate.
In accordance with the Randles–Sevick equation (eq ), we calculated the active surface
area of surface-functionalized electrodes.where D refers to the typical
diffusion coefficient of ferricyanide solution (0.76 × 10–5 cm2 s–1), Ipa, ν, A, C and n refer to the anodic peak current, scan rate (V s–1), active surface area, the concentration of the ferricyanide in
bulk solution, and the number of electron transfers (n = 1). Accordingly, the active surface area of LSG, Fe3O4/LSG, and Nafion/Fe3O4/LSG is
calculated to be 0.129, 0.224, and 0.379 cm2, respectively.The electron transfer characteristics and interfacial properties
of surface-modified LSG electrodes (LSG, Fe3O4/LSG, Nafion/LSG, and Nafion/Fe3O4/LSG) were
investigated using the electrochemical impedance spectroscopy (EIS)
technique. EIS experiments were carried out in a 1.0 mM ferro/ferricyanide
redox system containing 0.1 M KCl at an open-circuit potential ranging
from 100 kHz to 1 Hz, and the results are shown in the Nyquist plot
(Figure d). In addition,
the data were fitted to an equivalent circuit depicted in the inset
of Figure d, where W, C, Rs, Qdl, and Rct represent
Warburg impedance, capacitance, electrolyte solution resistance, double-layer
capacitance, and charge transfer resistance, respectively. The electron
transfer resistance is proportional to the semicircular diameter of
the impedance curve (Rct).[23] The semicircle domain is found in the bare LSG,
indicating a high electron transfer resistance (978.4 Ω). The Rct was reduced to 728.3 Ω after modification
with Fe3O4, which was attributed to Fe3O4’s excellent conductivity. By modifying the LSG
electrode with Nafion membranes, the semicircle diameter was significantly
increased (Rct = 1435.4 Ω). This
is due to the fact that the Nafion membrane acts as an electron barrier,
introducing resistance into the electrode/electrolyte system, interfering
with electron transfer. Besides, the Rct of Nafion/Fe3O4/LSG declined obviously (Rct = 453.3 Ω) when compared to that of
LSG, indicating a very good electron-conducting composite. The EIS
data were found to be consistent with the CV results (Figure a). These observations revealed
that the Fe3O4 NPs can help facilitate electron
transport, whereas the negatively charged group (SO3–) present in Nafion provides stronger resistance to
anions and bulky molecules, allowing the electrode to more effectively
preconcentrate the target cation.
Electrochemical
Oxidation of CQL on the Nafion/Fe3O4/LSG Strip
Electrode
The electrochemical
characteristics of various fabricated strip electrodes toward the
oxidation of CQL were evaluated by using CV at a scan rate of 100
mV/s in the absence (dotted line) and presence (solid line) of 400
nM CQL in 0.1 M NaOH (Figure a). Interestingly, in the absence of CQL, all fabricated electrodes
did not show any anodic peak current response. In contrast, in the
presence of CQL, a sharp and well-defined anodic peak current for
CQL oxidation was obtained for all electrodes. There is a significant
oxidation peak observed in the CV of bare LSG, corresponding to a
peak current of 11.31 μA at 0.42 V. The LSG modified with Fe3O4 exhibits improved electrocatalytic performance,
with the lowest peak potential at 0.41 V and a higher anodic current
intensity of 13.5 A. The outstanding catalytic activity and high electrical
conductivity of Fe3O4 are responsible for this
significant increase in current. In addition, the Nafion/Fe3O4/LSG CV curve displays a clearly pronounced and enhanced
oxidation response with a peak intensity of 20.14 μA at 0.42
V. It is noteworthy that the anodic peak current of CQL at Nafion/Fe3O4/LSG is two times greater than that at bare LSG.
The Nafion and Fe3O4 composite combined showed
a clear synergistic effect that transduced electron transfer and intensified
specific surface areas at the electrode interface, thus enhancing
the current response signal.
Figure 5
(a) Cyclic voltammograms obtained for absence
(dotted line) and
presence (solid line) of 400 nM CQL in 0.1 M NaOH at LSG, Fe3O4/LSG, and Nafion/Fe3O4/LSG electrodes.
