Adenosine triphosphate (ATP) is an important rapid signaling molecule involved in a host of pathologies in the body. Historically, ATP is difficult to directly detect electrochemically with fast-scan cyclic voltammetry (FSCV) due to limited interactions at bare carbon-fibers. Systematic investigations of how ATP interacts at electrode surfaces is necessary for developing more sensitive electrochemical detection methods. Here, we have developed gold nanoparticle (AuNP), and platinum nanoparticle (PtNP) modified carbon-fiber microelectrodes coupled to FSCV to measure the extent to which ATP interacts at metal nanoparticle-modified surfaces and to improve the sensitivity of direct electrochemical detection. AuNP and PtNPs were electrodeposited on the carbon-fiber surface by scanning from -1.2 to 1.5 V for 30 s in 0.5 mg/mL HAuCl4 or 0.5 mg/mLK2PtCl6. Overall, we demonstrate an average 4.1 ± 1.0-fold increase in oxidative ATP current at AuNP-modified and a 3.5 ± 0.3-fold increase at PtNP-modified electrodes. Metal nanoparticle-modified surfaces promoted improved electrocatalytic conversion of ATP oxidation products at the surface, facilitated enhanced adsorption strength and surface coverage, and significantly improved sensitivity. ATP was successfully detected within living murine lymph node tissue following exogenous application. Overall, this study demonstrates a detailed characterization of ATP oxidation at metal nanoparticle surfaces and a significantly improved method for direct electrochemical detection of ATP in tissue.
Adenosine triphosphate (ATP) is an important rapid signaling molecule involved in a host of pathologies in the body. Historically, ATP is difficult to directly detect electrochemically with fast-scan cyclic voltammetry (FSCV) due to limited interactions at bare carbon-fibers. Systematic investigations of how ATP interacts at electrode surfaces is necessary for developing more sensitive electrochemical detection methods. Here, we have developed gold nanoparticle (AuNP), and platinum nanoparticle (PtNP) modified carbon-fiber microelectrodes coupled to FSCV to measure the extent to which ATP interacts at metal nanoparticle-modified surfaces and to improve the sensitivity of direct electrochemical detection. AuNP and PtNPs were electrodeposited on the carbon-fiber surface by scanning from -1.2 to 1.5 V for 30 s in 0.5 mg/mL HAuCl4 or 0.5 mg/mLK2PtCl6. Overall, we demonstrate an average 4.1 ± 1.0-fold increase in oxidative ATP current at AuNP-modified and a 3.5 ± 0.3-fold increase at PtNP-modified electrodes. Metal nanoparticle-modified surfaces promoted improved electrocatalytic conversion of ATP oxidation products at the surface, facilitated enhanced adsorption strength and surface coverage, and significantly improved sensitivity. ATP was successfully detected within living murine lymph node tissue following exogenous application. Overall, this study demonstrates a detailed characterization of ATP oxidation at metal nanoparticle surfaces and a significantly improved method for direct electrochemical detection of ATP in tissue.
Adenosine-5′-triphosphate
(ATP) is an important purine extracellular
signaling molecule and has an essential role in a variety of neurological
and immunological disorders.[1−4] The ability to monitor ATP signaling on rapid time
scales is important for elucidating the mechanism of action and quantitating
the availability of ATP to interact at nearby receptors and transporters
during both health and disease. Many methods have been developed to
detect ATP including electrochemical aptamer-based sensors, enzyme
sensors and surface-modified electrodes,[5−8] and fluorescence;[9,10] however,
many of these methods suffer from slow temporal resolution or the
inability to quantitate rapid concentration changes over time. Fast-scan
cyclic voltammetry (FSCV) at carbon-fiber microelectrodes is an electroanalytical
technique which enables millisecond sensing of electroactive neurochemicals;
however, direct electrochemical detection of ATP is difficult due
to limited interactions at carbon-fiber surfaces.[5,11] Recently,
our lab developed a method to functionalize the surface of carbon-fiber
with amines to enable significant improvements in direct ATP detection
as facilitated by an increase in electrostatic interactions at the
surface.[5] Prior work from our lab has also
suggested that ATP interaction at carbon surfaces is directly impacted
by the surface topology and functionality.[6] Despite these advancements, ATP detection on carbon surfaces is
challenging, and an in-depth characterization of ATP interaction at
other electrode surfaces is necessary to advance understanding of
ATP electrochemistry. Here, we developed a method to electrodeposit
both Au and Pt nanoparticles on carbon-fiber surfaces to provide a
critical analysis of ATP interaction at metal nanoparticle-modified
carbon-fiber surfaces. We show that Au and Pt nanoparticles significantly
enhance ATP oxidative current, the catalytic conversion of ATP oxidation
products at the electrode surface, and promote adsorption-controlled
interactions. We demonstrate that metal nanoparticle-modified electrodes
enable subsecond detection of ATP in the lymph node which will provide
a platform for future work to investigate sympathetic ATP modulation
of immunity, a critically understudied area in immunomodulation.[12−14]Metal nanoparticles (NPs) have unique catalytic and electronic
properties that have made them attractive for electrochemical sensors.[15] A few reports have demonstrated AuNP-modified
carbon-fibers for improved dopamine detection with FSCV and amperometry.[16,17] The high surface energy of Au and Pt can promote adsorption with
amines, carboxyls, and other metal ions.[18] Adsorption of small molecules on metal nanoparticle surfaces can
positively impact electrochemical sensing. The catalytic behavior
of both Au and Pt leads to faster redox reactions and enables expansion
to difficult targets.[19−25] A recent study has shown that ATP adsorption onto AuNPs competes
with binding to aptamers on Au surfaces providing evidence that ATP
can interact with Au-based electrodes.[22,26] The high catalytic
activities and strong binding affinity of AuNP and PtNP facilitate
conjugation with biomolecules.[27] To further
understand and quantify ATP adsorption onto metal nanoparticles, we
compare and calculate the extent to which Au and Pt nanoparticle-modified
carbon-fiber microelectrodes affect adsorption strength, surface coverage,
and electrocatalytic behavior to inform the development of improved
methods to directly oxidize ATP at electrode surfaces.Here,
we optimized the electrodeposition of metal nanoparticles
on carbon-fiber surfaces to investigate the extent to which ATP interacts
at metal-modified surfaces to enable significant improvements in direct
real-time ATP detection with FSCV. Careful examination of how metal
surfaces modify the interactions of ATP at the surface provided a
detailed understanding of optimal methods for direct electrochemical
detection. Overall, we provide an easy and robust method to significantly
improve direct real-time ATP detection in tissue, which will have
far-reaching impacts on probing ATP signaling.
