Rapid, inexpensive, and precise water salinity testing remains indispensable in water quality monitoring applications. Despite many sensors and commercialized devices to monitor seawater salinity, salt detection and quantification at very low levels of drinking water (below 120 ppm) have been overlooked. In this paper, we report on optimization of a low-cost microfluidic sensor to measure water salinity in the range of 1-120 ppm. The proposed design employs two copper microbridge wires suspended orthogonally in a PDMS microchannel to measure salinity based on the electrical resistance between the wires. The preliminary design of the sensor microchannel with a rectangular cross-section width (w) of 900 μm and height (h) of 500 μm could measure the water salinity in the range of 1-20 ppm in less than 1 min with detection sensitivity, limit of detection (LOD), and limit of quantification (LOQ) of 17.1 ohm/ohm·cm, 0.31 ppm, and 0.37 ppm, respectively. Data from the preliminary design was used for developing and validating a numerical model which was subsequently used for parametric studies and optimization to improve the sensor's performance. The optimized design demonstrated an order of magnitude increase in sensitivity (385 ohm/ohm·cm), a 6-fold wider detection range (1-120 ppm), and a 15-fold enhancement in miniaturization of the microfluidic channel (w = 200 μm and h = 150 μm) with LOD and LOQ of 0.39 and 0.44 ppm, respectively. In the future, the sensor can be integrated into a hand-held device to remove present impediments for low-cost and ubiquitous salinity surveillance of drinking water.
Rapid, inexpensive, and precise water salinity testing remains indispensable in water quality monitoring applications. Despite many sensors and commercialized devices to monitor seawater salinity, salt detection and quantification at very low levels of drinking water (below 120 ppm) have been overlooked. In this paper, we report on optimization of a low-cost microfluidic sensor to measure water salinity in the range of 1-120 ppm. The proposed design employs two copper microbridge wires suspended orthogonally in a PDMS microchannel to measure salinity based on the electrical resistance between the wires. The preliminary design of the sensor microchannel with a rectangular cross-section width (w) of 900 μm and height (h) of 500 μm could measure the water salinity in the range of 1-20 ppm in less than 1 min with detection sensitivity, limit of detection (LOD), and limit of quantification (LOQ) of 17.1 ohm/ohm·cm, 0.31 ppm, and 0.37 ppm, respectively. Data from the preliminary design was used for developing and validating a numerical model which was subsequently used for parametric studies and optimization to improve the sensor's performance. The optimized design demonstrated an order of magnitude increase in sensitivity (385 ohm/ohm·cm), a 6-fold wider detection range (1-120 ppm), and a 15-fold enhancement in miniaturization of the microfluidic channel (w = 200 μm and h = 150 μm) with LOD and LOQ of 0.39 and 0.44 ppm, respectively. In the future, the sensor can be integrated into a hand-held device to remove present impediments for low-cost and ubiquitous salinity surveillance of drinking water.
Water salinity measurement
is critical in water quality monitoring
and surveillance to preserve water safety.[1] Uncontrolled salinity levels can have detrimental effects on human
health, agriculture, industry, and the environment.[2] The salt concentrations of water sources are continually
changing due to various natural and human-induced environmental phenomena.[3] Exceeding certain salt thresholds could pose
health threats to humans, primarily through drinking water.[4]Among 14 different salt types,[5] sodium
chloride (NaCl) is the dominant dissolved salt leading to water salinity.[5] Sodium plays a vital role in maintaining the
osmotic pressure of body fluid and preventing excessive fluid loss.
However, high sodium levels in the body could cause hypertension,
cardiovascular diseases, renal diseases, and Meniere’s disease.[4,6] High blood pressure alone is the leading cause of death and the
second major cause of disability in children worldwide.[7,8] According to the World Health Organization (WHO), acceptable salt
intake is recognized to be 5 g/day and 500 mg/day for adults on regular
and sodium-restricted diets, respectively.[9] Canada’s national health and welfare department requires
a salt limit of 20 ppm or mg/L on drinking water to maintain a sodium-restricted
diet, assuming a daily water consumption of 1.5 L.[10]Sodium levels in Canadian drinking water supplies
vary seasonally
and geographically.[6] The main sources of
water include tap water, groundwater, and surface water with sodium
levels in the ranges of 0.3–242,[11] 6–130,[12] and 1–300 ppm,[13] respectively. More than 75% of 2100 studied
water supply cases in the U.S. had sodium concentrations below 50
ppm.[14] Salinity levels within these low
limits must be monitored frequently, rapidly, and preferably specific
to the salt type to prevent drastic effects on humans and the ecosystem.There are several well-established methods for measuring water
salinity, among which measuring the electrical conductivity of water
remains the most prevalent.[15,16] Most of the current
conductivity-based salinity sensing technologies are restricted to
measuring high salt levels at the scale of seawater salinity (ranging
from 2000 to 42000 ppm, with a typical standard of 35000 ppm) while
being expensive, bulky, and not suitable for on-site drinking water
salinity monitoring.[16−18] Several fiber optic sensors[19−22] based on interference, fiber
Bragg grating, long period grating, and surface plasmon resonance
have been proposed for on-field salinity measurement. A key limitation
of this technology is its dependence on the change in the refractive
index of water which is not sensitive enough for reliable quantification
of salinity at drinking water levels.[23]Microfluidics allows for low-cost mass production of compact
and
field-deployable salinity sensors that would require a short time
and small sample volume for measurement.[24] Moreover, the surface to volume ratio in microfluidic devices increases
substantially due to the scaling laws of physics, which can lead to
attaining high levels of sensitivity and low levels of detection.