(b) CV responses of various scan rates (10–100 mV/s) at Nafion/Fe3O4/LSG electrodes with the presence of 400 nM CQL
in 0.1 M NaOH. (c) Linear plot of oxidation peak current vs scan rates. (d) Relationship of peak potential vs scan rates. (e) Proposed electro-oxidation mechanism for CQL on
Nafion/Fe3O4/LSG electrodes.
(a) Cyclic voltammograms obtained for absence
(dotted line) and
presence (solid line) of 400 nM CQL in 0.1 M NaOH at LSG, Fe3O4/LSG, and Nafion/Fe3O4/LSG electrodes.
(b) CV responses of various scan rates (10–100 mV/s) at Nafion/Fe3O4/LSG electrodes with the presence of 400 nM CQL
in 0.1 M NaOH. (c) Linear plot of oxidation peak current vs scan rates. (d) Relationship of peak potential vs scan rates. (e) Proposed electro-oxidation mechanism for CQL on
Nafion/Fe3O4/LSG electrodes.
Influence of the Scan Rate
The effect
of the scan rate (10–100 mV/s) was investigated on the electrochemical
oxidation of CQL (400 nM) at the Nafion/Fe3O4/LSG-modified electrode (Figure b). In the examined ranges, the oxidation peak current
for CQL was directly related to the scan rate. Furthermore, it is
observed that a good linear relationship is observed when plotting
the current against the scan rate (Figure c). Ipa (μA)
= 0.163 [mV/s] + 5.65 was found to be the linear regression equation,
with a correlation coefficient of R2 =
0.993, suggesting a diffusion-controlled electrode reaction of CQL
at the Nafion/Fe3O4/LSG electrodes.[76,77]Epa and υ relationship can be
defined using Laviron’s equation (eq ) in the context of a diffusion-controlled
electrochemical irreversible process.[60]Based on the Epa versus υ relationship shown in Figure d, the value of E0′
is anticipated to be 0.407 V by prolonging the curve to υ =
0. The electron transfer coefficient (α) and electron transfer
rate constant (K0) were calculated from
the slope of the Epa versus ln υ
plot to be 0.40 and 0.618 cm/s, respectively. This K0 value is twofold higher than that of pristine LSG (0.232
cm/s), indicating rapid electron transfer at the developed interface.The number of electrons (n) consumed in the electro-chemical
oxidation of CQL on Nafion/Fe3O4/LSG was calculated
from the equation (eq ) for irreversible reactions[78]where α is the charge transfer coefficient, n is the number of electrons consumed in the oxidation process,
and Ep/2 is potential at the half-peak
current. Thus, the number of electrons consumed in the process could
be calculated to be 0.94 (∼1 electron). A voltammetric electrochemical
mechanism for the electro-oxidation of CQL at Nafion/Fe3O4/LSG in an alkaline solution is represented in Figure e. Initially, the
CQL was oxidized by one electron and one proton, resulting in phenoxy
radicals, followed by an addition reaction with OH that generated
unstable semiquinone anion radicals. As a result of the fast nature
of the addition reaction, it was not possible to detect the cathodic
voltammetry peak associated with the reduction of the phenoxy radical.
Finally, the semiquinone intermediate underwent a disproportionation
reaction to generate CQL and a p-quinone intermediate.[3,4,11,79]
Influence of Concentration
Figure S3a shows the SWV (square-wave voltammetry)
response of CQL (10–100 nM) on Nafion/Fe3O4/LSG with consecutive injection of 10 nM CQL to 0.1 M NaOH at a scan
rate of 20 mV/s. The amplitude of the anodic current response peaks
with an increase in CQL from 10 to 100 nM would suggest that the sensor
could recognize the CQL without fouling. Anodic peak currents and
CQL concentrations are linearly related, as shown in Figure S3b. The good correlation coefficient (R2) 0.996 and the linear regression equation Ipa (μA) = 0.276 [CQL (nM)] + 8.626 imply that CQL
oxidation is controlled, which is consistent with an earlier report.[2−4] Furthermore, the plot of the logarithmic peak current against the
concentration of CQL (Figure S3c) is a
straight line and corresponds to a regression equation of Ipa (μA) = 0.455 [CQL (nM)] + 0.536 (R2 = 0.992). Based on these findings, the catalytic
reaction in the Nafion/Fe3O4/LSG sensor is driven
by first-order kinetics.
