Yumi Sugo1, Shin-Ichi Ohira2, Hinako Manabe3, Yo-Hei Maruyama3, Naoaki Yamazaki4, Ryoma Miyachi2, Kei Toda2, Noriko S Ishioka1, Masanobu Mori3. 1. Department of Radiation-Applied Biology Research, Takasaki Advanced Radiation Research Institute, National Institutes for Quantum Science and Technology, 1233 Watanuki, Takasaki, Gunma 370-1292, Japan. 2. Department of Chemistry, Kumamoto University, 2-39-1 Kurokami, Kumamoto 860-8555, Japan. 3. Faculty of Science and Technology, Kochi University, 2-5-1 Akebono-cho, Kochi 780-8520, Japan. 4. Graduate School of Engineering, Gunma University, 1-5-1 Tenjin-cho, Kiryu, Gunma 376-8515, Japan.
Abstract
Preparing compounds containing the radioisotope 64Cu for use in positron emission tomography cancer diagnostics is an ongoing area of research. In this study, a highly efficient separation method to recover 64Cu generated by irradiating the target 64Ni with a proton beam was developed by employing a flow electrolysis cell (FE). This system consists of (1) applying a reduction potential for the selective adsorption of 64Cu from the target solution when dissolved in HCl and (2) recovering the 64Cu deposited onto the carbon working electrode by desorbing it from the FE during elution with 10 mmol/L HNO3, which applies an oxidation potential. The 64Cu was selectively eluted at approximately 30 min under a flow rate of 0.5 mL/min from the injection to recovery. The newly developed flow electrolysis system can separate the femtomolar level of ultratrace radioisotopes from the larger amount of target metals as an alternative to conventional column chromatography.
Preparing compounds containing the radioisotope 64Cu for use in positron emission tomography cancer diagnostics is an ongoing area of research. In this study, a highly efficient separation method to recover 64Cu generated by irradiating the target 64Ni with a proton beam was developed by employing a flow electrolysis cell (FE). This system consists of (1) applying a reduction potential for the selective adsorption of 64Cu from the target solution when dissolved in HCl and (2) recovering the 64Cu deposited onto the carbon working electrode by desorbing it from the FE during elution with 10 mmol/L HNO3, which applies an oxidation potential. The 64Cu was selectively eluted at approximately 30 min under a flow rate of 0.5 mL/min from the injection to recovery. The newly developed flow electrolysis system can separate the femtomolar level of ultratrace radioisotopes from the larger amount of target metals as an alternative to conventional column chromatography.
Cancer is
a major cause of death
in economically developed countries.[1,2] Thus, there
have been many studies on cancer treatment methods such as surgery,[3] radiotherapy,[4] and
chemotherapy.[5] Advanced cancer is characterized
by invasion and metastasis; therefore, the early detection of cancer
is important to improve outcomes. Many techniques that enable an early
diagnosis of cancer have been developed.[6−10] However, when patients are exposed to X-rays and radiation for diagnosis,
it sometimes places a heavy burden on the body. Positron emission
tomography (PET)[11] is one of the less burdensome
diagnostic imaging methods. PET can measure primary and metastatic
lesions of cancer throughout the whole body at one time with relatively
small exposure to the radiation, while X-ray examination measures
partial parts of the human body such as lungs and stomach with a single
measurement. In PET, l-methionine labeled with 11C (11C-MET) or glucose labeled with 18F (18F-FDG) is injected into the body as a radioactive isotope.
These compounds can be prepared with a small cyclotron in hospitals.
