| Literature DB >> 35496770 |
William J Snelling1, Arsalan Afkhami2, Hannah L Turkington3, Claire Carlisle3, S Louise Cosby3, Jeremy W J Hamilton2, Nigel G Ternan1, Patrick S M Dunlop2.
Abstract
There is strong evidence that SARS-CoV-2 is spread predominantly by airborne transmission, with high viral loads released into the air as respiratory droplets and aerosols from the infected subject. The spread and persistence of SARS-CoV-2 in diverse indoor environments reinforces the urgent need to supplement distancing and PPE based approaches with effective engineering measures for microbial decontamination - thereby addressing the significant risk posed by aerosols. We hypothesized that a portable, single-pass UVC air treatment device (air flow 1254 L/min) could effectively inactivate bioaerosols containing bacterial and viral indicator organisms, and coronavirus without reliance on filtration technology, at reasonable scale. Robust experiments demonstrated UVC dose dependent inactivation of Staphylococcus aureus (UV rate constant (k) = 0.098 m2/J) and bacteriophage MS2, with up to 6-log MS2 reduction achieved in a single pass through the system (k = 0.119 m2/J). The inclusion of a PTFE diffuse reflector increased the effective UVC dose by up to 34% in comparison to a standard Al foil reflector (with identical lamp output), resulting in significant additional pathogen inactivation (1-log S. aureus and MS2, p < 0.001). Complete inactivation of bovine coronavirus bioaerosols was demonstrated through tissue culture infectivity (2.4-log reduction) and RT-qPCR analysis - confirming single pass UVC treatment to effectively deactivate coronavirus to the limit of detection of the culture-based method. Scenario-based modelling was used to investigate the reduction in risk of airborne person-to-person transmission based upon a single infected subject within the small room. Use of the system providing 5 air changes per hour was shown to significantly reduce airborne viral load and maintain low numbers of RNA copies when the infected subject remained in the room, reducing the risk of airborne pathogen transmission to other room users. We conclude that the application of single-pass UVC systems (without reliance on HEPA filtration) could play a critical role in reducing the risk of airborne pathogen transfer, including SARS-CoV2, in locations where adequate fresh air ventilation cannot be implemented.Entities:
Keywords: Air sterilisation; Air treatment; MS2; PTFE; SARS-CoV-2; UVC
Year: 2022 PMID: 35496770 PMCID: PMC9040443 DOI: 10.1016/j.jaerosci.2022.106003
Source DB: PubMed Journal: J Aerosol Sci ISSN: 0021-8502 Impact factor: 4.586
Fig. 1Schematic representation of the UVC air treatment device.
Fig. 2Irradiance profile within the air treatment chamber demonstrating a significant enhancement in UVC irradiance upon inclusion of PTFE liner in comparison to Al foil.
Calculated UV dose relative to number of lamps and reflector material.
| Lamps | Active chambers | Exposure time (s) | UV dose Al (mJ/cm2) | UV dose PTFE (mJ/cm2) |
|---|---|---|---|---|
| 1 | 1 | 0.29 | 3.72 | 5.00 |
| 2 | 1 | 0.29 | 7.42 | 8.27 |
| 3 | 1 & 2 | 0.59 | 11.14 | 13.27 |
| 4 | 1 & 2 | 0.59 | 14.84 | 16.54 |
| 5 | 1 & 2 & 3 | 0.86 | 18.56 | 21.54 |
Reduction in number of coliphage MS2 with respect to UVC dose and reflector material.
| Reflective surface | UV dose (mJ/cm2) | Mean log count (PFU/mL) ± SD | Log reduction |
|---|---|---|---|
| – | 0 | 6.48 ± 0.17 | – |
| Al foil | 3.72 | 4.43 ± 0.05 | 2.05 |
| PTFE | 5.00 | 3.16 ± 0.11 | 3.32 |
| Al foil | 7.42 | 3.51 ± 0.10 | 2.97 |
| PTFE | 8.27 | 2.32 ± 0.09 | 4.16 |
| Al foil | 11.14 | 0 | 6.48 |
| PTFE | 13.27 | 0 | 6.48 |
| Al foil | 14.84 | 0 | 6.48 |
| PTFE | 16.54 | 0 | 6.48 |
| Al foil | 18.56 | 0 | 6.48 |
| PTFE | 21.54 | 0 | 6.48 |
Reduction in number of S. aureus with respect to UVC dose and reflector material.
| Reflective surface | UV dose (mJ/cm2) | Mean log count (CFU/mL) ± SD | Log reduction |
|---|---|---|---|
| – | 0 | 4.70 ± 0.16 | – |
| Al foil | 3.72 | 4.29 ± 0.01 | 0.41 |
| PTFE | 5.00 | 3.56 ± 0.03 | 1.14 |
| Al foil | 7.42 | 1.54 ± 0.34 | 3.16 |
| PTFE | 8.27 | 0 | 4.70 |
| Al foil | 11.14 | 0 | 4.70 |
| PTFE | 13.27 | 0 | 4.70 |
| Al foil | 14.84 | 0 | 4.70 |
| PTFE | 16.54 | 0 | 4.70 |
| Al foil | 18.56 | 0 | 4.70 |
| PTFE | 21.54 | 0 | 4.70 |
Fig. 3Survival curve for MS2 and S. aureus bioaerosols under UVC exposure. Trendlines represent curve fits to the exponential decay equation (Eq (1)).
Fig. 4Example optical microscopy images of the BCoV tissue infectivity assay (×10 magnification), A) Negative control; B) Positive control; C) Air only (0 mJ/cm2), note the mix of viable cells and floating inactivated cells; D) UVC treated sample (5.0 mJ/cm2), no evidence of cytopathic effect due to BCoV inactivation. (Scale bar = 0.1 mm)
Viability assessment of bovine CoV following UVC treatment.
| UV dose (mJ/cm2) | Mean TCID50/mL | Log reduction | Mean qRT-PCR Ct ± SD |
|---|---|---|---|
| 0 | 2.40 ± 0.14 | – | |
| 5.00 | 0.00 | 2.40 | 29.80 ± 0.99 |
| 8.27 | 0.00 | 2.40 | 25.20 ± 7.35 |
| 13.27 | 0.00 | 2.40 | 30.65 ± 0.07 |
Fig. 5Modelled effect of UVC prototype to reduce aerosolised SARS-CoV-2 from a single infector in a 15 m3 room. A: Device turned off; B: Device switched on at T0; C) Device switched on at T 2.5 h (Modelling conditions: Well-mixed air, shedding rate of 29.2 RNA copies/s, removal due to biological decay, natural particle deposition and UVC treatment, UV dose of 5 mJ/cm2, fan flow rate of 1254 L/min)