Effect of Modifier Amount
The volume
of nanomaterials immobilized on the LSG electrode is crucial because
it affects the electrode’s sensitivity significantly. To investigate
the impact of loading-level optimization, different amounts of Nafion/Fe3O4 ranging from 2 to 12 μL were used to modify
LSG to fabricate the sensor, and their sensing ability was tested
toward 400 nM CQL drug detection (Figure S3d). It is clear that increasing the amount of catalyst loading up
to 8 μL increased the oxidation peak current density due to
an increase in recognition sites on the electrode surface. Furthermore,
increasing the catalyst loading on the electrode above the optimum
value reduces the peak current density response. These findings indicated
that using more than 8 μL results in a thicker modifier layer,
which has a negative impact on interfacial electrode transport between
solution and electrode surface. Therefore, 8 μL of drop-coated
Nafion/Fe3O4/LSG was used for further electrochemical
experiments.
Effect of Applied Potential
Working
potential is yet another important factor that influences the electrochemical
response of sensors. To optimize the working potential, the influence
of applied potential was initially explored from the amperometric
response of CQL on the Nafion/Fe3O4/LSG electrode
in the range of 0.15–0.30 V, and the corresponding amperometric
responses upon successive injection of 0.01 μM CQL in 0.1 M
NaOH solution under stirred circumstances are recorded in Figure S3e. As can be seen, as the applied potential
increases from 0.15 to 0.30 V, the anodic current step increases.
When compared to 0.25 V, the current response at 0.30 V decreases
significantly. An ideal applied potential for amperometric detection
of CQL was chosen based on the most intense current response for the
oxidation of CQL at 0.25 V for subsequent studies.
Quantitative Assay of CQL
The quantitative
assay of CQL was investigated by recording the differential pulse
voltammetry (DPV) and chronoamperometry (CA) response of various concentrations
of CQL at Nafion/Fe3O4/LSG. Figure a shows the DPV performance
of Nafion/Fe3O4/LSG as a function of various
concentrations of CQL (from 0.001 to 100 μM) in a potential
window of 0–0.6 V. The oxidation currents rose directly in
proportion to the CQL concentration. Calibration curves for CQL concentrations
versus oxidation peak currents are linear (Figure b). As illustrated in Figure b, two linear ranges (0.001–10 μM
and 10–100 μM) for the oxidation of CQL can be observed
at the Nafion/Fe3O4/LSG. The corresponding linear
equations are I (μA) = 0.929 [CQL (μM)]
+ 17.4 (R2 = 0.990) and I (μA) = 0.050 [CQL (μM)] + 13.1 (R2 = 0.991). Based on the standard deviation of the control
sample (RSD) and the slope of the fitted calibration plots [LOD =
3 (RSD/S)], we estimated the limit of detection (LOD) to be 0.73 nM.
Figure 6
(a) DPV
signals obtained for successive additions of CQL (0.001–100
μM) in 0.1 M NaOH at 50 mV/s on Nafion/Fe3O4/LSG. Inset: magnified view of the lower CQL concentration vs oxidation peak currents. (b) Linear calibration plots
of oxidation peak currents vs the CQL concentrations.
Inset: Linearity plot for lower CQL concentration vs peak current. The error bars represent ± SD and n = 3 samples, with a maximum RSD of 2.41%. (c) CA responses of the
Nafion/Fe3O4/LSG sensor in 0.1 M NaOH with successive
additions of CQL at 0.25 V applied potential. Inset: the amplified i–t curve for CQL at low concentrations.
(d) Calibration curves for current vs concentration
of CQL. Inset: expanded ranges of selected concentrations. The error
bars represent ±SD and n = 3 samples, with a
maximum RSD of 2.68%.
(a) DPV
signals obtained for successive additions of CQL (0.001–100
μM) in 0.1 M NaOH at 50 mV/s on Nafion/Fe3O4/LSG. Inset: magnified view of the lower CQL concentration vs oxidation peak currents. (b) Linear calibration plots
of oxidation peak currents vs the CQL concentrations.
Inset: Linearity plot for lower CQL concentration vs peak current. The error bars represent ± SD and n = 3 samples, with a maximum RSD of 2.41%. (c) CA responses of the
Nafion/Fe3O4/LSG sensor in 0.1 M NaOH with successive
additions of CQL at 0.25 V applied potential. Inset: the amplified i–t curve for CQL at low concentrations.
