Literature DB >> 35380063

Cut-off values in newborn screening for inborn errors of metabolism in Saudi Arabia.

Adbul Rafiq Khan1, Ali Alothaim1, Ahmed Alfares2, Adil Jowed1, Souad Marwan Al Enazi1, Saad Mohammed Al Ghamdi1, Ahmed Al Seneid1, Areej Algahtani1, Saleh Al Zahrani1, Majid AlFadhel3, Omar Aldibasi4, Lamya Abdulaziz AlOmair4, Rafah Bajudah5, Abeer Nawaf Alanazie5.   

Abstract

BACKGROUND: Newborn screening identifies individuals affected by a specific disorder within an apparently healthy population prior to the appearance of symptoms so that appropriate interventions can be initiated in time to minimize the harmful effects. Data on population based cut-off values, disease ranges for true positive cases, false positive rates, true positive rates, cut-off verification and comparisons with international cut-off ranges have not been done for Saudi Arabia.
OBJECTIVE: Establish population-based cut-off values and analyte ratios for newborn screening assays and clinically validate the values.
DESIGN: Population-based screening.
SETTING: Tertiary care hospitals and laboratories.
METHODS: After method verification, initial cut-off values were established by analyzing 400-500 dry blood spot (DBS) samples which were further evaluated after one year. About 74 000 patient results were reviewed to establish cut-off ranges from DBS samples received from five different hospitals during 2013-2020. Analysis was performed by tandem mass spectrometry (TMS) and a genetic screening processor. Confirmation of initial positive newborn screening results for different analytes were carried out using gas chromatography-mass spectrometry, high performance liquid chromatography and TMS. MAIN OUTCOME MEASURES: Cut-off values, ratios, positive predictive values, false positive rate, true positive rate and disease range. SAMPLE SIZE: 74 000 samples.
RESULTS: Population based cut-off values were calculated at different percentiles. These values were compared with 156 true positive samples and 80 proficiency samples. The false positive rate was less than 0.04 for all the analytes, except for valine, leucine, isovalerylcarnitine (C5), biotinidase (BTD), 17-hydroxyprogesterone and thyroid stimulating hormone. The highest false positive rate was 0.14 for BTD which was due to pre-analytical errors. The analytical positive predictive values were greater than 80% throughout the eight years.
CONCLUSION: We have established clinical disease ranges for most of the analytes tested in our lab and several ratios which gives excellent screening specificity and sensitivity for early detection. The samples were representative of the local populations. LIMITATIONS: Need for wider, population-based studies. CONFLICT OF INTEREST: None.

Entities:  

Mesh:

Substances:

Year:  2022        PMID: 35380063      PMCID: PMC8982004          DOI: 10.5144/0256-4947.2022.107

Source DB:  PubMed          Journal:  Ann Saudi Med        ISSN: 0256-4947            Impact factor:   1.526


INTRODUCTION

Newborn screening (NBS) of infants shortly after birth (24-72 hours) for conditions recommended by the national Newborn Screening Committee in Saudi Arabia can prevent disabilities and possibly death.[1,2] The objective of the NBS program is to diagnose infants born with certain genetic, metabolic and functional disorders.[3] With the increasing number of disorders for which screening methods are available, there is a greater need for a clinical and differential diagnosis of genetic metabolic disorders. Tandem mass spectrometry (TMS) has the capacity to screen for a wide range of inborn errors of metabolism (IEM) in a single test with a dry blood spot (DBS) sample.[4] The IEM disorder profile includes aminoacidemias, fatty acid oxidation disorders, and organic acidemias, as well as endocrine and enzymatic disorders. Early screening and diagnosis may decrease the mortality and morbidity rates in children with IEM. More than 50 different disorders can be screened with TMS, which has many advantages such as rapid testing, high sensitivity, high specificity, high throughput, low sample volume, as well as low maintenance and operational cost per sample.[5,6] The genetic screening processor (commercial name: GSP Instrument) from Perkin Elmer can be used to screen for endocrine disorders such as congenital adrenal hyperplasia and congenital hypothyroidism and enzymatic disorders by measuring biotinidase (BTD) and galactosemia (GALT). The GSP instrument is designed to provide a rapid diagnosis and to overcome most of the disadvantages due to changes in chemistry and the handling process.[7,8] NBS is not diagnostic, but determines whether the baby has a high or low risk of having an inherited metabolic disease. As in many scientific tests, cut-off values are used to determine which levels are normal and abnormal. NBS searches for markers of disease and the cut-off values inform the assessment of a high or low risk of disease. It is important to determine precise cut-off values for the local population screened, and the values are re-evaluated when required.[9-12] In this study, we describe how to determine initial cut-off values, and then evaluate, validate and monitor the values through the data analysis and perform external quality control of samples and confirm positive samples. The DBS samples were referred to our laboratory from several hospitals located in major cities including Riyadh, Jeddah, Dammam, Al-Ahsa and Madina under the Ministry of National Guard Health Affairs. We also received and analyzed NBS samples from other hospitals and laboratories. This study focused on the analytical procedure useful for all newborn screening laboratories to assess high risk, but clinically asymptomatic, newborns. It is also helpful for clinicians to evaluate the laboratory NBS results with population based cutoff values and ratios. The parents and other siblings of the affected child were also screened upon the request of the NBS committee, but NBS was not routinely performed for all newborns. We know of 7-8 hospitals and laboratories performing newborn screening in Saudi Arabia. No one has published data on population based cut-off values, disease ranges for true positive cases, false positive rates, true positive rates, cut-off verification and comparisons with international cut-off ranges, which indicates the clinical importance of our work. This study is one of the largest to determine NBS cut-off ranges and the results will be helpful for all laboratories and physicians across the country in the detection of IEM. We followed approved guidelines for newborn screening by TMS from the Clinical And Laboratory Standard Institute (CLSI) for cut-off determination and methods of performance evaluation.[13]

