| Literature DB >> 35514682 |
Jie-Qiong Zhang1, Ming-Hui Zhou1, Wei Tian1, Yan-Xiang Wu1, Xi Chen1, Song-Xue Wang1.
Abstract
In China, the production has not realized intensive cultivation and the problem of cadmium (Cd)-contaminated rice is salient, so it is important to classify rice with different degrees of Cd pollution by rapid detection method in situ. This paper established a method with a combination of dilute acid extraction pretreatment and electrochemical devices. Cd was extracted from rice using 3% HCl for 5 min. A standard curve was obtained based on a certified reference material in the rice matrix with different concentrations of Cd, which was fitted with the Cd concentration (μg kg-1) against the stripping peak current value (μA), and the linear correlation coefficient was 0.9997. To analyze the applicability of the method, three factors including substrate diluents, particle diameter of the sample, and stability towards the method were evaluated. The limit of detection (LOD) was 2.02 μg kg-1, and the repeatability and accuracy were satisfactory. Cd was determined in 142 samples collected from three major grain-producing provinces of China, and the results have good consistence with the microwave digestion-ICP-MS method. The developed method combined dilute acid extraction with a matrix matching standard curve in ASV for the first time, and it was significantly satisfactory for the detection requirements in China. This journal is © The Royal Society of Chemistry.Entities:
Year: 2019 PMID: 35514682 PMCID: PMC9065540 DOI: 10.1039/c9ra03073e
Source DB: PubMed Journal: RSC Adv ISSN: 2046-2069 Impact factor: 4.036
Fig. 1The schematic diagram of the working principle of this method.
Dialysis results of acid extraction solution (n = 3)
| Sample number | Reference values (μg kg−1) | Solid : liquid ratio | Cd concentration (ng mL−1) | Ratio of dialysis (%) | |
|---|---|---|---|---|---|
| Before dialysis | After dialysis | ||||
| GBW(E)100377 | 261 ± 20 | 1 : 10 | 26.3 ± 2.1 | 2.2 ± 0.1 | 91.6 |
| GBW(E)100348 | 240 ± 10 | 24.4 ± 0.4 | 0.4 ± 0.1 | 98.4 | |
| IRMM-804 | 1610 ± 70 | 164.0 ± 4.0 | 13.6 ± 0.1 | 91.7 | |
Selection of acid type (n = 3)
| Acid type | 1% HCl | 1% HNO3 | 1% HClO4 |
|---|---|---|---|
| Deposition potential/mv | −592 | −574 | −578 |
| Stripping peak current/μA | 2.37 ± 0.02 | 1.78 ± 0.04 | 1.56 ± 0.04 |
Fig. 2The effects of acid concentration on the determination results (n = 3).
Fig. 3The effects of extraction time on the determination results (n = 3).
Cd contents in GBW10045 as determined by different curves (n = 3)
| Stripping peak current (μA) | Matrix matching standard curve (μg kg−1) | Solution standard curve (μg kg−1) | Reference values (μg kg−1) |
|---|---|---|---|
| 1.84 ± 0.04 | 194.0 ± 3.50 | 74.7 ± 1.60 | 190.0 ± 20.0 |
Fig. 4Anodic stripping voltammograms and standard curve. A: anodic stripping voltammograms for extracted solutions of reference materials containing different concentrations of Cd. B: the corresponding fitted linear curve.
Fig. 5Matrix interferences.
Determination of IRRM-804 diluted by various sample diluents (n = 3)
| Dilution ratio | Determination of Cd by ASV (mg kg−1) | |
|---|---|---|
| Rice matrix diluent | Aqueous solvent diluent | |
| 1 : 2 | 1.52 ± 0.08 | 2.08 ± 0.04 |
| 1 : 5 | 1.61 ± 0.05 | 2.11 ± 0.05 |
| 1 : 10 | 1.59 ± 0.04 | 2.09 ± 0.06 |
| 1 : 20 | 1.55 ± 0.07 | 2.14 ± 0.04 |
| Reference value of IRRM-804 | 1.61 ± 0.07 | |
Fig. 6Comparison of the determination results of samples with different particle sizes by ASV, AAS, and ICP-MS. s1 denotes rice sample 1 and s2 denotes rice sample 2.
Fig. 7Effects of different sedimentation patterns on detection.
Comparison of LODs and LOQs for different methods
| Methods | LOD (μg L−1) | LOQ (μg L−1) |
|---|---|---|
| ASV | 0.20 | 0.62 |
| ICP-MS | 0.05 | 0.17 |
| AAS | 0.025 | 0.085 |
Results of ASV detection compared with the values for the reference material (n = 3)
| CRM no. | Matrix | Results ± SD (mg kg−1) | Reference values (mg kg−1) |
|---|---|---|---|
| GBW(E)100378 | Brown rice | 0.164 ± 0.01 | 0.169 ± 0.015 |
| GBW10045 | Rice | 0.199 ± 0.01 | 0.190 ± 0.02 |
| IRMM-804 | Rice | 1.63 ± 0.04 | 1.61 ± 0.07 |
Fig. 8Comparison of Cd results between ICP-MS and ASV.