Eliel Ben-David1,2, Abraham J Domb3, Haixing Liao4, Awanish Kumar3, Issac Nissenbaum1, Matthias Stechele5, Peter Siman6, Natalie Greenbaum1,7, Naama Lev Cohain1,7, S Nahum Goldberg1,7. 1. Faculty of Medicine, Hebrew University of Jerusalem, Jerusalem 9112102, Israel. 2. The Department of Radiology, Shaare Zedek Medical Center, Jerusalem 9103102, Israel. 3. Institute of Drug Research, School of Pharmacy-Faculty of Medicine, The Hebrew University of Jerusalem, Jerusalem 9112102, Israel. 4. Department of Ultrasonography, The First Affiliated Hospital of Guangzhou Medical University, Guangzhou 9112102, China. 5. der Klinik und Poliklinik für Radiologie Ludwig-Maximilians-Universität München, Munich 81377, Germany. 6. Intragel, Nazareth Industrial Area, Wadi El Haj 13, P.O. Box 1252, Nazareth 17111, Israel. 7. The Department of Radiology, Hadassah-Hebrew University Medical Center, Jerusalem 91121, Israel.
Abstract
Biodegradable polymer clips as multidimensional soft tissue biopsy markers were developed with better biocompatibility and imaging features. Unlike the commercially available metallic biopsy markers, the developed polymer clips are temporary implants with similar efficacies as metal markers in imaging and detection and get absorbed within the body with time. Herein, we evaluate the degradation rate of three resorbable polymer-based marker compounds in an in vivo murine model. Three polymers, abbreviated as Polymer A (PLGA poly(lactic-co-glycolic acid)50:50), Polymer B (PLGA (poly(lactic-co-glycolic acid)) 75:25), and Polymer C (polycaprolactone (PCL)), mixed with 20% lipiodol and 0.2% iron oxide and a control polymer were implanted into nine mice, followed by CT and MRI imaging. Images were evaluated for conspicuity. Specimens were examined for tissue analysis of iodine and iron contents. Significant differences in polymer resorption and visualization on CT were noted, particularly at 8 weeks (p < 0.027). Polymers A, B, and C were visible by CT at 4, 6, and 8 weeks, respectively. All marker locations were detected on MRI (T1 and SWI) after 24 weeks, with tattooing of the surrounding soft tissue by iron deposits. CT and MR visible polymer markers can be constructed to possess variable resorption, with stability ranging between 4 and 14 weeks post placement, making this approach suitable for distinct clinical scenarios with varying time points.
Biodegradable polymer clips as multidimensional soft tissue biopsy markers were developed with better biocompatibility and imaging features. Unlike the commercially available metallic biopsy markers, the developed polymer clips are temporary implants with similar efficacies as metal markers in imaging and detection and get absorbed within the body with time. Herein, we evaluate the degradation rate of three resorbable polymer-based marker compounds in an in vivo murine model. Three polymers, abbreviated as Polymer A (PLGA poly(lactic-co-glycolic acid)50:50), Polymer B (PLGA (poly(lactic-co-glycolic acid)) 75:25), and Polymer C (polycaprolactone (PCL)), mixed with 20% lipiodol and 0.2% iron oxide and a control polymer were implanted into nine mice, followed by CT and MRI imaging. Images were evaluated for conspicuity. Specimens were examined for tissue analysis of iodine and iron contents. Significant differences in polymer resorption and visualization on CT were noted, particularly at 8 weeks (p < 0.027). Polymers A, B, and C were visible by CT at 4, 6, and 8 weeks, respectively. All marker locations were detected on MRI (T1 and SWI) after 24 weeks, with tattooing of the surrounding soft tissue by iron deposits. CT and MR visible polymer markers can be constructed to possess variable resorption, with stability ranging between 4 and 14 weeks post placement, making this approach suitable for distinct clinical scenarios with varying time points.