METHODS

This study was a population-based screening study to establish cut off ranges for all newborn screening analytes including amino acids, acylcarnitines, analyte ratios, 17-hydroxyprogesterone (17-OHP), thyroid stimulating hormone (TSH), biotinidase activity (BTD) and galactose-1-phosphate uridyltransferase deficiency (GALT). The initial cut-off values were established with 400-500 DBS samples followed by a yearly review. The whole data sets of about 74 000 normal patient samples were reviewed over the 8 years from 2013 to 2020, and the statistical parameters and disease ranges were calculated. All DBS samples that passed preanalytical and analytical requirements for clinical analysis (i.e. sample collection, storage, transportation, pass quality control) were included. Samples were excluded if the wrong sample was collected, samples were not filed properly in filter cards, samples were hemolyzed, or had no identifier or did not meet all other acceptable criteria established for routine clinical DBS. The non-derivatized amino acid and acylcarnitine reagent kit was purchased from Chromsystem Germany (Part #57000) for the tandem mass spectrometry analysis. The reagents for 17-OHP, TSH, BTD and GALT for the genetic screening processor were obtained from Perkin Elmer. Whatman 903 filter paper was used for the blood collection. Screening for IEMs was done with the Waters TQD mass spectrometer, integrated with an ACQUITY UPLC system (Waters, MA, USA), the GSP Instrument from Perkin Elmer (GSP Instrument, https://www.perkinelmer.com/product/genetic-screening-processor-2021-2021-0010 and the Panthera DBS puncher. The DBS samples were collected by the heel prick method from 74 000 infants in five hospitals during 2013 to 2020. The barcoded filter paper card (Whatman 903), with instructions for collection, were distributed from the Biochemical Metabolic Laboratory (BML) to all the participating hospitals, located in Riyadh, Jeddah, Dammam, Al-Ahsa and Madinah. A few drops of blood were collected on the filter paper, and allowed to thoroughly saturate the five circles on the card, followed by air-drying for 4-5 hours. All the samples were transferred to our lab within 3 days of collection, stored in a polystyrene box packed with ice to keep the samples cold during transportation. All samples were collected within 24 to 72 hours of birth. DBS were checked for acceptability against the standard protocol before processing. Five 96-well plates were punched with Panthera. Four plates were loaded into the GSP Instrument. The other plate was extracted with an internal standard and loaded into the TMS. Mass spectrometry detection was performed in the multiple reaction monitoring mode. Each compound was quantified by calculating the signal intensity ratio of the compound to its internal standard. The parameters of the GSP are built in, and it performs the analysis without any external setting. All positive results were reported to the hospital newborn screening committee, which consisted of several molecular geneticists and physicians. All abnormal results were communicated to the healthcare provider who requested additional diagnostic testing to determine if the newborn had the disorder in question. The newborn samples were screened for 22 types of treatable IEM by tandem MS and the GSP, using the DBS as recommended by the Ministry of Health. There is no benefit for screening of untreatable IEMs. For confirmation, urinary organic acids were determined by gas chromatography-mass spectrometry, and the plasma amino acids by ion exchange chromatography on the Biochrom 30 (http://www.biochrom.co.uk/). The serum level of 17-hydroxyprogesterone and the BTD activity were also measured in our laboratory. As a final confirmatory testing, positive samples were sent to an international specialized laboratory performing molecular and enzymatic analysis.