Experimental Section
Reagents
All chemicals were purchased
from Fisher Scientific (USA) unless otherwise noted. A Tris buffer
(15 mM Tris, 1.25 mM NaH2PO4, 2.0 mM Na2SO4, 3.25 mM KCl, 140 mM NaCl, 1.2 mM CaCl2 dehydrate, and 1.2 mM MgCl2) at pH 7.4 was used
in all flow injection analysis experiments. Dopamine (DA), Norepinephrine
(NE), and ATP were dissolved in 0.1 M HCl solution to make 10 mM DA,
10 mM NE, and 10 mM ATP stock solutions (Sigma-Aldrich, St. Louis,
MO) and were stored at 4 °C. Stock solutions were diluted daily
in Tris buffer for experiments. Chloroauric acid (HAuCl4) and potassium hexachloroplatinate (K2PtCl6) were purchased from Sigma-Aldrich. All aqueous solutions and electrodeposition
solutions were made from deionized water (Milli-Q, Millipore, Billerica
MA). For biological experiments in the lymph node, a bicarbonate-buffer
was used (130 mM NaCl, 2.5 mM KCl, 1.3 mM NaH2PO4, 26 mM NaHCO3, 1 mM MgCl2, 2 mM CaCl2, and 10 mM glucose at pH 7.4).
Electrode
Fabrication and Modification
Cylindrical carbon-fiber microelectrodes
were made from 7-μm
in diameter T650 carbon-fibers (gift from Mitsubishi Chemical Carbon
Fiber and Composites Inc., Sacramento, CA). Carbon-fibers were vacuum
aspirated into a capillary glass tube (1.2 mm × 0.68 mm, A-M
Systems, Sequim, WA) and pulled into two using a vertical micropipet
puller (Narishige PE-22, Tokyo, Japan). To create a cylinder electrode,
fibers were trimmed to approximately 50–100 μm from the
glass seal using a microscope (Fisher Education). Electrodes were
pretested prior to modification. Electrodes were cleaned with isopropyl
alcohol (IPA) and water to remove salt and allowed to dry prior to
modification. To modify electrodes, either 0.5 mg/mL chloroauric acid
(HAuCl4) or 0.5 mg/mL potassium hexachloroplatinate was
electrodeposited on the carbon-fiber surface and this electrodeposition
created formation of metal nanoparticles on the surface. Electrodeposition
was done by applying a potential sweep to the surface scanning from
−1.2 to 1.5 V at a rate of 5 V/s against a Ag/AgCl reference
electrode for 30 s. See Figure S-1 for
electrodeposition optimization data.
Fast-Scan
Cyclic Voltammetry (FSCV)
Fast-scan cyclic voltammograms
were collected using the WaveNeuro
potentiostat with a 1 MΩ headstage (Pine Instruments, Durham,
NC). Data was collected using a National Instruments PCIe-6363 interface
board (Austin, TX) and HDCV software (UNC-Chapel Hill, Mark Wightman).
Cyclic voltammograms (CVs) were background subtracted to remove nonfaradaic
currents. The electrode was scanned from −0.4 to 1.45 V (vs
Ag/AgCl) and back with a 400 V/s scan rate and a repetition rate of
10 Hz. The electrode was equilibrated for 10 min prior to testing,
and the average current for each analyte was recorded. Electrodes
were tested using a home-built flow injection analysis system using
a six-port HPLC actuator (Valco Instruments, Houston, TX). A syringe
pump (Chemyx, Stafford, TX) set to a flow rate of 1 mL min–1 was used to deliver buffer and sample to the electrode. All experiments
were performed at room temperature.
Surface
Characterization
The physical
and chemical properties of unmodified and metal nanoparticle-modified
carbon-fiber surfaces were analyzed. To visualize and qualitatively
assess the electrode, scanning electron micrographs (SEM) were collected
using an FEI XL30 SEM coupled to an EDAX detector for energy dispersive
spectroscopy (EDS). SEM images were gathered at an accelerating voltage
of 5.00 kV and 610 mm away. ImageJ Fiji was used to analyze the metal
nanoparticles size after electrodeposition on the carbon fiber surface.
Exogenous Delivery of ATP to Lymph Node Slices
All the animal experiments were approved by the Institutional Animal
Care and Use Committee (IACUC) at the University of Cincinnati. Female
C57BL/6J mice (Jackson Laboratories, Bar Harbor, ME; Research Resource
Identifiers, RRID: IMSR_JAX:000664) between the ages of 8–12
weeks were used in the experiments. Mice were group housed in the
vivarium at the Laboratory Animal Medical Services (LAMS) Department
at the University of Cincinnati, where food and water was given ab
libitum, with 12-h light/dark cycles. For the experiment, mice were
anesthetized with isoflurane (Henry Shrein) and euthanized by cervical
dislocation. Lymph nodes were harvested, sliced, and collected as
previously reported.[28−30] Briefly, mesenteric lymph nodes from the gut were
harvested near the intestinal ileum and immediately placed in ice-cold
1× DPBS (without calcium and magnesium, Gibco) with 2% heat-inactivated
fetal bovine serum (premium grade FBS, VWR ≤ 20 EU/mL) for
2 min. Lymph nodes were embedded in 6% low melting point agarose (Lonza,
NJ) prepared in DPBS, and were allowed to harden on ice. A 10 mm tissue
punch was used to remove each lymph node embedded in the agarose (Robbins
Instrument, NJ). The agarose block was mounted with superglue onto
a specimen disc and placed in the buffer tray with slightly oxygenated
(5% CO2 with 95% O2) ice-cold bicarbonate buffer.