Fiducial marker placement is often used
to improve upon the accuracy
of cutting-edge, state-of-the-art techniques, such as minimally invasive
surgery,[1,2] interventional procedures,[3] and precision brachytherapy.[4] Clinical scenarios include the need for marking of abnormal imaging
findings or tumors, most commonly in the breast, to permit identification
at surgery[5,6] and the need for multiple three-dimensional
markers for precise serial application of stereotactic radiation.[7] Currently, these markers are most often metallic,
making them permanent. Yet, permanency of the markers is often not
required. For example, if management is surgical, markers are required
for in situ identification for at most a range of a few weeks, and
if radiation treatment is considered, the marker is usually required
for only a 2–3-month period and most often not beyond a 6-month
course of therapy. Nevertheless, retention of a foreign body beyond
this time frame is extraneous and sometimes potentially undesirable.
Particularly, it is crucial to minimize obscuring image artifacts,
particularly in the region most likely to demonstrate recurrence,
and metallic implants are known to exhibit a strong blooming artifact
on MR (magnetic resonance) imaging.To overcome this permanent
nondegradable nature of metallic clips,
slowly self-resorbing polymer-based markers have been developed.[8] One early example is a product that incorporates
a water-soluble polyethylene glycol-based hydrogel polymer, which
is readily identified by ultrasound detection but poorly visualized
on other modalities.[9] Recently, a biodegradable
implant has been designed and formulated incorporating clinically
approved, commonly used contrast agents visible at both CT (computer
tomography) and MR (magnetic resonance).[10] In a preliminary in vitro and a short-term in vivo study, the visibility,
shape, and degradation of these biodegradable implants containing
Lipiodol (an X-ray contrast medium) were evaluated by CT, and it was
concluded that this Lipiodol-containing poly(ricinoleic acid-co-sebacic acid) polymer is visible on CT, enabling polymer
degradation to be potentially monitored noninvasively.[10] Visualization of this material on both ultrasound
and MR has been characterized for this product to determine the optimal
polymer composition for conspicuity and biodegradability in vitro
and ex vivo for up to 2 weeks.[10] It was
further noted that alteration of the polymer formulation could affect
the duration of visualization of these potentially degradable products.
Nevertheless, longer-term studies over a time frame of weeks to months
needed for clinical markers have yet to be achieved. This includes
visualization in the setting of anticipated resorption over time and
the safety and biological reaction to these implants. Accordingly,
in this study, we evaluate the conspicuity and biodegradability properties
of an injectable multimodality polymer over a six-month interval in
an in vivo murine model.
Materials and Methods
Based on clinical need, the compounds for evaluation needed to
achieve CT and MR conspicuity lasting between 4 and 24 weeks to cover
a wide range of scenarios. Thus, based on a previous study that demonstrated
conspicuity and degradation over several weeks,[10] we chose three potential compounds for evaluation.
Materials
Materials used for the preparation of Lipiodol
implants included PCL (polycaprolactone) with Mw 14 000 Da; iron (II,III) oxide nanopowder with 50–100
nm particle size (Lot# MKBR5062V) (Sigma Aldrich, Israel); PLGA (poly(lactic-co-glycolic acid)) with a 50:50 ratio of lactic acid to
glycolic acid and Mw 17 kDa; PLGA 75:25
with Mw 18 kDa (PURAC, The Netherlands);
lipiodol ultrafluid with an iodine content of 0.49 mg/mL Ch-B:17LU602A
(Guerbet, Villepinte, France); and chloroform (CHCl3) (BioLab,
Israel).
Polymer Marker Preparation
Each of the three polymers
used in this study contained 20% Lipiodol and 0.2% w/w iron oxide
based on prior studies documenting the adequate conspicuity of these
concentrations on CT and MR, respectively.[10] The three polymer compounds were composed of Polymer A–PLGA
50:50, Polymer B–PLGA 75:25, and Polymer C–PCL. The
polymers were formed into a rod shape of an adequate 1 mm diameter
to fit into a biopsy syringe, similar to other commercially available
markers used in clinical practice.[11] Lipiodol,
iron oxide, and polymer were added to a 5 mL glass vial, and the mixture
was heated to 80 °C for 30 min while hand mixing with a spatula
to form a viscous liquid. The molten mixture was then transferred
to a hot glass syringe connected to a 17 G needle and cast by pressing
through the needle to form black cylindrical rods. The rods were then
cut into 2–3 mm implants using an 11-blade scalpel and used
for further studies.
