Raphaela Fritsche-Guenther1, Yoann Gloaguen1,2, Alina Eisenberger1, Jennifer A Kirwan3.
Abstract
The AbsoluteIDQ p400 HR kit is a commercial product for targeted metabolomics. While the kit has been validated for human plasma and serum, adherent cell lysates have not yet been evaluated. We have optimized the detection of polar and lipid metabolites in cell lysates using the kit to enable robust and repeatable analysis of the detected metabolites. Parameters optimized include total cell mass, loading volume and extraction solvent. We present a cell preparation and analytical method and report on the performance of the kit with regard to detectability of the targeted metabolites and their repeatability. The kit can be successfully used for a relative quantification analysis of cell lysates from adherent cells although validated only for human plasma and serum. Most metabolites are below the limit of the Biocrates' set quantification limits and we confirmed that this relative quantification can be used for further statistical analysis. Using this approach, up to 45% of the total metabolites in the kit can be detected with a reasonable analytical performance (lowest median RSD 9% and 13% for LC and FIA, respectively) dependent on the method used. We recommend using ethanol as the extraction solvent for cell lysates of osteosarcoma cell lines for the broadest metabolite coverage and 25 mg of cell mass with a loading volume of 20 µL per sample.Entities:
Mesh:
Substances:
Year: 2022 PMID: 35562573 PMCID: PMC9106690 DOI: 10.1038/s41598-022-11118-7
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Number of detected compounds for different cell masses (mg) and loading volumes (µL) for both liquid chromatography (LC) and flow injection analysis (FIA) mass spectrometry using the Biocrates p400 HR kit.
| Group | Measurement type | Compounds in total | Cell mass [mg] | |||||
|---|---|---|---|---|---|---|---|---|
| 5 | 15 | 25 | 5 | 15 | 25 | |||
| Loading volume [µL] | ||||||||
| 10 | 10 | 10 | 20 | 20 | 20 | |||
| Amino acids | LC | 21 | 16 | 18 | 19 | 16 | 18 | 19 |
| Biogenic amines | LC | 21 | 5 | 6 | 7 | 6 | 7 | 9 |
| 42 | 21 | 23 | 26 | 22 | 25 | 28 | ||
| Hexose | FIA | 1 | 0 | 0 | 0 | 0 | 0 | 1 |
| Acylcarnitines | FIA | 55 | 2 | 3 | 3 | 2 | 3 | 3 |
| Diclycerides | FIA | 18 | 3 | 4 | 4 | 4 | 5 | 5 |
| Triglycerides | FIA | 42 | 5 | 4 | 1 | 5 | 7 | 0 |
| Lysophosphatidylcholines | FIA | 24 | 5 | 5 | 4 | 4 | 5 | 8 |
| Phosphatidylcholines | FIA | 172 | 102 | 116 | 97 | 96 | 120 | 119 |
| Sphingomyelins | FIA | 31 | 14 | 16 | 9 | 10 | 16 | 15 |
| Ceramides | FIA | 9 | 1 | 0 | 0 | 1 | 0 | 0 |
| Cholesteryl Esters | FIA | 14 | 3 | 2 | 3 | 3 | 2 | 2 |
| 366 | 135 | 150 | 121 | 125 | 158 | 153 | ||
| 408 | 156 | 173 | 146 | 147 | 183 | 179 | ||
Five replicates were measured in each condition; outliers were removed from the final analysis as described in the materials and methods. BR: Biological replicates. A strict noise threshold filter was applied to data such that in some cases, compounds were detected but filtered out before these results were compiled.
Significant values are in bold.
The median relative standard deviation (RSD) of individual compounds in HOS cells and the percentage of detected compounds with RSD < 15% (in total and in % from detected number) as detected with the Biocrates p400 HR kit using either liquid chromatography (LC) for flow injection analysis (FIA) mass spectrometry.
| Measurement type | Criteria | Cell mass [mg] | |||||
|---|---|---|---|---|---|---|---|
| 5 | 15 | 25 | 5 | 15 | 25 | ||
| Loading volume [µL] | |||||||
| 10 | 10 | 10 | 20 | 20 | 20 | ||
| LC | Median RSD [%] | 19 | 22 | 13 | 25 | 9 | 10 |
| Number of compounds with RSD < 15% | 4/20 (20%) | 3/23 (13%) | 16/26 (62%) | 5/22 (23%) | 21/25 (84%) | 22/27 (81%) | |
| FIA | Median RSD [%] | 23 | 13 | 16 | 24 | 31 | 13 |
| Number of compounds with RSD < 15% | 26/136 (20%) | 92/150 (61%) | 52/121 (44%) | 21/125 (17%) | 12/158 (8%) | 91/153 (59%) | |
Five replicates were measured in each condition; outliers were removed from the final analysis as described in the materials and methods.
Figure 1Boxplots of individual compounds’ relative standard deviation (RSD) using different cell masses (mg) and loading volumes (µL). Dashed line represents the quality assurance threshold of 15%. Five replicates were measured in each condition; outliers were removed from the final analysis as described in the materials and methods. Highlighted in green is the combination we consider the best. LC: liquid chromatography. FIA: flow injection analysis.
Figure 4Summary of the obtained results from both the cell and volume determination (study 1) and the extraction solvent determination (study 2) in HOS cells. Red stars indicate the best condition either in study 1 or study 2 for the analyzed objects. LC: liquid chromatography. FIA: flow injection analysis.
Figure 2Number of detected compounds per biological group using different cell lysis solvents. Five replicates were measured in each condition; outliers were removed from the final analysis as described in the materials and methods. A strict noise threshold filter was applied to data such that in some cases, compounds were detected but filtered out before these results were compiled.
The median relative standard deviation (RSD) of the individual compounds in HOS cells and the percentage of compounds with a RSD < 15% threshold (in total and in % from detected number) for liquid chromatography (LC) and flow injection analysis (FIA) mass spectrometry.
| Measure type | Lysis buffer | EtOH-P | EtOH | MeOH |
|---|---|---|---|---|
| LC | Median RSD [%] | 25 | 9 | 11 |
| Compounds with RSD < 15% [%] | 3/26 (12%) | 23/27 (85%) | 23/26 (88%) | |
| FIA | Median RSD [%] | 43 | 23 | 30 |
| Compounds with RSD < 15% [%] | 3/24 (13%) | 16/77 (21%) | 4/55 (7%) |
Five replicates were measured in each condition; outliers were removed from the final analysis as described in the materials and methods.
Figure 3Results of an analysis of HOS cells using the Biocrates p400 HR kit using different cell lysis extraction solvents. Median relative standard deviation (RSD) of the individual compounds per biological group. Both liquid chromatography (LC) and flow injection analysis (FIA) mass spectrometry measurements are shown. Five replicates were measured in each condition; outliers were removed from the final analysis as described in the materials and methods.