The agarose blocks containing lymph nodes were sliced at 300-μm
thickness using a Leica VT1000S vibratome (Chicago, IL). The slices
were immediately transferred using a camel-hair paint brush (TedPella,
CA) into a 6-well culture plate. Each well contained 3 mL of complete
RPMI culture media including RPMI (Hyclone) with 10% FBS, 1× l-glutamine (Gibco), 50 U/mL of Pen/Strep (Gibco), 50 μM
beta-mercaptoethanol (Gibco), 1 mM sodium pyruvate (Hyclone), 1×
nonessential amino acids (Hyclone), and 20 mM HEPES (Gibco). Immediately
after slicing, the culture plate was placed in a sterile incubator
set to 37 °C with 5% CO2. The tissue slices were allowed
to recover for approximately an hour prior to the experiment.During testing, slices were placed in a Warner Instruments standard
perfusion chamber (Hamden, CT) held at 37 °C and perfused with
oxygenated bicarbonate buffer at a rate of 2 mL min–1 using an Ismatec Reglo ICC 4-channel digital peristaltic pump (Cole-Parmer,
Vernon Hills, IL). A metal-modified electrode was then implanted within
the T-cell zone of the lymph node, approximately 75 μm beneath
the surface, using an MM-3 micromanipulator (Narishige) and allowed
to equilibrate. A Parker Hannifin Picospritzer III (Hollis, NH) was
used for exogenous delivery. ATP (150 μM) was administered approximately
100 μm away from the working electrode using a glass micropipet
with a 15–20 μm opening at the tip, similar to previous
reports.[5,31] The picospritzer pressure was set to 30
psi with 800 ms ejections. Exogenous ATP was successfully detected
in tissue using both AuNP and PtNP-modified electrodes.To test
the extent to which the electrodes experience biofouling,
we harvested lymph nodes from mice and homogenized the tissue samples
in bicarbonate buffer using a tissue probe sonicator (QSonica, LLC
model Q55, Newtown, CT). Electrodes (bare carbon-fiber, AuNP- and
PtNP-modified carbon-fiber) were pretested for both 1 μM DA
and 5 μM ATP. Electrodes were then soaked in the homogenized
tissue solution for 1 h, followed by immediate post-testing. The ratio
of post-tested oxidative current to pretested oxidative current for
each analyte was determined to quantify the impact of biofouling on
the electrode surface.
Statistics
All
data were analyzed
with GraphPad Prism V. 9.0 (GraphPad Software Inc., La Jolla, CA).
Statistical p-values were significant at the 95%
confidence level (p < 0.05). Values are reported
as the mean ± standard error of the mean (SEM), and n represents the number of electrodes or slices.
Results and Discussion
ATP Detection at Unmodified
Carbon-Fiber Microelectrodes
ATP undergoes three two-electron/two-proton
oxidation reactions;
however, only the first two oxidation products are ever detected with
FSCV due to slow electron transfer kinetics.[6,32−34] To observe ATP oxidation peaks with FSCV, an electrochemical
waveform scanning from −0.4 to 1.45 V and back at a rate of
400 V/s is used. ATP’s primary oxidation peak is observed at
1.4 V on the cathodic scan due to slowed electron transfer, and the
secondary peak is observed at 0.8 V[11,32] and is always
significantly smaller than the primary oxidation peak. The tertiary
oxidation peak is rarely observed on bare carbon-fibers. Detection
of ATP at carbon-fiber microelectrodes is significantly more challenging
than detecting dopamine (DA), a popular analyte detected with FSCV
(Figure ).[35−37]Figure shows an
example of false-color plots and cyclic voltammograms (CVs) for 5
μM ATP (Figure A,C) and 1 μM DA (Figure B,D) collected on the same electrode to demonstrate
a significant difference in observed oxidative current. Despite ATP
being 5-fold higher in concentration, an approximately 6-fold decrease
in oxidative current for ATP is detected compared to DA. This has
been well documented in the literature and is mostly attributed to
ATP’s negative charge inhibiting interaction at oxide-rich
carbon surfaces and ATP’s slow electron transfer kinetics.[6,32−34] Because of this, it is essential to develop electrode
surfaces which are specifically tuned to improve ATP detection. Fundamental
understanding of direct ATP interaction at the electrode surfaces
is limited but is crucial for developing sensitive ATP detection methods.
Figure 1
The carbon-fiber
microelectrode is significantly less sensitive
to ATP compared to dopamine. (A) A false-color plot demonstrates potential
on the y-axis, time on the x-axis,
and current in false color for 5 μM ATP. (B) A false-color plot
for 1 μM DA. (C) The representative CV of 5 μM ATP and
(D) 1 μM DA. The secondary peak for ATP (0.8 V) is almost negligible
on this electrode. Data shown were collected on the same electrode.
The carbon-fiber
microelectrode is significantly less sensitive
to ATP compared to dopamine. (A) A false-color plot demonstrates potential
on the y-axis, time on the x-axis,
and current in false color for 5 μM ATP. (B) A false-color plot
for 1 μM DA. (C) The representative CV of 5 μM ATP and
(D) 1 μM DA. The secondary peak for ATP (0.8 V) is almost negligible
on this electrode. Data shown were collected on the same electrode.