Murine In Vivo Model
A total of
nine male BaLB/C OLAHSD
mice, 8–9 weeks of age, weighing approximately 20 g, were obtained
from Harlan Laboratories (Rehovot, Israel). Each mouse had four markers
inserted, the three study polymers and a control substance (the polymer
marker without Lipiodol or iron). Mice were housed in cages with free
access to food and water. Animal care and the test injections were
conducted at a good laboratory practice (GLP)-certified site (Sharett
institute SPF unit, Hadassah Medical School), in accordance with the
National Institutes of Health guide for the care and use of laboratory
animals. The animals were anesthetized with a ketamine–xylazine
cocktail: 87.5 mg/kg ketamine (Ketaset, 100 mg/mL, Fort Dodge, Iowa)
and 12.5 mg/kg xylazine (20 mg/mL, Biob, France) administrated intramuscularly
(IM) at a dose of 5 mL/kg bodyweight. Mice were anesthetized and underwent
subcutaneous insertion of polymer markers using a coaxial needle technique
(17-gauge) to the back (right and left upper and lower regions, respectively).
Three 3 mm × 1 mm markers containing the polymer formulas (A,
B, and C) and one control marker, PSA/RA polymer (polymer without
contrast agents), were inserted. The location of each polymer marker
type was random for each mouse. Euthanasia was achieved by means of
carbon dioxide, according to institutional animal care and use committee
guidelines.
Imaging
Imaging was performed at
baseline 24 h after
implantation and at 2-week scheduled intervals to 6 months (13 scans
per mouse). Imaging protocols included serial CT and a final MRI.
CT Imaging
CT was performed on anesthetized mice, in
the prone position, imaged on a 64-detector scanner (Brilliance 64
CT scanner, Philips Medical Systems, Cleveland, OH). Scans were performed
with the following parameters: 120 kV, 70 mAs, collimation 64 ×
0.625 slice thickness 0.9 mm, increment 0.45 mm, rotation time 0.5
s, and pitch of 0.641. Images were reconstructed using bone and soft
tissue algorithms. Follow-up scans were performed every 2 weeks, up
to 24 weeks.
MR Imaging
MR imaging was performed
at 6 months for
all implanted mice immediately after sacrifice. Mice were placed in
the prone position and imaged using a 1.5 T clinical scanner (Avanto,
Siemens Healthcare, Belgium) with a 16-channel body coil placed above
the animals. Based on prior experience [10], two relevant sequences
were acquired: T1 (TR = 2300 ms, TE = 3.05 ms) weighted images, with
a 256 mm field of view with a 265 × 265 matrix, a section thickness
of 1 mm, and 7.3 mm spacing; and susceptibility weighted images (SWIs),
with a 230 × 230 mm matrix on a 230 mm field of view.
Radiologic/Pathologic
Evaluation
Radiologic–pathologic
correlation was performed following mice sacrifice at 6 months. Radiology
images of all scans for all mice were reviewed for conspicuity by
three readers (SNG, EBD, and HL) in consensus and then subsequently
compared over time on mouse-by-mouse and marker-by-marker bases. For
the purpose of evaluating clinically acceptable conspicuity of the
polymer markers, a five-point scale was used to determine the visualization
quality of the markers and the degree of resorption, until deemed
not useful clinically. The categories of the scale were as follows:
(1) baseline (post insertion); (2) mild resorption (clearly visible);
(3) substantial resorption (<50% of initial marker); (4) near-total
elimination (barely visible); and (5) total elimination (Figure ). The baseline,
mild, and substantial resorptions were regarded as clinically acceptable,
and the fourth and fifth (near-total elimination and total elimination)
were not considered adequate for clinical use.
Figure 1
Representative serial
axial CT scans of the Lipiodol-containing
polymer rod marker with grading of conspicuity over time. Arrows point
to polymer C (PCL (polycaprolactone)). Progressive resorption and
polymer degradation are noted. Grades of conspicuity (red numbers):
1, baseline; 2, mild decrease; 3, substantial decrease; 4, near-total
elimination; and 5, total elimination.