When 11C-MET[12] and 18F-FDG[13] accumulate in cancer cells, they
emit radiation, allowing PET to measure the location and size of the
tumor. However, because these radioactive elements have a short half-life
(18F: 109.77(5) m; 11C: 20.364(14) m), they
are not suitable for use with drugs, such as antibodies, that take
a long time to accumulate in tumors. If radiopharmaceuticals could
be rapidly prepared and, moreover, entirely in a closed flow system,
i.e., from the separation through drug synthesis, then the loss of
radioactivity could be alleviated, thereby contributing to more effective
diagnosis and treatment methods.The use of 64Cu,
a copper radioisotope with a moderately
long half-life (12.701(2) h), has recently been investigated as one
of RI alternating to 18F or 11C.[14−16] Because copper can be stably complexed with a variety of compounds,
including 1,4,7,10-tetraazacyclododecane-1,4,7,10-tetraacetic acid
(DOTA), it has been widely used in clinical trials of PET diagnostics.[17] Also, Yoshii et al. reported that Cu-cetuximab
was injected into a model mouse and successfully visualized pancreatic
cancer smaller than 1 cm by a PET system equipped with a three-dimensional
(3D) radiation detector.[18] Thus, utilizing 64Cu has increased as the attractive RI metal in PET diagnosis.However, because the radiation intensity of the produced nuclides
has decreased by the time of clinical use, the shortest possible process
time from 64Cu production to drug synthesis (as described
below) has been required. The production steps include:Production of pico-
or nanomole levels
of 64Cu by irradiating 64NiO or 64Ni metal with a proton beam.Dissolution of the irradiated samples
with hydrochloric acid and heating.Separation of 64Cu from
the 64Ni matrix by column chromatography.Removal of hydrochloric acid by evaporation
to dryness.The process (1) takes 1–5
h, depending on the amount of
RI–Cu required.[19,20] The processes (2)–(4)
related to the separation and purification of 64Cu take
approximately 2.5–4 h, i.e., 1–2 h in process (2), 15–30
min in process (3), and 1–2 h in process (4), though the required
times for the processes are different among the research groups.[19,20] The attenuation factor of the 64Cu (f) during processes (2)–(4) estimated from eq is reduced to 0.80–0.87,
assuming an initial radioactivity of 1where T is half-life, t is the attenuation time, A0 is the initial radioactivity, and A is the radioactivity
to attenuation time.The manufacturing process can be fully
automated to avoid radioactive
contamination and radiation exposure to workers.[21] However, to prevent the loss of radiation intensity, rapid
and highly efficient separation, purification, and drug synthesis
methods are required.Our group recently developed a drug synthesis
process using an
electrodialytic ion-transfer device (ITD).[22] This device was first developed by Ohira et al.[23] and has been applied to extract inorganic anions,[24] enable chromium speciation,[25] and separate weak acids.[26] The
ITD was used to demonstrate in-line probe synthesis using DOTA and
1.0 pmol/L 64Cu2+. The advantage of this process
is that the desired drug can more rapidly be obtained because it eliminates
the evaporation to dryness step, which takes time to completely remove
the strong acid in the aqueous solution. However, in the previous
work, the separation of 64Cu from a large amount of 64Ni matrix in the irradiated solid sample was still required.In the current study, we developed an effective technique for separating 64Cu from the 64Ni matrix using electrolysis instead
of the conventional method in which 64Cu is separated from 64Ni via a chelating or an ion-exchange resin column; however,
the 64Cu product still needs to be concentrated after separation.
Electrolysis can selectively separate 64Cu from the 64Ni matrix by controlling the redox potential in a handheld-sized
flow electrolysis cell (FE). The separation is based on the electrochemical
nature of Cu, which preferentially undergoes reduction compared with
Ni.[27] The FE adsorbs zerovalent Cu (Cu0) and eliminates the Ni matrix by applying a reduction potential;
divalent Cu (Cu2+) is then desorbed by applying an oxidation
potential during which acid eluent is flowed through the FE. This
study preliminarily reports that the flow electrolysis method can
achieve a highly efficient separation of ultratrace 64Cu
produced from irradiating 64NiO with a proton beam under
conditions that were optimized for the stable isotope Cu2+.