Metal Nanoparticle Carbon-Fiber
Microelectrode
Modification Procedure
The concentration of metal nanoparticles
and the duration of electrodeposition were optimized to modify carbon-fiber
microelectrodes for detection with FSCV. Previously, the Zestos lab
showed that AuNP-modified carbon-fiber microelectrodes increase DA’s
oxidative current approximately 2-fold and was stable for 4 h.[17] Here, we further optimized the concentration
and deposition time and used an extended deposition waveform, at faster
scan rates than what was previously demonstrated, to measure the extent
to which AuNP deposition procedures impact ATP detection, with DA
serving as a “control”. All optimization procedures
were completed using AuNPs, and this method was later used for PtNP
modification. The waveform for electrodeposition scans from −1.2
to 1.5 V, at a rate of 5 V/s. Deposition times for 0.5 mg/mL AuNPs
ranging from 15 s to 5 min were tested (Figure S-1A,B, n = 10–12) for both 1 μM
DA and 5 μM ATP. The ratio of oxidative current postmodification
to premodification was calculated and plotted as a function of deposition
time. A ratio above 1.0 indicates an increase in current due to modification,
whereas a ratio below 1.0 indicates a decrease in current. Overall,
the oxidative current for DA minimally changed at AuNP-modified electrodes;
however, oxidative current for DA was significantly higher at 30 s
compared to 3 min deposition times (one-way ANOVA F(4,27) = 2.852
and p = 0.043, Bonferroni post-test p = 0.0195, n = 7–12) (Figure S-1A). Despite the significant difference between 30
s and 3 min deposition, improvements in dopamine current never increased
above 1.7-fold. Our observed change in dopamine oxidative current
is slightly lower than prior reports; however, the deposition waveform
and time of deposition are different making these results difficult
to accurately compare.[17] Overall, our results
indicate that AuNP-modified surfaces increase DA interactions but
in general, changes are minimal. Additionally, prior work has demonstrated
that Au is less sensitive to DA with FSCV.[38,39] Conversely, the oxidative current for ATP increased at all deposition
times, and 30 s was significantly different compared to 5 min deposition
(one-way ANOVA F(4,35) = 1.943 and p = 0.021, Bonferroni
post-test p = 0.0195, n = 7–12, Figure S-1B). Much larger increases in oxidative
current was observed for ATP compared to DA. Based on these results,
a deposition time of 30 s was chosen as optimal for ATP detection.
Deposition times greater than 1 min produced measurable noise at the
electrode, with less improvements in oxidative current, likely due
to agglomerations of AuNPs on the surface.The concentration
of metal nanoparticles was optimized by keeping the deposition time
constant at 30 s while varying the HAuCl4 concentration
from 0.1 mg/mL, 0.5 mg/mL, and 1 mg/mL (Figure S-1C,D, n = 10–12). Small increases
in DA current were observed at all concentrations tested; however,
no significant differences between concentrations were observed (one-way
ANOVA F(4,26) = 0.534 and p = 0.593). On average,
the ATP oxidative current increased between 3 and 4-fold at all concentrations
tested. Again, no significant differences between concentrations were
observed (one-way ANOVA F(4,27) = 0.804, p = 0.923).
Despite this result, it was qualitatively observed that larger concentrations
resulted in an increase in the variance in the data set with decreases
in the signal/noise (S/N). Overall, 0.5 mg/mL was chosen as optimal.
At the optimized procedure, DA increased on average 1.7 ± 0.2-fold
and ATP increased 4.1 ± 1.0-fold at AuNP-modified electrodes
(Figure , n = 7–12). The improvements in detected current after
AuNP-modification are significantly different between DA and ATP (unpaired t test, p < 0.01). As a control, we
tested whether applying the electrodeposition waveform alone (no metal
nanoparticles in solution) could increase observed oxidative current
for DA and ATP due to electrooxidation of the surface at high potentials.
On average, we observed 1.3 ± 0.1- and 1.5 ± 0.2-fold increases
for DA and ATP, respectively, demonstrating that the majority of the
changes in ATP current arises from the presence of metal nanoparticles
on the electrode surface, whereas the increases observed for DA were
likely due solely to the electrodeposition waveform (n = 6, Figure S-2). The electrodeposition
waveform could be exposing more surface oxides on the carbon surface
due to the expanded potential window applied.[40]
Figure 2
ATP
oxidative current increases at both Au and Pt nanoparticle
modified carbon-fiber microelectrodes. Electrodes were electrodeposited
with either Au or Pt nanoparticles by applying a waveform that scans
from −1.2 to 1.5 at 5 V/s. Significantly higher increases in
current are observed for ATP compared to DA at AuNP (A) and PtNP (B)
modified electrodes (n = 7–12). Representative
CVs are shown for bare carbon (CF, black), AuNP-modified (orange),
and PtNP-modified (blue) for ATP (C) and DA (D).
ATP
oxidative current increases at both Au and Pt nanoparticle
modified carbon-fiber microelectrodes. Electrodes were electrodeposited
with either Au or Pt nanoparticles by applying a waveform that scans
from −1.2 to 1.5 at 5 V/s. Significantly higher increases in
current are observed for ATP compared to DA at AuNP (A) and PtNP (B)
modified electrodes (n = 7–12). Representative
CVs are shown for bare carbon (CF, black), AuNP-modified (orange),
and PtNP-modified (blue) for ATP (C) and DA (D).
Comparison of Oxidative Current Increases
at AuNP and PtNP-Modified Carbon-Fiber Microelectrodes
Electrodeposition
of 0.5 mg/mL potassium hexachloroplatinate resulted in significantly
larger increases in ATP current compared to DA (Figure B, paired t test, p < 0.0001, n = 7–12). Electrodes
were tested prior to and after deposition. The ratio of postmodification
current to premodification/bare electrode current (ip/ib) was calculated and plotted
as a function of the analyte (Figure ). On average, the DA oxidative current increased 1.7
± 0.1-fold and 1.8 ± 0.1-fold for AuNP and PtNP-modified
electrodes, and ATP increased 4.1 ± 0.3-fold and 3.5 ± 0.3-fold
for AuNP and PtNP-modified electrodes, respectively (n = 7–12). Example representative CVs for ATP (Figure C) and DA (Figure D) comparing each electrode
type (bare carbon fiber, AuNP, and PtNP-modified) show that metal
nanoparticle modification is beneficial for improving ATP detection
with FSCV; whereas, minimal improvements are observed for DA detection.