Representative serial
axial CT scans of the Lipiodol-containing
polymer rod marker with grading of conspicuity over time. Arrows point
to polymer C (PCL (polycaprolactone)). Progressive resorption and
polymer degradation are noted. Grades of conspicuity (red numbers):
1, baseline; 2, mild decrease; 3, substantial decrease; 4, near-total
elimination; and 5, total elimination.Post sacrifice, gross and histopathology specimens were extracted
by resecting the region of the marker placement. The specimens were
stained with hematoxylin–eosin. All tissue specimens were examined
and evaluated for polymer degradation, iron retention, and inflammatory
response.Representative remaining identified material in the
region of marker
implantation was sent for biochemical analysis including an iodine/iron
analysis by four separate analyses. This included energy-dispersive
X-ray analysis (EDX), an X-ray technique used to identify the elemental
composition of the materials. The EDX instrument with scanning electron
microscopy (Quanta 200, FEI Company) was equipped with an EDAX detector.[12] To prevent burning, specimens were sputter-coated
with palladium at 40 mV for 40 s prior to analysis. Scanning electron
microscopy (SEM) (VEGA3, TESCAN) was used for imaging and analyzing
the size and morphology of specimens.[13] The specimens were visualized under a vacuum (upper limit of 6 ×
10–6 mbar), and images were taken with a beam excitation
energy of 20 kV. Next, Fourier transform infrared spectroscopy (FTIR)
spectra were recorded with a ThermoScientific FTIR spectrometer (Smart
iTR Nicolet iS10 FTIR) with a diamond crystal.[14] A 5–10 mg sample was placed in a crystal window,
and the spectrum was recorded. The scanning range was 400–4000
cm–1 with a resolution of 4 cm–1. The number of scans for each sample was set to 10. Finally, nuclear
magnetic resonance spectroscopy (1H NMR) was performed
using a Varian Mercury 300 MHz NMR spectrometer. A 10 mg polymer sample
was dissolved in 2 mL of deuterated chloroform (CDCl3). The sample
solution was transferred to NMR tubes of 5 mm diameter.
Statistics
For each arm, results were analyzed for
each time point comparing the resorption rates for each marker in
each mouse, using multivariate ANOVA with t-tests performed for various
time points if p < 0.05. Additionally, multiple
Kaplan–Meier plots with a 95% CI were constructed using the
transition between the different grades as the defining endpoint event
using MedCalc Statistical Software version 19.5.3 (MedCalc Software
Ltd., Ostend, Belgium) for analysis.Approval for this study
was given by the Animal Care and Use Committees of The Hebrew University
and the National Council for Animal Experimentation, Israel, in accordance
with National Institutes of Health guidelines.
Results and Discussion
Eight mice survived to the 6-month study
endpoint, with one mouse expiring at week 16 upon anesthetic injection
(autopsy showed no pathologic abnormalities or signs of infection
or excessive inflammation). Conspicuity on CT, based on the grading
system described in the methods section, is
summarized in Table and Figure a,b.
All polymers had similar maximal conspicuity at baseline and showed
evidence of degradation over the 6-month study period. However, significant
differences in the rate of degradation were observed among the polymers
overall (ANOVA; p < 0.01), particularly at 8 weeks
(p < 0.027). Overall, Polymer A began to demonstrate
diminution of conspicuity at 4 weeks. However, all markers with this
compound remained sufficiently perceivable (grades 1–3) for
six weeks. Subsequently, a decline to inadequate levels of visualization
(grades 4–5) was seen in 44% by 14 weeks (Figure a,b). Polymer B retained excellent
conspicuity for at least 8 weeks (grades 1–2), with all animals
demonstrating clinical relevance (grades 1–3) for up to 12
weeks. Thereafter, a rapid decline in conspicuity was noted, rendering
the markers not clinically usable. Polymer C demonstrated rapid loss
of contrast visualization by the 4th week. At six months, there was
no visibility of Polymer A in all animals, borderline visibility (grade
3) in 33% of Polymer B, and 11% in Polymer C. As anticipated, the
control polymer was not visualized on CT discreetly from the subcutaneous
tissues.