Experimental Section
Flow Electrolysis Cell
In this initial
investigation
of separating Cu from Ni via electrolysis, we employed the VF2 flow
electrolysis cell from EC Frontier, Co., Ltd. (Kyoto, Japan) (Figure a). The working electrode
was a carbon felt electrode (Φ18 mm, volume: 130 mm3, effective surface area: 1900 cm2) because it enabled
the efficient adsorption of the target element. The electrode was
fabricated from high-density carbon fibers; the fibers are intertwined
with each other, as shown in the scanning electron microscopy (SEM)
image in Figure S1. Copper that is passed
through the FE when a reduction potential is applied adsorbs to the
surface of the fiber electrode (Figures and S1). The
current variations when applying the constant voltage to the FE for
each faction were monitored by connecting the digital multimeter (PC7000,
Sanwa Electric Instrument, Co., Ltd. Tokyo, Japan) to the electrodes
of the FE. The variations were displayed by a PC that installed the
PC Link 7 version 4.1 (Sanwa Electric Instrument, Co., Ltd.).
Figure 1
Schematic of
the flow electrolysis cell: (a) overview of the FE
and (b) electrodeposition and recovery of Cu2+ from the
Ni2+ matrix.
Schematic of
the flow electrolysis cell: (a) overview of the FE
and (b) electrodeposition and recovery of Cu2+ from the
Ni2+ matrix.
Procedures
Applying
a reduction potential to the FE
reduced the Cu2+ and deposited it onto the working electrode
as a simple substance. Because of the difference in standard potentials
(ECu° = +0.340 V, ENi° = −0.257 V),[28] the Ni matrix was
kept as its ionic form, Ni2+. Applying an oxidation potential
with flowing 10 mmol/L HNO3 eluted the deposited Cu. A
schematic of the procedure is provided in Figure b.In this study, we defined a four-step
procedure as follows: the adsorption of Cu2+ to the FE
(fraction 1, F1); flushing remaining sample solution from the FE (fraction
2, F2); recovery of Cu2+ from the FE (fraction 3, F3);
and washing the residue from the FE (fraction 4, F4). The effluent
of each fraction was separately collected and analyzed to evaluate
the performance.The electrodeposition rates (ED) of
Cu2+ and Ni2+ were estimated by eq where nint (mol)
is the initial amount of metal (Cu2+ or Ni2+), and nF1 (mol) is the amount in the
effluent that passed through the FE (i.e., F1).The amount of
Cu2+ adsorbed by the FE (nads, mol) was estimated by eqThe recovery (R) of Cu2+ that was adsorbed
by the FE was calculated using eq where nF3 (mol)
is the amount of metal in the effluent when eluting with 10 mmol/L
HNO3 (F3).
Optimization of Cu Separation Conditions
in a Cold Test
The experimental conditions applied voltage
and flow rate were optimized
with a stable isotope experiment cold test. The sample solutions containing
stable isotopes were prepared from a copper standard solution (Cu(NO3)2 in 0.1 mol/L HNO3) and a nickel standard
solution (Ni(NO3)2 in 0.1 mol/L HNO3) (Fujifilm Wako Pure Chemical Industries, Tokyo, Osaka). The stable
isotope samples were 0.1 mol/L HCl solutions containing 200 μg/L
Cu2+ and 200 mg/L Ni2+. The samples were flowed
into the FE using a peristaltic pump at a rate of 0.5–1.0 mL/min.
For Cu2+ adsorption, −0.8–0 V (vs Ag/AgCl)
was applied to the FE; for its recovery, 0–1.0 V (vs Ag/AgCl)
was applied. The voltage applied to the FE was controlled by a potentiostat
(HA-151B, Hokuto Denko, Co., Ltd. Tokyo, Japan). The concentrations
of Cu2+ and Ni2+ in the solution eluted from
the FE were determined by a flame atomic absorption spectrometry (AAnalyst
200, Perkin Elmer, Waltham, MA) and inductively coupled plasma mass
spectrometry (ICP-MS, Agilent 7500cx ICP-MS, Agilent Technology, Santa
Clara, CA). Their analytical conditions are summarized in Table S1.
