Diagnostic efficacy of the magnetic resonance T1w/T2w ratio for the middle cerebellar peduncle in multiple system atrophy and spinocerebellar ataxia: A preliminary study.

Jiaqi Wang1, Atsuhiko Sugiyama1, Hajime Yokota2, Shigeki Hirano1, Graham Cooper3,4,5,6,7, Hiroki Mukai8, Kenji Ohira8, Kyosuke Koide1, Shoichi Ito1,9, Carsten Finke5,7,10, Alexander U Brandt4,11, Friedemann Paul3,4,5, Satoshi Kuwabara1.   

Abstract

BACKGROUND: The standardized T1-weighted/T2-weighted (sT1w/T2w) ratio for the middle cerebellar peduncle (MCP) has been reported to be sensitive for detecting degenerative changes in the cerebellar subtype of multiple system atrophy (MSA-C), even in the early stages. We aimed to investigate the diagnostic value of the MCP sT1w/T2w ratio for differentiating between MSA-C and spinocerebellar ataxia (SCA).
METHODS: We included 32 MSA-C, 8 SCA type 3 (SCA3), 16 SCA type 6 (SCA6) patients, and 17 controls, and the MCP sT1w/T2w ratio was analyzed using a region-of-interest approach. The diagnostic performance of the MCP sT1w/T2w ratio in discriminating among MSA-C, SCA3, and SCA6 was assessed and compared with diagnosis based on visual interpretation of MCP hyperintensities and the "hot cross bun" (HCB) sign.
RESULTS: MCP sT1w/T2w ratio values were markedly lower in patients with MSA-C than in those with SCA3, those with SCA6, and controls (p < 0.001). The MCP sT1w/T2w ratio showed high diagnostic accuracy for distinguishing MSA-C from SCA3 (area under curve = 0.934), SCA6 (area under curve = 0.965), and controls (area under curve = 0.980). The diagnostic accuracy of the MCP sT1w/T2w ratio for differentiating MSA-C from SCA3 or SCA6 (90.0% for MSA-C vs. SCA3, and 91.7% for MSA-C vs. SCA6) was comparable to or superior than that of visual interpretation of MCP hyperintensities (80.0-87.5% in MSA-C vs. SCA3 and 87.6-97.9% in MSA-C vs. SCA6) or the HCB sign (72.5-80.0% in MSA-C vs. SCA3 and 77.1-93.8% in MSA-C vs. SCA6).
CONCLUSIONS: The MCP sT1w/T2w ratio might be a sensitive imaging-based marker for detecting MSA-C-related changes and differentiating MSA-C from SCA3 or SCA6.

Entities:  

Mesh:

Substances:

Year:  2022        PMID: 35427382      PMCID: PMC9012356          DOI: 10.1371/journal.pone.0267024

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.752


Introduction

Multiple system atrophy (MSA) is an adult-onset, sporadic degenerative disease of the nervous system that is characterized by a combination of autonomic nervous dysfunction, parkinsonism, cerebellar ataxia, and pyramidal signs. MSA is principally divided into two clinical subtypes: MSA with parkinsonism as the predominant manifestation (MSA-P) and MSA with cerebellar ataxia as the predominant manifestation (MSA-C). According to the second consensus statement on the diagnosis of MSA, diagnostic criteria for MSA-C span three categories: definite MSA-C, probable MSA-C, and possible MSA-C [1]. Pons or middle cerebellar peduncle (MCP) atrophy seen on brain magnetic resonance imaging (MRI) is defined as one of the additional features required for a diagnosis of possible MSA-C [1]. Other MRI features, such as the “hot cross bun” (HCB) sign and hyperintensities in the MCP on T2-weighted images (MCP hyperintensities), have also been described in MSA-C patients [2], and it has been proposed that these MRI features can be added to the list of additional features of possible MSA-C depending on their diagnostic utility [3]. MCP hyperintensities are thought to reflect myelin loss in the MCP [4] and have been repeatedly reported to be useful in the diagnosis of MSA-C [4-10]; however, their evaluation is qualitative, is subject to interpreter bias, and have limited sensitivity. Recent studies have demonstrated the utility and unique quantitative contrast provided by the ratio of the signal intensity of T1-weighted and T2-weighted images (T1w/T2w ratio) as they possess high test-retest reliability and are sensitive to neurodegenerative changes [11-13]. Recently, Misaki et al., have proposed standardization of the T1w/T2w ratio (sT1w/T2w ratio), as it would allow meaningful comparison between subjects and scanners by not only creating scaled intensity values but also correcting for inhomogeneities in receiver coil sensitivity [14]. Evidence on the sensitivity of the sT1w/T2w ratio for detecting disease related changes, for example, in multiple sclerosis and MSA, continues to increase, [15-17], and we have previously reported that using the sT1w/T2w ratio for evaluating the MCP in MSA-C patients can help detect early MSA-C related degenerative changes as it shows extremely high diagnostic accuracy in distinguishing between MSA-C and healthy individuals [16]. However, it is not known if the MCP sT1w/T2w ratio is useful for differentiating between MSA-C and spinocerebellar ataxia (SCA), as cerebellar ataxia is a predominant motor symptom in both conditions. SCA types 3 (SCA3) and 6 (SCA6) are common subtypes of autosomal dominant cerebellar ataxia [18], and, in Japan, SCA3 is the most common subtype in which MCP hyperintensities and the HCB sign can be observed [19]. Therefore, to assess the diagnostic utility of the MCP sT1w/T2w ratio in detecting MSA-C and SCA, we calculated the MCP sT1w/T2w ratio for MSA-C, SCA type 3 (SCA3), and SCA type 6 (SCA6), and compared the diagnostic accuracy of MCP sT1w/T2w ratio with that of visual interpretation of MCP hyperintensities and the HCB sign.

Materials and methods

Subjects

This retrospective study was approved by the Institutional Review Board of the Chiba University Graduate School of Medicine and the need for informed consent was waived. The inclusion criteria for patients with MSA-C were as follows: 1.5-T MR images acquired between April 2012 and March 2020, and a diagnosis of clinically confirmed, probable MSA-C based on criteria listed in the second consensus statement [1]. The inclusion criteria for patients with SCA3 and SCA6 were genetically confirmed SCA3 or SCA6 diagnosis and 1.5-T MR images acquired between April 2012 and March 2020. SCA3 and SCA6 patients with a disease duration of 10 years or more were excluded because differentiation from MSA-C is not needed such patients due to reduced life expectancy in MSA. The inclusion criteria for controls were as follows: patients who were referred to our hospital with complaints of headache, dizziness, or lightheadedness; those who had no neurological abnormalities; and those who underwent 1.5-T MRI between April 2012 and March 2020. The study cohort included 32 patients with MSA-C, 8 patients with SCA3, 16 patients with SCA6, and 17 controls. None of the patients had any other central nervous system disorders. Medical records of all patients were reviewed and data on age at MRI, age at onset, and disease duration from onset to MRI were retrieved.

MRI acquisition

All MR images were acquired using a 1.5-T MRI system (Genesis Signa; GE Healthcare) with a 12-channel head coil. The MRI parameters for T1WI were as follows: 3D-IR-SPGR; sagittal plane; TR, 8 ms; TE, 3 ms; flip angle, 15°; FOV, 220 × 220 mm; matrix, 256 × 256; voxel size, 0.86 × 0.86 × 1.5 mm; number of slices, 248; and number of averages, 1. Meanwhile, those for T2WI were as follows: 2D-TSE; TR, 4000 ms; TE, 106 ms; FOV, 220 × 220 mm; matrix, 320 × 256; voxel size, 0.43 × 0.43 × 6 mm; interslice gap, 1.5 mm; number of slices 24; and number of averages 1.

MRI preprocessing

All T1w and T2w images were preprocessed as previously described before the sT1w/T2w ratio was calculated [16]; specifically, the intensity inhomogeneity correction was applied to the T1w and T2w images using N4BiasFieldCorrection [20]. 3D T1w images were linearly co-registered with 2D axial T2w images using SPM12. Brain masks were created by skull-stripping the co-registered T1w images using the Brain Extraction Tool with FSL (version 5.0.11) [21] and binarizing it with FSLmaths. White matter and gray matter brain masks were generated using FMRIB Automatic Segmentation Tool (FSL FAST) on the co-registered T1w image [22].

T1w/T2w ratio and standardized T1w/T2w ratio

Median intensity values of T1w and T2w images, in both white and gray matter masks from each subject, were calculated using FSLstats. To obtain sT1w/T2w ratio, first, a scaling factor was calculated by dividing median gray matter intensity value in T1w images by the median gray matter intensity value in the T2w image. A scaled T2w image (sT2) was then created by multiplying the T2w image by the scaling factor. Finally, the sT1w/T2w ratio was calculated using the following equation developed by Misaki et al [14]. Schematic representation of the pipeline for creating a sT1w/T2w ratio map is described in Fig 1. The sT1w/T2w ratio map for each subject was registered in the Montreal Neurological Institute 152 space [23] using Advanced Normalization Tools (ANTs) [24]. Each image was spatially smoothed with an 8-mm full-width at half-maximum Gaussian Kernel. Regions of interest in the MCP were defined bilaterally on normalized sT1w/T2w ratio maps using a validated probabilistic 3D atlas of the cerebellar white matter structure [25] and SPM in Matlab 2014a, as described previously [14]. Parcellation at a 90% probability threshold was used. Atlas registration accuracy was visually verified using the registration tool of SPM12 in Matlab 2014a. The mean value of median sT1w/T2w ratio values in the left and right MCP regions was used as the MCP sT1w/T2w ratio value for each subject.
Fig 1

Schematic representation of the pipeline for creating a sT1w/T2w ratio map.

T1w, T1-weighted; T2w, T2-weighted; GM, gray matter: WM, white matter; sT2, a scaled T2-weighted image.

Schematic representation of the pipeline for creating a sT1w/T2w ratio map.

T1w, T1-weighted; T2w, T2-weighted; GM, gray matter: WM, white matter; sT2, a scaled T2-weighted image.