RESULTS

Initial cut-off ranges were determined by analyzing 400-500 DBS samples with the tandem MS and GSP. The positive predictive values, negative predictive values, sensitivity and accuracy were determined by using proficiency testing samples provided from the United States Centers for Disease Control (CDC) (). After one year, six percentiles (95%, 97%, 98%, 99%, 99.5%, 99.9%) were calculated from the normal patient samples to re-evaluate cut off ranges. The values at 99% were very close to initial cut-off values with slightly higher false positive rates for some analytes (data not shown). To establish new ranges based on a larger sample size, NBS results of about 74 000 patients samples (from 2013 to 2020) were reviewed and different percentiles (95%, 97%, 98%, 99%, 99.5%, 99.9%) were calculated for all analytes except the C0 (low), methionine (low), BTD and GALT, where lower percentiles (0.1%, 0.25%, 0.50%, 0.75%, 1.0%) were calculated (). The percentile disorder ranges of different analytes were calculated from the results of 156 confirmed positive cases for amino acids, amino acid ratios, acylcarnitines, acylcarnitines ratios, 17-OHP, TSH, BTD and GALT (). High and low target ranges of specific analytes for different diseases and their ratios were determined from true positive cases (). The number of true positives, the true positive prevalence rates (disease prevalence per 100 000), the false positives and false positive rates are shown in Table 5. True positive rates (sensitivity) with 95% confidence interval limit are shown in . The data are shown in . Initial and the new cut-off values for some analytes are shown in . Validation of initial cut-off values by CDC proficiency samples. For abbreviations, see footnote Table 2. PPV: positive predictive value, NPV: negative predictive value.
Table 2.

Percentiles and cut-off selection (μmol/L) and comparison of Region for the Stork Study (R4S) and Centers for Disease Control cutoff values.[14]

Analyte95%97%98%99%99.5%99.9%Cut offR4S Cut-offCDC Cut-off
GLY465502532585635733<733767NA
ALA327357381419463536<536507NA
VAL113124133148166208<208212300
LEU/ILEU152166178199221258<258235290
ORN119130139153164177<177NANA
ASP8395104119131138<138NANA
MET (high)313537435082<824475
PHE71848999107143<10797150
CIT252831364073<732855
TYR140184200227241269<241207350
GLU484521525530533536<536551NA
ARG253033363842<423270
C0 (high)253235394144<4459NA
C2283639424549<4952NA
C32.83.74.04.65.35.9<5.94.745.65
C40.40.60.60.80.91.0<1.00.751.30
C50.160.250.300.370.460.70<0.700.390.7
C60.060.080.090.100.100.11<0.100.180.4
C5DC0.150.200.220.250.280.38<0.380.170.35
C80.060.080.090.110.140.27<0.270.210.45
C100.090.120.130.150.170.55<0.170.260.45
C120.120.160.180.200.240.33<0.330.41NA
C140.220.280.300.340.380.56<0.380.500.75
C163.544.344.625.095.626.57<6.576.07.50
C181.031.301.401.581.792.51<1.791.72.3
C5:1000.010.020.030.05<0.250.080.25
C5OH0.250.320.360.420.500.52<0.500.380.80
C8:10.070.110.120.160.210.32<0.21NANA
C10:10.050.060.070.080.090.10<0.100.180.30
C14:10.130.180.200.250.330.37<0.370.370.60
C14OH0.010.020.020.020.040.04<0.04NANA
C16:10.240.300.330.400.460.51<0.53NANA
C16:1OH0.050.060.060.070.080.12<0.120.13NA
C16OH0.020.030.040.060.080.11<0.100.080.13
C18:11.752.252.472.903.383.76<3.762.53.5
C18:1OH0.020.030.030.050.080.12<0.120.07NA
C18OH0.020.030.030.040.040.09<0.090.060.10
PHE/TYR1.041.351.461.701.962.98<1.96NANA
C3/C20.150.190.200.230.250.41<0.25NANA
C5/C00.010.020.020.030.030.06<0.03NANA
C8/C101.001.501.502.002.502.5<2.05NANA
TSH (μU/mL)91012152138<21NA13.6
17OHP (nmol/L)222833445793<93NA47.6
Analyte 0.1% 0.25% 0.50% 0.75% 1.00% Cut off ----------
C0 (Low)0.02.44.505.205.6>4.50NANA8.0
Meth (Low)5.716.056.456.786.97>6.05NANANA
GALT (U/dL)3.34.65.56.16.5>3.3NANANA
BTD (U/dL)18.528.237.043.349>50NANANA

GLY: glycine, ALA: alanine, VAL: valine, LEU/ILEU: leucine/isoleucine, ORN: ornithine, ASP: aspartic acid, MET: methionine, PHE: phenylalanine, CIT: citrulline, TYR: tyrosine, GLU: glutamic acid, ARG: arginine, C0: free carnitine, C2: acetylcarnitine, C3: propionylcarnitine, C4: butyryl-/isobutyrylcarnitine, C5: isovaleryl-/2-methylbutyrylcarnitine, C6: hexanoylcarnitine, C5DC: glutarylcarnitine, C8: octanoylcarnitine, C10:decanoylcarnitine, C12: dodecanoylcarnitine, C14: tetradecanolycarnitine, C16: pamitoylcarnitine, C18: stearylcarnitine, C5:1: tiglylcarnitine, C5OH: hydroxyl-isovalerylcarnitine, C8:1: otenylcarnitine, C10:1: dodecenylcarnitine, C14:1: tetradecenylcarnitine, C14OH: hydroxyl tetradecanoylcarnitine, C16:1: hexadecenoylcarnitine, C16:1OH: hydroxyl-hexadecenoylcarnitine, C16OH: hydroxypalmitoylcarnitine, C18:1: oleylcarnitine, C18:1OH: hydroxyl-oleylcarnitine, C18OH: hydroxyl-stearylcarnitine: TSH: thyroid stimulating harmone, 17-OHP: 17-hydroxy progesterone, GALT: galactosemia: BTD: biotinidase deficiency