(d) Calibration curves for current vs concentration
of CQL. Inset: expanded ranges of selected concentrations. The error
bars represent ±SD and n = 3 samples, with a
maximum RSD of 2.68%.In addition, the amperometric
(i–t) method
was adopted for the quantitative analysis of CQL because it possesses
good selectivity, excellent sensitivity, wide linear range, trace
level threshold, and minimal electrode passivation. Figure c shows the amperometric response
(i–t curve) of Nafion/Fe3O4/LSG to successive injections of CQL into 0.1
M NaOH solution at an optimized applied potential (0.25 V). Each addition
of CQL results in well-defined and steady amperometric responses.
As shown in Figure c, response currents progressively increase as CQL concentration
increases from 0.001 to 10 μM. As CQL concentrations increase
from 0.001 to 10 μm at 0.25 V, it is revealed that the sensor
produces a larger response current, and significant background noise
causes an aggrandized step current. This behavior is consistent with
earlier reports.[3,36] Similarly, two dynamic linear
ranges of CQL from 0.001 to 1.0 and 1.0 to 10 μM (Figure d), the linear regression equation
could be expressed as Ip (μA) =
0.212 [CQL (μM)] + 6.60 (R2 = 0.998)
and Ip (μA) = −0.176 [CQL
(μM)] + 6.96 (R2 = 0.993), respectively.
Furthermore, based on the abovementioned formula, the LOD corresponding
to three times the average blank value’s RSD (relative standard
deviation) was calculated to be 0.73 nM, and sensitivity was predicted
to be 17.4 μA μM–1 cm–2. The electro-catalytic performance of our sensor was superior or
comparable to that of other modified electrodes reported previously,
as shown in Table . Based on the comparison table, the fabricated sensor offers commendable
performance for detecting CQL in simple processing, high stability,
and low detection limits. Moreover, these different sensor platforms
for CQL quantification typically used macroelectrodes [glassy carbon
electrode (GCE)[11,12,36] and screen-printed carbon electrode (SPCE)][3,4] that
were decorated with graphene/metal or metal oxide NPs to improve sensitivity.
Most of the preparation processes for these similar nanostructures
are complicated due to their multistep fabrication process with less
controlled repeatability, and the high preparation costs involved.
Although they cannot be used for the POC testing due to the larger
size of the electrochemical system consisting of external reference
and counter electrodes.
Table 1
Analytical Comparison
of the Proposed
Method and Some of the Previously Reported Electrochemical CQL Biosensors
modified
electrode
technique
linear range
(μM)
LOD (nM)
references
aBSA-AuNC-Cu2+
fluorescence
1–12
630
(1)
Sn(MoO4)2/bGCE
DPV
0.05–234
14
(2)
MCO/S-rGO/cSPCE
i–t
0.001–84
0.9
(3)
CeVO4/SPCE
DPV
0.02–215
4.0
(4)
dSQD
fluorescence
0.024–30
15
(10)
GO@LaVO4–NCs/GCE
DPV
0.025–438.25
2.44
(11)
N-eCQD@Gd2O3/GCE
i–t
0.3–220
2.1
(36)
Nafion/Fe3O4/fLSG
DPV, i–t
0.001–100
0.73
this work
Bovine serum albumin–gold
nanoclusters quenched by Cu2+.
In the analysis
of real samples, selectivity plays a key role. Therefore, we investigated
the selectivity of our fabricated sensor by measuring its response
(100 nM) via DPV by introducing several interferences,
as shown in Figure a and b. The results illustrated that the presence of 100-fold excess
of various common metal ions (Zn2+, Mg2+, Cu2+, Pb2+, Cd2+, K+, and Na+) and some biologically active molecules (dopamine, uric acid,
ascorbic acid, glucose, uric acid, glycine, hydroquinone, and caffeine)
in 50-fold excess, were not significantly affected. Furthermore, 20-fold
excess concentration of some pharmaceutical drugs (flutamide, metronidazole,
acetaminophen, and amoxicillin) also exhibited negligible (<5%)
interference with the CQL test. Therefore, Nafin/Fe3O4/LSG electrodes can be used in electrochemical sensors or
pharmaceutical formulations to detect CQL in a selective way.
Figure 7
(a) DPV response
of CQL oxidation in the presence of interfering
species with 100 times excess of common metal ions (Zn2+, Mg2+, Cu2+, Pb2+, Cd2+, K+, and Na+); a 50-fold excess of physiological
interference (dopamine, uric acid, ascorbic acid, glucose, uric acid,
glycine, hydroquinone, and caffeine); and a 20-fold excess of widely
prescribed drugs (flutamide, metronidazole, acetaminophen, and amoxicillin).