Visual interpretation of MCP hyperintensities and the HCB sign

Two board-certified neuroradiologists (H.M. and K.O.) and a board-certified neurologist (A.S.), who were blinded to the clinical data, independently evaluated MCP hyperintensities and the HCB sign for each subject, except for controls. MCP hyperintensities were marked as present or absent, where present was defined as “when high intensities relative to that of the brainstem and cerebellum in the adjacent white matter were observed on at least one side of the MCP” (Fig 2). The HCB sign was graded as described in a previous study [26], where 0 denoted no changes; 1 indicated initial appearance or presence of a clear vertical T2 high-intensity; and 2 indicated initial appearance of a horizontal line along with a vertical line or the presence of clear horizontal and vertical lines in the ventral pons (Fig 2). Here both grade 1 and grade 2 were defined as positive for HCB sign.
Fig 2

Typical Middle Cerebellar Peduncle (MCP) hyperintensities and Hot Cross Bun (HCB) sign on magnetic resonance imaging in cerebellar-type multiple system atrophy (MSA-C) and spinocerebellar ataxia type 3 (SCA3) and 6 (SCA6).

(A) Grade 0 (negative) HCB sign and absent MCP hyperintensities in a 74-year-old man with a SCA6 disease duration of 2.3 years. (B) Grade 1 HCB sign and absent MCP hyperintensities in a 66-year-old man with SCA3 and a disease duration of 3.8 years. (C) Grade 1 HCB sign and MCP hyperintensities in a 57-year-old man with MSA-C and a disease duration of 0.9 years. (D) Grade 2 HCB sign and MCP hyperintensities in a 62-year-old woman with MSA-C and a disease duration of 2.8 years. Upper row: axial T2-weighted images. Lower row: standardized T1-weighted/T2-weighted ratio maps.

Typical Middle Cerebellar Peduncle (MCP) hyperintensities and Hot Cross Bun (HCB) sign on magnetic resonance imaging in cerebellar-type multiple system atrophy (MSA-C) and spinocerebellar ataxia type 3 (SCA3) and 6 (SCA6).

(A) Grade 0 (negative) HCB sign and absent MCP hyperintensities in a 74-year-old man with a SCA6 disease duration of 2.3 years. (B) Grade 1 HCB sign and absent MCP hyperintensities in a 66-year-old man with SCA3 and a disease duration of 3.8 years. (C) Grade 1 HCB sign and MCP hyperintensities in a 57-year-old man with MSA-C and a disease duration of 0.9 years. (D) Grade 2 HCB sign and MCP hyperintensities in a 62-year-old woman with MSA-C and a disease duration of 2.8 years. Upper row: axial T2-weighted images. Lower row: standardized T1-weighted/T2-weighted ratio maps. A board-certified neuroradiologist (H. Y.), who was blinded to the clinical data, semiquantitatively evaluated the extent of MCP hyperintensities in the patients with MSA by applying the following scores: 0 = absent, 1 = mild, 2 = moderate, and 3 = severe.

MCP volume analysis

The MCP volume was measured in all subjects by exploiting a validated probabilistic 3D atlas of the cerebellar white matter structure [26]. Parcellation at a 90% probability threshold was used. The atlas of the MCP was transformed from the Montreal Neurological Institute 152 space [23] to the individual subject’s space using the deformation parameters obtained from the normalization procedure using ANTs [24]. The mean volume value in the left and right MCPs was used as the MCP volume for each subject. For further analyses, the MCP volume was normalized by each subject’s intracranial volume. Values represent the ratio of the volume of the MCP divided by the subject-specific total intracranial volume (written as m × 10−3).

Statistical analyses

All statistical analyses, except for receiver operating characteristic (ROC) curve analyses, were performed using Statistical Package for the Social Sciences (version 25.0; SPSS Inc., IBM Corp., Chicago, IL, USA). The ROC curve analyses and Smirnov–Grubb’s test were performed using EZR (Saitama Medical Center, Jichi Medical University, Saitama, Japan), which is a graphical user interface for R (The R Foundation for Statistical Computing, Vienna, Austria, version 2.13.0) [27]. More precisely, EZR is a modified version of R commander (version 1.6–3) designed to add statistical functions frequently used in biostatistics. Demographic data of the patients with MSA-C, SCA3, and SCA6 and controls were compared using the chi-square (χ2) test for sex and univariate one-way analysis of variance for age at MRI and onset. Differences in disease duration among the three groups of patients were analyzed using the Kruskal–Wallis test, followed by post-hoc Mann–Whitney U tests adjusted for multiple comparisons (Bonferroni correction). For evaluating differences in MCP sT1w/T2w ratio values and MCP volume among the four groups, a one-way analysis of covariance was used using age as the covariate. The ability of the MCP sT1w/T2w ratio values and MCP volume to differentiate between groups of patients was assessed using ROC curve analyses, and the optimal cutoff point was determined as the point on the ROC curve closest to the upper left corner, as used by Holmes [28]. The DeLong test was used to compare the area under curves (AUCs) between the MCP sT1w/T2w ratio and MCP volume to differentiate among the four groups. Diagnostic accuracy of the MCP sT1w/T2w ratio and visual interpretation of MCP hyperintensities and HCB sign provided by three reviewers were compared using binomial tests. The intraclass correlation coefficient (ICC) was calculated to assess inter-reviewer variability in visual interpretation of the MCP hyperintensities and the HCB sign. Spearman correlation analysis was performed to evaluate the relationship between the extent of MCP hyperintensities assessed semiquantitatively and the MCP sT1w/T2w ratio values in patients with MSA-C. Pearson correlation analysis was performed to evaluate the relationship between the MCP sT1w/T2w ratio values and MCP volume in patients with MSA-C. The Smirnov–Grubb’s test was used for evaluating the outliers of the MCP sT1w/T2w ratio value in each subject group. P < 0.05 was considered statistically significant.

Results

Demographic and clinical data of patients with MSA-C, SCA3, and SCA6 are summarized in Table 1.
Table 1

Demographic and clinical data of the patients with cerebellar subtype multiple system atrophy (MSA-C), spinocerebellar ataxia type 3 (SCA3), and type 6 (SCA6) and controls.

Group (n)Sex DistributionaAge at MRIbDisease DurationcAge at Onsetb
(Male/Female)(years, mean ± SD)(years, median, range)(years, mean ± SD)
MSA-C (n = 32)19/1362.8 ± 9.42.3 (0.8–5.9)60.4 ± 9.4
SCA3 (n = 8)2/651.4 ± 16.74.5 (0.8–5.7)47.3 ± 17.4
SCA6 (n = 16)10/658.3 ± 13.73.5 (0.3–9.8)54.1 ± 12.3
Control (n = 17)9/862.4 ± 10.8NANA
P-value for group comparisons0.3500.078 0.026 0.069
P-value for post-hoc group comparisons
MSA-C vs. SCA3NANA0.030NA
MSA-C vs. SCA6NANA0.034NA
MSA-C vs. controlNANANANA
SCA3 vs. SCA6NANA0.742NA
SCA3 vs. controlNANANANA
SCA6 vs. controlNANANANA

MRI, magnetic resonance imaging; NA, not applicable.

a Chi-squared test.

b Parametric tests (univariate 1-way analysis of variance [ANOVA]).

c Nonparametric tests (Kruskal-Wallis 1-way ANOVA with post-hoc Mann-Whitney U tests adjusted for multipe comparisons; P value for comparisons of Disease Duration: P < 0.05/3 = 0.0167).

MRI, magnetic resonance imaging; NA, not applicable. a Chi-squared test. b Parametric tests (univariate 1-way analysis of variance [ANOVA]). c Nonparametric tests (Kruskal-Wallis 1-way ANOVA with post-hoc Mann-Whitney U tests adjusted for multipe comparisons; P value for comparisons of Disease Duration: P < 0.05/3 = 0.0167). The MCP sT1w/T2w ratio values were significantly lower in patients with MSA-C than those in patients with SCA3 and SCA6 and controls (0.07 ± 0.06 vs. 0.17 ± 0.03, 0.18 ± 0.03, and 0.19 ± 0.03, respectively; p < 0.001) (Fig 3). The AUC for differentiating MSA-C from SCA-3 was 0.934 (95% confidence interval [CI], 0.854–1.000), with a sensitivity of 0.906 and specificity of 0.875. The AUC value for differentiating between MSA-C and SCA-6 was 0.965 (95% CI, 0.922–1.000), with a sensitivity of 0.906 and a specificity of 0.938. The AUC for differentiating MSA-C from controls was 0.980 (95% CI, 0.951–1.000), with a sensitivity of 0.906 and specificity of 0.941. The AUC for differentiating SCA-3 from SCA-6 was 0.633 (95% CI, 0.388–0.878), with a sensitivity of 0.750 and specificity of 0.625 (Fig 4). Even after excluding one patient with MSA-C with an outlier MCP sT1w/T2w ratio value, the MCP sT1w/T2w ratio values in patients with MSA-C were significantly lower than those in patients with SCA3 and SCA6 and controls (0.07 ± 0.05 vs. 0.17 ± 0.03, 0.18 ± 0.03, and 0.19 ± 0.03, respectively; p < 0.001).
Fig 3

Comparison of the Middle Cerebellar Peduncle (MCP) standardized T1-weighted/T2-weighted (sT1w/T2w) ratio among patients with cerebellar subtype multiple system atrophy (MSA-C), spinocerebellar ataxia type 3 (SCA-3), and spinocerebellar ataxia type 6 (SCA-6) and controls.

The MCP sT1w/T2w ratio values were significantly lower in patients with MSA-C than those in patients with SCA-3 and SCA-6 and controls (*p < 0.001).

Fig 4

Receiver Operating (ROC) curves for the MCP sT1w/T2w ratio showing area under the curve (AUC), sensitivity (Sen), and specificity (Spec).

A, B, and C are the ROC curves distinguishing cerebellar subtype multiple system atrophy (MSA-C) from spinocerebellar ataxia type 3 (SCA3) (A), spinocerebellar ataxia type 6 (SCA-6) (B), and controls (C). D is the ROC curve distinguishing SCA-3 from SCA-6.

Comparison of the Middle Cerebellar Peduncle (MCP) standardized T1-weighted/T2-weighted (sT1w/T2w) ratio among patients with cerebellar subtype multiple system atrophy (MSA-C), spinocerebellar ataxia type 3 (SCA-3), and spinocerebellar ataxia type 6 (SCA-6) and controls.

The MCP sT1w/T2w ratio values were significantly lower in patients with MSA-C than those in patients with SCA-3 and SCA-6 and controls (*p < 0.001).

Receiver Operating (ROC) curves for the MCP sT1w/T2w ratio showing area under the curve (AUC), sensitivity (Sen), and specificity (Spec).

