| Literature DB >> 35492834 |
Alex J Nusbickel1, Steven J Ross2, William M Miles2, Kun Xiang2.
Abstract
Entities:
Keywords: Bradycardia; Capture threshold; Electrophysiology devices; Leadless pacemaker; Pacemaker implantation; Pacing
Year: 2021 PMID: 35492834 PMCID: PMC9039530 DOI: 10.1016/j.hrcr.2021.12.004
Source DB: PubMed Journal: HeartRhythm Case Rep ISSN: 2214-0271
Figure 1The Micra leadless pacemaker (Medtronic Inc, Minneapolis, MN): before (A) and after (B) pull-and-hold test. Note the 4 tines (black arrows) on the distal end of the Micra device that adhere to the myocardium. Opening of the tines after the pull-and-hold test, as seen in Figure 2B, is an indicator of adequate fixation.
Heart rate vs capture threshold of Micra pacemaker, during the periprocedural period and 2 weeks following procedure
| Heart rate (beats/min) | Capture threshold (V) | ||||
|---|---|---|---|---|---|
| Intraprocedural | Immediately postprocedure | Three hours postprocedure | One day postprocedure | Two weeks postprocedure | |
| 100 | 0.5 (@ 0.24 ms) | 0.5 (@ 0.24 ms) | 0.5 (@ 0.24 ms) | 1.25 (@ 0.24 ms) | |
| 90 | 1.75 (@ 0.4 ms | ||||
| 80 | 1.5 (@ 0.24 ms) | ||||
| 60 | 2.5 (@ 0.24 ms) | 2.5 (@ 0.4 ms) | 2.5 (@ 0.4 ms) | 1.13 (@ 0.4 ms), 0.88 (@ 0.4 ms) | |
Figure 2Chest radiographs demonstrating location of the Micra pacemaker (Medtronic Inc, Minneapolis, MN), both immediately (A) and 1 day (B) following implantation. The position of the Micra pacemaker remained unchanged on chest radiograph (arrow).