Sample Preparation in a
Hot Test
64Cu was
obtained by irradiating the isotopically enriched 64NiO
target (Isoflex, San Francisco, CA) with a proton beam at Takasaki
Ion Accelerators for Advanced Radiation Application facility in the
National Institute for Quantum Science and Technology, Takasaki, Japan.
The irradiated sample was dissolved in HCl with heating, and the generated 64Cu was separated as described previously.[29,30] In the current study, as a tracer of the Ni matrix, 57Ni (half-life: 35.60(6) h) was prepared as follows: (1) proton beam
irradiation of natural Ni metal, (2) dissolution of the irradiated
Ni in 6 mol/L HCl with heating, (3) evaporation to dryness, and (4)
dissolution in ultrapure water. The initial concentrations of the
test solution were 300 fmol/L 64Cu and 3.1 mmol/L Ni (including
400 fmol/L 57Ni as the tracer) in 0.1 mol/L HCl.
Separation
of 64Cu in a Hot Test
Prior to
the electrodeposition of 64Cu, 0.1 mol/L HCl was flowed
into the FE, applying −0.8 V (vs Ag/AgCl) to the FE by a potentiostat.
Next, the test solution of 5 mL was flowed into the FE using a peristaltic
pump at a flow rate of 0.5 mL/min. Subsequently, 5 mL of 0.01 mol/L
HNO3 was flowed into the FE under the same conditions;
this procedure gave F1. The sample solution remaining in the FE was
then flushed out with 10 mmol/L HNO3 (5 mL) at a flow rate
of 0.5 mL/min while applying −0.8 V. This procedure produced
F2. The Cu deposited in the FE was eluted with 0.01 mol/L HNO3 (5 mL) at a flow rate of 0.5 mL/min while applying +1.0 V
(vs Ag/AgCl). This procedure gave F3. Finally, 0.1 mol/L HCl (5 mL)
was used to flush any residue from the FE in preparation for the next
test. This procedure generated F4.The radioactivity of 64Cu and 57Ni in each fraction was measured by a
NaI (Tl) scintillation detector (AccuFLEX γ 7001, Hitachi Co.,
Ltd.). The chemical compositions of the effluent samples that were
collected after electrodeposition (F1) and after recovery (F3) were
measured by a Ge semiconductor detector coupled to a multichannel
analyzer (7800 MCA, Seiko EG&G).
Results and Discussion
Selective
Deposition and Recovery of Stable Cu in the Cold Test
In
the present method, the applied potential is the key to enable
the separation. Thus, the applied potential was first optimized by
the cold test. The electrodeposition of 200 μg/L Cu2+ and 200 mg/L Ni2+ onto the working electrode in the FE
was observed when applying a reduction potential in the range of −0.8–0
V. As shown in Figure a, more than 97% of the Cu2+ was adsorbed to the FE over
the range of −0.8 to −0.5 V, while almost all of the
Ni2+ was excluded from adsorption. As mentioned above,
this result was expected because zerovalent Cu is adsorbed onto the
working electrode when applying a reduction potential, while divalent
Ni flows through the FE.
Figure 2
Effects of applied potential on (a) electrodeposition
rate (ED) of Ni2+ and Cu2+ by
applying a negative potential and (b) recovery (R). Injected solution: (a) 200 μg/L Cu(NO3)2 and 200 mg/L Ni(NO3)2 in 0.1 mol/L HCl, (b)
10 mmol/L HNO3. Flow rate: 0.5 mL/min in both the electrodeposition
and recovery operations. Sample volume: 5 mL. A volume of 10 mmol/L
HNO3 was used to flush out the residue (F2): 5 mL. A volume
of 10 mmol/L HNO3 was used to desorb the Cu from the FE
(F3): 5 mL.