Sun et al.[25] used grayscale readouts from
a microscope to detect salinity concentration changes in water on
a microfluidic device. This method was restricted to seawater samples
as changes at lower salt concentrations did not change the visual
readouts. Kim et al.[26] developed an integrated
temperature and conductivity sensor for monitoring the product water
of reverse osmosis desalination. Their hand-held microfluidic device
was calibrated to detect conductivities in the range of 0.33–14.64
mS/cm, corresponding to relatively high salt levels of 165–7320
ppm, which is not suitable for drinking water monitoring. In another
effort, Xie et al.[27] presented a salinity
sensor based on integrating microfluidics and optical fiber-assisted
Mach–Zehnder interferometry. The study showed a high sensitivity
detection for a 40–31000 ppm range of NaCl but was inept below
40 ppm, which is needed in the drinking water application. As such,
the need to measure low water salinity ranges of 1–120 ppm,
which is more relevant to drinking waters, is not fully addressed
by existing technologies.In this paper, we introduce a low-cost
and sensitive microfluidic
sensor, in two primary and optimized design versions, to detect water
salinity levels in the range of 1–120 ppm. Water samples with
various NaCl concentrations were infused into a microfluidic channel.
The electric resistance of the sample was measured using two copper
microwires suspended in the middle of the channel and orthogonal to
the flow direction (called microbridges).[28] The measured resistance values were correlated to NaCl concentrations,
and the sensor calibration curve was established. Subsequently, a
numerical model of the salinity sensor was developed, verified, and
validated using preliminary experimental findings. A validated model
was then employed for a parametric study of geometrical and physical
properties to reach an optimized sensor design with 1 order of magnitude
higher sensitivity and 6-fold wider detection range, which was experimentally
evaluated. We envision that the integration of this sensing paradigm
into a hand-held device would allow for addressing the existing unmet
need of low-cost and on-field salinity surveillance of drinking waters.
Materials
and Methods
Experimental Study
The experimental setup is shown
in Figure a. It consisted
of a microfluidic sensor, a DC electrical source-meter (Model 2410,
Keithley Instruments Inc., USA), a syringe pump (Legato 110, KD Scientific
Inc., USA), a DMIL LED conventional inverted fluorescence microscope
(Leica, Germany), and a PC to control and record electrical signals
of the source-meter using the manufacturer’s software (KickStart,
Keithley Instruments Inc., USA).
Figure 1
Experimental setup and microfluidic salinity
sensor. (a) Various
components of the experimental setup consisting of the microfluidic
sensor, electric source-meter, syringe pump, microscope, and PC. (b)
Microfluidic device consisting of a sample flow microchannel with
an inlet and an outlet and two copper microwire bridges acting as
terminal and ground electrodes. (c) Schematic of the microfluidic
device and the geometrical parameters contributing to the salinity
sensor’s response, including the channel height (h), channel length (L), channel width (w), interwire distance (g), and microwire diameter
(dw). (d) The two wires were connected
to the source meter. Samples with different NaCl concentrations dissolved
in DI water were flown in the channel, DC electric current was swept
between the wires from 10 nA to 1 μA, and the corresponding
voltages were recorded to calculate the electric resistances based
on ohm’s law. The resistance consists of the solution resistance
(Rsol.) and the electrode–electrolyte
interface resistances (Rint.).
Experimental setup and microfluidic salinity
sensor. (a) Various
components of the experimental setup consisting of the microfluidic
sensor, electric source-meter, syringe pump, microscope, and PC. (b)
Microfluidic device consisting of a sample flow microchannel with
an inlet and an outlet and two copper microwire bridges acting as
terminal and ground electrodes. (c) Schematic of the microfluidic
device and the geometrical parameters contributing to the salinity
sensor’s response, including the channel height (h), channel length (L), channel width (w), interwire distance (g), and microwire diameter
(dw). (d) The two wires were connected
to the source meter. Samples with different NaCl concentrations dissolved
in DI water were flown in the channel, DC electric current was swept
between the wires from 10 nA to 1 μA, and the corresponding
voltages were recorded to calculate the electric resistances based
on ohm’s law. The resistance consists of the solution resistance
(Rsol.) and the electrode–electrolyte
interface resistances (Rint.).The microfluidic sensor shown in Figure b,c comprised two polydimethylsiloxanes (PDMS)
layers, with mirrored channel designs, cast on 3D printed molds. PDMS
prepolymer (Sylgard 184, Dow Corning Co., USA) was prepared with a
1:10 cross-linker to monomer ratio, poured on 3D printed molds containing
