Jingyuan Li1, Weishi Lin2, Pibo Du3, Wei Liu2, Xiong Liu2, Chaojie Yang2, Ruizhong Jia2, Yong Wang2, Yong Chen2, Leili Jia2, Li Han2, Weilong Tan4, Nan Liu3, Junjie Du5, Yuehua Ke6, Changjun Wang7.
Abstract
Accurate detection of severe acute respiratory syndrome coronavirus 2 is not only necessary for viral load monitoring to optimize treatment in hospitalized coronavirus disease 2019 patients, but also critical for deciding whether the patient could be discharged without any risk of viral shedding. Digital droplet PCR (ddPCR) is more sensitive than reverse transcription quantitative real-time polymerase chain reaction (RT-qPCR) and is usually considered the superior choice. In the current study, we compared the clinical performance of RT-qPCR and ddPCR using oropharyngeal swab samples from patients hospitalized in the temporary Huoshenshan Hospital, Wuhan, Hubei, China. Results demonstrated that ddPCR was indeed more sensitive than RT-qPCR. Negative results might be caused by poor sampling technique or recovered patients, as the range of viral load in these patients varied significantly. In addition, both methods were highly correlated in terms of their ability to detect all three target genes as well as the ratio of copies of viral genes to that of the IC gene. Furthermore, our results evidenced that both methods detected the N gene more easily than the ORF gene. Taken together, these findings imply that the use of ddPCR, as an alternative to RT-qPCR, is necessary for the accurate diagnosis of hospitalized coronavirus disease 2019 patients.Entities:
Keywords: COVID-19; Clinical performance; Quantitative detection; RT-qPCR; SARS-CoV-2; ddPCR
Mesh:
Substances:
Year: 2022 PMID: 35417835 PMCID: PMC8933867 DOI: 10.1016/j.diagmicrobio.2022.115677
Source DB: PubMed Journal: Diagn Microbiol Infect Dis ISSN: 0732-8893 Impact factor: 2.983
Comparison between the ddPCR and RT-qPCR results of the clinical samples.
| RT-qPCR | ||||
|---|---|---|---|---|
| Negative | Suspected | Positive | ||
| ddPCR | Negative | 16 | 0 | 0 |
| Suspected | 13 | 8 | 0 | |
| Positive | 3 | 1 | 89 | |
ddPCR = digital droplet PCR; RT-qPCR = real-time polymerase chain reaction.
Comparison between the results of ddPCR and RT-qPCR performed for different targets, using clinical samples.
| RT-qPCR | ||||
|---|---|---|---|---|
| Negative | Positive | |||
| ddPCR | IC | Negative | 0 | 0 |
| Positive | 0 | 130 | ||
| ORF gene | Negative | 35 | 0 | |
| Positive | 6 | 89 | ||
| N gene | Negative | 18 | 0 | |
| Positive | 14 | 98 | ||
ddPCR = digital droplet PCR; RT-qPCR = real-time polymerase chain reaction.
Fig. 1Comparison of the distribution of results obtained with RT-qPCR and ddPCR. (A) CT values of the IC gene in negative, suspected, and positive samples determined using RT-qPCR. (B) -log2 values of the IC gene in negative, suspected, and positive samples determined using ddPCR. (C) CT values of the ORF and N genes determined using RT-qPCR. (D) CT values of the ORF and N genes determined using ddPCR. P values are included in the box. CT = cycle threshold; ddPCR = digital droplet PCR; IC gene = internal control gene; RT-qPCR = real-time polymerase chain reaction.
Fig. 2Correlation analysis between the CT values and -log2 values determined by RT-qPCR and ddPCR, respectively, for the IC gene (A), ORF gene (B), and N gene (C). Comparison of deviations of values obtained for each sample ran using the same method. (A) CT values of the three target genes determined using RT-qPCR. (B) -log2 values of the three target genes determined using ddPCR. Correlation analysis between the ORF and N genes. (A) CT values determined with RT-qPCR. (B) -log2 values determined by ddPCR. CT = cycle threshold; ddPCR = digital droplet PCR; IC gene = internal control gene; RT-qPCR = real-time polymerase chain reaction.
Fig. 3Correlation analysis among ratios of the three target genes determined using the same or different method. (A) Correlation between CT/ CT and CT/ CT determined using RT-qPCR. (B) Correlation between −log2(value/ value) and -log2(value/ value) determined using ddPCR. (C) Correlation between CT/ CT and −log2(value/ value) determined by RT-qPCR and ddPCR, respectively. (D) Correlation between CT/ CT and −log2(value/value) determined by RT-qPCR and ddPCR, respectively. (E) Correlation between CT/ CT and −log2(value/value) determined by RT-qPCR and ddPCR, respectively. CT = cycle threshold; ddPCR = digital droplet PCR; IC gene = internal control gene; RT-qPCR = real-time polymerase chain reaction.