Frédéric Cornaz1,2,3, Marie-Rosa Fasser2,3, Jess Gerrit Snedeker1,2, José Miguel Spirig1, Mazda Farshad1, Jonas Widmer4,5.
Abstract
Posterior screw-rod constructs can be used to stabilize spinal segments; however, the stiffness is not absolute, and some motion can persist. While the effect of crosslink-augmentation has been evaluated in multiple studies, the fundamental explanation of their effectiveness has not been investigated. The aim of this study was to quantify the parameters "screw rotation" and "parallelogram deformation" in posterior instrumentations with and without crosslinks to analyze and explain their fundamental effect. Biomechanical testing of 15 posteriorly instrumented human spinal segments (Th10/11-L4/L5) was conducted in axial rotation, lateral bending, and flexion-extension with ± 7.5 Nm. Screw rotation and parallelogram deformation were compared for both configurations. Parallelogram deformation occurred predominantly during axial rotation (2.6°) and was reduced by 60% (-1.45°, p = 0.02) by the addition of a crosslink. Simultaneously, screw rotation (0.56°) was reduced by 48% (-0.27°, p = 0.02) in this loading condition. During lateral bending, 0.38° of parallelogram deformation and 1.44° of screw rotation was measured and no significant reduction was achieved by crosslink-augmentation (8%, -0.03°, -p = 0.3 and -13%, -0.19°, p = 0.7 respectively). During flexion-extension, parallelogram deformation was 0.4° and screw rotation was 0.39° and crosslink-augmentation had no significant effect on these values (-0.12°, -30%, p = 0.5 and -0°, -0%, p = 0.8 respectively). In axial rotation, crosslink-augmentation can reduce parallelogram deformation and with that, screw rotation. In lateral bending and flexion-extension parallelogram deformation is minimal and crosslink-augmentation has no significant effect. Since the relatively large screw rotation in lateral bending is not caused by parallelogram deformation, crosslink-augmentation is no adequate countermeasure. The fundamental understanding of the biomechanical effect of crosslink-augmentation helps better understand its potential and limitations in increasing construct stiffness.Entities:
Mesh:
Year: 2022 PMID: 35538122 PMCID: PMC9090827 DOI: 10.1038/s41598-022-11719-2
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.996
Figure 1(A) Segmental deformation after posterior instrumentation during physiological loading (± 7.5 Nm)[1]. (B) Hypothezides construct deformations due to the bending forces in the three major rotational motion planes (figure adapted from[3]). (C) Illustration of the angular displacement of the pedicle screw in relation to the vertebral body (“screw rotation”) and the relative motion between one side of the screw-rod-construct to the other (“parallelogram deformation”).
Overview on the spinal segments used for this study and the pedicle screw rajectories used for instrumentation.
| # | Demographics | Th9/10 | Th10/11 | Th11/12 | Th12/L1 | L1/2 | L2/3 | L3/4 | L4/5 |
|---|---|---|---|---|---|---|---|---|---|
| 1 | 65 years, female | – | TT | – | CBT | – | TT | – | CBT |
| 2 | 45 years, female | – | CBT | – | TT | – | CBT | – | TT |
| 3 | 62 years, male | CBT | – | TT | – | CBT | – | TT | – |
| 4 | 64 years, male | Excluded | – | CBT | – | TT | – | CBT | – |
TT traditional trajectory, CBT cortical bone trajectory.
Figure 2Image of a dorsally instrumented spinal segment (A) without and (B) with crosslink-augmentation.
Figure 3(A) Image of the biomechanical test setup with the mounting position to perform flexion–extension loading. (B) Telecentric camera system. (C) Telecentric image of an illustrative specimen with the six crosshair-labels to measure translational and rotational displacement of the two vertebral bodies and the four screw heads.
Figure 4Effect on (A) parallelogram deformation and (B) screw rotation by the addition of a crosslink to the construct. Significant differences (p < 0.05) are marked with an asterisk (*), wo-CL configuration without crosslink, CL configuration with crosslink.
Comparison of parallelogram deformation and screw rotation in segments instrumented with the traditional trajectory (TT) and segments instrumented with the cortical bone trajectory (CBT).
| Without crosslink | Effect of crosslink | ||||||
|---|---|---|---|---|---|---|---|
| Loading | TT | CBT | p-value | TT | CBT | p-value | |
| Parallelogram deformation | AR | 2.76° (0.97°, 2.84°) | 2.39° (1.37°, 5.24°) | 0.69 | 0.94° (0.39°, 1.40°) | 0.88° (0.58°, 2.51°) | 0.45 |
| LB | 0.36° (0.10°, 0.38°) | 0.63° (0.26°, 1.13°) | 0.02 | 0.03° (−0.17°, 0.29°) | 0.11° (−0.12°, 0.59°) | 0.45 | |
| FE | 0.41° (0.11°, 1.01°) | 0.33° (0.08°, 0.87°) | 0.69 | 0.26° (−0.15°, 0.60°) | −0.03° (−0.34°, 0.33°) | 0.52 | |
| Screw rotation | AR | 0.75° (0.24°, 1.18°) | 0.54° (0.21°, 1.63°) | 0.69 | 0.48° (−0.15°, 0.73°) | 0.23° (−0.23°, 1.40°) | 0.86 |
| LB | 1.25° (0.88°, 2.52°) | 1.60° (0.66°, 1.89°) | 0.60 | 0.08° (−0.23°, 0.31°) | 0.10° (−0.63°, 0.55°) | 0.95 | |
| FE | 0.40° (0.30°, 0.71°) | 0.26° (0.15°, 0.54°) | 0.06 | −0.04° (−0.44°, 0.26°) | 0.01° (−0.31°, 0.14°) | 0.86 | |
The median (minimum, maximum) values are reported for each group.
AR axial rotation, LB lateral bending, FE flexion–extension.