the microchannel design, and maintained at 80 °C for 2 h. After
peeling off the cured PDMS layers, two copper microwires (also called
microbridges) with a diameter of 90 μm were placed in the middle
of the channel. In the primary and optimized designs of the microfluidic
sensor, the device geometries such as channel width (w), height (h), and interwire spacing (g) were changed. Oxygen plasma bonding was used to seal the two PDMS
layers and then to bond them to a glass substrate. Afterward, the
device was placed on the inverted microscope to monitor the microchannel
while conducting the experiments.Four and three replicates
of the primary and optimized microfluidic
sensors were manufactured and tested, respectively. Experiments were
repeated for each replicate of the primary and optimized devices seven
and five times, respectively, to measure NaCl concentration levels
at a statistical significance of 0.05. Samples with salinities ranging
from 1 to 20 ppm and 1 to 120 ppm of NaCl dissolved in DI water were
used in the primary and optimized sensor studies, respectively. The
20 and 120 ppm samples were prepared by dissolving 40 and 240 mg of
NaCl, from Sigma-Aldrich Inc. (cat. no. S7653), in 2 L of DI water,
respectively. Through serial dilution with deionized water (DI), concentrations
of 15, 10, 7.5, 5, 3, 2, and 1 ppm of NaCl were produced and used
in the primary device experiments. For the optimized design, serial
dilution was performed to obtain 100, 80, and 60 ppm of NaCl in addition
to the above concentrations. The samples were infused into the microchannels
with the syringe pump at a flow rate of 1 and 0.2 mL/min in the primary
and optimized phases, respectively. As the optimized device had a
smaller cross-section channel, the sample influx rate was reduced
to maintain a similar Reynolds number for both devices.The
terminal and ground copper microbridge wires were connected
to the DC source meter. As this study aims to pave the way for developing
a hand-held sensor that runs by a battery in the future, a DC supply
was used in our tests. The source-meter swept the DC electric current
from I = 10 nA to I = 1 μA
in 56 s during each experiment while measuring the electric voltage
(V) across the copper microwires (Figure d). The corresponding voltages
were recorded using the KickStart software. The measured current and
voltage values were used to calculate the NaCl sample resistance, R = 2Rint + Rsol in Figure d, using Ohm’s law (R = V/I). Here, 2Rint is
the electrode–electrolyte interface resistances of the two
wires due to their electric double layers in series with the nonuniform
solution resistances near the wires.[29]Rsol. is the bulk solution resistance which is
relatively high for the cases in this study with samples containing
low salt concentrations, i.e., high resistivities.[30,31] All of the measurements were performed at standard ambient temperature
and pressure (approximately 25 °C and 100 kPa, respectively),
and the results would be valid for these conditions.To eliminate
the variability between devices, each set of results
was normalized with a measured baseline resistance, R0, at 0 ppm of NaCl (blank DI water). The fold change
method (R/R0) was used
for normalization, and data points at each concentration were divided
to the mean resistance value of the baseline.All of the statistical
analyses in this study were conducted using
Minitab 20. Levene’s test was applied to examine the homogeneity
of variances prior to conventional one-way ANOVA. If the test showed
a significant result (p < 0.05), then Welch’s
ANOVA was conducted instead of one-way ANOVA. Also, when ANOVA showed
a significant difference between the mean values, a post hoc pairwise
analysis was performed to identify which pairs of means are significantly
different. Tukey HSD posthoc pairwise comparison was conducted for
data sets with homogeneous variances, and a Games–Howell post
hoc test was adapted for the ones with inhomogeneous variances.The limits of detection (LOD) and quantification (LOQ) were determined
as salinity concentrations with resistances equal to DI water resistance
(RDI) minus 3 and 10 times its standard
deviation (SD), respectively (i.e., RLOD = RDI – 3SDDI and RLOQ = RDI −10SDDI). The NaCl concentrations at which these resistances would
be obtained were found using the calibration curves established for
the primary and optimized sensors.
Numerical Study
A numerical model was developed in
COMSOL Multiphysics based on the experimental configuration of the
primary sensor (Figure c,d) which was then verified and validated against the experimental
measurements. Figure c depicts the geometrical parameters of the salinity sensor, such
as the channel length (L), height (h), width (w), wire diameter (dw), and interwire distance (g) that were investigated
numerically. After validation, the model was employed to find an optimized
sensor geometry to increase its sensitivity for salinity measurement
applications. The parameters used in this study and their units are
summarized in Table .