Effects of applied potential on (a) electrodeposition
rate (ED) of Ni2+ and Cu2+ by
applying a negative potential and (b) recovery (R). Injected solution: (a) 200 μg/L Cu(NO3)2 and 200 mg/L Ni(NO3)2 in 0.1 mol/L HCl, (b)
10 mmol/L HNO3. Flow rate: 0.5 mL/min in both the electrodeposition
and recovery operations. Sample volume: 5 mL. A volume of 10 mmol/L
HNO3 was used to flush out the residue (F2): 5 mL. A volume
of 10 mmol/L HNO3 was used to desorb the Cu from the FE
(F3): 5 mL.The deposition and recovery of
Cu in the FE did not significantly
change over the flow rate range of 0.2–1.0 mL/min (Figure S2). Actually, even at a flow rate of
2.0 mL/min, the effective electrodeposition of Cu2+ was
achieved (Figure S3). However, the slight
deposition of Ni2+ to FE or the unstable recovery of Cu2+ was obtained. In the hot test, the flow rate was set to
0.5 mL/min to ensure quantitative and stable Cu2+ adsorption
onto the working electrode.We next attempted to desorb the
Cu2+ that was deposited
onto the working electrode. The eluent was 10 mmol/L HNO3 and was flowed at 0.5 mL/min. Cu was eluted even without the applied
potential because the Cu deposited in the FE was dissolved by the
HNO3. The best recovery efficiency of Cu2+ was
obtained when +1.0 V was applied to the FE (Figure b), where the R estimated
from eq was 95 ±
5%.The concentration of HNO3 as the eluent was tested
at
1 and 100 mmol/L, and the Cu was completely eluted in the range from
10 to 100 mmol/L solution. However, the HNO3 concentration
should be as low as possible considering the electrical load on the
FE and the drug synthesis after recovery. From these results, the
optimal applied voltage (Vapp) was concluded
to be −0.8 V for the deposition of Cu in the FE, a Vapp of +1.0 V for recovery, 10 mmol/L HNO3 as the eluent, and a sample and eluent flow rate of 0.5 mL/min.
In addition, the eluent to recover Cu2+ deposited into
the FE can be used by a lower acid concentration compared to the column
chromatographic method (e.g., >0.1 M HCl).[19,20,31] In this method, the Cu recovery from the
FE can also be obtained by HCl, citric acid, or acetic acid (10 mM
each), as shown in Figure S4.Next,
we optimized the volumes flowed during flushing (F2), Cu
recovery (F3), and washing (F4) when 5 mL of sample containing Cu2+ and Ni2+ was flowed through the FE (F1). The
solution volume directly affects the concentration of 64Cu in the separated solution. It is also important to minimize the
radioactive waste. Each fraction was separately collected every 5
mL. As shown in Figure , no Cu2+ was eluted in F1 or F2 while Ni2+ flowed through the system and was present in F1 and F2. The Cu2+ deposited in the FE was almost entirely eluted in the first
5 mL of HNO3 (F3). The Ni2+ and Cu2+ in the effluents collected in the other fractions were below the
quantification limits in the ICP-MS used in this study. The volume
flowed for each fraction was 5 mL each from F2 to F4. The average
concentrations of metals for each fraction were 195.1 ± 4.1 mg/L
of Ni2+ in F1, 5.2 ± 0.8 mg/L of Ni2+ in
F2, and 198.4 ± 1.1 μg/L of Cu2+ in F3 at three
consecutive operations (Table S2).
Figure 3
Recoveries
(R) of Ni2+ and Cu2+ eluted
in the four fractionation procedures. A sample volume of
5 mL (0.1 mol/L HCl containing 20 μg/L Cu2+ and 200
mg/L Ni2+) was flowed into the FE (F1); 20 mL of 10 mmol/L
HNO3 was used to flush the sample residue from the FE (F2);
20 mL of 10 mmol/L HNO3 was used to elute the Cu from the
FE (F3), and 10 mL of 0.1 mol/L HCl was used to wash the FE (F4).
