| Literature DB >> 35418785 |
Tomoto Kasahara1, Hiroshi Toshida1, Kohei Ichikawa1, Yusuke Matsuzaki1, Junji Ono2.
Abstract
YOUSOFT is a new soft contact lens (CL) specifically designed for the correction of astigmatism. It is a conventional FDA Group II soft contact lens (SCL) and the optical zone in its centre is thick, and on its rear surface it has a prism ballast structure to prevent rotation. This lens was prescribed for a 59-year-old male on the eye with irregular corneal astigmatism after receiving penetrating keratoplasty (PKP) for refractive correction, whose eye was difficult to wear hard CLs (HCLs). After following the directions in the fitting manual provided by the manufacturer, fitting was good in the right eye and corrected visual acuity was 20/25 in Snellen Chart. He could continue wearing without CL discomfortable YOUSOFT may be useful in patients who have received PKP.Entities:
Keywords: irregular astigmatism; keratoconus; penetrating keratoplasty; soft contact lens
Year: 2022 PMID: 35418785 PMCID: PMC8999709 DOI: 10.2147/IMCRJ.S362072
Source DB: PubMed Journal: Int Med Case Rep J ISSN: 1179-142X
Characteristics of YOUFOFT
| Base Curve | 7.80 to 8.40 (0.20 Steps) |
|---|---|
| Diameter | 14.50mm |
| Power range | Sphere: +30.00 D to −30.00D (in 0.25 steps) |
| Cylinder: −0.25 to −6.00 D (in 0.25 steps) | |
| Axis | 5°- 180° (in 5° steps) |
| Thickness in the centre of lens | 0.4mm |
| FDA group | Group 2, 80% Water |
| Dk | 44 x 10−11 (cm2/sec) [ml02/ (mL x mmHg)] |
| Materials | N, N-Dimethylacrylamide, N-vinyl-pyrrolidone, methyl methacrylate, Alkyl methacrylate-based compounds, Ethylene glycol dimethacrylate |
Figure 1(A) Figure 1: Scheme of YOUSOFT. The circle in the middle is the optical zone. Prizm ballast design (asterisk) is adopted in the lower area of the lens. (B) Corneal videokeratograph after cataract surgery in the right eye. Irregularities are shown. (C) Ocular images in the right eye wearing YOUSOFT. The fitting parameters of the base curve, total diameter, spherical and cylinder power, and axis were 7.80 mm, 14.5 mm, −2.00 D, −0.25D, 180° for YOUSOFT, respectively. (D) Vertical Meridian image of YOUSOFT on the cornea obtained by anterior segment optical coherence tomography. Thicker part of the lens on the lower area of the cornea (Arrow).