Table 1
Parameters Used in the Numerical Model
with Their Symbols, Descriptions, And Units
symbol
description
units
symbol
description
units
ρ
density
kg/m3
E
electric field
V/m
u
mean velocity
m/s
ρel
electrical resistivity
ohm. m
p
pressure
N/m2
J
current
density
A/m2
μ
viscosity
Pa·s
V
electrical potential
V
Re
Reynolds number
σ
electrical conductivity
S/m
Dh
hydraulic diameter
m
Qj
current source
A/m3
A
microchannel cross-section
area
m2
Je
external
current density
A/m2
P
microchannel wetted perimeter
m
zi
charge
number of ions (e.g.,
+1 for Na+)
C
dw
wire diameter
m
T
temperature
K
w
width (along the z-axis in Figure 1)
m
Cp
specific heat capacity
J/kg·K
h
height (along the y-axis in Figure 1)
m
k
thermal conductivity
W/m·K
L
microchannel length (along
the x-axis in Figure 1)
m
Qe
resistive dissipation
W/m3
g
interwire distance (along
the x-axis in Figure 1)
m
Qted
thermoelastic dissipation
W/m3
Ji
diffusive flux
mol/cm2.s
Qvd
viscous dissipation
W/m3
Fc
Faraday’s
constant
C/mol1
Qp
pressure work
W/m3
um,i
ionic mobility
m2/s·volt
Q
heat source
W/m3
Λ
molar conductivity
S/cm2.mol
R
molar gas constant
J/mol.K
c
molar concentration
mol/L
Ri
reaction rate
M/s
α
correlation factor
F
other forces
N/m3
Λ0
the infinite dilution
S/cm2·mol
I
identity matrix
κ
specific conductance
S/m
The model of the primary sensor consisted
of a 2D rectangular microfluidic
channel, cross-sectioned across the x–y plane of Figure c at the middle of the channel, with two 90 μm diameter
circles representing the wire electrodes in the middle. The microchannel
height (along the y-axis) was h =
500 μm. The width of the microchannel along the z-axis was set to w = 900 μm using COMSOL’s
out-of-plane thickness feature, which allowed achieving similar results
to 3D simulation in a 2D setup. Interwire distance was set to g = 1500 μm. Simulation runtime was minimized by reducing
the microchannel length along the x-axis while maintaining
the model domain independence (see Figure S1). For the optimized sensor simulation, the geometrical characteristics
were h = 150 μm, w = 200 μm,
and g = 2000 μm.The boundary conditions
were selected based on the experimental
conditions. A fully developed flow profile with a constant flow rate
of 1 or 0.2 mL/min was imposed in the x-direction
for the primary and optimized sensors, respectively. The actual devices
had a longer channel length in the x-direction than
the numerical study that allowed for the safe assumption of a fully
developed flow. The velocity in the y-axis direction
was zero, and a no-slip boundary condition was applied throughout
the microchannel and wire walls. A static gage pressure of zero (p = 0) was considered at the channel outlet. The two copper
microwires were considered as the electrodes that probe and measure
the resistance in the salinity sensor. A ramping DC current of 10
nA to 1 μA was maintained in the terminal while the ground was
assumed to have no voltage (V = 0). The PDMS walls
were insulated (J = 0).Two important parameters
of the numerical model were the dynamic
viscosity and density of the sample. According to previous studies,[32,33] upon adding 10000 ppm of salt to water at 25 °C, an increase
of 0.7% and 2% would be observed in the dynamic viscosity and density
of water, respectively. These slight changes were safe to be neglected
in our models with 1–120 ppm salt concentrations. Therefore,
the water dynamic viscosity of μ = 8.90 × 10–4 Pa·s and density of ρ = 1000 kg/m3 were considered
in the numerical analyses.The simulation model was governed
by the physics of fluid flow,
electric field, transport of diluted species, and heat transfer (discussed
in the Supporting Information Section 1). To obtain a better convergence in our simulation, we first computed
the fluid flow and electric potential profiles in the microchannel,
and then the electrochemical and heat transfer behaviors were developed.The flow regime in this study would fall into the laminar flow
category with the Reynolds number, Re in equation , being 16 < Re < 64. Thereby, the simplified
incompressible non-Newtonian continuity equation and Navier–Stokes equation were employed to solve for
the electrolyte flow under steady-state conditions.The hydraulic diameter of
the channel was obtained from eq .The electrolyte
sample containing sodium and chloride ions flows
at a constant flow rate in the microchannel. The applied electric
field between the wires causes the ions to migrate toward the electrodes
with the opposite charge, and an electric double layer would be shaped.
Steady-state charge conservation eq governs the electric field in the microchannel and
electric current passing from the terminal to the ground electrode
through the electrically conductive electrolyte sample.The electrical resistance, R, between the wires
was assumed to be related to the sample electrical resistivity, ρel, and other geometrical properties of
the channel as defined in eq .Mass conservation equation was solved to investigate
the diffusion, convection, and
ionic migration of the dissolved NaCl ions for one or more solutes
(i).[34] Fick’s law governs the diffusion
term.The reaction rate is for chemical reactions in the model, which
does not apply in our study. The diffusive flux vector in the case
of applying the electric field is expressed below.For a very dilute
solution, the ionic mobility of solute i can be obtained
from the Nernst–Einstein (eq ).The Debye–Hückel–Onsager
(eq ) was used to
calculate the electrolyte
conductivity at different concentrations of NaCl below 0.001 mol/L
(∼60 ppm).[35] This equation is suitable
for very dilute solutions of strong electrolytes (e.g., NaCl) with
high solubility in their solvent, as expressed below.The values of the A and B constants
depend on the viscosity and dielectric constant of the solvent, temperature,
and charge. In the case of NaCl and water at 25 °C, A = 60.20, B = 0.229, and Λ0 = 126.39
× 10–14.[35] For salinity
concentrations between 60 and 120 ppm, a correlation factor of α
= 0.55 was used to correlate the salt concentration to electrical
conductivity.[36,37] The electrical resistivity for
the salty water samples was then calculated from eq .
Results and Discussion
Primary Sensor Experimental Evaluation
We measured
the primary sensor’s electric resistances for salty water samples
with NaCl concentrations of 1–20 ppm flown at a flow rate of
1 mL/min in the channel, as shown in Figure (four replicate devices, each tested 7 times).
