| Literature DB >> 35497695 |
Balbir Kumar1, Brajesh K Ratre2, Rakesh Garg2, Jitendra K Meena3, Ram Singh1, Sushma Bhatnagar2.
Abstract
Background and Aims: The risk of contracting infection while intubating a coronavirus disease 2019 (COVID-19)-positive patient can be reduced by the use of personal protective equipment (PPE), video laryngoscope (VL) and aerosol-preventing intubation box. We compared two VLs (C-MAC and King Vision laryngoscope [KVL]) for ease of intubation and time taken to intubate the manikin using an intubation box.Entities:
Keywords: COVID-19; endotracheal intubation; laryngoscope; manikin; personal protective equipment
Year: 2022 PMID: 35497695 PMCID: PMC9053892 DOI: 10.4103/ija.ija_832_21
Source DB: PubMed Journal: Indian J Anaesth ISSN: 0019-5049
Figure 1CONSORT flow diagram. CONSORT = Consolidated standards of reporting trials, KVL = King Vision laryngoscope
Figure 2Intubation of manikin, wearing proper personal protection equipment
Figure 3Manikin with transparent aerosol box/intubation box
Study parameters in the usage of two devices for tracheal intubation: C-MAC versus KVL
| Study parameters | Attempt number | Group C-MAC ( | Group KVL ( |
|
|---|---|---|---|---|
| Time taken for successful tracheal intubation, s (mean±SD) | A | 22.65±6.04 | 31.97±28.88 | 0.652 |
| B | 20.35±4.71 | 26.06±19.43 | 0.211 | |
| C | 19.26±4.90 | 21.48±8.22 | 0.562 | |
| POGO score (mean±SD) | A | 85.48±10.90 | 88.06±21.51 | 0.554 |
| B | 86.13±19.77 | 83.55±26.90 | 0.631 | |
| C | 87.74±21.55 | 87.74±22.76 | 0.747 | |
| Optimisation manoeuvres (not performed), | A | 25 (80.6%) | 25 (83.3%) | 0.957 |
| B | 30 (96.8%) | 27 (90.0%) | 0.285 | |
| C | 29 (93.5%) | 29 (96.7%) | 0.573 | |
| Successful tracheal intubation with a single attempt, | A | 31 (100%) | 27 (90%) | 0.196 |
| B | 27 (87.1%) | 28 (93.3%) | 0.414 | |
| C | 27 (87.1%) | 29 (96.7%) | 0.173 | |
| Dental click (nil), | A | 28 (90.3%) | 25 (80.6%) | 0.319 |
| B | 31 (100%) | 28 (90.3%) | 0.412 | |
| C | 27 (87.1%) | 29 (93.5%) | 0.744 | |
| Participant reported difficulty in using the device (mean±SD) | A | 1.52±0.76 | 2.19±1.13 | 0.011 |
| B | 1.35±0.75 | 2.03±1.16 | 0.010 | |
| C | 1.42±0.72 | 2.10±1.10 | 0.006 | |
| Ease of insertion of laryngoscope blade (yes: no), | A | 25 (80.6) | 7 (22.6) | <0.001 |
| B | 23 (74.2) | 8 (25.8) | <0.001 | |
| C | 23 (74.2) | 8 (25.8) | <0.001 |
KVL=King Vision laryngoscope, POGO=percentage of glottic opening, SD=standard deviation