Kaikai Liu1,2,3, Xiaojing Jiang2,3, Yuehai Song1, Rongning Liang2,3.
Abstract
Currently, nanomaterial-based all-solid-state ion-selective electrodes (ASS-ISEs) have become attractive tools for ion sensing in environmental and biological applications. However, nanomaterial solid contact can easily fall off the electrode surface owing to poor adhesion. This poses serious limits to the wide use of these sensors. Herein, we report a general and facile method for the robust fabrication of nanomaterial-based ASS-ISEs. It is based on the silver-based conductive adhesive (CA) with excellent electronic conductivity and strong adhesion ability as the binder to construct nanomaterial-based solid contact. The solid-contact Ca2+-ISE based on single-walled carbon nanotubes (SWCNTs) is chosen as a model. The proposed electrode based on CA-SWCNTs shows a linear response in the concentration range of 10-6 to 10-3 M with a slope of 25.96 ± 0.36 mV per decade and a detection limit of 1.7 × 10-7 M. In addition, the CA-SWCNT-based Ca2+-ISE exhibits an improved potential stability and reduced water film compared to the coated-wire ISE. Above all, experiments also show that the CA-SWCNT-based electrode exhibits nearly the same electrochemical characteristics as the classical only SWCNT-based electrode in term of resistance, capacitance and potential stability. We believe that CA-nanomaterial-based solid contacts provide an appealing substitute for traditional solid contacts based on nanomaterials. This journal is © The Royal Society of Chemistry.Entities:
Year: 2019 PMID: 35516404 PMCID: PMC9064413 DOI: 10.1039/c9ra02770j
Source DB: PubMed Journal: RSC Adv ISSN: 2046-2069 Impact factor: 3.361
Fig. 1SEM images of CA/SWCNT- (a) and SWCNT-based (b) solid contacts.
Fig. 2Impedance spectra date for the proposed Ca2+-ISE based on GC/Ca2+-ISE (a), GC/CA-SWCNTs/Ca2+-ISE (b) and GC/SWCNTs/Ca2+-ISE (c). The inset is for the low frequency data of line (a).
Fig. 3Effects of O2, CO2 and light on the potential stability of the GC/CA-SWCNTs/Ca2+-ISE in 1.0 × 10−3 M CaCl2.
Fig. 4Chronopotentiograms for GC/CA-SWCNTs/Ca2+-ISE (a), GC/SWCNTs/Ca2+-ISE (b) and GC/Ca2+-ISE (c) recorded in a 1.0 × 10−3 M CaCl2 solution. Applied current, ±1 nA for 60 s, respectively.
Fig. 5Time traces (A) and calibration curves (B) of the GC/Ca2+-ISE (a), GC/CA-SWCNTs/Ca2+-ISE (b) and the GC/SWCNTs/Ca2+-ISE (c) in CaCl2 in the concentration range of 10−8 to 10−3 M.
Fig. 6Water layer tests for GC/CA-SWCNTs/Ca2+-ISE (a) and GC/Ca2+-ISE (b).