Similar increases in oxidative current were observed for both AuNP
and PtNP-modified electrodes. Additionally, we validate that electrodeposition
of metal nanoparticles on the surface does not impact the temporal
resolution of the technique by comparing the shapes of the current
vs time traces before and after modification (Figure S3). No change in peak shape was observed.
AuNP and PtNP-Modified Electrodes Increase
Electrocatalytic Behavior for ATP
Detection of ATP’s
secondary and tertiary oxidation peaks are difficult to detect on
bare carbon-fiber electrodes due to slow kinetics and minimal interaction
at the electrode surface. We observed increases in current for ATP’s
secondary and even tertiary oxidation peaks at metal-nanoparticle
modified electrodes providing evidence that metal surfaces improve
the electrocatalytic activity of ATP at the surface (Figure ). ATP’s oxidation scheme
is represented in Figure A. Figure shows example color plots and CVs for 5 μM ATP at bare carbon-fiber
(Figure B), AuNP-modified
carbon-fiber (Figure C), and PtNP-modified carbon-fiber (Figure D). Three dotted lines (1,2,3) are drawn
on the color plots at different time points for each of the electrodes.
At each of these dotted traces, the CV is shown for this time point
and the oxidation peaks are labeled providing evidence that the secondary
and tertiary oxidation peaks are occurring sequentially in time. Having
three oxidation peaks instead of just one or two could improve the
selectivity of ATP detection over other neurochemical analytes with
FSCV. To quantitate and compare the change in conversion from the
primary to the secondary oxidation product on a bare electrode vs
a modified electrode, we calculated the ratio of the (secondary) 2°/1°
(primary) peak current for each the bare electrode and the modified
electrode. We divided these two ratios (Ratiopost/Ratiopre, Figure S-4). A number above
1.0 indicates that there was an increase in conversion between the
primary and the secondary oxidation peak, whereas a number below 1.0
indicates less conversion. On average, there was a 2.4 ± 0.5-
fold and 2.6 ± 1.3-fold increase in conversion of the primary
to the secondary oxidation product at AuNP and PtNP-modified electrodes,
respectively (Figure S-4, n = 8–10). There was no significant difference between the
two metal nanoparticle treatments (unpaired t test, p = 0.8483). This result indicates that metal-nanoparticle
modified carbon-fiber microelectrodes significantly improve the electrocatalytic
behavior of ATP at the surface.
Figure 3
Metal nanoparticle electrode modification
enhances the electrocatalytic
conversion of ATP oxidation products. The oxidation scheme for ATP
(A) and a comparison of the color plots and representative CVs of
5 μM ATP over time are shown for bare carbon-fiber (B), AuNP-modified
carbon-fiber (C), and PtNP-modified carbon-fiber (D). ATP was manually
injected at approximately 5 s and washed away at approximately 9 s.
Dashed lines indicate at different time points (time points 1, 2,
and 3) CVs were analyzed to visualize the oxidation reaction over
time, and the oxidation peaks are labeled on the CVs. At line 1 (5
s, black line), only the primary oxidation product is observed. By
line 2 (7 s, red line), the secondary oxidation product is observed,
and by line 3 (9 s, purple line), the tertiary oxidation product is
clearly observed. The representative CVs at each of these lines (1,
2, and 3) demonstrate oxidation peaks growing over time.
Metal nanoparticle electrode modification
enhances the electrocatalytic
conversion of ATP oxidation products. The oxidation scheme for ATP
(A) and a comparison of the color plots and representative CVs of
5 μM ATP over time are shown for bare carbon-fiber (B), AuNP-modified
carbon-fiber (C), and PtNP-modified carbon-fiber (D). ATP was manually
injected at approximately 5 s and washed away at approximately 9 s.
Dashed lines indicate at different time points (time points 1, 2,
and 3) CVs were analyzed to visualize the oxidation reaction over
time, and the oxidation peaks are labeled on the CVs. At line 1 (5
s, black line), only the primary oxidation product is observed. By
line 2 (7 s, red line), the secondary oxidation product is observed,
and by line 3 (9 s, purple line), the tertiary oxidation product is
clearly observed. The representative CVs at each of these lines (1,
2, and 3) demonstrate oxidation peaks growing over time.We tested whether metal nanoparticle modified carbon-fibers
could
improve electrochemical detection of other purines with FSCV (Figure S-5). Overall, we observed a 3.6 ±
0.5-fold and 3.6 ± 0.2-fold increase in adenosine (AD) current
at AuNP- and PtNP-modified electrodes, respectively, indicating that
the nucleoside of ATP interacts favorably at metal surfaces. Despite
this observation, we only observed a 1.5 ± 0.3-fold and 1.4 ±
0.3-fold increase for guanosine (GN) and GTP, respectively, at AuNP-modified
electrodes. Likewise, we observed only a 1.8 ± 0.1-fold and 2.1
± 0.2-fold increase for GN and GTP at PtNP-modified electrodes
(Figure S-5). In previous work from our
lab, we observed less changes in oxidative current for GN and GTP
compared to ATP and AD at plasma-treated electrodes with FSCV.[6] Additionally, we have observed that small changes
in the purine structure can significantly impact the extent to which
it interacts at the electrode surface.[41] Guanine-based purines are easier to oxidize[41] than adenine-based purines and may be less impacted by surface modifications.
This provides additional evidence which supports many prior reports
that the finite interactions at the electrode surface are most definitely
controlled by not only the electrode surface but also the analyte
structure.
Comparison of Detection
Stability at Bare
and Metal Nanoparticle Modified Carbon-Fibers
The stability
of detection over time is important for monitoring neurochemicals
in tissue and to ensure that the modification procedure is stable.