Percentiles and cut-off selection (μmol/L) and comparison of Region for the Stork Study (R4S) and Centers for Disease Control cutoff values.[14] GLY: glycine, ALA: alanine, VAL: valine, LEU/ILEU: leucine/isoleucine, ORN: ornithine, ASP: aspartic acid, MET: methionine, PHE: phenylalanine, CIT: citrulline, TYR: tyrosine, GLU: glutamic acid, ARG: arginine, C0: free carnitine, C2: acetylcarnitine, C3: propionylcarnitine, C4: butyryl-/isobutyrylcarnitine, C5: isovaleryl-/2-methylbutyrylcarnitine, C6: hexanoylcarnitine, C5DC: glutarylcarnitine, C8: octanoylcarnitine, C10:decanoylcarnitine, C12: dodecanoylcarnitine, C14: tetradecanolycarnitine, C16: pamitoylcarnitine, C18: stearylcarnitine, C5:1: tiglylcarnitine, C5OH: hydroxyl-isovalerylcarnitine, C8:1: otenylcarnitine, C10:1: dodecenylcarnitine, C14:1: tetradecenylcarnitine, C14OH: hydroxyl tetradecanoylcarnitine, C16:1: hexadecenoylcarnitine, C16:1OH: hydroxyl-hexadecenoylcarnitine, C16OH: hydroxypalmitoylcarnitine, C18:1: oleylcarnitine, C18:1OH: hydroxyl-oleylcarnitine, C18OH: hydroxyl-stearylcarnitine: TSH: thyroid stimulating harmone, 17-OHP: 17-hydroxy progesterone, GALT: galactosemia: BTD: biotinidase deficiency Percentile disorder ranges (μmol/L). For abbreviations, see footnote Table 2. Target range calculations (μmol/L). Calculated from normal population Calculated from true positive cases. For abbreviations, see footnote Table 2. MSUD: Maple syrup urine disease. Analytical false positive rates and true positive prevalence. For abbreviations, see footnote Table 2. True positive rates (sensitivity) for analytes. For abbreviations, see footnote Table 2