Flow rate: 0.5 mL/min. Applied voltage: (F1 and F2) −0.8 V,
(F3) +1.0 V, and (F4) 0 V.
Recoveries
(R) of Ni2+ and Cu2+ eluted
in the four fractionation procedures. A sample volume of
5 mL (0.1 mol/L HCl containing 20 μg/L Cu2+ and 200
mg/L Ni2+) was flowed into the FE (F1); 20 mL of 10 mmol/L
HNO3 was used to flush the sample residue from the FE (F2);
20 mL of 10 mmol/L HNO3 was used to elute the Cu from the
FE (F3), and 10 mL of 0.1 mol/L HCl was used to wash the FE (F4).
Flow rate: 0.5 mL/min. Applied voltage: (F1 and F2) −0.8 V,
(F3) +1.0 V, and (F4) 0 V.In the present method, the time from the injection (F1) to recovery
(F3) of 64Cu was 30 min at a flow rate of 0.5 mL/min, which
is almost unchanged compared to that of the conventional method with
column chromatography. By contrast, the elution of 64Cu
from the FE could easily be achieved by eluent with a lower concentration
of strong acid or a weaker acid (Figure S4), compared to the eluent used in the column chromatographic method.[19,20,31] The electrochemical oxidation/reduction
may facilitate copper adsorption and desorption of Cu to the FE. The
property may contribute to omitting evaporation to dryness for conversion
to weak acids for radiolabeling in the conventional method, which
requires the most time for separation and recovery. If the evaporation
to dryness is skipped, the attenuation of 64Cu (f) estimated by eq from the time of the dissolution of the irradiated sample
and separation to recovery by the present method can be suppressed
to 0.87–0.88.We further monitored the variation in current
as the applied potential
during the entire operation. As shown in Figure S5, the current was greatly reduced when flowing 0.1 mol/L
HCl containing Cu2+ and Ni2+ (F1). The reduction
was larger than when flowing the blank solution (0.1 mol/L HCl only).
The reduction of Cu2+ is related to the presence of protons
in an acidic solution. When flushing the residue from the FE by flowing
10 mmol/L HNO3 (F3), the current was increased to −200
μA, which was similar to that of the blank. A rapid increase
in current was observed when switching the polarity from negative
to positive potentials to desorb the Cu from the FE (F3), but the
current exponentially decreased and finally plateaued at approximately
0.5 μA.
Hot Test for Electrodeposition and Recovery
of the Radioisotope 64Cu
A hot test with the radioisotope
was performed.
Radioisotopes can be detected even at the femtomolar level. In the
present study, we found that the separation mechanism was effective
at ultratrace levels of Cu. The 0.1 mol/L HCl solutions (5 mL) containing
300 fmol/L 64Cu2+ (1.5 fmol) and 3.1 mmol/L
Ni2+ (including 400 fmol/L 57Ni2+ as the tracer) were injected into the FE under the optimized conditions
found in the stable isotope experiments.During 64Cu2+ electrodeposition (F1), Ni2+ was quantitatively
eluted, while 64Cu2+ was selectively adsorbed
in the FE (Figure ). When the residue was flushed from the FE (F2), no 64Cu2+ or Ni2+ elution was observed. This indicated
that the quantitative deposition of 64Cu and the elimination
of 64Ni quantitatively occurred during the F1 procedure.
During recovery (F3), 64Cu was quantitatively desorbed
from the FE by changing the voltage to +1.0 V (1st 64Cu
recovery: 99.3%, 2nd: 100.3%). Finally, during the washing procedure
without any applied voltage (F4), no 64Cu2+ or
Ni2+ elution was observed.
Figure 4
Recoveries of Ni matrix (57Ni as the tracer) and 64Cu in four fractions collected
from the FE. Injected sample
solutions: 300 fmol/L 64Cu2+ and 3.1 mmol/L 64Ni2+ (including 400 fmol/L 57Ni2+ as the tracer) in 0.1 mol/L HCl. Eluent: 10 mmol/L HNO3. Flow rate: 0.5 mL/min in both the electrodeposition and
recovery operations. Sample volume: 5 mL. Volumes for F2–F4:
5 mL each.