A, B, and C are the ROC curves distinguishing cerebellar subtype multiple system atrophy (MSA-C) from spinocerebellar ataxia type 3 (SCA3) (A), spinocerebellar ataxia type 6 (SCA-6) (B), and controls (C). D is the ROC curve distinguishing SCA-3 from SCA-6. Sensitivity, specificity, positive and negative predictive values, and overall correct classification for each reviewer for visual evaluation of MCP hyperintensities and the HCB sign are shown in Table 2. MCP hyperintensities were not found in SCA6 patients and the sensitivity of MCP hyperintensities in MSA-C patients ranged from 84.4%–96.9%. Grade 2 HCB sign was highly specific for differentiating MSA-C from SCA3 and SCA6 but with moderate sensitivity. Grade 1 or 2 HCB sign (vertical or cruciform hyper-intensity) was highly sensitive (87.5%–100%) for MSA-C. However, even though grade 1 or 2 HCB sign could be observed in SCA3 patients, it was characterized by relatively low specificity for differentiating between MSA-C and SCA3. ICC values demonstrated substantial inter-reviewer agreement for visual interpretation of the MCP hyperintensities and the HCB sign (0.762 and 0.729, respectively; all p < 0.001). In patients with MSA-C, a significant inverse correlation between the extent of MCP hyperintensities and MCP sT1w/T2w ratio was observed (r = −0.832; p < 0.001).
Table 2

Visual evaluation of the Middle Cerebellar Peduncle (MCP) hyperintensities and hot cross bun sign for cerebellar subtype multiple system atrophy (MSA-C) and spinocerebellar ataxia type 3 (SCA3) and 6 (SCA6).

MSA vs. SCA3MSA vs. SCA6SCA3 vs. SCA6
Reviewer 1Reviewer 2Reviewer 3Reviewer 1Reviewer 2Reviewer 3Reviewer 1Reviewer 2Reviewer 3
MCP hyperintensities
 Sensitivity, % (95% CI)90.696.984.490.696.984.425.087.512.5
(75.0–98.0)(83.8–99.9)(67.2–94.7)(75.0–98.0)(83.8–99.9)(67.2–94.7)(3.2–65.1)(47.3–99.7)(0.3–52.7)
 Specificity, % (95% CI)75.012.587.5100100100100100100
(34.9–96.8)(0.3–52.7)(47.3–99.7)(71.3–100)(71.3–100)(71.3–100)(71.3–100)(71.3–100)(71.3–100)
 Positive predictive value, % (95% CI)93.581.696.4100100100100100100
(78.6–99.2)(65.7–92.3)(81.7–99.9)(82.8–100)(83.8–100)(81.7–100)(9.4–100)(47.3–100)(1.3–100)
 Negative predictive value, % (95% CI)66.750.058.384.294.176.272.794.169.6
(29.9–92.5)(1.3–98.7)(27.7–84.8)(60.4–96.6)(71.3–99.9)(52.8–91.8)(49.8–89.3)(71.3–99.9)(47.1–86.8)
 Overall correct classification, % (95% CI)87.580.085.093.897.989.675.095.870.8
(73.2–95.8(64.4–90.9)(70.2–94.3)(82.8–98.7)(88.9–99.9)(77.3–96.5)(53.3–90.2)(78.9–99.9)(48.9–87.4)
Grade 2 HCB sign (cruciform hyper-intensity)
 Sensitivity, % (95% CI)59.481.27559.481.2750012.5
(40.6–76.3)(63.6–92.8)(56.6–88.5)(40.6–76.3)(63.6–92.8)(56.6–88.5)(0–48.2)(0–48.2)(0.3–52.7)
 Specificity, % (95% CI)10010087.5100100100100100100
(51.8–100)(51.8–100)(47.3–99.7)(71.3–100)(71.3–100)(71.3–100)(71.3–100)(71.3–100)(71.3–100)
 Positive predictive value, % (95% CI)10010096.010010010000100
(75.1–100)(81.0–100)(79.6–99.9)(75.1–100)(81.0–100)(79.6–100)(0–100)(0–100)(1.3–100)
 Negative predictive value, % (95% CI)38.157.146.755.272.766.766.766.769.6
(18.1–61.6)(28.9–82.3)(21.3–73.4)(35.7–73.6)(49.8–89.3)(44.7–84.4)(44.7–84.4)(44.7–84.4)(47.1–86.8)
 Overall correct classification, % (95% CI)67.585.077.572.987.583.366.766.770.8
(50.9–81.4)(70.2–94.3)(61.5–89.2)(58.2–84.7)(74.8–95.3)(69.8–92.5)(44.7–84.4)(44.7–84.4)(48.9–87.4)
Grade 1 or 2 HCB sign (vertical or cruciform hyper-intensity)
 Sensitivity, % (95% CI)87.593.810087.593.810087.5100100
(71.0–96.5)(79.2–99.2)(84.2–100)(71.0–96.5)(79.2–99.2)(84.2–100)(47.3–99.7)(51.8–100)(51.8–100)
 Specificity, % (95% CI)12.50093.893.831.293.893.831.2
(0.3–52.7)(0–48.2)(0–48.2)(69.8–99.8)(69.8–99.8)(11.0–58.7)(69.8–99.8)(69.8–99.8)(11.0–58.7)
 Positive predictive value, % (95% CI)80.078.980.096.696.874.487.588.942.1
(63.1–91.6)(62.7–90.4)(64.4–90.9)(82.2–99.9)(83.3–99.9)(58.8–86.5)(47.3–99.7)(51.8–99.7)(20.3–66.5)
 Negative predictive value, % (95% CI)20.00078.988.210093.8100100
(0.5–71.6)(0–90.6)(0–100)(54.4–93.9)(63.6–98.5)(35.9–100)(69.8–99.8)(69.8–100)(35.9–100)
 Overall correct classification, % (95% CI)72.575.080.089.693.877.191.795.854.2
(56.1–85.4)(58.8–87.3)(64.4–90.9)(77.3–96.5)(82.8–98.7)(62.7–88.0)(73.0–99.0)(78.9–99.9)(32.8–74.4)
Diagnostic accuracy of the MCP sT1w/T2w ratio, visual interpretation of the MCP hyperintensities, and the HCB sign are listed in Table 3, and we found that the diagnostic performance of the MCP sT1w/T2w ratio for differentiating MSA-C from SCA3, SCA6, and controls was comparable to or better than that of visual interpretation of MCP hyperintensities and the HCB sign for all three reviewers. However, the diagnostic performance of visual interpretation of MCP hyperintensities and the HCB sign for differentiating SCA3 from SCA6 was comparable to or better than that of the MCP sT1w/T2w ratio.
Table 3

Comparison of accuracy between the MCP sT1w/T2w ratio and visual evaluation by three reviewers.

MSA vs. SCA3MSA vs. SCA6SCA3 vs. SCA6
AccuracyP value (compared with MCP sT1w/T2w ratio)AccuracyP value (compared with MCP sT1w/T2w ratio)AccuracyP value (compared with MCP sT1w/T2w ratio)
MCP sTw/T2w ratio90.0 (76.3–97.2)91.7 (80.0–97.7)66.7 (44.7–84.4)
MCP hyperintensities
 Reviewer 187.50.37193.80.42875.00.125
(73.2–95.8)(82.8–98.7)(53.3–90.2)
 Reviewer 280.0 0.042 97.90.08395.8 0.001
(64.4–90.9)(88.9–99.9)(78.9–99.9)
 Reviewer 385.00.20689.60.36870.80.161
(70.2–94.3)(77.3–96.5)(48.9–87.4)
Grade 2 HCB sign
 Reviewer 167.5 < 0.001 72.9 < 0.001 66.70.171
(50.9–81.4)(58.2–84.7)(44.7–84.4)
 Reviewer 285.00.20687.50.20566.70.171
(70.2–94.3)(74.8–95.3)(44.7–84.4)
 Reviewer 377.5 0.015 83.3 0.043 70.80.161
(61.5–89.2)(69.8–92.5)(48.9–87.4)
Grade 1 or 2 HCB sign
 Reviewer 172.5 0.001 89.60.36891.7 0.004
(56.1–85.4)(77.3–96.5)(73.0–99.0)
 Reviewer 275.0 0.005 93.80.42895.8 0.001
(58.8–87.3)(82.8–98.7)(78.9–99.9)
 Reviewer 380.00.04277.1 0.002 54.20.072
(64.4–90.9)(62.7–88.0)(32.8–74.4)
The MCP volume, normalized by each subject’s intracranial volume, was significantly lower in patients with MSA-C patients than those in patients with SCA6 and controls (0.20 ± 0.06 vs. 0.31 ± 0.05 and 0.34 ± 0.05, respectively; p < 0.001). Moreover, the MCP volume was significantly lower in patients with SCA3 than those in patients with SCA6 and controls (0.23 ± 0.04 vs. 0.31 ± 0.05 and 0.34 ± 0.05, respectively; p < 0.001; p = 0.01 and p < 0.001). The AUC value for differentiating MSA-C from SCA3 was 0.664 (95% CI, 0.486–0.842), with a sensitivity of 0.750 and specificity of 0.625. The AUC for differentiating MSA-C from SCA6 was 0.924 (95% CI, 0.852–0.995), with a sensitivity of 0.844 and specificity of 0.875. The AUC for differentiating MSA-C from controls was 0.967 (95% CI, 0.925–1.000), with a sensitivity of 0.906 and specificity of 0.941). The AUC for differentiating SCA3 from SCA6 was 0.914 (95% CI, 0.802–1.000), with a sensitivity of 0.875 and specificity of 0.875. The AUC of the MCP sT1w/T2w ratio was greater than that of the MCP volume for differentiating MSA-C from SCA3 (p < 0.001). No significant differences were observed between the AUC of the MCP sT1w/T2w ratio and that of MCP volume for differentiating MSA-C from SCA6 and controls (p = 0.234 and 0.577, respectively). The AUC of the MCP volume was greater than that of the MCP sT1w/T2w ratio for differentiating SCA3 from SCA6 (p = 0.020). In patients with MSA-C, a significant correlation was observed between the MCP volume and MCP sT1W/T2W ratio (r = 0.619; p < 0.001).