Table 1
Grading of Degree of Contrast Resorption
on CT for the Three Polymersa (Numbers Represent
Percentages)
time
post implantation
polymer A (PLGA 50:50)
resorption
day 1
4 weeks
6 weeks
8 weeks
10 weeks
12 weeks
14 weeks
16 weeks
24 weeks
1. baseline
100
56
44
22
0
0
0
0
0
2. mild
0
22
33
33
56
11
11
0
0
3. substantial
0
22
22
33
22
67
44
56
0
4. near total
0
0
0
11
11
0
22
11
56
5. total
0
0
0
0
11
22
22
33
33
polymer B (PLGA 75:25)
day 1
4 weeks
6 weeks
8 weeks
10 weeks
12 weeks
14 weeks
16 weeks
24 weeks
1. baseline
100
89
22
11
0
0
0
0
0
2. mild
0
11
78
89
67
33
0
0
0
3. substantial
0
0
0
0
33
67
56
44
33
4. near total
0
0
0
0
0
0
44
33
0
5. total
0
0
0
0
0
0
0
22
56
polymer C (PCL)
day 1
4 weeks
6
weeks
8 weeks
10 weeks
12
weeks
14 weeks
16 weeks
24 weeks
1. baseline
100
22
22
22
0
0
0
0
0
2. mild
0
44
44
33
44
22
11
11
0
3. substantial
0
11
0
11
11
22
0
0
11
4. near total
0
11
0
0
0
11
22
22
11
5. total
0
11
33
33
44
44
67
67
67
Numbers
represent percentages.
Figure 2
(A, B). Changes in conspicuity over time for three polymers: Polymer
A–PLGA 50:50, Polymer B–PLGA 75:25, and Polymer C–PCL.
PCL, polycaprolactone; PLGA, poly(lactic-co-glycolic
acid). (A) Results for all mice graphically. (B) Plot of the % of
mice with visible marker over time, for the three polymers. Resorption
was defined as reaching a score of 4 (near-total elimination), the
point at which visualization is challenging. Statistically significant
differences among the polymers are noted (p <
0.0001). The categories of the scale were as follows: (1), baseline
(post insertion); (2), mild resorption (clearly visible); (3), substantial
resorption (<50% of the initial marker); (4), near-total elimination
(barely visible); and (5), total elimination.
(A, B). Changes in conspicuity over time for three polymers: Polymer
A–PLGA 50:50, Polymer B–PLGA 75:25, and Polymer C–PCL.
PCL, polycaprolactone; PLGA, poly(lactic-co-glycolic
acid). (A) Results for all mice graphically. (B) Plot of the % of
mice with visible marker over time, for the three polymers. Resorption
was defined as reaching a score of 4 (near-total elimination), the
point at which visualization is challenging. Statistically significant
differences among the polymers are noted (p <
0.0001). The categories of the scale were as follows: (1), baseline
(post insertion); (2), mild resorption (clearly visible); (3), substantial
resorption (<50% of the initial marker); (4), near-total elimination
(barely visible); and (5), total elimination.Numbers
represent percentages.Kaplan–Meier
plots for the three polymers were significant,
with a 95% CI (p < 0.0001, χ2 = 23.25) for resorption from grade 3 (substantial resorption, but
still clinically visible) to grade 4 (trace visualization, deemed
challenging for good visualization in clinical practice) and additionally
from grade 4 to grade 5 (trace visualization vs total resorption)
(p < 0.0001, χ2 = 32.58). Transitions
from grades 1–2 and 2–3 were not statistically significant
(p > 0.05).At
24 weeks, there was a susceptibility
signal at the injection site on T1 and to an even greater degree on
the SWI images (Figure ), without significant distortion of the image at all insertion sites,
even in mice where there was total elimination of conspicuity on CT.
No MR signal was detected at the injection site of the control marker.
Figure 3
(A–C).
CT/MR gross pathologic correlation for Polymer B
(PLGA 75:25). (A) CT demonstrates total elimination of the polymer.