Recoveries of Ni matrix (57Ni as the tracer) and 64Cu in four fractions collected
from the FE. Injected sample
solutions: 300 fmol/L 64Cu2+ and 3.1 mmol/L 64Ni2+ (including 400 fmol/L 57Ni2+ as the tracer) in 0.1 mol/L HCl. Eluent: 10 mmol/L HNO3. Flow rate: 0.5 mL/min in both the electrodeposition and
recovery operations. Sample volume: 5 mL. Volumes for F2–F4:
5 mL each.The obtained solutions in each
protocol were also measured using
a Ge semiconductor. The γ-ray spectra were obtained from the
sample solution before injection into the FE (Figure a). The solution eluted as F1 contained 57Ni (Figure b), while not containing 64Cu. The effluent when 10 mmol/L
HNO3 was used to flush the FE (F2) did not contain either 64Cu or 57Ni (Figure c). The F3 effluent contained only 64Cu
(Figure d). These
results show that the selective and highly efficient separation of
ultratrace 64Cu was successfully achieved. On the other
hand, 55Co, which is a byproduct generated by producing 57Ni, was eluted in the F1 but was not eluted in F2 or F3.
Because E° of Co2+ is −0.277
V,[28]55Co2+ was eluted
by the procedures in the F1 as well as Ni2+. In addition,
stable Ni in the F3 was below the limit of quantification (<0.27
μg L–1) by the ICP-MS used in this study.
Figure 5
γ-ray
spectra of FE effluents as determined by a Ge semiconductor
detector. (a) A mixture of 57Ni and 64Cu in
0.1 mol/L HCl before injecting into the FE, (b) effluent after passing
the solution through the FE by applying −0.8 V (F1), (c) effluent
after flushing any remaining sample by flowing 10 mmol/L HNO3 through the FE with applying −0.8 V (F2), and (d) effluent
when flowing 10 mmol/L HNO3 with applying +1.0 V (F3).
γ-ray
spectra of FE effluents as determined by a Ge semiconductor
detector. (a) A mixture of 57Ni and 64Cu in
0.1 mol/L HCl before injecting into the FE, (b) effluent after passing
the solution through the FE by applying −0.8 V (F1), (c) effluent
after flushing any remaining sample by flowing 10 mmol/L HNO3 through the FE with applying −0.8 V (F2), and (d) effluent
when flowing 10 mmol/L HNO3 with applying +1.0 V (F3).Accordingly, in the results of the hot test, 64Cu was
not eluted in F2, and the separation is more efficient compared with
those of the cold test. The high purity of 64Cu was concluded
to be recovered in F3 of the present procedures and without the need
for heating and evaporation to dryness.
Conclusions
The
flow electrolysis cell was able to selectively separate and
recover femtomolar levels of 64Cu. This method can thus
be used as an alternative electrochemical separation method to conventional
chromatographic separation. The developed method will contribute to
the efficient manufacturing of PET diagnostic materials because the
electrochemical procedures do not require evaporation to dryness.
Therefore, we believe that this electrochemical technique will be
applied in radioactive isotope medical technologies.
Authors: Nicholas P van der Meulen; Roger Hasler; Alain Blanc; Renata Farkas; Martina Benešová; Zeynep Talip; Cristina Müller; Roger Schibli Journal: J Labelled Comp Radiopharm Date: 2019-06-30 Impact factor: 1.921
Authors: Megan N Stewart; Matthew F L Parker; Salma Jivan; Justin M Luu; Tony L Huynh; Brailee Schulte; Youngho Seo; Joseph E Blecha; Javier E Villanueva-Meyer; Robert R Flavell; Henry F VanBrocklin; Michael A Ohliger; Oren Rosenberg; David M Wilson Journal: ACS Infect Dis Date: 2019-11-21 Impact factor: 5.084