The sensor had a transient response at the beginning of the experiment,
followed by a steady-state plateau. The transient effect is mainly
a result of sudden implementation of an external electric field which
is intrinsic to measurement systems.[38] It
usually dies out over time and the device would reach a steady state
that is suitable for measurement.[38] Another
reason for the transient mode presumably resulted from changes in
the ionic composition of the medium around the wires when the current
was applied. Samples with NaCl concentrations lower than 5 ppm exhibited
shorter transient modes (∼10 s), and the duration was increased
at higher concentrations (∼25 s). A relatively stable plateau
was observed for all of the concentrations once a steady-state regime
was reached (after 30 s; see Figure insets). The resistance values in the range of 30–56
s, corresponding to the current range of 0.5 to 1 μA, showed
a gradual decrease. Analysis of this data showed that the resistance
changes in the last 30 s of the experiments were smaller than the
standard deviations (SD) in the same duration (Figure S2). Thus, the recorded data in the 30–56 s
of each experiment was inferred as the steady state plateau and used
to obtain the resistance means and standard deviations for correlation
to NaCl concentration in the samples.
Figure 2
Dose–response measurements of the
primary salinity sensor.
Electrical resistances were recorded as the current was swept from
10 nA to 1 μA during 56 s. Results are shown for samples with
(a) 1, (b) 2, (c) 3, (d) 5, (e) 7.5, (f) 10, (g) 15, and (h) 20 ppm
of NaCl in DI water. Samples were flown in the microchannel at a flow
rate of 1 mL/min. The insets depict the 30–56 s intervals of
the resistance measurements. Each panel consists of 28 recorded measurements
from four replicates and seven measurements. Repetition experiments
in each plot are represented with different colors.
Dose–response measurements of the
primary salinity sensor.
Electrical resistances were recorded as the current was swept from
10 nA to 1 μA during 56 s. Results are shown for samples with
(a) 1, (b) 2, (c) 3, (d) 5, (e) 7.5, (f) 10, (g) 15, and (h) 20 ppm
of NaCl in DI water. Samples were flown in the microchannel at a flow
rate of 1 mL/min. The insets depict the 30–56 s intervals of
the resistance measurements. Each panel consists of 28 recorded measurements
from four replicates and seven measurements. Repetition experiments
in each plot are represented with different colors.Figure a
illustrates
the electrical resistance means and SDs from the four replicated devices
at different NaCl concentrations. The results show that different
replicates had similar readouts at each concentration as the error
bars of replicates overlapped. Results qualitatively demonstrate that
measurements were reproducible and reliable with no overlap between
the error bars of any two concentrations. However, there are wider
gaps between resistances of the samples with lower salinities (e.g.,
ΔR = ∼1.6 MΩ between 2 and 3 ppm of NaCl vs ΔR
= ∼300 kΩ between 15 and 20 ppm of NaCl). To compare
the device readings quantitatively, we performed a one-way analysis
of variance (ANOVA) test to examine if the sensor could statistically
distinguish the difference among the samples. Statistically significant
differences between the resistance mean values of different concentrations
were observed (p < 0.0001). Games–Howell
post hoc test was adopted to conduct a pairwise comparison for different
concentrations. An effect size of 3.58 with a statistical power of
0.9 was obtained at a significance level of 0.05 and a sample size
of 7 for each concentration. It was then established that the sensor
could distinguish different tested concentrations, and the statistical
difference for every two concentrations would be sufficiently large
with at least seven repetitions.
Figure 3
Dose–response measurements of the
primary salinity sensor.
(a) Averaged electric resistances of four replicate devices at different
NaCl concentrations. To eliminate interdevice variability, we normalized
the results of each replicate in (a) with their baseline values at
0 ppm of NaCl, and the normalized mean values and standard deviations
are shown versus samples NaCl concentrations and resistivities in
(b). The calibration curve fits the experimental data. The numerical
simulation results are shown in (c). The error bars represent standard
deviations
Dose–response measurements of the
primary salinity sensor.
(a) Averaged electric resistances of four replicate devices at different
NaCl concentrations. To eliminate interdevice variability, we normalized
the results of each replicate in (a) with their baseline values at
0 ppm of NaCl, and the normalized mean values and standard deviations
are shown versus samples NaCl concentrations and resistivities in
(b). The calibration curve fits the experimental data. The numerical
simulation results are shown in (c). The error bars represent standard
deviationsThe raw resistances in Figure a were normalized
with respect to measured DI water
resistances to eliminate interdevice variabilities, as depicted in Figure b. As shown, the
normalized resistance decreases with an increase in salt concentration
due to the rise in the number of ions between the two wires. Similar
statistical analyses to the ones described for Figure a were performed with the normalized data
in Figure b, and identical
results were achieved in terms of distinguishing between various salt
concentrations (p < 0.0001).Figure b inset
depicts the normalized mean resistances correlated to the electrical
resistivities of the samples, calculated according to the Debye–Hückel–Onsager
equation (eq ). The
samples with lower NaCl concentrations would have higher resistivities.