DISCUSSION

After the verification of the method of performance specification (linearity, precision, comparison and sensitivity), the initial cut-off was determined by analyzing 400-500 DBS samples with the tandem MS and GSP. For dicarboxylic, unsaturated and hydroxylated forms of carnitines, the cut-off values were obtained from the literature because most of the results were very low. Validation of initial cut-off values was performed by participating in the international proficiency testing program from the CDC for newborn screening. After analyzing 5 positive and 25 negative proficiency-testing samples, the results were submitted to the CDC. Our performance was 100% satisfactory, except for a few false positive results for valine and oleoylcarnintine (C18:1). The positive predictive values, negative predictive values, sensitivity and accuracy were calculated. TSH, 17 OH-progesterone, GALT and BTD were not included in the proficiency testing program (). After one year, the results of normal patient samples were reviewed to evaluate the initial cut-off ranges. As the patient data did not have a Gaussian distribution, the cut-off ranges were estimated by calculating different percentiles. Typically, six percentiles (95%, 97%, 98%, 99%, 99.5%, 99.8%) were determined and compared with initial ranges. The values at 99% were very close to initial cut-off values with slightly higher false positive rates for some analytes. After 8 years, the newborn screening results of about 74 000 normal patient samples were reviewed. These samples were received from five different hospitals located in Riyadh, Jeddah, Dammam, Al-Ahsa and Madinah under the Ministry of National Guard during 2013 to 2020. To determine population-based cut-off values, six percentiles (95%, 97%, 98%, 99%, 99.5%, 99.9%) were calculated for all the analytes in our NBS panel except the C0(low), Methionine (low), BTD and GALT, where lower percentiles (0.1%, 0.25%, 0.50%, 0.75%, 1.0%) were calculated. These percentiles were estimated because the data were not of a Gaussian distribution as recommended by mass spectrometry guidelines from the National Committee for Clinical Laboratory Standards. The newly established initial cut-off values were compared with initial cut off values and necessary changes were made. For most of the analytes, the 99% and 99.9% percentile distribution were selected as the new cut-off values for the normal population (). The selection of the new values was assessed by reviewing the results of the 156 true positive cases, identified and confirmed in our laboratory. The selection is also based on the satisfactory performance evaluation from the CDC proficiency-testing programs. The results of the C5:1 in most of the samples were very low, and the percentile calculation was not possible. Based on the literature, we used 0.25 μmol/L as the cut-off value, which was verified by analyzing the true positive proficiency samples from the CDC. The values corresponding to the 0.5% and 0.1% percentile (C0=4.5 μmol/L and GALT=3.3 U/dL) were selected for the C0 (low) and GALT, respectively, because these cut-off values were used for the evaluation of the CDC proficiency testing results, which gave 100% satisfactory performance for both analytes. For BTD, four true positive patients were identified and confirmed with results ranging between 40 – 50 U/dL. The calculated cut-off value with the 0.75% percentile was 43.3 U/dL. All the proficiency sample results were much lower, from 5–15 U/dL, which did not help to differentiate between a false positive from a true positive patient. As a result, we used >50 U/dL (normal) as the cut-off value for BTD, which caused an increased number of false positive results per year. The newly established cut-off values were compared with the cut-off values determined by Region for the Stork Study (R4S) and CDC.[14] A comparison of the three values (our lab, CDC and R4S) is presented in . The established cut-off values are greater than the R4S and lower than the CDC cut-off values for more than half of the analytes of amino acids and acylcarnitine. The cut off of citrulline (<73 μmol/L), methionine (<82 μmol/L), C4 (<1.0 μmol/L) and C18:1 (3.76 μmol/L) are higher than R4S and CDC values. For the TSH and 17-hydroxyprogesterone, our cut-off values are higher than the CDC values. These variations are due to age, health status of the newborn at the time of collection, environmental conditions during transport, DBS collection procedures, methods used, instrument platform and stability of the analyte measured. The results of the 156 true positive cases and 80 proficiency-testing samples were evaluated against new cut off ranges and no false negative results were obtained, verifying the accuracy of these ranges. The percentile disorder ranges were calculated from confirmed positive cases for amino acids, amino acid ratios, acylcarnitines, acylcarnitines ratios, 17-OHP, TSH, BTD and GALT (). The high target disorder ranges were obtained from the 99th percentile of the normal population low range was taken from 5th percentile of all the disorder ranges of the same analyte. We included positive proficiency sample results in the calculation of the target range for MSUD, tyrosinemia and PKU () due to the low number of true positive samples. Most of the false positive cases were within this high target range. We observed that most of the positive proficiency samples were near the lowest 5th percentile of the disorder ranges, and the false positive patient samples were near the 99th percentile of the normal population, but not close to the lowest 5th percentile of the disorder ranges. From the number of true positive, true positive prevalence (disease prevalence) and false positive rates, the disease prevalence was calculated per 100 000. TSH had the highest prevalence (27.2) and PKU had lowest prevalence (1.0). The disease prevalence related to BTD deficiency, 17-OHP, GALT, C14:1, C5OH and C3 were all greater than 14. The false positive rate was less than 0.04 for all the analytes, except for valine, leucine, C5, BTD, 17-OHP and TSH (). The highest false positive rate of BTD (0.144) was connected to a pre-analytical error where the DBS samples were packed in biohazard bags and sent to the laboratory without drying for 3-4 hours. We observed that the BTD activity decreased 10% to 20% if the DBS were not completely dried. Decreased enzyme activity can also be caused by insufficient cooling of the samples in case of high external temperatures. Valine (0.068) and C5 (0.086) also had a high false positive rate, which was due to lower cut-off values. The high positive rate for TSH (0.107) and 17-OHP (0.049) was due to early sampling. In terms of leucine (0.068), we could not find any pre-analytical or analytical reason for the high false positive rate. It was not related to the lower cut-off values, as some of the true positive proficiency testing samples were just above the initially set cut-off value. In the 8 years, no questions were raised by physicians from any of the five tertiary care hospitals about the cut-off values except the higher false positive results for BTD. We also did not observe any false negative results since the start of the NBS program. The true positive rate (sensitivity) with 95% confidence intervals were calculated. During the last 8 years, we had no false negative results after reporting as normal. There were no recalls from our hospital newborn screening committee, which is responsible for treatment and management of all true positive cases. Therefore, we were unable to make receiver operating characteristic curves and calculate area under the curve; however, true positive rates (sensitivity) were calculated for all analytes having positive results with 95% confidence intervals (). About 80 unknown proficiency-testing samples were analyzed and evaluated against new cut-off values during 2013 to 2020. A few false positive results were reported, but there were no false negative results. The overall performance was satisfactory with 100% accuracy. Distribution of true positive (red), false positive (green) and true positive proficiency (blue) samples around the cut-off (red Line) are shown in for few analytes. The true positive samples were far from the cut-off ranges for most of the analytes. The comparison of initial and the new cut-off values for some analytes was also prepared. Number of false positives was significantly reduced for valine, pentanylcarnitine (C5) and methionine (). The new cut-off values for C3 were lower than the initial values, which did not result in an increase in the false positive rate. Similarly, many changes were made in the initial cut-off values for the amino acids, acylcarnitines and ratios to obtain accurate normal ranges, representing the local population. We used to collect dry blood spot samples early during 24-72 hours after birth for several reasons including family demands to discharge early and hospital capacity. However, some metabolites may not always be sufficiently high/low for a reliable diagnosis at this early age. In most NBS programs blood sampling between 48-72 hours is preferred. While this study is the largest study in the region, a wider population-based study is still desirable with the different types of reagent kits used in the country. Therefore, we considered our established cut-off ranges as representative of the local populations, but a wider population-based study would still be desirable. In conclusion, establishing specific and population-based cut-off values and analyte ratios are imperative for quick and accurate diagnoses of IEM. Established cut-off values also support a reduction in false positive rates, increases the positive predictive values, and prevents unnecessary testing as well as family anxiety. Using population-based data with true positive samples, we established cut-off values for each analyte tested in our laboratory. We calculated several ratios for various analytes, which provides excellent screening specificity and sensitivity.
Table 1.