Discussion

We showed that the MCP sT1W/T2W ratio is highly sensitive and specific for distinguishing MSA-C from SCA3, SCA6, or controls. Additionally, the diagnostic accuracy of the MCP sT1w/T2w ratio for differentiating MSA-C from SCA3 or SCA6 was comparable or superior to that of visual interpretation of MCP hyperintensities or the HCB sign. Given the variability in visual interpretation of MCP hyperintensities and the HCB sign among the three reviewers seen in the current study, the MCP sT1w/T2w ratio appears to be a feasible, reliable, and clinically valuable imaging biomarker for differentiating MSA-C from SCA3 or SCA6. Decreased MCP sT1W/T2W ratio values in patients with MSA-C might reflect demyelination and gliosis in the MCP due to MSA-C-related degeneration. Pathologically, the loss of axons, myelin, and reactive gliosis was observed in the MCP of patients with MSA-C [29]. The sT1W/T2W ratio is sensitive to myelin and decreases in demyelinating lesions [30, 31]. In contrast, inconsistent correlations with histology [32], myelin water imaging [12, 33], simultaneous tissue relaxometry, and magnetization transfer saturation index [34] suggest that the sT1W/T2W ratio is sensitive to not only myelin content but also other microstructural factors. Moreover, gliosis is presumed to affect the sT1W/T2W ratio in a downward direction, mainly by prolonged T1 relaxation time [35]. Therefore, although the sT1W/T2W ratio might not be specific for myelin content, it might be more sensitive than evaluating MCP hyperintensities on T2WI to detect degenerative processes in the MCP, such as demyelination and gliosis associated with MSA-C. In this study, the MCP sT1W/T2W ratio had better diagnostic accuracy in discriminating MSA-C from SCA3 than MCP volume. This result is consistent with those of previous pathological studies, showing that gliosis and demyelination precede atrophy in MSA [36, 37]. The usefulness of the MCP sT1W/T2W ratio in discriminating MSA-C from SCA3 in this study might be due to differences in the degree of MCP degeneration in the two diseases. Glial cytoplasmic inclusions (GCIs) are found in oligodendroglial cells, are recognized neuropathological hallmarks of MSA, and their density significantly correlates with disease progression [38, 39]. One of the regions where GCI pathology progresses in the earliest phases of MSA-C is the pontocerebellar fibers in the MCP, and regional severity of GCI pathology has been shown to correlate with the severity of myelin loss [40]. In SCA3, pathological changes are mainly seen in the efferent dentatorubral system, which passes through the superior peduncle, and in the pallidosubthalamic system [41]. Further, although degenerative changes can also be observed in the MCP, they are less severe than those seen in the superior peduncle [42, 43]. Previous MRI studies have also reported that MCP hyperintensities, which reflect degenerative changes in the MCP, are less frequently seen in patients with SCA3 than in patients with MSA, and that they were rarely observed during the early stages of SCAs, including SCA3 [10, 44]. The substantial discriminatory power of the MCP sT1w/T2w ratio and visual interpretation of MCP hyperintensities and the HCB sign described here implies that they may also be useful for differentiating between SCA3 and SCA6. The pathological background for this may be the difference in MCP degeneration, which is observed to some extent in SCA3, as described above, but is rarely observed in SCA6. Previous pathoanatomical studies have identified the cerebellar cortex, the dentate nucleus, and the inferior olive as predilection sites in SCA6 [45, 46]. Although recent studies have shown that neurodegeneration in SCA6 is more widely distributed, it is less widespread and less severe than in SCA1, SCA2, SCA3, and SCA7 [47, 48]. Concurring with our results, a previous MRI study has reported that MCP hyperintensities were not observed in any of the patients with SCA6 [49]. The cohort of SCA3 and SCA6 patients included in this study is not large, and therefore, further studies that include data from a large cohort of SCA3 and SCA6 patients are needed to verify the usefulness of MCP sT1w/T2w ratio values and visual interpretation of MCP hyperintensities and the HCB sign in differentiating between SCA3 and SCA6. This study has several limitations. The first is that 2D T2W images were co-registered with 3D T1W images, which do not have the same image resolution, and a 1.5-T MRI scanner with an inherently low signal-to-noise ratio was used; therefore, the results of the MCP sT1W/T2W ratio may be biased by noise. However, the MCP sT1W/T2W ratio values in patients with MSA-C correlated well with the extent of MCP hyperintensities and MCP volume, suggesting that the MCP sT1W/T2W ratio values measured in this study well reflected the degeneration of the MCP in MSA-C. Moreover, the diagnostic accuracy of the MCP sT1W/T2W ratio in discriminating MSA-C from controls in this study was comparable to that observed in a study using a 3-T MRI scanner [16]. Further studies employing 3D T1WI and T2WI sequences using a 3-T MRI scanner, which might allow better quantification of the sT1W/T2W ratio, are needed to confirm the results of this study. Second, we did not pathologically confirm the diagnosis of MSA, and thus, the possibility of a misdiagnosis in some cases cannot be excluded. Third, owing to the retrospective nature of this study, the relationship between the MCP sT1W/T2W ratio and rapid eye movement sleep behavior disorder or clinical scales for cerebellar ataxia, such as the International Cooperative Ataxia Rating Scale or Scale for the Assessment and Rating of Ataxia, could not be evaluated. Finally, the sample size of the SCA3 group was small (n = 8), and future prospective studies with larger sample sizes that evaluate clinical scales for cerebellar ataxia are needed to validate the findings of this study and clarify the relationship between clinical severity of cerebellar ataxia and the MCP sT1W/T2W ratio.

Conclusions

In conclusion, the MCP sT1W/T2W ratio might be a sensitive imaging-based marker for detecting MSA-C-related changes and useful for differentiating MSA-C from SCA3 or SCA6.

Details of the clinical characteristics of subjects.

(XLSX) Click here for additional data file. 20 Sep 2021
PONE-D-21-24426
Diagnostic efficacy of the magnetic resonance T1w/T2w ratio for the middle cerebellar peduncle in multiple system atrophy and spinocerebellar ataxia
PLOS ONE Dear Dr. Sugiyama, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.
 
