Impact of right ventricular pacing site on the subcutaneous ICD sensing-a step towards personalised device therapy?

Mohamed ElRefai1,2, Mohamed Abouelasaad3, Christina Menexi3, John Morgan4, Paul R Roberts3,4.   

Abstract

BACKGROUND: Patients with an existing subcutaneous implantable cardiac defibrillator (S-ICD) may develop a pacing indication. When transvenous pacing is not feasible, combining an S-ICD and a leadless pacemaker (LP) can be a reasonable option. There are reports of concomitant use of both devices. However, the effect of pacing on the S-ICD sensing is not well studied. We hypothesise that pacing changes R and T-wave amplitudes, causing changes in R:T ratios as perceived by a S-ICD, increasing the risk for T wave oversensing (TWO) during paced rhythm with a subsequent risk of inappropriate shocks.
METHODS: This is a prospective study in patients undergoing electrophysiological studies. Participants were fitted with a Holter®, and the leads were placed to correspond to the vectors of an S-ICD. The right ventricle was paced at four positions for 10 beats each at 8 mA/2 ms. The Holter® traces were analysed, using two-way analysis of variance (ANOVA) to assess the effect of pacing on the R:T ratio.
RESULTS: Forty-seven patients (age 56.02 ± 16.02, 72% male) were enrolled (81% structurally normal heart, 15% dilated cardiomyopathy, 2% ischaemic cardiomyopathy, and 2% adult congenital heart disease). Age, sex, and aetiology had no effect on the R:T ratio. Pacing caused significant changes in the R:T ratio. There was no significant difference in the R:T ratios between the pacing sites (p < 0.001).
CONCLUSIONS: Pacing alters the R:T ratio significantly in most patients, theoretically increasing the risk for TWO and inappropriate shocks. Tailored programming for both devices is important for concomitant use of LPs and S-ICDs.
© 2022. The Author(s).

Entities:  

Keywords:  Cardiac implantable devices; Leadless pacemakers; Personalised medicine; Subcutaneous implantable cardiac defibrillators

Year:  2022        PMID: 35460502     DOI: 10.1007/s10840-022-01218-9

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  10 in total

1.  A Leadless Intracardiac Transcatheter Pacing System.

Authors:  Dwight Reynolds; Gabor Z Duray; Razali Omar; Kyoko Soejima; Petr Neuzil; Shu Zhang; Calambur Narasimhan; Clemens Steinwender; Josep Brugada; Michael Lloyd; Paul R Roberts; Venkata Sagi; John Hummel; Maria Grazia Bongiorni; Reinoud E Knops; Christopher R Ellis; Charles C Gornick; Matthew A Bernabei; Verla Laager; Kurt Stromberg; Eric R Williams; J Harrison Hudnall; Philippe Ritter
Journal:  N Engl J Med       Date:  2015-11-09       Impact factor: 91.245

2.  Clinical experience with pacemaker pulse generators and transvenous leads: an 8-year prospective multicenter study.

Authors:  Robert G Hauser; David L Hayes; Linda M Kallinen; David S Cannom; Andrew E Epstein; Adrian K Almquist; Susan L Song; G Frank O Tyers; Stephen C Vlay; Marleen Irwin
Journal:  Heart Rhythm       Date:  2006-10-12       Impact factor: 6.343

3.  Leadless pacing with Micra TPS: A comparison between right ventricular outflow tract, mid-septal, and apical implant sites.

Authors:  Christophe Garweg; Bert Vandenberk; Stefaan Foulon; Peter Haemers; Joris Ector; Rik Willems
Journal:  J Cardiovasc Electrophysiol       Date:  2019-08-05

4.  Incidence and predictors of short- and long-term complications in pacemaker therapy: the FOLLOWPACE study.

Authors:  Erik O Udo; Nicolaas P A Zuithoff; Norbert M van Hemel; Carel C de Cock; Thijs Hendriks; Pieter A Doevendans; Karel G M Moons
Journal:  Heart Rhythm       Date:  2011-12-17       Impact factor: 6.343

5.  Which patients are not suitable for a subcutaneous ICD: incidence and predictors of failed QRS-T-wave morphology screening.

Authors:  Louise R A Olde Nordkamp; Joppe L F Warnaars; Kirsten M Kooiman; Joris R de Groot; Boudewijn R A M Rosenmöller; Arthur A M Wilde; Reinoud E Knops
Journal:  J Cardiovasc Electrophysiol       Date:  2014-01-07

6.  Inappropriate implantable cardioverter-defibrillator shocks in MADIT II: frequency, mechanisms, predictors, and survival impact.

Authors:  James P Daubert; Wojciech Zareba; David S Cannom; Scott McNitt; Spencer Z Rosero; Paul Wang; Claudio Schuger; Jonathan S Steinberg; Steven L Higgins; David J Wilber; Helmut Klein; Mark L Andrews; W Jackson Hall; Arthur J Moss
Journal:  J Am Coll Cardiol       Date:  2008-04-08       Impact factor: 24.094

7.  Updated performance of the Micra transcatheter pacemaker in the real-world setting: A comparison to the investigational study and a transvenous historical control.

Authors:  Mikhael F El-Chami; Faisal Al-Samadi; Nicolas Clementy; Christophe Garweg; Jose Luis Martinez-Sande; Jonathan P Piccini; Saverio Iacopino; Michael Lloyd; Xavier Viñolas Prat; Michael Dilou Jacobsen; Philippe Ritter; Jens Brock Johansen; Claudio Tondo; Fang Liu; Dedra H Fagan; Alyssa K Eakley; Paul R Roberts
Journal:  Heart Rhythm       Date:  2018-08-10       Impact factor: 6.343

8.  Combination of a leadless pacemaker and subcutaneous defibrillator: First in-human report.

Authors:  Blandine Mondésert; Marc Dubuc; Paul Khairy; Peter G Guerra; Gilbert Gosselin; Bernard Thibault
Journal:  HeartRhythm Case Rep       Date:  2015-07-30

9.  Simultaneous leadless pacemaker and subcutaneous implantable cardioverter-defibrillator implantation-When vascular options have run out.

Authors:  Jit Beng Ng; Kelvin Chua; Wee Siong Teo
Journal:  J Arrhythm       Date:  2018-11-20

10.  Complications after cardiac implantable electronic device implantations: an analysis of a complete, nationwide cohort in Denmark.

Authors:  Rikke Esberg Kirkfeldt; Jens Brock Johansen; Ellen Aagaard Nohr; Ole Dan Jørgensen; Jens Cosedis Nielsen
Journal:  Eur Heart J       Date:  2013-12-17       Impact factor: 29.983

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.