| Literature DB >> 35449225 |
Milad Massoudifarid1, Amin Piri1, Jungho Hwang2.
Abstract
Development of efficient virus aerosol monitoring and removal devices requires aerosolization of the test virus using atomizers. The number concentration and size measurements of aerosolized virus particles are required to evaluate the performance of the devices. Although diffusion dryers can remove water droplets generated using atomizers, they often fail to remove them entirely from the air stream. Consequently, particle measurement devices, such as scanning mobility particle sizer (SMPS), can falsely identify the remaining nanosized water droplets as virus aerosol particles. This in turn affects the accuracy of the evaluation of devices for sampling or removing virus aerosol particles. In this study, a plaque-forming assay combined with SMPS measurement was used to evaluate sufficient drying conditions. We proposed an empirical equation to determine the total number concentration of aerosolized particles measured using the SMPS as a function of the carrier air flow rate and residence time of the particles in the diffusion dryers. The difference in the total number concentration of particles under sufficient and insufficient diffusion drying conditions was presented as a percentage of error.Entities:
Mesh:
Substances:
Year: 2022 PMID: 35449225 PMCID: PMC9022418 DOI: 10.1038/s41598-022-10440-4
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.996
Figure 1Schematic of the experimental design used to generate aerosol particles.
Characteristics of different types of diffusion dryers used during experiments.
| Diffusion dryer type | Tube diameter | Tube length |
|---|---|---|
| 0.02 | 0.28 | |
| 0.02 | 0.24 | |
| 0.015 | 0.48 |
Figure 2Effects of air flow rate and residence time in diffusion dryers on number concentration and size distribution of the aerosolized water droplets. (A) Case 1: one diffusion dryer is used (Table 2). (B) Air flow rate is 2 L/min, and residence time of 2.64, 4.90, 7.16, and 9.70 s represent Case 1, Case 2, Case 3, and Case 4, respectively (Table 2). (C) Air flow rate is 3 L/min, and residence time of 1.76, 3.27, 4.77, and 6.47 s represent Case 1, Case 2, Case 3, and Case 4, respectively (Table 2). (D) Air flow rate is 4 L/min, and residence time of 1.13, 2.45, 3.58, and 4.85 s represent Case 1, Case 2, Case 3, and Case 4, respectively (Table 2). (E) Air flow rate is 5 L/min, and residence time of 1.06, 1.96, 2.86, and 3.88 s represent Case 1, Case 2, Case 3, and Case 4, respectively (Table 2). (F) Effect of the air flow rate on RH is shown for each case.
Overview of different dryer combinations, air flow rates, and corresponding residence times.
| Experiment case | Case 1 | Case 2 | Case 3 | Case 4 |
|---|---|---|---|---|
| Combination of diffusion dryers | ||||
| Air flow rate (L/min) | ||||
| 2 | 2.64 | 4.9 | 7.16 | 9.7 |
| 3 | 1.76 | 3.27 | 4.77 | 6.47 |
| 4 | 1.13 | 2.45 | 3.58 | 4.85 |
| 5 | 1.06 | 1.96 | 2.86 | 3.88 |
Values of and k coefficients and R2 for the non-linear, exponential decay regression .
| Flow rate (L/min) | 2 | 3 | 4 | 5 |
|---|---|---|---|---|
| 1.492e + 5 | 6.425e + 5 | 9.997e + 5 | 1.15e + 6 | |
| 0.3774 | 0.6007 | 0.9703 | 1.1219 | |
| R2 | 0.99 | 0.99 | 0.99 | 0.99 |
Figure 3Effects of different MS2 virus solution concentrations and residence times on number concentration and size distribution of the aerosolized MS2 bacteriophage. (A) Effect of different MS2 virus solution concentrations on size distribution and concentration of the aerosolized particles at = 7.16 s (Table 2). (B) Effect of different residence times on size distribution and concentration of the aerosolized particles under a constant air flow rate of 2 L/min. (C) Effects of different residence times on total number concentration of aerosolized MS2 virus. (D) Plaque analysis results for the sampled MS2 bacteriophage virus under different residence times.
Figure 4Particle loss per particle diameter for 2 L/min air flow rate.