Purines have been shown to polymerize on the electrode surface under
specific circumstances.[42,43] We demonstrate that
ATP detection is not stable over time on bare carbon-fiber microelectrodes
(Figure ) by repeatedly
injecting 5 μM ATP at the electrode 25 times. The current was
normalized to the first injection, and deviation from 100% indicates
the signal is changing over time. On average, ATP current decreased
by 54 ± 1% by the 25th injection (n = 7, Figure ). ATP detection
stability with FSCV has not been measured prior and our result indicates
potential electropolymerization of ATP in solution leading to chemical
fouling on the surface of the electrode; however, the mechanism of
ATP polymerization and how it adsorbs to the carbon surface are unclear.[8,44] Conversely, when repeating this experiment at AuNP and PtNP-modified
carbon-fiber microelectrodes, we show that ATP detection is more stable
on the electrode surface (Figure ). AuNP-modified electrodes are more stable than PtNP-modified
electrodes with only 13 ± 1% loss (n = 5) compared
to 29 ± 2.2% (n = 8) at PtNP-modified. On average,
the loss of signal was significantly different between AuNP-modified
and CFME and between PtNP-modified and CFME (one-way ANOVA, Bonferroni
post-test, p < 0.001).
Figure 4
Metal nanoparticle-modified
electrodes are stable with repeated
injections of ATP. 5 μM ATP was repeatedly injected at the electrode
25 times. The current for the primary oxidation peak for ATP was normalized
to the first injection and compared over time. Deviation from 100%
indicates instability of detection. ATP current decreased by 54% by
the 25th injection at bare carbon-fiber microelectrodes (open circles),
with only a 29% and 13% decrease in ATP current at AuNP (orange squares)
and PtNP (blue circles), respectively. (n = 5–8).
Metal nanoparticle-modified
electrodes are stable with repeated
injections of ATP. 5 μM ATP was repeatedly injected at the electrode
25 times. The current for the primary oxidation peak for ATP was normalized
to the first injection and compared over time. Deviation from 100%
indicates instability of detection. ATP current decreased by 54% by
the 25th injection at bare carbon-fiber microelectrodes (open circles),
with only a 29% and 13% decrease in ATP current at AuNP (orange squares)
and PtNP (blue circles), respectively. (n = 5–8).
Surface Characterization
Scanning
electron microscopy (SEM) with EDAX was used to compare qualitatively
and quantitatively the surface of the bare and modified electrodes.
SEM images of a bare carbon-fiber (Figure A), AuNP-modified carbon-fiber (Figure B), and PtNP-modified
carbon-fiber (Figure C) demonstrate differences in the particle size between Au and Pt.
SEM images were taken for four to six electrodes per group, and Figure shows examples for
each condition. Topographical differences are evident between the
traditional CFME and metal nanoparticles treated CFME. The average
nanoparticle diameter for AuNPs was calculated to be 39.8 ± 0.45
nm and 27.4 ± 0.44 nm for PtNPs using ImageJ analysis (Figure S-6). Particle size differences can impact
the electroactive area; however, we did not observe significant differences
in oxidative current improvements between these two metals.[38,45] We confirmed the presence of Au and Pt on the surface by also collecting
EDS spectra of the surface (Figure S-7).
Figure 5
Scanning
electron microscopy (SEM) images of bare and metal nanoparticle
modified carbon-fibers. (A) Bare carbon fiber, (B) 0.5 mg/mL AuNP-modified
carbon-fibers (30 s deposition), and (C) 0.5 mg/mL PtNP-modified carbon-fibers
(30 s deposition). Images on the right are at higher magnification
to reveal NPs on the surface. Scale bars are shown on the image.
Scanning
electron microscopy (SEM) images of bare and metal nanoparticle
modified carbon-fibers. (A) Bare carbon fiber, (B) 0.5 mg/mL AuNP-modified
carbon-fibers (30 s deposition), and (C) 0.5 mg/mL PtNP-modified carbon-fibers
(30 s deposition). Images on the right are at higher magnification
to reveal NPs on the surface. Scale bars are shown on the image.
Adsorption of ATP at Metal
Nanoparticle-Modified
Surfaces
Clear differences in surface topology compared to
bare carbon-fibers were observed due to the presence of nanoparticles
on the surface. Varying topologies can affect the surface adsorption
behavior.[6] Likewise, the nanoparticle can
become a nanoelectrode upon contact with the electrode substrate.[46] In this study, we observe that changes in analyte
structure (DA vs ATP) can directly impact adsorption behavior on the
metal surface.[47−50] Many studies suggest that Au and Pt surfaces offer high sensitivity
for dopamine detection; however, we did not observe significant enhancements
in current for DA compared to traditional carbon-fibers which implies
negligible changes in adsorption behavior for DA at metal nanoparticle-modified
surfaces. Previous literature has suggested that the ATP moiety is
both adsorption and diffusion-controlled at unmodified carbon-based
electrodes.[6] To test the extent to which
ATP interacts at bare and metal-modified surfaces, we monitored the
change in current as a function of scan rate. Scan rates tested ranged
from 50 to 800 V s–1. The log of the current vs
the log of scan rate was plotted, and the slope of the linear regression
provides insight into the limiting reactions at the surface (Figure ). A slope of 0.5
describes a diffusion-controlled interaction, whereas a slope of 1.0
indicates an adsorption-controlled process.[42] We observed increases in the slope from 0.634 at bare carbon-fiber
electrodes to 0.855 at AuNP-modified carbon and 0.815 at PtNP-modified
carbon (Figure , n = 6–8). At the bare carbon-fiber, ATP interactions
are governed primarily by diffusion; however, at metal nanoparticle-modified
carbon, ATP’s interactions become more adsorption-limited.
Figure 6
Metal
nanoparticles change the interaction of ATP at the electrode
surface. The oxidative current for ATP was recorded as a function
of scan rate. Scan rates tested ranged from 50 to 800 V/s. The log
of the peak oxidative current (ip) is plotted vs the log
of scan rate. A slope of the line describes the dominating electrode
interaction. A slope closer to 1.0 indicates adsorption-controlled
processes, and a slope closer to 0.5 indicates diffusion-controlled
processes. The slope of the line changes from 0.634 to 0.855 (AuNP-modified
carbon) and 0.815 (PtNP-modified carbon), indicating changes in the
surface interaction to more adsorption-limited processes (n = 6–8).
