Comparisons of glandular breast dose between digital mammography, tomosynthesis and breast CT based on anthropomorphic patient-derived breast phantoms.

Antonio Sarno1, Giovanni Mettivier2, Kristina Bliznakova3, Andrew M Hernandez4, John M Boone4, Paolo Russo2.   

Abstract

PURPOSE: To evaluate the bias to the mean glandular dose (MGD) estimates introduced by the homogeneous breast models in digital breast tomosynthesis (DBT) and to have an insight into the glandular dose distributions in 2D (digital mammography, DM) and 3D (DBT and breast dedicated CT, BCT) x-ray breast imaging by employing breast models with realistic glandular tissue distribution and organ silhouette.
METHODS: A Monte Carlo software for DM, DBT and BCT simulations was adopted for the evaluation of glandular dose distribution in 60 computational anthropomorphic phantoms. These computational phantoms were derived from 3D breast images acquired via a clinical BCT scanner.
RESULTS: g·c·s·T conversion coefficients based on homogeneous breast model led to a MGD overestimate of 18% in DBT when compared to MGD estimated via anthropomorphic phantoms; this overestimate increased up to 21% for recently computed DgNDBT conversion coefficients. The standard deviation of the glandular dose distribution in BCT resulted 60% lower than in DM and 55% lower than in DBT. The glandular dose peak - evaluated as the average value over the 5% of the gland receiving the highest dose - is 2.8 times the MGD in DM, this factor reducing to 2.6 and 1.6 in DBT and BCT, respectively.
CONCLUSIONS: Conventional conversion coefficients for MGD estimates based on homogeneous breast models overestimate MGD by 18%, when compared to MGD estimated via anthropomorphic phantoms. The ratio between the peak glandular dose and the MGD is 2.8 in DM. This ratio is 8% and 75% higher than in DBT and BCT, respectively.
Copyright © 2022 Associazione Italiana di Fisica Medica e Sanitaria. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Anthropomorphic computational breast models; BCT; DBT; Dosimetry; Mammography

Mesh:

Year:  2022        PMID: 35395535     DOI: 10.1016/j.ejmp.2022.03.016

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  1 in total

1.  Patient-derived heterogeneous breast phantoms for advanced dosimetry in mammography and tomosynthesis.

Authors:  Marco Caballo; Carolina Rabin; Christian Fedon; Alejandro Rodríguez-Ruiz; Oliver Diaz; John M Boone; David R Dance; Ioannis Sechopoulos
Journal:  Med Phys       Date:  2022-06-08       Impact factor: 4.506

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.