Validation of initial cut-off values by CDC proficiency samples.

AnalytesCut-off (μmol/L)Positive samples (n)Negative samples (n)Analytical PPV (%)NPV (%)Sensitivity (%)Accuracy (%)
Valine<174525100100100100
Leucine/Isoleucine<281525100100100100
Methionine<70525100100100100
Phenylalanine<165525100100100100
Citrulline<49525100100100100
Tyrosine<217525100100100100
Arginine<99525100100100100
C0<57525100968096
C3<6.0525100100100100
C4<0.95525100100100100
C5<0.56525100100100100
C5DC<0.47525100100100100
C5OH<0.5525100100100100
C6<0.13525100100100100
C8<0.14525100100100100
C10<0.17525100100100100
C14<0.6525100100100100
C14:1<0.39525100100100100
C16<9.30525100100100100
C18<2.005258310010096

For abbreviations, see footnote Table 2. PPV: positive predictive value, NPV: negative predictive value.

Table 3.

Percentile disorder ranges (μmol/L).

AnalyteConditionsn1%5%10%25%50%75%90%99%
ValineMSUD7352357362383401428459492
Leucine/IsoleucineMSUD567633714801847113011651292
MethHCY71221291371541642396601020
CitrullineCIT-1/ASA139511013415820139414622323.6
TyrosineTYT-11713713713713713713713713
PhenylalaninePKU1274274274274274274274274
C3MMA/PA167.167.507.7310.3513.3021.1536.2044.82
C3/C2MMA/PA0.640.680.811.171.483.474.157.97
C5IVA51.121.141.211.612.012.967.628.77
C5/C00.090.090.090.090.090.091.162.12
C5/C20.130.130.140.140.140.160.500.81
C5OH3MCC/HMG Co Lyase140.820.830.901.482.606.028.7225.24
C5DCGA-130.880.900.931.021.172.944.004.64
C8MCAD50.830.901.001.223.359.5115.0018.69
C100.060.070.070.090.190.550.931.19
C8/C1013.5013.5013.5013.8115.2116.6918.4019.63
C14:1VLCAD90.570.680.821.452.072.973.133.52
17-OHPCAH1588104121219300300318338
TSHCH2529293058133226300300
BTDBTD16811122130404546
GALTGALT190.71.11.11.52.22.52.62.9

For abbreviations, see footnote Table 2.

Table 4.

Target range calculations (μmol/L).

AnalyteConditionsNo. of casesHigh target ranges
Low (99th)[a]High (5th)[b]
ValineMSUD7256356
Leu/Isoleucine263633
MethHCY784117.6
CitrullineCIT-11372110
PhenylalaninePKU7104274
TyrosineTYT-18238631
C3MMA/PA165.17.5
C3/C20.240.68
C5IVA50.801.14
C5/C00.030.09
C5/C20.040.13
C5OH3-HMG CoA/3MCC140.500.84
C5DCGA-130.290.61
C8MCAD50.270.93
C100.070.17
C8/C102.0513.5
C14:1VLCAD90.390.63
17OHPCAH1580103.9
TSHCH252129

Calculated from normal population

Calculated from true positive cases. For abbreviations, see footnote Table 2. MSUD: Maple syrup urine disease.

Table 5.

Analytical false positive rates and true positive prevalence.

ParametersTrue positives (TP) (n)TP prevalence (Per 100 000) (n)False positives (n)False positive rate (%)
VAL45.7700.068
LEU45.7700.068
MET710200.019
PHE57.110.001
CIT1014.3130.013
TYR45.750.005
C868.640.004
ARG34.340.004
C1068.640.004
C8/C1068.640.004
C14:11115.7210.020
C31724.380.008
C3/C21724.380.008
C568.6890.086
C5/C068.6890.086
C5/C268.6890.086
C5OH1521.4160.016
C5DC34.3270.026
BTD1827.71480.144
GALT2028.6390.038
17OHP1724.3500.049
TSH28401100.107

For abbreviations, see footnote Table 2.

Table 6.

True positive rates (sensitivity) for analytes.