Major methodological concerns have been raised, especially by Reviewer 2. It is unlikely that these limitations/issues can be fully addressed with the data that has been presented. It would be highly desirable if you could include additional scans, though, that are acquired with a protocol that is more similar to what is typically done with T1w/T2w ratio to show that your results are comparable.
Please submit your revised manuscript by Nov 04 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript:
A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'. A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'. An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'. If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols. We look forward to receiving your revised manuscript. Kind regards, Niels Bergsland Academic Editor PLOS ONE Journal Requirements: When submitting your revision, we need you to address these additional requirements. 1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf 2. We noticed you have some minor occurrence of overlapping text with the following previous publication, which needs to be addressed: - https://link.springer.com/article/10.1007%2Fs00330-020-07521-1 The text that needs to be addressed involves the materials and methods section. In your revision ensure you cite all your sources (including your own works), and quote or rephrase any duplicated text. Further consideration is dependent on these concerns being addressed. 3. In your Data Availability statement, you have not specified where the minimal data set underlying the results described in your manuscript can be found. PLOS defines a study's minimal data set as the underlying data used to reach the conclusions drawn in the manuscript and any additional data required to replicate the reported study findings in their entirety. All PLOS journals require that the minimal data set be made fully available. For more information about our data policy, please see http://journals.plos.org/plosone/s/data-availability. Upon re-submitting your revised manuscript, please upload your study’s minimal underlying data set as either Supporting Information files or to a stable, public repository and include the relevant URLs, DOIs, or accession numbers within your revised cover letter. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories. Any potentially identifying patient information must be fully anonymized. Important: If there are ethical or legal restrictions to sharing your data publicly, please explain these restrictions in detail. Please see our guidelines for more information on what we consider unacceptable restrictions to publicly sharing data: http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions. Note that it is not acceptable for the authors to be the sole named individuals responsible for ensuring data access. We will update your Data Availability statement to reflect the information you provide in your cover letter. 4. Your ethics statement should only appear in the Methods section of your manuscript. If your ethics statement is written in any section besides the Methods, please delete it from any other section. [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: Partly ********** 2. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes Reviewer #2: No ********** 3. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: Yes ********** 4. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: The authors wrote an outstanding manuscript entitled "Diagnostic efficacy of the magnetic resonance T1w/T2w ratio for the middle cerebellarpeduncle in multiple system atrophy and spinocerebellar ataxia". They concluded that the MCP sT1w/T2w ratio is a sensitive imaging-based marker for detecting MSA-C related changes and might be useful for differentiating MSA-C from SCA3 or SCA6. I agree with the authors and I have some questions and comments: 1- Was there a correlation between MRI findings suggestive of MSA-C and the presence of a REM sleep behavior disorder? 2-Was there a correlation between MRI findings suggestive of MSA-C and the SARA ataxia scale? 3-Without a doubt, a prospective study would be very suitable to analyze the results of this study in greater depth, as well as with a greater number of cases of MSA-C and the group of patients with SCAs 3 and 6. Reviewer #2: This study sought to investigate the diagnostic value of standardized T1w/T2w ratio in the middle cerebellar peduncle (MCP) for differentiating between the cerebellar subtype of multiple system atrophy (MSA-C) and spinocerebellar ataxia (SCA). The authors conducted this study on 32 patients with MSA-C, 8 patients with SCA3and 16 patients with SCA6 at 1.5 T MRI. They claimed that sT1w/T2w in MCP is a highly sensitive imaging-based marker for detecting MSA-C related changes and might be useful for differentiating MSA-C from SCA3 or SCA6.  However, a couple of major and minor concerns need to be addressed as listed below: 1.     One of the major concerns of the paper is the data used. The results might be biased mostly by noise. Data used in this work is 2D and image resolutions arenot the same for T1w and T2w images and data was collected with 1.5 T MRI whichinherently has a low signal-to-noise ratio. 2.     The authors haven’t clearly mentioned what they are measuring with T1w/T2w and how sensitive the method is to tissue microstructure especially the myelin. T1w/T2w method is considered to be sensitive to myelin but depends on how T1w and T2w images are acquired as well as the image processing pipeline. For better quantification of T1w/T2w, 3D T1w and 3D T2w SPACE sequences are recommended. A recent work by Uddin et al (NMR in Biomedicine 2018) reported that TSE based T1w/T2w ratio is not a good measure to quantify myelin. 3.     The authors mentioned about T2 hyperintensities in MCP. I was wondering if they have any scores for WM hyperintensities or disease severity scores that can be correlated with T1w/T2w. 4.     MRI acquisition: data were acquired with a 1.5 T scanner. Please add information about the RF coil (i.e, 32-channel head coil), number of averages, number of slices. 5.     MRI preprocessing: “3D T1w images were linearly co-registered with 2D  axial T2w images using SPM12. Brain masks were created by skull-stripping the co-registered T1w images using the BrainExtraction Tool with FSL (version 5.0.11) [20] and binarizing it with FSLmaths.”I was wondering why did the authors use SPM for coregistration only. It's better to use FSL or SPM, sometime the header information is not properly copied from one platform to other. However, it's possible to copy the headers from spm to fsl properly. 6.     T1w/T2w calculation: The authors mentioned that for T1w/T2w quantification they used previously published technique by Misaki et al MRM 2014, but it is not clear how did they perform bias correction which is important for better quantification ofT1w/T2w. Did they use their script or just follow their approach (it looks there is no publicly available tool online by Misaki et al)? Depending on the bias correction technique, T1w/T2w maps may contain residual radiofrequency transmit field (B1+) biases, which maybe correlated with these variables of interest, leading to potentially spurious results. 7.     To measure diagnostic accuracy by Youden ROC method, I suspect about the accuracy of the measurement with this small sample size. 8.     Figure1: you haven’t mentioned what type of images (i.e., T2-weighted images) are in the figure. Please include the corresponding T1w/T2w maps with the signal intensity scale. 9.     Figure2: significant difference between MSA-C with SCA3 and SCA6 might be derived from the outliers of MSA-C data. I would use the outlier removal tool and compare the results. 10.  Discussion section should be more organized. 11.  It would be great to show the comparison of volumetric measure of the MCP indifferent disease conditions. 12.  Please add the small sample size as a limitation. 13.  “In conclusion, the MCP sT1w/T2w ratio is a highly sensitive imaging-based marker for detecting MSA-C related changes and might be useful for differentiating 274 MSA-C from SCA3 or SCA6.” With a relatively small sample size and poor quality of data (especially 2D T1w, T2w data, 1.5 mm vs. 6mm slice thickness collected at 1.5 T), it is tough to claim “highly sensitive imaging-based marker ….” ********** 6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: Yes: Md Nasir Uddin [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. 5 Nov 2021 04/11/2021 Prof. Emily Chenette Editor-in-Chief PLOS ONE Dear Prof. Chenette: We greatly appreciate your review of our manuscript and the helpful suggestions. We have now revised the manuscript according to the reviewers’ comments. Changes made in accordance with the reviewers’ comments are indicated by red text in the revised manuscript. In your email of September 20, you recommended including additional scans acquired with a protocol that is more similar to what is typically done with T1w/T2w ratio. However, scans with such protocol have not been taken in the patients with SCA3 or SCA6 in our department so far. Instead, we have added a control group to the current study and have shown that diagnostic accuracy in differentiating MSA from controls is comparable to a previous study using 3T MRI (Sugiyama A, et al. Eur Radiol 2021;31:4277-4284). We hope that you will consider this revised version suitable for publication. Sincerely, Dr. Atsuhiko Sugiyama, MD, PhD Department of Neurology, Graduate School of Medicine, Chiba University Phone No: +81-43-226-2129 Fax No: +81-43-226-2160 Email Address: asugiyama@chiba-u.jp Reply to Reviewer 1 We greatly appreciate your helpful comments and suggestions. Changes made in accordance with reviewers’ comments are indicated in red font in the revised manuscript. #1 Was there a correlation between MRI findings suggestive of MSA-C and the presence of a REM sleep behavior disorder? Reply: Unfortunately, there was no systematic evaluation of RBD using questionnaires or PSG in the patients with MSA included in this study. Moreover, some patients were not asked about their RBD since little attention had been paid to the relationship between RBD and MSA in the past. We have added a note to the Discussion section about this as a limitation, as follows. Page 21, line 348-351 “Third, owing to the retrospective nature of this study, the relationship between the MCP sT1W/T2W ratio and rapid eye movement sleep behavior disorder or clinical scales for cerebellar ataxia, such as the International Cooperative Ataxia Rating Scale or Scale for the Assessment and Rating of Ataxia, could not be evaluated.” #2 Was there a correlation between MRI findings suggestive of MSA-C and the SARA ataxia scale? Reply: We agree with the reviewer’s point that assessing the correlation between MRI findings and cerebellar symptom rating scales, including SARA, is important. Unfortunately, no systematic assessment of cerebellar symptoms, including SARA, has been performed in the patients with MSA included in this study. We have added a note to the Discussion section about this as a limitation, as follows. We hope to evaluate the correlation between SARA and MRI findings in future prospective studies. Page 21, line 348-351 “Third, owing to the retrospective nature of this study, the relationship between the MCP sT1W/T2W ratio and rapid eye movement sleep behavior disorder or clinical scales for cerebellar ataxia, such as the International Cooperative Ataxia Rating Scale or Scale for the Assessment and Rating of Ataxia, could not be evaluated.” #3 Without a doubt, a prospective study would be very suitable to analyze the results of this study in greater depth, as well as with a greater number of cases of MSA-C and the group of patients with SCAs 3 and 6. Reply: We agree with the reviewer’s comment and have revised and added sentences in the Discussion section, as follows. Page 20, line 336 “This study has several limitations.” Page 21, line 352-355 “Fourth, the sample size of the SCA3 group was small (n = 8), and future prospective studies with larger sample sizes that evaluate clinical scales for cerebellar ataxia are needed to validate the findings of this study and clarify the relationship between clinical severity of cerebellar ataxia and the MCP sT1W/T2W ratio.” Reply to Reviewer 2 We greatly appreciate your helpful comments and suggestions. Changes made in accordance with reviewers’ comments are indicated in red font in the revised manuscript. #1 One of the major concerns of the paper is the data used. The results might be biased mostly by noise. Data used in this work is 2D and image resolutions are not the same for T1w and T2w images and data was collected with 1.5 T MRI which hinherently has a low signal-to-noise ratio. Reply: We agree with the reviewer’s concerns about bias in the data due to the MRI sequences and magnetic field strength of the MRI scanner used in this study. We have added sentences in the Discussion section as a limitation on this point, as follows. Page 20, line 336 to Page 21 line 346 “The first is that 2D T2W images were co-registered with 3D T1W images, which do not have the same image resolution, and a 1.5-T MRI scanner with an inherently low signal-to-noise ratio was used; therefore, the results of the MCP sT1W/T2W ratio may be biased by noise. However, the MCP sT1W/T2W ratio values in patients with MSA-C correlated well with the extent of MCP hyperintensities and MCP volume, suggesting that the MCP sT1W/T2W ratio values measured in this study well reflected the degeneration of the MCP in MSA-C. Moreover, the diagnostic accuracy of the MCP sT1W/T2W ratio in discriminating MSA-C from controls in this study was comparable to that observed in a study using a 3-T MRI scanner [16]. Further studies employing 3D T1WI and T2WI sequences using a 3-T MRI scanner, which might allow better quantification of the sT1W/T2W ratio, are needed to confirm the results of this study.” #2 The authors haven’t clearly mentioned what they are measuring with T1w/T2w and how sensitive the method is to tissue microstructure especially the myelin. T1w/T2w method is considered to be sensitive to myelin but depends on how T1w and T2w images are acquired as well as the image processing pipeline. For better quantification of T1w/T2w, 3D T1w and 3D T2w SPACE sequences are recommended. A recent work by Uddin et al (NMR in Biomedicine 2018) reported that TSE based T1w/T2w ratio is not a good measure to quantify myelin. Reply: We want to thank the reviewer for this insightful comment. According to this reviewer’s comment, we have added sentences in the Discussion section, as follows. Page 19, line 294-308 “Decreased MCP sT1W/T2W ratio values in patients with MSA-C might reflect demyelination and gliosis in the MCP due to MSA-C-related degeneration. Pathologically, the loss of axons, myelin, and reactive gliosis was observed in the MCP of patients with MSA-C [28]. The sT1W/T2W ratio is sensitive to myelin and decreases in demyelinating lesions [29, 30]. In contrast, inconsistent correlations with histology [31], myelin water imaging [12, 32], simultaneous tissue relaxometry, and magnetization transfer saturation index [33] suggest that the sT1W/T2W ratio is sensitive to not only myelin content but also other microstructural factors. Moreover, gliosis is presumed to affect the sT1W/T2W ratio in a downward direction, mainly by prolonged T1 relaxation time [34]. Therefore, although the sT1W/T2W ratio might not be specific for myelin content, it might be more sensitive than evaluating MCP hyperintensities on T2WI to detect degenerative processes in the MCP, such as demyelination and gliosis associated with MSA-C. In this study, the MCP sT1W/T2W ratio had better diagnostic accuracy in discriminating MSA-C from SCA3 than MCP volume. This result is consistent with those of previous pathological studies, showing that gliosis and demyelination precede atrophy in MSA [35, 36].” In line with this revision, we have added references. 