| Literature DB >> 3548322 |
M T Kelly, J A Smith, S Jaffer, L Pearce, A Clarke.
Abstract
A latex agglutination test (Marion Laboratories) was compared with standard culture methods for the detection of Group A streptococci in two studies of 500 throat swabs each. Swabs were first inoculated to sheep blood agar and then tested for Group A streptococcal antigen. The direct test performed with nearly identical sensitivity and specificity in the two phases of the study. Overall, Group A streptococci were isolated from 91 specimens, and 81 (89%) of these were detected by the direct latex test. The predictive value positive for the latex test was 90%, and the accuracy was 98%. The sensitivity of the latex test for detection of specimens having ten or more colonies of Group A streptococci was 95%. Non-Group A beta-hemolytic streptococci were isolated from 65 specimens, and all of these specimens had negative latex tests. The authors' findings suggest that this direct latex agglutination test is a reliable screening method for rapid detection of Group A streptococci in outpatient throat specimens.Entities:
Mesh:
Year: 1987 PMID: 3548322 DOI: 10.1093/ajcp/87.4.522
Source DB: PubMed Journal: Am J Clin Pathol ISSN: 0002-9173 Impact factor: 2.493