(b) Percentage of interfering-species current signals compared to
CQL. (c) CV response of the Nafion/Fe3O4/LSG
sensor in 0.1 M NaOH containing 400 nM CQL at different time intervals.
(d) Steady-state response observed at Nafion/Fe3O4/LSG for 100 nM CQL in 0.1 M NaOH up to 3000 s. The applied potential
is +0.25 V. (e) Reproducibility of Nafion/Fe3O4/LSG for six separate replicates and (f) repeatability of Nafion/Fe3O4/LSG for nine repeated measurements at a single
electrode in 400 nM CQL in 0.1 M NaOH at 100 mV/s. (g) Typical CV
curves for Nafion/Fe3O4/LSG for 400 nM CQL at
50 mV/s in 0.1 M NaOH as supporting electrolyte in flat, convex, and
concave bending modes. Inset: Schematic representation of Nafion/Fe3O4/LSG in flat, convex, and concave bending stress
modes, respectively. (h,i) SWVs recorded at Nafion/Fe3O4/LSG sensors after sequential addition of CQL-spiked human
blood serum (10–100 nM) and human urine samples (5–100
nM) into 0.1 M NaOH at 50 mV/s.
(a) DPV response
of CQL oxidation in the presence of interfering
species with 100 times excess of common metal ions (Zn2+, Mg2+, Cu2+, Pb2+, Cd2+, K+, and Na+); a 50-fold excess of physiological
interference (dopamine, uric acid, ascorbic acid, glucose, uric acid,
glycine, hydroquinone, and caffeine); and a 20-fold excess of widely
prescribed drugs (flutamide, metronidazole, acetaminophen, and amoxicillin).
(b) Percentage of interfering-species current signals compared to
CQL. (c) CV response of the Nafion/Fe3O4/LSG
sensor in 0.1 M NaOH containing 400 nM CQL at different time intervals.
(d) Steady-state response observed at Nafion/Fe3O4/LSG for 100 nM CQL in 0.1 M NaOH up to 3000 s. The applied potential
is +0.25 V. (e) Reproducibility of Nafion/Fe3O4/LSG for six separate replicates and (f) repeatability of Nafion/Fe3O4/LSG for nine repeated measurements at a single
electrode in 400 nM CQL in 0.1 M NaOH at 100 mV/s. (g) Typical CV
curves for Nafion/Fe3O4/LSG for 400 nM CQL at
50 mV/s in 0.1 M NaOH as supporting electrolyte in flat, convex, and
concave bending modes. Inset: Schematic representation of Nafion/Fe3O4/LSG in flat, convex, and concave bending stress
modes, respectively. (h,i) SWVs recorded at Nafion/Fe3O4/LSG sensors after sequential addition of CQL-spiked human
blood serum (10–100 nM) and human urine samples (5–100
nM) into 0.1 M NaOH at 50 mV/s.
Stability, Reliability, and Repeatability
Measurements
In order to make the developed sensor feasible
and applicable, the stability and reproducibility are of paramount
importance. The stability of the Nafion/Fe3O4/LSG sensor was assessed by storing the sensor for 15 days under
ambient conditions and then performing voltammetry measurements in
400 nM CQL solution (Figure c). The RSD of the sensor was 3.46%, indicating that it is
very stable over time. In addition, another method for examining the
stability of the sensor response toward CQL oxidation was evaluated
by amperometric technique with CQL at 100 nM for 3000 s (Figure d), implying that
the strip electrode exhibited a long-term electrochemical stability
and had no surface fouling. Reliability of the Nafion/Fe3O4/LSG sensor was assessed through repeated voltammetric
measurements on six replicate sensors (Figure e). They had good reproducibility with an
RSD of 2.37% for 400 nM CQL solution. Furthermore, a series of nine
CV measurements using the same Nafion/Fe3O4/LSG
electrode in 0.1 M NaOH containing 400 nM CQL (Figure f) revealed an RSD value of 2.51%, indicating
that the modified electrode has good reproducibility.