28. Itoh K, Kasai T, Tsuji Y, Saito K, Mizuta I, Harada Y, et al. Definite familial multiple system atrophy with unknown genetics. Neuropathology. 2014;34:309-13. 29. Glasser MF, Goyal MS, Preuss TM, Raichle ME, Van Essen DC. Trends and properties of human cerebral cortex: correlations with cortical myelin content. NeuroImage. 2014;93:165-75. 30. Nakamura K, Chen JT, Ontaneda D, Fox RJ, Trapp BD. T1-/T2-weighted ratio differs in demyelinated cortex in multiple sclerosis. Ann Neurol. 2017;82:635-9. 31. Righart R, Biberacher V, Jonkman LE, Klaver R, Schmidt P, Buck D, et al. Cortical pathology in multiple sclerosis detected by the T1/T2-weighted ratio from routine magnetic resonance imaging. Ann Neurol 2017;82:519-29. 32. Uddin MN, Figley TD, Marrie RA, Figley CR; For the CCOMS Study Group. Can T1w/T2w ratio be used as a myelin-specific measure in subcortical structures? Comparisons between FSE-based T1w/T2w ratios, GRASE-based T1w/T2w ratios and multi-echo GRASE-based myelin water fractions. NMR Biomed. 2018;31:e3868. 33. Hagiwara A, Hori M, Kamagata K, Warntjes M, Matsuyoshi D, Nakazawa M, et al. Myelin measurement: comparison between simultaneous tissue relaxometry, magnetization transfer saturation index, and T1w/T2w ratio methods. Sci Rep. 2018;8:10554. 34. Barnes D, McDonald WI, Landon DN, Johnson G. The characterization of experimental gliosis by quantitative nuclear magnetic resonance imaging. Brain. 1988;111:83-94. 35. Jellinger KA, Seppi K, Wenning GK. Grading of neuropathology in multiple system atrophy: proposal for a novel scale. Mov Disord. 2005;20:29-36. 36. Kon T, Mori F, Tanji K, Miki Y, Wakabayashi K. An autopsy case of preclinical multiple system atrophy (MSA-C). Neuropathology. 2013;33:667-72. #3 The authors mentioned about T2 hyperintensities in MCP. I was wondering if they have any scores for WM hyperintensities or disease severity scores that can be correlated with T1w/T2w. Reply: We agree with the reviewer’s point that assessing the correlation between MCP sT1W/T2W ratio values and scores for MCP hyperintensities or disease severity scores, such as SARA, is important. We have assessed the MCP hyperintensities semiquantitatively and evaluated the correlation between the MCP sT1W/T2W ratio and MCP hyperintensities. In line with this revision, we have added sentences in the Methods and Results sections. Unfortunately, no systematic assessment of cerebellar symptoms, including SARA, has been performed in the patients with MSA included in this study. We have added a note to the Discussion section about this as a limitation. Page 7, line 159-161 “A board-certified neuroradiologist (H. Y.), who was blinded to the clinical data, semiquantitatively evaluated the extent of MCP hyperintensities in the patients with MSA by applying the following scores: 0 = absent, 1 = mild, 2 = moderate, and 3 = severe.” Page 9, line 208-210 “Spearman correlation analysis was performed to evaluate the relationship between the extent of MCP hyperintensities assessed semiquantitatively and the MCP sT1W/T2W ratio values in patients with MSA-C.” Page 12, line 256-258 “In patients with MSA-C, a significant inverse correlation between the extent of MCP hyperintensities and MCP sT1W/T2W ratio was observed (r = −0.745; p < 0.001).” Page 21, line 348-351 “Third, owing to the retrospective nature of this study, the relationship between the MCP sT1W/T2W ratio and rapid eye movement sleep behavior disorder or clinical scales for cerebellar ataxia, such as the International Cooperative Ataxia Rating Scale or Scale for the Assessment and Rating of Ataxia, could not be evaluated.” #4 MRI acquisition: data were acquired with a 1.5 T scanner. Please add information about the RF coil (i.e, 32-channel head coil), number of averages, number of slices. Reply: According to the reviewer’s comments, we have revised sentences in the Discussion section, as follows. Page 5, line 114-120 “All MR images were acquired using a 1.5-T MRI system (Genesis Signa; GE Healthcare) with a 12-channel head coil. The MRI parameters for T1WI were as follows: 3D-IR-SPGR; sagittal plane; TR, 8 ms; TE, 3 ms; flip angle, 15°; FOV, 220 × 220 mm; matrix, 256 × 256; voxel size, 0.86 × 0.86 × 1.5 mm; number of slices, 248; and number of averages, 1. Meanwhile, those for T2WI were as follows: 2D-TSE; TR, 4000 ms; TE, 106 ms; FOV, 220 × 220 mm; matrix, 320 × 256; voxel size, 0.43 × 0.43 × 6 mm; interslice gap, 1.5 mm; number of slices 24; and number of averages 1.” #5 MRI preprocessing: “3D T1w images were linearly co-registered with 2D axial T2w images using SPM12. Brain masks were created by skull-stripping the co-registered T1w images using the BrainExtraction Tool with FSL (version 5.0.11) [20] and binarizing it with FSLmaths.”I was wondering why did the authors use SPM for coregistration only. It's better to use FSL or SPM, sometime the header information is not properly copied from one platform to other. However, it's possible to copy the headers from spm to fsl properly. Reply: We originally planned to run the same pipeline as described by Cooper G, et al. (Front Neurol 2019;10:334). However, when testing the pipeline with our available data, we found that FSL did not coregister 3D and 2D images together very accurately, whereas SPM12 was able to do this very successfully. #6 T1w/T2w calculation: The authors mentioned that for T1w/T2w quantification they used previously published technique by Misaki et al MRM 2014, but it is not clear how did they perform bias correction which is important for better quantification of T1w/T2w. Did they use their script or just follow their approach (it looks there is no publicly available tool online by Misaki et al)? Depending on the bias correction technique, T1w/T2w maps may contain residual radio frequency transmit field (B1+) biases, which may be correlated with these variables of interest, leading to potentially spurious results. Reply: We thank the reviewer for pointing out that bias correction is missing our pipeline. We are aware of the growing literature demonstrating the importance of B1+ correction for T1W/T2W at 3 T and 7 T, where B1+ is more inhomogeneous. However, this study used a 1.5-T scanner, where B1+ is more homogenous and the effect could be assumed to be smaller (see, for example, Figure 3 of Dieringer et al. 2014 PLOS ONE, https://journals.plos.org/plosone/article?id=10.1371/journal.pone.0091318. We have added a statement indicating a limitation that bias correction is missing in our pipeline, as follows. Page 21, line 355-359 “Finally, bias correction was missing in our pipeline of sT1W/T2W calculation. Bias correction is important for the sT1W/T2W ratio on high-magnetic-field MRI to avoid the effects of the inhomogeneity of the radiofrequency transmit (B1+) field. However, this study used a 1.5-T scanner, where B1+ is more homogenous and the effect could be assumed to be smaller [49].” In line with this revision, we have added a reference. 49. Dieringer MA, Deimling M, Santoro D, Wuerfel J, Madai VI, Sobesky J, et al. Rapid parametric mapping of the longitudinal relaxation time T1 using two-dimensional variable flip angle magnetic resonance imaging at 1.5 Tesla, 3 Tesla, and 7 Tesla. PLoS One. 2014;9:e91318. #7 To measure diagnostic accuracy by Youden ROC method, I suspect about the accuracy of the measurement with this small sample size. Reply: According to the reviewer’s comment, we have changed the method for determining the cutoff point for ROC analysis from the Youden method to the point on the ROC curve closest to the upper left corner. Moreover, we have added the small sample size as a limitation. With these changes, we have revised sentences in the Methods section and added sentences in the Results and Discussion sections, as follows. Page 8, line 186-192 “All statistical analyses, except for receiver operating characteristic (ROC) curve analyses, were performed using Statistical Package for the Social Sciences (version 25.0; SPSS Inc., IBM Corp., Chicago, IL, USA). The ROC curve analyses and Smirnov–Grubb’s test were performed using EZR (Saitama Medical Center, Jichi Medical University, Saitama, Japan), which is a graphical user interface for R (The R Foundation for Statistical Computing, Vienna, Austria, version 2.13.0) [26]. More precisely, EZR is a modified version of R commander (version 1.6-3) designed to add statistical functions frequently used in biostatistics.” Page 9, line 197-202 “For evaluating differences in MCP sT1W/T2W ratio values and MCP volume among the four groups, a one-way analysis of covariance was used using age as the covariate. The ability of the MCP sT1W/T2W ratio values and MCP volume to differentiate between groups of patients was assessed using ROC curve analyses, and the optimal cutoff point was determined as the point on the ROC curve closest to the upper left corner, as used by Holmes [27].” Page 11, line 221-223 “The AUC for differentiating MSA-C from SCA-3 was 0.895 (95% confidence interval [CI], 0.794–0.995), with a sensitivity of 0.844 and specificity of 1.0.” Page 11, line 224-227 “The AUC for differentiating MSA-C from controls was 0.960 (95% CI, 0.908–1.000), with a sensitivity of 0.906 and specificity of 0.941. The AUC for differentiating SCA-3 from SCA-6 was 0.898 (95% CI, 0.775–1.000), with a sensitivity of 0.875 and specificity of 0.812 (Fig. 3).” Page 20, line 336 “This study has several limitations.” Page 21, line 352-355 “Fourth, the sample size of the SCA3 group was small (n = 8), and future prospective studies with larger sample sizes that evaluate clinical scales for cerebellar ataxia are needed to validate the findings of this study and clarify the relationship between clinical severity of cerebellar ataxia and the MCP sT1W/T2W ratio.” In line with these revisions, we have revised Figure 3 and added references. 26. Kanda Y. Investigation of the freely available easy-to-use software ‘EZR’ for medical statistics. Bone Marrow Transplant 2013;48:452-8. 27. Holmes W. A short, psychiatric, case-finding measure for HIV seropositive outpatients. Medical Care 1998;36:237-43. #8 Figure1: you haven’t mentioned what type of images (i.e., T2-weighted images) are in the figure. Please include the corresponding T1w/T2w maps with the signal intensity scale. Reply: According to the reviewer’s comments, we have revised Figure 1 and added a sentence in the Figure Legends section, as follows. Page 7, line 171 to Page 8, line 172 “Upper row: axial T2-weighted images. Lower row: standardized T1-weighted/T2-weighted ratio maps” #9 Figure2: significant difference between MSA-C with SCA3 and SCA6 might be derived from the outliers of MSA-C data. I would use the outlier removal tool and compare the results. Reply: We thank the reviewer for the comment. We have added Smirnov–Grubb’s test for evaluating the outliers. Moreover, we have revised sentences in the Methods section and added a sentence in the Results section, as follows. Page 8, line 188-191 “The ROC curve analyses and Smirnov–Grubb’s test were performed using EZR (Saitama Medical Center, Jichi Medical University, Saitama, Japan), which is a graphical user interface for R (The R Foundation for Statistical Computing, Vienna, Austria, version 2.13.0) [26].” Page 9, line 212-213 “The Smirnov–Grubb’s test was used for evaluating the outliers of the MCP sT1W/T2W ratio value in each subject group.” Page 11, line 227-231 “Even after excluding one patient with MSA-C with an outlier MCP sT1W/T2W ratio value, the MCP sT1W/T2W ratio values in patients with MSA-C were significantly lower than those in patients with SCA3 and SCA6 and controls (0.03 ± 0.06 vs. 0.13 ± 0.02, 0.17 ± 0.03, and 0.17 ± 0.03, respectively; p < 0.001).” #10 Discussion section should be more organized. Reply: According to the reviewer’s suggestion, we have revised sentences in the Discussion section, as follows. Page 18, line 286-288 “We showed that the MCP sT1W/T2W ratio is highly sensitive and specific for not only distinguishing MSA-C from SCA3, SCA6, or controls but also differentiating SCA3 from SCA6.” Page 19, line 309-310 “The usefulness of the MCP sT1W/T2W ratio in discriminating MSA-C from SCA3 in this study might be due to differences in the degree of MCP degeneration in the two diseases.” Page 20, line 324-326 “The pathological background for this may be the difference in MCP degeneration, which is observed to some extent in SCA3, as described above, but is rarely observed in SCA6.” #11 It would be great to show the comparison of volumetric measure of the MCP indifferent disease conditions. Reply: We thank the reviewer for this important suggestion. The volume of the MCP was measured and compared between the groups. Moreover, we evaluated the correlation between MCP volume and MCP sT1w/T2w ratio value and compared the diagnostic accuracy between MCP volume and MCP sT1w/T2w ratio. Along with this revision, we have added sentences in the Methods and Results sections, as follows. Page 8, line 174-183 “MCP volume analysis The MCP volume was measured in all subjects by exploiting a validated probabilistic 3D atlas of the cerebellar white matter structure [24]. Parcellation at a 90% probability threshold was used. The atlas of the MCP was transformed from the Montreal Neurological Institute 152 space [20] to the individual subject’s space using the deformation parameters obtained from the normalization procedure using ANTs [23]. The mean volume value in the left and right MCPs was used as the MCP volume for each subject. For further analyses, the MCP volume was normalized by each subject’s intracranial volume. Values represent the ratio of the volume of the MCP divided by the subject-specific total intracranial volume (written as m × 10-3).” Page 9, line 197-202 “For evaluating differences in MCP sT1W/T2W ratio values and MCP volume among the four groups, a one-way analysis of covariance was used using age as the covariate. The ability of the MCP sT1W/T2W ratio values and MCP volume to differentiate between groups of patients was assessed using ROC curve analyses, and the optimal cutoff point was determined as the point on the ROC curve closest to the upper left corner, as used by Holmes [27].” Page 9, line 202-204 “The DeLong test was used to compare the area under curves (AUCs) between the MCP sT1W/T2W ratio and MCP volume to differentiate among the four groups.” Page 9, line 210-212 “Pearson correlation analysis was performed to evaluate the relationship between the MCP sT1W/T2W ratio values and MCP volume in patients with MSA-C.” Page 18, line 265-283 “The MCP volume, normalized by each subject’s intracranial volume, was significantly lower in patients with MSA-C patients than those in patients with SCA6 and controls (0.20 ± 0.06 vs. 0.31 ± 0.05 and 0.34 ± 0.05, respectively; p < 0.001). Moreover, the MCP volume was significantly lower in patients with SCA3 than those in patients with SCA6 and controls (0.23 ± 0.04 vs. 0.31 ± 0.05 and 0.34 ± 0.05, respectively; p < 0.001; p = 0.01 and p < 0.001). The AUC value for differentiating MSA-C from SCA3 was 0.664 (95% CI, 0.486–0.842), with a sensitivity of 0.750 and specificity of 0.625. The AUC for differentiating MSA-C from SCA6 was 0.924 (95% CI, 0.852–0.995), with a sensitivity of 0.844 and specificity of 0.875. The AUC for differentiating MSA-C from controls was 0.967 (95% CI, 0.925–1.000), with a sensitivity of 0.906 and specificity of 0.941). The AUC for differentiating SCA3 from SCA6 was 0.914 (95% CI, 0.802–1.000), with a sensitivity of 0.875 and specificity of 0.875. The AUC of the MCP sT1W/T2W ratio was greater than that of the MCP volume for differentiating MSA-C from SCA3 (p = 0.0049). No significant differences were observed between the AUC of the MCP sT1W/T2W ratio and that of MCP volume for differentiating MSA-C from SCA6 and controls (p = 0.304 and 0.781, respectively). Moreover, no significant difference was observed between the AUC of the MCP sT1W/T2W ratio and that of MCP volume for differentiating SCA3 from SCA6 (p = 0.855). In patients with MSA-C, a significant correlation was observed between the MCP volume and MCP sT1W/T2W ratio (r = 0.732; p < 0.001).” #12 Please add the small sample size as a limitation. Reply: We agree with the reviewer’s concerns about the small sample sizes in our study. We have revised and added sentences in the Discussion section, as follows. Page 20, line 336 “This study has several limitations.” Page 21, line 352-355 “Fourth, the sample size of the SCA3 group was small (n = 8), and future prospective studies with larger sample sizes that evaluate clinical scales for cerebellar ataxia are needed to validate the findings of this study and clarify the relationship between clinical severity of cerebellar ataxia and the MCP sT1W/T2W ratio.” #13 “In conclusion, the MCP sT1w/T2w ratio is a highly sensitive imaging-based marker for detecting MSA-C related changes and might be useful for differentiating 274 MSA-C from SCA3 or SCA6.” With a relatively small sample size and poor quality of data (especially 2D T1w,T2w data, 1.5 mm vs. 6mm slice thickness collected at 1.5 T), it is tough to claim “highly sensitive imaging-based marker ….” Reply: We thank the reviewer for the comment. Accordingly, we have revised sentences in the Abstract and Conclusions sections, as follows. Page 3, line 50-51 “The MCP sT1W/T2W ratio might be a sensitive imaging-based marker for detecting MSA-C-related changes and differentiating MSA-C from SCA3 or SCA6.” Page 21, line 362 to Page 22, line 363 “In conclusion, the MCP sT1W/T2W ratio might be a sensitive imaging-based marker for detecting MSA-C-related changes and useful for differentiating MSA-C from SCA3 or SCA6.” Submitted filename: Response to Reviewers.docx Click here for additional data file. 18 Nov 2021
