E Randy Craven1, Vikas Chopra2, Jeffrey L Goldberg3, Kenneth M Marion4, Xiaoming Chen5, Cheng-Tao Chang6, Michelle Y Chen6.
Abstract
Purpose: To quantitatively compare iridocorneal angle assessments using gonioscopy and anterior segment optical coherence tomography (AS-OCT). Patients: US and Chinese patients with open-angle glaucoma (OAG) and/or ocular hypertension (OHT).Entities:
Keywords: angle assessment; anterior segment optical coherence tomography; gonioscopy; open-angle glaucoma
Year: 2022 PMID: 35510274 PMCID: PMC9058653 DOI: 10.2147/OPTH.S322962
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Description of the Shaffer Grading System
| Shaffer Grade | Angular Width | Description | Angle-Closure Risk |
|---|---|---|---|
| 0 | 0° | Closed: No angular structures visible and the iris is against the trabecular meshwork | Closed |
| 1 | <10° | Very narrow angle: Schwalbe’s line is visible | High |
| 2 | 10–20° | Narrow angle: Trabecular meshwork is visible | Possible, but not likely |
| 3 | 20–34° | Wide open: Scleral spur is visible | No risk |
| 4 | >35° | Wide open: All structures clearly visible | No risk |
Figure 1(A) An AS-OCT image of the iridocorneal angle measured using AOD-SL, which can vary depending on the visibility of the image, identification of multiple landmarks, and slope of the curvature post Schwalbe’s line, (B) and the AS-OCT angle width measurement, the diameter of the largest circle that would fit between Schwalbe’s line and iris.
Equations Used to Determine Sensitivity, PPV, and NPV
| True positive/(true positive + false negative) | |
| True positive/(true positive + false positive) | |
| True negative/(true negative + false negative) |
Example of the 2×2 Table
| Shaffer Grade <3 | Shaffer Grade ≥3 | |
|---|---|---|
| False Negative (FN) | True Positive (TP) | |
| True Negative (TN) | False Positive (FP) |
Patient Characteristics of the Pooled Study Population with at Least 1 Goniometric Measurement and Anterior Segment Optical Coherence Tomography
| Characteristic | Study Population (N = 239) |
|---|---|
| Age (years) | |
| Mean (SD) | 52.2 (16.6) |
| Range | 20–92 |
| Male, n (%) | 127 (53.1%) |
| Race, n (%) | |
| Asian | 161 (67.4) |
| White | 65 (27.2) |
| Black | 13 (5.4) |
| Ocular diagnosis, n (%) | |
| Open-angle glaucoma | 190 (80.5) |
| Ocular hypertension | 46 (19.5) |
| Unknowna | 3 (1.0) |
| Pseudophakic, n (%) | |
| Yes | 20 (8.4) |
Note: aDiagnosis not recorded on the clinical report form.
The Range (Min, Max), Median, Mean (SD), and Number of AS-OCT Angle Width Measurements in Each Shaffer Grade from the Pooled Study
| Shaffer Grade | AS-OCT Angle Width (μm) | Number of OCT Measurements in Each Shaffer Grade | ||
|---|---|---|---|---|
| Min, Max | Median | Mean (SD) | ||
| 175.82, 500.47 | 336.18 | 357 (117) | 6 | |
| 152.51, 1221.89 | 347.92 | 427 (215) | 37 | |
| 73.11, 1321.47 | 619.80 | 636 (245) | 196 | |
Abbreviation: AS-OCT, anterior segment optical coherence tomography.
Figure 2AS-OCT angle width measurements (µm) vs Shaffer Grade (2x2 table). Visualization of TP, TN, FP, and FN when the AS-OCT angle width measurement 300 µm corresponds to Shaffer Grade ≥3. Table overlayed on a visualization of the distribution of the AS-OCT angle width measurements in each Shaffer grade in the pooled study.
The Sensitivity, PPV, and NPV at Different AS-OCT Angle Width Measurements (µm) Corresponding to Shaffer Grade ≥3
| OCT Angle Width (µm) | ≤300 | ≤400 | ≤500 | ≤650 |
|---|---|---|---|---|
| 97.6 | 98.9 | 99.3 | 100.0 | |
| 88.0 | 75.1 | 63.1 | 41.6 | |
| 16.7 | 66.7 | 83.3 | 100.0 |
Abbreviation: AS-OCT, anterior segment optical coherence tomography.
Distribution (Percentage) of Shaffer Grade by Quadrant in the Single-Center Study
| Quadrant | Shaffer Grade | Total | ||
|---|---|---|---|---|
| 2 | 3 | 4 | ||
| 2 (9.5%) | 2 (9.5%) | 17 (81.0%) | 21 | |
| 3 (12.0%) | 2 (8.0%) | 20 (80.0%) | 25 | |
| 2 (11.8%) | 0 (0.0%) | 15 (88.0%) | 17 | |
| 5 (19.2%) | 4 (15.4%) | 17 (65.4%) | 26 | |
Figure 3A visualization of the distribution of the AS-OCT angle width measurements in each Shaffer grade in the single-center study.
The Median, and Mean (SD) AS-OCT Angle Width Measurements for Each Shaffer Grade by Each Quadrant in the Single-Center Study
| Inferior | Nasal | Superior | Temporal | |
|---|---|---|---|---|
| 122.7 | 110.3 | 143.4 | 212.2 | |
| 122.7 (4.5) | 109.3 (2.0) | 143.4 (19.5) | 189.7 (50.6) | |
| 485.3 | 252.3 | – | 303.3 | |
| 485.3 (319.2) | 252.4 (11.9) | 315.4 (84.4) | ||
| 679.5 | 517.0 | 625.7 | 613.8 | |
| 686.9 (296.6) | 587.6 (241.4) | 667.1 (280.4) | 673.4 (298.1) |
Abbreviation: AS-OCT, anterior segment optical coherence tomography.