Metal
nanoparticles change the interaction of ATP at the electrode
surface. The oxidative current for ATP was recorded as a function
of scan rate. Scan rates tested ranged from 50 to 800 V/s. The log
of the peak oxidative current (ip) is plotted vs the log
of scan rate. A slope of the line describes the dominating electrode
interaction. A slope closer to 1.0 indicates adsorption-controlled
processes, and a slope closer to 0.5 indicates diffusion-controlled
processes. The slope of the line changes from 0.634 to 0.855 (AuNP-modified
carbon) and 0.815 (PtNP-modified carbon), indicating changes in the
surface interaction to more adsorption-limited processes (n = 6–8).A Langmuir isotherm was used to quantitate and compare the surface
coverage of ATP on bare carbon-fiber microelectrodes vs metal nanoparticle-modified
electrode surfaces at equilibrium. The equations used are described
below:[34]ΓA is surface concentration,
Γs is the saturation surface coverage, and β
is the thermodynamic equilibrium constant for adsorption. The surface
coverage for purines was determined using eq since ATP’s oxidation involves both
reversible and irreversible steps.[34,43] For eq , ip is the peak oxidative current, R is the
ideal gas constant, T is the temperature (25 °C), F is Faraday’s constant, v is the
scan rate, n′ is the number of electrons transferred
prior to the chemically irreversible steps, θ is the total number
of electrons transferred in the redox reaction divided by n′, and A is the area of the electrode
surface (1.68 × 10–5 cm2 for the
75 μm length cylinder carbon fiber microelectrode and for ATP n′ = 2 and θ = 3).[34] The linear region of the isotherm can be used to calculate b (eq ), the
equilibrium coefficient which dictates adsorption.The surface
coverage explains the monolayer surface saturation
coverage of the adsorbate on the electrode. The average surface coverage
for ATP increased from 7.16 pmol/cm2 to 42.6 pmol/cm2 after introducing AuNPs on the carbon-fiber surface, equating
to an approximate 495% increase in surface coverage (Figure , Table ). At PtNP-modified electrodes, the surface
coverage increased to 74.3 pmol/cm2 resulting in an average
increase of 938% (Figure and Table ). The thermodynamic equilibrium constant (β) for adsorption
helps describe the rate of desorption at the electrode surface. β
decreased by 12-fold at PtNP-modified electrodes and by 8-fold at
AuNP-modified, which suggests that the rate of desorption of ATP is
faster at metal nanoparticle modified electrodes. This result may
help elucidate why less ATP fouling is observed at metal nanoparticle-modified
carbon electrodes. For the adsorption strength (b), the number increased 5-fold and 6.9-fold for AuNP and PtNP, respectively,
compared to bare carbon fiber. This result shows that ATP adsorbs
strongly to metal nanoparticle-modified surfaces, but the oxidation
product desorbs quickly. The sensitivity (Figure B, Table ) improved 4.7-fold at PtNP-modified electrodes and
4.3-fold at AuNP-modified electrodes compared to the traditional bare
carbon-fiber. Higher sensitivity measurements of ATP are crucial for
applications in vivo and demonstrates that metal-nanoparticle-modified
surfaces could be a critical advance in real-time ATP sensing.
Figure 7
Higher sensitivity, surface coverage, and stronger adsorption strength
are observed for ATP at metal-nanoparticle modified carbon-fiber microelectrodes.
(A) The oxidative current for ATP was measured for concentrations
ranging from 1 μM to 100 μM (B). The linear region of
the curve spanned from 1 μM to 10 μM ATP and was used
to compare the sensitivity of each electrode type for ATP (r2 Bare CF = 0.9384, r2 AuNP = 0.9588, r2 PtNP = 0.9864). (C)
Langmuir isotherms for ATP at the bare carbon-fiber (CF, black), AuNP
modified (orange), and PtNP modified (blue) electrodes (n = 6–8).
Table 1
Metal Nanoparticle Modified Carbon-Fiber
Increase Sensitivity and Adsorption Strength of ATP on the Surface
(n = 6–8)
electrode
Γs (pmol/cm2)
β (× 10–2cm3/pmol)
sensitivity of
ATP
adsorption strength (b, cm)
bare CFME
7.2 ± 1.1
0.48
0.9 ± 0.1
0.30 ± 0.06
PtNP-modified
74.3 ± 7.8
0.04
4.2 ± 0.2
2.07 ± 0.01
AuNP-modified
43.0 ± 4.7
0.06
3.8 ± 0.2
1.51 ± 0.02
Higher sensitivity, surface coverage, and stronger adsorption strength
are observed for ATP at metal-nanoparticle modified carbon-fiber microelectrodes.
(A) The oxidative current for ATP was measured for concentrations
ranging from 1 μM to 100 μM (B). The linear region of
the curve spanned from 1 μM to 10 μM ATP and was used
to compare the sensitivity of each electrode type for ATP (r2 Bare CF = 0.9384, r2 AuNP = 0.9588, r2 PtNP = 0.9864). (C)
Langmuir isotherms for ATP at the bare carbon-fiber (CF, black), AuNP
modified (orange), and PtNP modified (blue) electrodes (n = 6–8).
Detection
of Exogenous ATP in Lymph Node Slices
Using Metal Nanoparticle Electrodes
The AuNP and PtNP-modified
electrodes were used to detect exogenous ATP in slices of live murine
mesenteric lymph node. ATP is known to be the “currency”
of biological processes and has been monitored within the immune system
to assess sensitivity to allergens and characterize T-cell motility.[51,52] Likewise, ATP is copackaged with norepinephrine in sympathetic neurons
which innervate lymphoid organs and can be rapidly released to modulate
immune signaling. Despite this knowledge, subsecond detection of ATP
release in the lymph node has not been explored, introducing a fundamental
gap in our understanding of rapid ATP regulated immune processes.