PONE-D-21-24426R1
Diagnostic efficacy of the magnetic resonance T1w/T2w ratio for the middle cerebellar peduncle in multiple system atrophy and spinocerebellar ataxia
PLOS ONE Dear Dr. Sugiyama, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. While one of the Reviewers feels that your manuscript is ready to be accepted, the other Reviewer still has major concerns that I invite you to address. While the sample size limitation can be acknowledged in the manuscript, I strongly recommend that you follow the Reviewer's advice to implement bias field correction in the analysis. Please submit your revised manuscript by Jan 02 2022 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript:
A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'. A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'. An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'. If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols. We look forward to receiving your revised manuscript. Kind regards, Niels Bergsland Academic Editor PLOS ONE [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #1: All comments have been addressed Reviewer #2: (No Response) ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: Partly ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes Reviewer #2: Yes ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: The revised manuscript is now OK for publication. I have no additional comments or suggestions fo rthe authors ! Reviewer #2: The authors have partly responded to my questions and made changes to the manuscript. I would thank the authors to add healthy controls. But the major concern is the data used and sample size. Therefore, it can be reported as preliminary findings and further investigation with a large sample size and standard T1 and T2 images is required to validate these findings. I have some comments: 1. The authors followed the paper by Misaki et al 2015 but they didn’t perform bias correction. As I mentioned before that bias correction is an important step for T1w/T2w ratio mapping and depending on the bias correction technique, T1w/T2w maps may contain residual radiofrequency B1+ biases, which may be correlated with these variables of interest, leading to potentially spurious results. It's true that B1+ is more homogeneous at 1.5T compared to higher magnetic field strengths such as 3T or 7T (say center brightness artifact is ~30% at 3T while ~1.5% at 1.5T; Bernstein et al 2006), but not perfectly homogeneous and it’s still required to perform bias correction for T1w/T2w mapping even at 1.5 T MRI (see previous works at 1.5 T; Ganzetti et al 2016). 2. It would be great to add a figure for a pipeline for creating a T1w/T2w image (i.e., summarize key steps with representative images in native and standard spaces). 3. Figure 1: Why did the authors use images at different slice locations in Figure 1? I would recommend using the same slice locations for all cases in Figure 1. Also, show the corresponding images for T1w/T2w maps with the same image size. 4. Abstract: “Methods” section was not updated for healthy controls. ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: Yes: Md Nasir Uddin [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. 24 Mar 2022 Reply to Reviewer 1 We greatly appreciate your helpful comments and suggestions. Changes made in accordance with reviewers’ comments are indicated in red font in the revised manuscript. #1 The authors have partly responded to my questions and made changes to the manuscript. I would thank the authors to add healthy controls. But the major concern is the data used and sample size. Therefore, it can be reported as preliminary findings and further investigation with a large sample size and standard T1 and T2 images is required to validate these findings. I have some comments: Reply: We thank the Reviewer for their important comment. Accordingly, we have revised the title. Page1, line 1-3 “Diagnostic efficacy of the magnetic resonance T1w/T2w ratio for the middle cerebellar peduncle in multiple system atrophy and spinocerebellar ataxia: a preliminary study” #2 The authors followed the paper by Misaki et al 2015 but they didn’t perform bias correction. As I mentioned before that bias correction is an important step for T1w/T2w ratio mapping and depending on the bias correction technique, T1w/T2w maps may contain residual radiofrequency B1+ biases, which may be correlated with these variables of interest, leading to potentially spurious results. It's true that B1+ is more homogeneous at 1.5T compared to higher magnetic field strengths such as 3T or 7T (say center brightness artifact is ~30% at 3T while ~1.5% at 1.5T; Bernstein et al 2006), but not perfectly homogeneous and it’s still required to perform bias correction for T1w/T2w mapping even at 1.5 T MRI (see previous works at 1.5 T; Ganzetti et al 2016). Reply: We followed the reviewer's suggestion and added the bias collection step, which partially changed the results of the study. We strongly agree with the reviewer's comments on the importance of bias collection step and would like to express our deepest gratitude to the reviewer for suggesting the addition of bias collection. Accordingly, we have revised and added sentences in the Abstract, Methods, Results, and Discussion sections. Page 2, line 41-48 “The MCP sT1w/T2w ratio showed high diagnostic accuracy for distinguishing MSA-C from SCA3 (area under curve = 0.934), SCA6 (area under curve = 0.965), and controls (area under curve = 0.980). The diagnostic accuracy of the MCP sT1w/T2w ratio for differentiating MSA-C from SCA3 or SCA6 (90.0% for MSA-C vs. SCA3, and 91.7% for MSA-C vs. SCA6) was comparable to or superior than that of visual interpretation of MCP hyperintensities (80.0–87.5% in MSA-C vs. SCA3 and 87.6–97.9% in MSA-C vs. SCA6) or the HCB sign (72.5–80.0% in MSA-C vs. SCA3 and 77.1–93.8% in MSA-C vs. SCA6).” Page 6, line 123-125 “All T1w and T2w images were preprocessed as previously described before the sT1w/T2w ratio was calculated [16]; specifically, the intensity inhomogeneity correction was applied to the T1w and T2w images using N4BiasFieldCorrection [20].” Page 12, line 226-238 “The MCP sT1w/T2w ratio values were significantly lower in patients with MSA-C than those in patients with SCA3 and SCA6 and controls (0.07 ± 0.06 vs. 0.17 ± 0.03, 0.18 ± 0.03, and 0.19 ± 0.03, respectively; p < 0.001) (Fig. 3). The AUC for differentiating MSA-C from SCA-3 was 0.934 (95% confidence interval [CI], 0.854–1.000), with a sensitivity of 0.906 and specificity of 0.875. The AUC value for differentiating between MSA-C and SCA-6 was 0.965 (95% CI, 0.922–1.000), with a sensitivity of 0.906 and a specificity of 0.938. The AUC for differentiating MSA-C from controls was 0.980 (95% CI, 0.951–1.000), with a sensitivity of 0.906 and specificity of 0.941. The AUC for differentiating SCA-3 from SCA-6 was 0.633 (95% CI, 0.388–0.878), with a sensitivity of 0.750 and specificity of 0.625 (Fig. 4). Even after excluding one patient with MSA-C with an outlier MCP sT1w/T2w ratio value, the MCP sT1W/T2W ratio values in patients with MSA-C were significantly lower than those in patients with SCA3 and SCA6 and controls (0.07 ± 0.05 vs. 0.17 ± 0.03, 0.18 ± 0.03, and 0.19 ± 0.03, respectively; p < 0.001).” Page 13, line 263-265 “In patients with MSA-C, a significant inverse correlation between the extent of MCP hyperintensities and MCP sT1w/T2w ratio was observed (r = −0.832; p < 0.001).” Page 16, line 267-274 “Diagnostic accuracy of the MCP sT1w/T2w ratio, visual interpretation of the MCP hyperintensities, and the HCB sign are listed in Table 3, and we found that the diagnostic performance of the MCP sT1w/T2w ratio for differentiating MSA-C from SCA3, SCA6, and controls was comparable to or better than that of visual interpretation of MCP hyperintensities and the HCB sign for all three reviewers. However, the diagnostic performance of visual interpretation of MCP hyperintensities and the HCB sign for differentiating SCA3 from SCA6 was comparable to or better than that of the MCP sT1w/T2w ratio.” Page 19, line 286-292 “The AUC of the MCP sT1w/T2w ratio was greater than that of the MCP volume for differentiating MSA-C from SCA3 (p < 0.001). No significant differences were observed between the AUC of the MCP sT1w/T2w ratio and that of MCP volume for differentiating MSA-C from SCA6 and controls (p = 0.234 and 0.577, respectively). The AUC of the MCP volume was greater than that of the MCP sT1w/T2w ratio for differentiating SCA3 from SCA6 (p = 0.020).” Page 19, line 292-293 “In patients with MSA-C, a significant correlation was observed between the MCP volume and MCP sT1W/T2W ratio (r = 0.619; p < 0.001).” Page 19, line 296-297 “We showed that the MCP sT1W/T2W ratio is highly sensitive and specific for distinguishing MSA-C from SCA3, SCA6, or controls.” Page 21, line 332-336 “The substantial discriminatory power of the MCP sT1w/T2w ratio and visual interpretation of MCP hyperintensities and the HCB sign described here implies that they may also be useful for differentiating between SCA3 and SCA6. The pathological background for this may be the difference in MCP degeneration, which is observed to some extent in SCA3, as described above, but is rarely observed in SCA6.” Page 21, line 341-345 “The cohort of SCA3 and SCA6 patients included in this study is not large, and therefore, further studies that include data from a large cohort of SCA3 and SCA6 patients are needed to verify the usefulness of MCP sT1w/T2w ratio values and visual interpretation of MCP hyperintensities and the HCB sign in differentiating between SCA3 and SCA6.” Page 22, line 362-365 “Finally, the sample size of the SCA3 group was small (n = 8), and future prospective studies with larger sample sizes that evaluate clinical scales for cerebellar ataxia are needed to validate the findings of this study and clarify the relationship between clinical severity of cerebellar ataxia and the MCP sT1W/T2W ratio.” According to this revision, we have revised Figure 3 and Figure 4. According to this revision, we have revised S1 Table. According to this revision, we have added a reference. 20. Tustison NJ, Avants BB, Cook PA, Zheng Y, Egan A, Yushkevich PA, et al. N4ITK: improved N3 bias correction. IEEE Trans Med Imaging. 2010;29:1310-20. #3 It would be great to add a figure for a pipeline for creating a T1w/T2w image (i.e., summarize key steps with representative images in native and standard spaces). Reply: We thank the Reviewer for their important comment. Accordingly, we have added a pipeline for creating a sT1w/T2w ratio map as Figure 1. Page 6, line 140-141 “Schematic representation of the pipeline for creating a sT1w/T2w ratio map is described in Figure 1.” Page 7, line 152-154 “Figure 1. Schematic representation of the pipeline for creating a sT1w/T2w ratio map. T1w, T1-weighted; T2w, T2-weighted; GM, gray matter: WM, white matter; sT2, a scaled T2-weighted image” #4 Figure 1: Why did the authors use images at different slice locations in Figure 1? I would recommend using the same slice locations for all cases in Figure 1. Also, show the corresponding images for T1w/T2w maps with the same image size. Reply: We thank the Reviewer for their important comment. Accordingly, we have revised Figure 2 and a sentence in Figure legend. Page 8, line 173-175 “(A) Grade 0 (negative) HCB sign and absent MCP hyperintensities in a 74-year-old man with a SCA6 disease duration of 2.3 years.” #5 Abstract: “Methods” section was not updated for healthy controls. Reply: We thank the Reviewer for their careful observation and apologize for the oversight. The sentence has been revised as follows: Page 2, line 34-35 “We included 32 MSA-C, 8 SCA type 3 (SCA3), 16 SCA type 6 (SCA6) patients, and 17 controls, and the MCP sT1w/T2w ratio was analyzed using a region-of-interest approach.” Submitted filename: Response to Reviewers.docx Click here for additional data file. 1 Apr 2022 Diagnostic efficacy of the magnetic resonance T1w/T2w ratio for the middle cerebellar peduncle in multiple system atrophy and spinocerebellar ataxia: a preliminary study PONE-D-21-24426R2 Dear Dr. Sugiyama, We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements. Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication. An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org. If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. Kind regards, Niels Bergsland Academic Editor PLOS ONE Additional Editor Comments (optional): Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #2: All comments have been addressed ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #2: Yes ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #2: Yes ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #2: (No Response) ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #2: (No Response) ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #2: The authors have sufficiently improved the manuscript, and I think the manuscript is now ready for publication. ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #2: Yes: Md Nasir Uddin 7 Apr 2022 PONE-D-21-24426R2 Diagnostic efficacy of the magnetic resonance T1w/T2w ratio for the middle cerebellar peduncle in multiple system atrophy and spinocerebellar ataxia: a preliminary study Dear Dr. Sugiyama: I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department. If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org. If we can help with anything else, please email us at plosone@plos.org. Thank you for submitting your work to PLOS ONE and supporting open access. Kind regards, PLOS ONE Editorial Office Staff on behalf of Dr. Niels Bergsland Academic Editor PLOS ONE
  49 in total

