| Literature DB >> 35434406 |
Yusuke Takahashi1, Kazuki Okura1, Shin Minakata1, Motoyuki Watanabe1, Kazutoshi Hatakeyama1, Satoaki Chida1, Kimio Saito1, Toshiki Matsunaga1, Yoichi Shimada2.
Abstract
Objectives: Wearable devices such as fitness trackers have become popular in the healthcare field. Tracking heart rate and respiratory rate, in addition to physical activity, may provide an accurate picture of daily health. We believe that a combination of two types of devices can simultaneously measure and record physical activity, heart rate, and respiratory rate. However, the measurement accuracies of these two types of devices are not clear. This study aimed to determine the measurement accuracies of two wearable devices for heart and respiratory rate measurements.Entities:
Keywords: accuracy; exercise load test; heart rate; respiratory rate; wearable device
Year: 2022 PMID: 35434406 PMCID: PMC8983874 DOI: 10.2490/prm.20220016
Source DB: PubMed Journal: Prog Rehabil Med ISSN: 2432-1354
Fig. 1.Correlation between heart rate measured with wrist-worn device and that measured with electrocardiogram (ECG) during incremental load test. Sudden drops in the heart rate (<80 bpm measured with a wrist-worn device or a difference between measured heart rates >20 bpm) during ILT were excluded from the analysis as measurement errors because the wrist-worn device likely detected the pedal cadence. Visual observation clearly determined that such measurements were inappropriate. bpm, beats per minute.
Summary of the results of Bland–Altman analysis
| Fixed bias | Proportional bias | LOA | ||||||
| 95% CI | Y/N | r | P value | Y/N | ||||
| Wrist-worn device heart rate | Rest | −3.19, −0.07 | Y | −0.055 | 0.780 | N | −9.79, 6.53 | |
| ILT | −17.66, −7.12 | Y | −0.152 | 0.420 | N | −38.31, 12.93 | ||
| CLT | −7.04, −2.19 | Y | 0.382 | 0.004 | Y | −22.22, 12.98 | ||
| Respiratory tracking device respiratory rate | Rest | −0.18, 1.24 | N | −0.353 | 0.030 | Y | −3.79, 4.83 | |
| ILT | −2.10, −1.45 | Y | −0.278 | <0.001 | Y | −6.24, 2.69 | ||
| CLT | −1.49, −0.28 | Y | −0.279 | 0.030 | Y | −5.48, 3.71 | ||
CI, confidence interval; Y, yes; N, no.
Fig. 2.Bland–Altman plots and 95% limits of agreement for at-rest measurements of heart rate (HR) (left) and respiratory rate (right).
Fig. 4.Bland–Altman plots and 95% limits of agreement during constant load tests for measurement of heart rate (HR) (left) and respiratory rate (right).
Summary of the results of regression analysis
| R | R2 | Adjusted R2 | SEE | P value | 95% CI | ||
| Wrist-worn device | |||||||
| Rest | 0.939 | 0.881 | 0.877 | 4.20611 | <0.001 | 0.820, 1.093 | |
| ILT | 0.877 | 0.768 | 0.760 | 13.23761 | <0.001 | 0.740, 1.150 | |
| CLT | 0.942 | 0.887 | 0.885 | 7.72141 | <0.001 | 0.740, 0.901 | |
| Respiratory tracking device | |||||||
| Rest | 0.691 | 0.477 | 0.458 | 3.69535 | <0.001 | 0.498, 1.178 | |
| ILT | 0.679 | 0.461 | 0.456 | 5.24037 | <0.001 | 0.667, 1.034 | |
| CLT | 0.803 | 0.644 | 0.640 | 2.64804 | <0.001 | 0.815, 1.128 | |
SEE, standard error of estimates.