To validate detection within biological tissue and demonstrate the
relative power of the nanoparticle modifications,[12,13,53] both AuNP and PtNP-modified electrodes were
implanted in mesenteric lymph node slices approximately 100 μm
away from a picospritzing pipet. The pipet was backfilled with 150
μM ATP; large concentrations are common for this procedure due
to diffusional loss of analyte in tissue, with a few micromolar actually
reaching the electrode surface.[30] A short
bolus was administered to the tissue to mimic a rapid endogenous event.
Detection was successful using both the AuNP and PtNP-modified electrodes;
ATP was detected in four slices using seven AuNP-modified electrodes
and in five slices using seven PtNP-modified electrodes (Figure ). Example CVs and
color plots for ATP using the AuNP (Figure A) and PtNP-modified electrodes (Figure B) are comparable
to in vitro data. The primary (1°) and secondary
(2°) oxidation peaks are observed; small shifts in peak potential
are a common phenomenon when detecting adenine-based purines within
a biological matrix.[5,54,55] Likewise, differences in the current magnitude between the two electrodes
are likely due to the inability to precisely place the picospritzing
pipet at the same location in the X- and Z-directions from the implanted electrode across all trials.
Electrodes which are closer to the picospritzing pipet will measure
larger concentrations than ones further away. Because of this, the
oxidative current for ATP detected at each electrode was variable
ranging from 5 to 30 nA. Rapid decay of the signal over time indicates
native metabolism and uptake of ATP within the lymph node slice.
Figure 8
Exogenous
detection of ATP within mesenteric lymph node slices
at both AuNP and PtNP-modified electrodes. (A) Example color plot,
current vs time (top) and CV (bottom) for exogenous ATP at a AuNP-modified
electrode. The blue arrow indicates when ATP was pressure-ejected
into the tissue. Both the primary (1°, 1.2 V) and secondary (2°,
1.0 V) oxidation peaks are present on the CV for ATP. (B) Example
color plot, current vs time, and CV for exogenous ATP at a PtNP-modified
electrode. The primary (1°, 1.2 V) and secondary (2°, 1.0
V) peaks are visible on the CV for ATP.
Exogenous
detection of ATP within mesenteric lymph node slices
at both AuNP and PtNP-modified electrodes. (A) Example color plot,
current vs time (top) and CV (bottom) for exogenous ATP at a AuNP-modified
electrode. The blue arrow indicates when ATP was pressure-ejected
into the tissue. Both the primary (1°, 1.2 V) and secondary (2°,
1.0 V) oxidation peaks are present on the CV for ATP. (B) Example
color plot, current vs time, and CV for exogenous ATP at a PtNP-modified
electrode. The primary (1°, 1.2 V) and secondary (2°, 1.0
V) peaks are visible on the CV for ATP.
Multiplexed Detection and Biofouling at Metal-Modified
Carbon-Fiber Microelectrodes
ATP is copackaged with norepinephrine
(NE) in vesicles in the lymph node, which can be coreleased to modulate
immunity. To demonstrate that our metal-modified electrodes are capable
of codetecting both NE and ATP and this codetection is improved compared
to bare CFMEs, we tested a mixture of 5 μM ATP with 1 μM
NE at each electrode type (Figure S-8).
Minimal oxidative peak for ATP is observed when in a mixture with
NE at bare carbon-fibers; however, both oxidation peaks are clearly
observed at AuNP-modified (Figure S-8)
and PtNP-modified electrodes (Figure S-8). This demonstrates the utility of these electrodes for multiplexed
detection in the lymph node in the future.Nonspecific adsorption
of proteins and bioproducts on the electrode surface can lead to biofouling.
To test the stability of metal-modified and bare carbon-fiber electrodes
in tissue, we measured the change in oxidative current for both DA
and ATP after soaking in homogenized tissue. Electrodes were pretested
with 5 μM ATP and 1 μM DA and then soaked in homogenized
lymph node tissue for 1 h. Following soaking, the electrodes were
post-tested, and the ratio of the post-test current to pretest current
was plotted (ipost:ipre) as a function of electrode type (bare carbon
fiber, AuNP-, and PtNP-modified). Overall, we observed reductions
in measured current for all electrodes indicating that all electrodes
experience biofouling in tissue (Figure S-9). The ratio of post-tested current to pretested current for 5 μM
ATP and 1 μM DA is 0.37 ± 0.11 and 0.63 ± 0.03 on
bare carbon-fibers. On AuNP, the ratio is 0.38 ± 0.16 for ATP
and 0.62 ± 0.03 for DA. For the PtNP treated electrodes, the
ratio is 0.80 ± 0.10 for DA and 0.46 ± 0.13 for ATP. The
decreases in current for metal-modified electrodes were not significantly
different from bare carbon-fibers for both ATP (one-way ANOVA, F(2,18)
= 0.138, p = 0.873, n = 6) and DA
(one-way ANOVA, F(2,18) = 1.1817, p = 0.204, n = 6). Overall, these results indicate that metal-modified
surfaces do not provide an improvement for the impacts of biofouling
compared to traditional carbon-fibers; however, future work focusing
on nanostructuring the metal nanoparticle surfaces could provide better
biofouling resistance, according to previous reports.[56]
Conclusion
Electrodeposition
is an easy method to introduce metal nanoparticles
on carbon-fiber microelectrode surfaces to facilitate studies of analyte-metal
nanoparticle interfaces and to advance fundamental understanding of
nanoscale electrochemical reactions and kinetics. Here, we show that
the surface structure and catalytic properties of metal nanoparticles
enhance ATP detection with FSCV. Exogenous detection within living
lymph node tissue was shown to be feasible using both nanoparticle
types. This work provides critical insight into the mechanisms of
ATP interaction at electrode surfaces, which will enable strategic
design of ultrasensitive electrodes for real-time ATP sensing in the
future.
Authors: Audrey L Sanford; Stephen W Morton; Kelsey L Whitehouse; Hannah M Oara; Leyda Z Lugo-Morales; James G Roberts; Leslie A Sombers Journal: Anal Chem Date: 2010-06-15 Impact factor: 6.986