A linear relationship between the resistance and resistivity in eq allows the sensor’s
calibration curve to be established asA goodness of linear fit of 97.84% was achieved for eq . Disregarding the 1 ppm concentration
in the curve-fitting process (since it demonstrated more noise than
the other concentrations) would increase the goodness of the fit to
99.26%. Thus, the sensitivity of the sensor based on the primary design
was 17.1 ohm/ohm·cm. The blank DI water resistance in the primary
sensor was RDI = (30.1 ± 7.04) ×
106. Accordingly, the limit of detection (LOD) and limit
of quantification (LOQ) were calculated to be 0.31 and 0.37 ppm according
to 3SD and 10SD methods, respectively.
Numerical Optimization
of the Sensor
To increase the
sensor’s sensitivity and its detection range, further optimization
studies were warranted. For rapid optimization, a numerical model
of the primary sensor based on the experimental results of Figure was developed, verified,
and validated and then applied in a parametric study to determine
the most contributing device parameters.We simulated several
conditions to ensure the results were domain and mesh independent
while achieving the least possible computational load in developing
the model. The optimum channel length and mesh properties were found,
according to Figures S1 and S3, to be 3
mm and 191026 triangular elements, respectively.To verify the
model, we examined several basic physical phenomena
that were expected with our numerical method (results not shown).
For example, the fluid flow showed a parabolic pattern through the
microfluidic channel. The electrolyte conductivity increased at higher
salt concentrations, resulting in lower electric resistance values.
The electric potential decreased from its initial value to zero from
the terminal to the ground. The ions migration was observed, and Na+ and Cl– ions accumulated around the ground
and terminal electrodes due to potential differences, respectively.
The accumulation of ions around the wires was increased as the salt
concentration was raised in the sample. The fluid flow caused the
convection of accumulated ions around the wires and reduced the effective
ions in the current transfer. The above observations agreed well with
the sensor’s physics and verified the correct performance of
the model.We then validated our numerical model by comparing
our results
with the obtained experimental resistance values from the primary
sensor. Figure c depicts
the recorded electrical resistance mean values in the experimental
study and the numerical study in the range of 1–20 ppm, corresponding
to resistivities of 46.4 × 104 to 4.5 × 104 Ω·cm. As expected, the resulting resistances from
the simulation show a perfectly linear behavior (R = 2.52 × 105 ρel; R2 = 1). There was a deviation between the two data sets
as the simulation did not consider the experimental errors. As the
simulation could not completely mimic experimental conditions, a transfer
function was defined so that the user could obtain experimental values
using the simulated model and compensate for the observed deviation.
We derived the transfer function of system by dividing the line equations
of the two data sets.The numerical optimization
study was carried out with a full factorial
parametric approach. The effects of six parameters with three levels
were examined on the sensor performance as tabulated in Table . The minimum channel height
and width levels were selected on the basis of fabrication restrictions
using 3D printing. The wire diameters were chosen from several available
choices in the market. The effects of changing the fluid properties
and flow rate were also studied using the Reynolds number. The study
involved 36 = 729 combinations, and each combination was
simulated for 1–20 ppm of NaCl concentrations.
Table 2
Numerical Optimization Study Parameters
and Levelsa
levels
parameter
low
medium
high
P value
channel height (h) (μm)
150
250
500
<0.001
channel width
(w) (μm)
200
500
900
<0.001
interwire distance (g) (μm)
1000
1500
2000
<0.001
wire diameter (dw)
(μm)
90
110
130
0.973
electric current (I) (nA)
1
100
1000
1.00
Reynolds number (Re)
16
32
64
0.998
The P values
were achieved through a full factorial design for each parameter.
The P values
were achieved through a full factorial design for each parameter.The results of the optimization
study are shown partially in Figure . Our simulations
showed that the channel width, channel height, and interwire distance
had the most significant effects on the sensor’s performance,
determined by examining the increase in resistance compared to the
primary sensor; these observations were also confirmed by ANOVA test
(Table ). Low levels
of the channel width and height caused the highest increase in resistance
value as the sensing medium was shrunk (h × w would decrease in eq ). On the other hand, the effect of increasing the interwire distance
was not as intense as the channel height and width. The reason is
that as g increased, there were two opposing effects
on the resistance, i.e., an increase as there was more electrolyte
with small conductivity and a decrease as the electric field was weakened
when the two electrodes moved further apart.
Figure 4
Effect of the most contributing
parameters at different levels
on the electrical resistance of the sensor. The increase in resistance
compared to the primary sensor is plotted. The optimized design parameter
configuration should be composed of low levels of channel width and
height and a high level of interwire distance, resulting in g = 2000 μm, w = 200 μm, and h = 150 μm.
Effect of the most contributing
parameters at different levels
on the electrical resistance of the sensor. The increase in resistance
compared to the primary sensor is plotted. The optimized design parameter
configuration should be composed of low levels of channel width and
height and a high level of interwire distance, resulting in g = 2000 μm, w = 200 μm, and h = 150 μm.We found that the electrode diameter, electric current, and Reynolds
number had no statistically significant effect on the resistance (P values close to 1 in Table ). Changing the electric current resulted in similar
changes in voltage that left the resistance intact. The fluid flow
was always in the laminar regime and did not significantly reduce
the effective concentration of the ions in the electric circuit to
cause changes in the resistance. It was counterintuitive that the
wire diameter had no significant impact on sensor performance as increasing
the electrode surface was expected to allow for better electron transfer.