Review 1.  Advances in functional and structural MR image analysis and implementation as FSL.

Authors:  Stephen M Smith; Mark Jenkinson; Mark W Woolrich; Christian F Beckmann; Timothy E J Behrens; Heidi Johansen-Berg; Peter R Bannister; Marilena De Luca; Ivana Drobnjak; David E Flitney; Rami K Niazy; James Saunders; John Vickers; Yongyue Zhang; Nicola De Stefano; J Michael Brady; Paul M Matthews
Journal:  Neuroimage       Date:  2004       Impact factor: 6.556

Review 2.  Clinical features, neurogenetics and neuropathology of the polyglutamine spinocerebellar ataxias type 1, 2, 3, 6 and 7.

Authors:  Udo Rüb; Ludger Schöls; Henry Paulson; Georg Auburger; Pawel Kermer; Joanna C Jen; Kay Seidel; Horst-Werner Korf; Thomas Deller
Journal:  Prog Neurobiol       Date:  2013-02-21       Impact factor: 11.685

3.  Grading of neuropathology in multiple system atrophy: proposal for a novel scale.

Authors:  Kurt A Jellinger; Klaus Seppi; Gregor K Wenning
Journal:  Mov Disord       Date:  2005-08       Impact factor: 10.338

4.  Magnetic resonance imaging findings of Machado-Joseph disease: histopathologic correlation.

Authors:  Aya M Tokumaru; Keiko Kamakura; Toshiyuki Maki; Shigeo Murayama; Ikuko Sakata; Tatsumi Kaji; Shinnya Kohyama; Shoichi Kusano; Seiji Hasegawa
Journal:  J Comput Assist Tomogr       Date:  2003 Mar-Apr       Impact factor: 1.826

5.  N4ITK: improved N3 bias correction.

Authors:  Nicholas J Tustison; Brian B Avants; Philip A Cook; Yuanjie Zheng; Alexander Egan; Paul A Yushkevich; James C Gee
Journal:  IEEE Trans Med Imaging       Date:  2010-04-08       Impact factor: 10.048

Review 6.  Multiple system atrophy: an update.

Authors:  Nadia Stefanova; Philipp Bücke; Susanne Duerr; Gregor Karl Wenning
Journal:  Lancet Neurol       Date:  2009-12       Impact factor: 44.182

7.  Symmetric diffeomorphic image registration with cross-correlation: evaluating automated labeling of elderly and neurodegenerative brain.

Authors:  B B Avants; C L Epstein; M Grossman; J C Gee
Journal:  Med Image Anal       Date:  2007-06-23       Impact factor: 8.545

8.  Olivopontocerebellar atrophy: MR diagnosis and relationship to multisystem atrophy.

Authors:  M Savoiardo; L Strada; F Girotti; R A Zimmerman; M Grisoli; D Testa; R Petrillo
Journal:  Radiology       Date:  1990-03       Impact factor: 11.105

9.  Vertical pons hyperintensity and hot cross bun sign in cerebellar-type multiple system atrophy and spinocerebellar ataxia type 3.

Authors:  Atsuhiko Sugiyama; Hajime Yokota; Yoshitaka Yamanaka; Hiroki Mukai; Tatsuya Yamamoto; Shigeki Hirano; Kyosuke Koide; Shoichi Ito; Satoshi Kuwabara
Journal:  BMC Neurol       Date:  2020-04-27       Impact factor: 2.474

10.  Longitudinal analysis of T1w/T2w ratio in patients with multiple sclerosis from first clinical presentation.

Authors:  Graham Cooper; Claudia Chien; Hanna Zimmermann; Judith Bellmann-Strobl; Klemens Ruprecht; Joseph Kuchling; Susanna Asseyer; Alexander U Brandt; Michael Scheel; Carsten Finke; Friedemann Paul
Journal:  Mult Scler       Date:  2021-04-15       Impact factor: 6.312

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.