(B) T1 and SWI show susceptibility artifact, causing a minor local
distortion of the image. (C) Gross pathology shows a residual tattooing
of the soft tissue with a <1 mm area of the amorphous material
(long arrow), with a smaller region of the residual marker noted by
the short arrow. PLGA, poly(lactic-co-glycolic acid).
(A–C).
CT/MR gross pathologic correlation for Polymer B
(PLGA 75:25). (A) CT demonstrates total elimination of the polymer.
(B) T1 and SWI show susceptibility artifact, causing a minor local
distortion of the image. (C) Gross pathology shows a residual tattooing
of the soft tissue with a <1 mm area of the amorphous material
(long arrow), with a smaller region of the residual marker noted by
the short arrow. PLGA, poly(lactic-co-glycolic acid).
Radiologic–Pathologic Correlation
Gross pathologic
inspection demonstrated marked dissolving of the markers in all cases
where no Lipiodol was detected at CT. Residual “blobs”
(2–3 mm) were identified for those four injection sites with
a residual CT signal (Figure ). In the remaining 20 injection sites with total or near-total
elimination, there was a 1–2 mm region of a blackened tattoo-like
appearance of the soft tissues surrounding the initial implant site
that contained clusters of highly pigmented macrophages (Figure ). Histopathologically,
the insertion site was surrounded by several layers of fibroblasts,
representing a characteristic inflammatory response similar to that
previously reported [10]. Control injection sites had no pockets of
macrophages and only limited inflammatory reaction adjacent to fat
and muscle.
Figure 4
(A–C). Histopathology. Hematoxylin–eosin staining
of a sample of Polymer B at 24 weeks. (A) (×20) demonstrates
partially resolved polymer residue adjacent to muscle tissue. A small
rim of inflammatory cells surrounds the residual polymer (yellow arrow).
(B) Higher power (×40) in a different mouse demonstrates a dark
granular pigmented appearance caused by iron deposits within macrophages.
(C) Control mouse, demonstrating inflammatory response (yellow arrow)
surrounding the insertion site, with no apparent residual polymer.
(A–C). Histopathology. Hematoxylin–eosin staining
of a sample of Polymer B at 24 weeks. (A) (×20) demonstrates
partially resolved polymer residue adjacent to muscle tissue. A small
rim of inflammatory cells surrounds the residual polymer (yellow arrow).
(B) Higher power (×40) in a different mouse demonstrates a dark
granular pigmented appearance caused by iron deposits within macrophages.
(C) Control mouse, demonstrating inflammatory response (yellow arrow)
surrounding the insertion site, with no apparent residual polymer.The sizes of the markers obtained after 6 months
were small and
weighed ∼1 mg for Polymer A and Polymer B markers, whereas
Polymer C markers weighed 3–5 mg. SEM for iron nanoparticles
demonstrated scattered clusters of iron nanoparticles over the surface
of the markers (Figure ). The detection of the iron content by EDX was not possible for
any of the markers. Since the amount of marker clips recovered after
6 months in vivo studies was very low, the total iron content in the
recovered materials was below the minimal detectable elemental concentration.
On the other hand, iodine was detected in Polymer A and Polymer B
markers, whereas the Polymer C marker did not show any sign of iodine
(Figure ).
Figure 5
Detection of
iron oxide nanoparticles using SEM analysis of the
markers after 6 months in vivo study, (a) PCL (Polymer C), (b) PLGA
50:50 (Polymer A), and (c) PLGA 75:25 (Polymer B) markers with Lipiodol
and iron oxide after 6 months in vivo degradation. (d) Original iron
oxide nanoparticles.
Figure 6
Elemental analysis of
PCL (Polymer C) did not detect Lipiodol;
however, it was seen in (b) PLGA 50:50 (Polymer A) after 6 months
degradation in vivo (red-dotted box and arrow). PLGA 75:25 (Polymer
B) polymer markers, not depicted here, demonstrated a profile similar
to Polymer A.