However, we hypothesize that as the salt concentrations were scarce,
the wires surface area was more than enough for the ions to transfer
electrons. To investigate this hypothesis, we simulated smaller wire
diameters in the range of 1–130 μm and studied the discharged
current density on the wire surfaces (Figure S4). There was less than 10% difference in current densities between
wire diameters of 90, 110, and 130 μm. This is also in agreement
with our findings from Table that wire diameter did not have significant contribution.According to Table and Figure , the
optimized sensor should have the dimensions of g =
2000 μm, w = 200 μm, and h = 150 μm. This finding is in agreement with eq as increasing and decreasing the
cross-section area and interwire distance, respectively, would result
in a wider range of resistances and, thereby, higher sensitivity.
The achieved resistances, calculated through simulation, were plotted
against NaCl solution resistivities for the optimized and primary
designs in Figure . It is estimated that the sensitivity of the optimized sensor would
increase 27-fold compared to the primary sensor. This estimation will
be further investigated in experimental analysis in the next section.
Figure 5
Numerically
calculated electrical resistance versus saltwater resistivity
for the primary and optimized sensors. The numerical model predicts
27-fold higher sensitivities with the optimized sensor compared to
the primary one.
Numerically
calculated electrical resistance versus saltwater resistivity
for the primary and optimized sensors. The numerical model predicts
27-fold higher sensitivities with the optimized sensor compared to
the primary one.
Experimental Analysis of
the Optimized Sensor
To evaluate
the optimized sensor after fabrication, we performed the same experimental
procedures as the preliminary device (see Figure S5). Figure a illustrates the electrical resistance means and SDs at different
NaCl concentrations in the optimized sensor (dw = 90 μm, g = 2000 μm, w = 200 μm, and h = 150 μm).
Similar to the primary sensor, the resistance decreases as the ions
population increases, and the electron transfer improves. Also, there
are wider resistance variations between replicates of samples with
less than 3 ppm of NaCl concentrations. The analysis of variance and
Games–Howell post hoc analyses were performed. It was obtained
that there is a statistically significant difference between all and
each pair of concentrations with a 0.01 significance level for both
analyses (p < 0.0001). An effect size of 2.89
with a statistical power of 0.99 was obtained at a significance level
of 0.01 and a sample size of 5 for each concentration. The analysis
demonstrated that the sensor could distinguish different tested concentrations,
and the statistical difference for every two concentrations would
be sufficiently large with at least five repetitions.
Figure 6
Dose–response
experimental measurements of the optimized
salinity sensor. (a) Measured electric resistances of three replicate
devices at different NaCl concentrations. To eliminate interdevice
variability, the results of each replicate in (a) were normalized
relative to their baseline values at 0 ppm of NaCl and the normalized
mean values and standard deviations are shown versus samples (b) NaCl
concentrations and (inset) resistivities. The calibration curve fit
to the experimental data is also shown in the inset of (b). The error
bars represent standard deviations.
Dose–response
experimental measurements of the optimized
salinity sensor. (a) Measured electric resistances of three replicate
devices at different NaCl concentrations. To eliminate interdevice
variability, the results of each replicate in (a) were normalized
relative to their baseline values at 0 ppm of NaCl and the normalized
mean values and standard deviations are shown versus samples (b) NaCl
concentrations and (inset) resistivities. The calibration curve fit
to the experimental data is also shown in the inset of (b). The error
bars represent standard deviations.The raw resistance values (R) in Figure a were normalized to eliminate
interdevice variability, as in Figure b. Moreover, the normalized values were plotted against
the resistivity to establish the sensor’s linear calibration
curve, expressed in Figure b inset and eq .The goodness of linear regression was found to be R2 = 96.49%. Neglecting 1, 2, and 3 ppm of NaCl concentrations
would increase R2 to 99.76%. Comparing
the primary and optimized sensors’ calibration curves, we found
that the sensitivity increased an order of magnitude to 385 ohm/ohm·cm
with averaged data (further discussed in Figure ). The higher sensitivity allowed the sensor’s
detection range to be widened by 6-fold, increasing it from 1 to 20
ppm to 1–120 ppm. The blank DI water resistance in the optimized
sensor was RDI = (491.2 ± 8) ×
106. Accordingly, the LOD and LOQ was calculated to be
0.39 and 0.44 ppm using 3SD and 10SD methods, respectively, which
was almost equivalent to the primary design.
Figure 7
Comparison of the dose–response
measurements of the primary
and optimized sensors in our experimental and numerical analyses.
(a) The optimized model has higher theoretical and experimental sensitivities
and allows a detection range of 1–120 ppm of NaCl (46.37 –
0.39 × 104 Ω·cm) compared to 1–20
ppm (46.37 – 2.34 × 104 Ω·cm) for
the primary sensor. (b) Otimized sensor results are illustrated in
the range of 5–120 ppm of NaCl (9.31 – 0.39 × 104 Ω·cm) that has the least deviation between the
simulation and experimental results. The inset magnifies the 7.5–120
ppm of NaCl (6.21 – 0.39 × 104 Ω·cm)
range for the optimized sensor.
Comparison of the dose–response
measurements of the primary
and optimized sensors in our experimental and numerical analyses.