AnalyteProportion95% CI
CIT0.45450.2465–0.6626
VAL, LEU, ILEU0.05330.0025–0.1042
METH0.24140.0857–0.3971
TYR0.40000.0960–0.7036
PHE1.001.000
C3, C3/C20.61900.4113–0.8267
C50.05380.0079–0.0996
C5OH0.50000.3268–0.6732
C5DC0.18180.0502–0.3134
C8 & C100.66670.3999–0.9334
C14:10.27270.1411–0.4043
17-OHP0.22390.1241–0.3422
TSH0.18380.1187–0.2489
GALT0.32200.2028–0.4413
BTD0.02320.0521–0.1432

For abbreviations, see footnote Table 2

  11 in total

1.  CDC Grand Rounds: Newborn screening and improved outcomes.

Authors: 
Journal:  MMWR Morb Mortal Wkly Rep       Date:  2012-06-01       Impact factor: 17.586

Review 2.  A biochemical perspective on the use of tandem mass spectrometry for newborn screening and clinical testing.

Authors:  Donald H Chace; Theodore A Kalas
Journal:  Clin Biochem       Date:  2005-04       Impact factor: 3.281

3.  Nine years of newborn screening for classical galactosemia in the Netherlands: Effectiveness of screening methods, and identification of patients with previously unreported phenotypes.

Authors:  Lindsey Welling; Anita Boelen; Terry G J Derks; Peter C J I Schielen; Maaike de Vries; Monique Williams; Frits A Wijburg; Annet M Bosch
Journal:  Mol Genet Metab       Date:  2016-12-29       Impact factor: 4.797

Review 4.  Treatable inborn errors of metabolism causing intellectual disability: a systematic literature review.

Authors:  Clara D M van Karnebeek; Sylvia Stockler
Journal:  Mol Genet Metab       Date:  2011-11-30       Impact factor: 4.797

5.  Introducing and Expanding Newborn Screening in the MENA Region.

Authors:  Victor Skrinska; Issam Khneisser; Peter Schielen; Gerard Loeber
Journal:  Int J Neonatal Screen       Date:  2020-02-19

6.  The adjustment of 17-hydroxyprogesterone cut-off values for congenital adrenal hyperplasia neonatal screening by GSP according to gestational age and age at sampling.

Authors:  Xiang Jiang; Fang Tang; Yi Feng; Bei Li; Xuefang Jia; Chengfang Tang; Sichi Liu; Yonglan Huang
Journal:  J Pediatr Endocrinol Metab       Date:  2019-11-26       Impact factor: 1.634

7.  Screening newborns for inborn errors of metabolism by tandem mass spectrometry.

Authors:  Bridget Wilcken; Veronica Wiley; Judith Hammond; Kevin Carpenter
Journal:  N Engl J Med       Date:  2003-06-05       Impact factor: 91.245

Review 8.  A review of newborn screening in the era of tandem mass spectrometry: what's new for the pediatric neurologist?

Authors:  Sara Copeland
Journal:  Semin Pediatr Neurol       Date:  2008-09       Impact factor: 1.636

9.  Clinical validation of cutoff target ranges in newborn screening of metabolic disorders by tandem mass spectrometry: a worldwide collaborative project.