Versatility Test
To determine the
impact of bending stress on electrochemical reactions in sensors,
CV testing was used to examine the rigidity of Nafion/Fe3O4/LSG electrodes. Hence, the electrochemical properties
of the flexible device were evaluated under flat, convex, and concave
bending stresses to simulate various deformations that may occur in
practice. CV measurements are performed on each sensor in 0.1 M NaOH
contains 400 nM CQL at 50 mV/s, as shown in Figure g. It was observed that the electrochemical
responses of Nafion/Fe3O4/LSG electrodes were
not affected by convex and concave bending. Furthermore, the electrode
responses remain almost the same, showing that the new electrode platform
is flexible.
Detection of CQL in Spiked
Human Urine and
Blood Serum Samples
A practical application of the Nafion/Fe3O4/LSG strip electrode has been explored by examining
the standard addition method to determine CQL in human blood serum
and urine samples. To accomplish this, human blood serum was purchased
from Sigma-Aldrich, Taiwan, and fresh urine samples were obtained
from healthy individuals at National Taipei University of Technology,
Taiwan. Human urine samples were centrifuged for 10 min at 5000 rpm,
filtered, and the supernatant urine was obtained. On other hand, the
serum sample was centrifuged at 5000 rpm for 10 min, and the supernatant
was collected. Both serum and urine supernatant samples (250 μL)
were diluted in 25 mL of 0.1 M NaOH and stored in a refrigerator for
further analysis. The dilution can actually help reduce the matrix
effect of real samples. A known quantity of blood serum and human
urine diluted samples was spiked separately with CQL at different
concentrations using the standard addition method, and SWV curves
are recorded (Figure h,i). The analytical results are presented in Table S1. The estimated recovery of CQL in the spiked urine
and blood serum samples were found to be in the range of 97.0–102
and 98.0–102.6%, respectively, with an RSD less than 2.88%
(n = 3), indicating that the electrode is feasible
and reliable for detecting CQL in blood serum and urine samples. Figure S4 shows the SWV response for the determination
of CQL in human urine samples spiked at 100 nM over commercial AuEs
(gold electrodes) and a Nafion/Fe3O4-modified
AuE to validate the accuracy of the newly proposed sensor. The results
showed that the Nafion/Fe3O4/LSG electrode had
a superior electrochemical performance over a commercial Au electrode
and a modified Nafion/Fe3O4/AuE electrode, indicating
that the new procedure is reliable and feasible.
Application of Portable CQL sensors
We demonstrate
the use of Nafion/Fe3O4/LSG
sensors for analysis of small sample volumes and POC testing for the
detection of CQL by the standard addition method using a portable
electrochemical workstation (Figure S5a). Furthermore, to determine the relationship between peak current
responses and CQL concentrations on a portable mini-workstation, SWV
was applied (Figure S5b). Results from
the portable mini-station are comparable to those from traditional
large-scale workstations. A linear relationship exists between peak
currents and CQL concentrations. From Table S2, we can see that satisfactory recoveries range from 91.2 to 99.35%,
which indicates that the proposed flexible sensor for intelligent
detection of CQL in real samples is satisfactory and applicable.
Conclusions
In this work, we report the development
of a new flexible strip
electrode by immobilizing Nafion/Fe3O4 nanohybrids
on a porous LSG electrode. LSG was fabricated on commercial PI using
a direct laser scribing process. A highly ambient stable Nafion/Fe3O4 nanohybrid was synthesized using an ultrasonic-assisted
technique. The proposed flexible electrode was fabricated by drop
coating Nafion/Fe3O4, which is electrostatically
self-assembled between positively charged Fe3O4[80,81] and negatively charged Nafion on LSG flexible substrates.
In addition, the electrodes were characterized by various spectroscopy
and analytical methods. Owing to their porous 3D structure, an enriched
surface area at the active edges and an excellent wettability in water
contributed to the impressive kinetics of electron transfer reactions.
Specifically, the 3D hierarchical structure of flexible LSG provides
a larger specific surface area and higher mechanical strength than
traditional electrodes. This electrode was used to construct a flexible
transducer for ultrasensitive clioquinol detection, and it performed
admirably in terms of a wide linear range (1 nM to 100 μM),
low detection limit (0.73 nM), and high stability. In addition, it
was demonstrated that this analytical method proved effective even
at interferences that were 50-fold greater than the measured value.
The detection of clioquinol in human serum and urine samples using
the fabricated sensor shows that it can be used as an effective clinical
diagnosis in in vivo studies.