(a) The optimized model has higher theoretical and experimental sensitivities
and allows a detection range of 1–120 ppm of NaCl (46.37 –
0.39 × 104 Ω·cm) compared to 1–20
ppm (46.37 – 2.34 × 104 Ω·cm) for
the primary sensor. (b) Otimized sensor results are illustrated in
the range of 5–120 ppm of NaCl (9.31 – 0.39 × 104 Ω·cm) that has the least deviation between the
simulation and experimental results. The inset magnifies the 7.5–120
ppm of NaCl (6.21 – 0.39 × 104 Ω·cm)
range for the optimized sensor.The overall performance of the two sensors was replotted under
both experimental and numerical conditions in Figure a. The first three concentrations of NaCl
at 1, 2, and 3 ppm (46.37 – 15.49 × 104 Ω·cm
resistivity range) had a lower signal-to-noise ratio and brought about
the highest deviation from the simulation findings. As previously
mentioned, this deviation resulted from simulation ideal intrinsic
quality that was not counted in the experimental errors. This deviation
is observed as the experiment and simulation results are compared
for the case of the optimized sensor (Figure a). However, the simulation results are in
better agreement with the experimental findings in the range of 5–120
ppm (corresponding to 9.31 – 0.39 × 104 Ω.cm
resistivity range), as illustrated in Figure b, with an average deviation of 12%. As the
simulation could not completely mimic experimental conditions, a transfer
function was calculated and applied to experimental results. The transfer
function was established as a result of dividing eq and optimized sensor numerical
equation (R = 6.75 × 106ρel) and is expressed in eq .As shown in Figure a,b, the slope of the primary
sensor fitted line is 17.1 ohm/ohm·cm
with the prenormalized data, which in comparison to the optimized
sensor slope of 385 ohm/ohm·cm (1–120 ppm) or 692.4 ohm/ohm·cm
(5–120 ppm) shows a significant improvement in the sensitivity
of the optimized sensor. The optimized salinity sensor demonstrated
promising potential in differentiating the samples with NaCl concentrations
in the range of 1–120 ppm. The sensor exhibited a LOD of 0.39
ppm, lower than the NaCl detectors reported before. Although previously
developed optical salinity sensors have reported lowest detected values
of 2,[39,40] 4,[41] 6.7,[20] 10,[42] and 40 ppm,[27] they lack the required resolution to quantify
salinity within drinking water ranges. This limitation occurs because
a 1000 ppm change in salinity normally causes an infinitesimal deviation
in the optical path length.[1] In addition,
these sensors were complex, laboratory-based, and costly due to their
reliance on prisms or delicate fiber optics but were able to detect
nonionic salts as well. Reported conductivity-based sensors offered
lowest detected values of 12,[43] 16,[44] 165,[26] and even higher
(e.g., 7800 ppm),[45−47] which are not suitable for sensitive detection in
the drinking water applications. Also, these sensors involved labor-intensive
and expensive fabrication procedures. In comparison, the proposed
optimized sensor was fabricated with a simple and low-cost technique
which allows for sensing in drinking water ranges on a miniaturized
and inexpensive platform.The current design of the developed
microfluidic salinity sensor
suffers from three main limitations. First, salinity detection is
performed through measuring conductivity which can be interfered by
nonspecific ionic entities in the sample. As such, the detection of
the current version of the sensor is not selective. Second, the discussed
characterizations of the sensors were performed in ambient temperature
and pressure and the established calibration curves are only accurate
in these conditions. Third, the present sensors rely on syringe pump
and source meter that impedes in situ measurement. Given the promising
results presented in this manuscript, we are carrying out research
toward addressing the above limitations.
Conclusions
In
this paper, a low-cost microfluidic sensor was developed and
optimized to measure water salinity based on the sample’s DC
electrical resistance. The sensor consisted of a microchannel and
two microbridge wires, with two different configurations resulted
from primary studies and numerical optimizations. Using the primary
sensor, we detected low NaCl levels in the range of 1–20 ppm
in less than 1 min by measuring the resistance between the two wires.
Our findings showed that the primary sensor sensitivity, LOD, and
LOQ were 17.1 ohm/ohm·cm, 0.31 ppm, and 0.37 ppm, respectively.
We developed, verified, and validated a numerical model against the
experimental findings to optimize the sensor and increase its sensitivity
and detection range. A parametric study was conducted to establish
the most contributing parameters of the device, which were the channel
width, channel height and interwire spacing among six investigated
parameters. Through experimental analysis with the optimized sensor,
it was found that the sensitivity and detection range were increased
by 1 order of magnitude and 6-fold (to 1–120 ppm), respectively,
while the sensor could be miniaturized further by 15-fold. The experimental
evidence showed the sensor’s accuracy and repeatability, making
it a promising candidate for different applications such as surveillance
of consumed water by individuals with salt-restricted diets. Although
the proposed device does not specifically detect sodium chloride,
it could be used as a preliminary surveillance system to warn the
water consumers or inspection officials to perform further tests on
water safety. We envision integrating it into a hand-held sensor for
on-site and point-of-care surveillance of salinity levels after miniaturization
of the DC source-meter with portable alternatives and replacing the
syringe pump with a passive on-chip or a battery-operated peristaltic
pump.
Authors: J M White; J G Wingo; L M Alligood; G R Cooper; J Gutridge; W Hydaker; R T Benack; J W Dening; F B Taylor Journal: J Am Diet Assoc Date: 1967-01