Authors:  David M S McHugh; Cynthia A Cameron; Jose E Abdenur; Mahera Abdulrahman; Ona Adair; Shahira Ahmed Al Nuaimi; Henrik Åhlman; Jennifer J Allen; Italo Antonozzi; Shaina Archer; Sylvia Au; Christiane Auray-Blais; Mei Baker; Fiona Bamforth; Kinga Beckmann; Gessi Bentz Pino; Stanton L Berberich; Robert Binard; François Boemer; Jim Bonham; Nancy N Breen; Sandra C Bryant; Michele Caggana; S Graham Caldwell; Marta Camilot; Carlene Campbell; Claudia Carducci; Sandra C Bryant; Michele Caggana; S Graham Caldwell; Marta Camilot; Carlene Campbell; Claudia Carducci; Rohit Cariappa; Clover Carlisle; Ubaldo Caruso; Michela Cassanello; Ane Miren Castilla; Daisy E Castiñeiras Ramos; Pranesh Chakraborty; Ram Chandrasekar; Alfredo Chardon Ramos; David Cheillan; Yin-Hsiu Chien; Thomas A Childs; Petr Chrastina; Yuri Cleverthon Sica; Jose Angel Cocho de Juan; Maria Elena Colandre; Veronica Cornejo Espinoza; Gaetano Corso; Robert Currier; Denis Cyr; Noemi Czuczy; Oceania D'Apolito; Tim Davis; Monique G de Sain-Van der Velden; Carmen Delgado Pecellin; Iole Maria Di Gangi; Cristina Maria Di Stefano; Yannis Dotsikas; Melanie Downing; Stephen M Downs; Bonifacio Dy; Mark Dymerski; Inmaculada Rueda; Bert Elvers; Roger Eaton; Barbara M Eckerd; Fatma El Mougy; Sarah Eroh; Mercedes Espada; Catherine Evans; Sandy Fawbush; Kristel F Fijolek; Lawrence Fisher; Leifur Franzson; Dianne M Frazier; Luciana R C Garcia; Maria Sierra García-Valdecasas Bermejo; Dimitar Gavrilov; Rosemarie Gerace; Giuseppe Giordano; Yolanda González Irazabal; Lawrence C Greed; Robert Grier; Elyse Grycki; Xuefan Gu; Fizza Gulamali-Majid; Arthur F Hagar; Lianshu Han; W Harry Hannon; Christa Haslip; Fayza Abdelhamid Hassan; Miao He; Amy Hietala; Leslie Himstedt; Gary L Hoffman; William Hoffman; Philis Hoggatt; Patrick V Hopkins; David M Hougaard; Kerie Hughes; Patricia R Hunt; Wuh-Liang Hwu; June Hynes; Isabel Ibarra-González; Cindy A Ingham; Maria Ivanova; Ward B Jacox; Catharine John; John P Johnson; Jón J Jónsson; Eszter Karg; David Kasper; Brenda Klopper; Dimitris Katakouzinos; Issam Khneisser; Detlef Knoll; Hirinori Kobayashi; Ronald Koneski; Viktor Kozich; Rasoul Kouapei; Dirk Kohlmueller; Ivo Kremensky; Giancarlo la Marca; Marcia Lavochkin; Soo-Youn Lee; Denis C Lehotay; Aida Lemes; Joyce Lepage; Barbara Lesko; Barry Lewis; Carol Lim; Sharon Linard; Martin Lindner; Michele A Lloyd-Puryear; Fred Lorey; Yannis L Loukas; Julie Luedtke; Neil Maffitt; J Fergall Magee; Adrienne Manning; Shawn Manos; Sandrine Marie; Sônia Marchezi Hadachi; Gregg Marquardt; Stephen J Martin; Dietrich Matern; Stephanie K Mayfield Gibson; Philip Mayne; Tonya D McCallister; Mark McCann; Julie McClure; James J McGill; Christine D McKeever; Barbara McNeilly; Mark A Morrissey; Paraskevi Moutsatsou; Eleanor A Mulcahy; Dimitris Nikoloudis; Bent Norgaard-Pedersen; Devin Oglesbee; Mariusz Oltarzewski; Daniela Ombrone; Jelili Ojodu; Vagelis Papakonstantinou; Sherly Pardo Reoyo; Hyung-Doo Park; Marzia Pasquali; Elisabetta Pasquini; Pallavi Patel; Kenneth A Pass; Colleen Peterson; Rolf D Pettersen; James J Pitt; Sherry Poh; Arnold Pollak; Cory Porter; Philip A Poston; Ricky W Price; Cecilia Queijo; Jonessy Quesada; Edward Randell; Enzo Ranieri; Kimiyo Raymond; John E Reddic; Alejandra Reuben; Charla Ricciardi; Piero Rinaldo; Jeff D Rivera; Alicia Roberts; Hugo Rocha; Geraldine Roche; Cheryl Rochman Greenberg; José María Egea Mellado; María Jesús Juan-Fita; Consuelo Ruiz; Margherita Ruoppolo; S Lane Rutledge; Euijung Ryu; Christine Saban; Inderneel Sahai; Maria Isabel Salazar García-Blanco; Pedro Santiago-Borrero; Andrea Schenone; Roland Schoos; Barb Schweitzer; Patricia Scott; Margretta R Seashore; Mary A Seeterlin; David E Sesser; Darrin W Sevier; Scott M Shone; Graham Sinclair; Victor A Skrinska; Eleanor L Stanley; Erin T Strovel; April L Studinski Jones; Sherlykutty Sunny; Zoltan Takats; Tijen Tanyalcin; Francesca Teofoli; J Robert Thompson; Kathy Tomashitis; Mouseline Torquado Domingos; Jasmin Torres; Rosario Torres; Silvia Tortorelli; Sandor Turi; Kimberley Turner; Nick Tzanakos; Alf G Valiente; Hillary Vallance; Marcela Vela-Amieva; Laura Vilarinho; Ulrika von Döbeln; Marie-Francoise Vincent; B Chris Vorster; Michael S Watson; Dianne Webster; Sheila Weiss; Bridget Wilcken; Veronica Wiley; Sharon K Williams; Sharon A Willis; Michael Woontner; Katherine Wright; Raquel Yahyaoui; Seiji Yamaguchi; Melissa Yssel; Wendy M Zakowicz
Journal:  Genet Med       Date:  2011-03       Impact factor: 8.822

10.  Diagnosis and therapeutic monitoring of inborn errors of metabolism in 100,077 newborns from Jining city in China.

Authors:  Chi-Ju Yang; Na Wei; Ming Li; Kun Xie; Jian-Qiu Li; Cheng-Gang Huang; Yong-Sheng Xiao; Wen-Hua Liu; Xi-Gui Chen
Journal:  BMC Pediatr       Date:  2018-03-13       Impact factor: 2.125

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.