Detection of
iron oxide nanoparticles using SEM analysis of the
markers after 6 months in vivo study, (a) PCL (Polymer C), (b) PLGA
50:50 (Polymer A), and (c) PLGA 75:25 (Polymer B) markers with Lipiodol
and iron oxide after 6 months in vivo degradation. (d) Original iron
oxide nanoparticles.Elemental analysis of
PCL (Polymer C) did not detect Lipiodol;
however, it was seen in (b) PLGA 50:50 (Polymer A) after 6 months
degradation in vivo (red-dotted box and arrow). PLGA 75:25 (Polymer
B) polymer markers, not depicted here, demonstrated a profile similar
to Polymer A.1H NMR spectroscopy
measurements were performed to identify
Lipiodol in the polymer blend. It should be noted that Lipiodol is
not a pure compound and hence its exact chemical structure is still
unknown. Lipiodol is an iodinated (480 mg iodine/ml) and ethylated
ester of poppy seed oil.[15] However, the
signals in its 1H NMR spectrum can be used to identify
its presence in the polymer clips. The characteristic peaks at 0.96
ppm, 2.25 ppm, and 4.27 ppm have been used to distinguish Lipiodol
from the polymers in the clip formulation.[16] NMR analysis of the markers also confirmed the presence of Lipiodol
remaining in the samples after 6 months. However, 1H NMR
splitting of the polymer and Lipiodol could not be determined in the
marker samples because of excess dissolved tissue samples along with
iron oxide nanoparticles. Further, for Polymer C markers, no NMR splitting
at 0.96 ppm was observed. For Polymer A markers, no characteristic
NMR signals were detected, whereas Polymer B markers showed weak but
characteristic peaks of Lipiodol at 0.96 ppm (Figure ).
Figure 7
1H NMR of (a) PLGA 75:25 (Polymer
B) markers with Lipiodol
and iron oxide after 6 months in vivo degradation. Polymer B markers
showed weak but characteristic peaks of iodine at 0.96 ppm. (b) FTIR
analyses of PLGA 50:50 (black) and PLGA 75:25 (red) polymer markers
with Lipiodol and iron oxide after 6 months in vivo degradation.
1H NMR of (a) PLGA 75:25 (Polymer
B) markers with Lipiodol
and iron oxide after 6 months in vivo degradation. Polymer B markers
showed weak but characteristic peaks of iodine at 0.96 ppm. (b) FTIR
analyses of PLGA 50:50 (black) and PLGA 75:25 (red) polymer markers
with Lipiodol and iron oxide after 6 months in vivo degradation.Finally, FTIR also detected the presence of Lipiodol
in the Polymer
B markers (Figure ). Absorption bands at ∼ 1744 cm–1 are assigned
to C=O stretching, which is similar to the blank PLGA clips.[17] The FTIR spectra of Lipiodol resemble the FTIR
of the poppy seed oil.[18] Like polymers,
the regions of 1700–1800 cm–1 for C=O
stretching existed in Lipiodol. The visible characteristic absorbance
that distinguishes Lipiodol from polymer spectra is only possible
at 2854 cm–1, which is due to symmetric C–H
from phythyl chains (tocopherols) and the band at ∼1371 cm–1 for the O–CH2 groups[19] (Figure ).In this study, we evaluated three potential degradable
polymers
formulated and designed to be used as implantable markers visible
by multimodality imaging in an in vivo model and note substantially
different rates of degradation for different formulations. It is well
known that the higher glycolic acid content of the PLGA polymer makes
the polymer degrade faster.[20] However,
it was unknown until seen that the PLGA 50:50 degradation also led
to a decrease in the sensitivity of the detection of the implants
in vivo. On the other hand, Polymer B (PLGA 75:25) achieved the slowest
resorption and generated the longest acceptable conspicuity on CT—at
least until week 12 post insertion. Hence, it is likely the most suitable
for clinical scenarios such as radiation therapy where a 3-month period
of visualization on CT and MRI may be considered ideal. Radiation
planning for most tumors (such as breast and liver cancers) falls
within the 12-week window. This conspicuity may also be beneficial
for CT-guided procedures, such as thermal and nonthermal ablation
of tumors and image-guided biopsies.[21−23] While the extended 12-week
conspicuity achieved with Polymer B may be beneficial in some clinical
instances, some, such as breast surgery, may favor a shorter, 6-week,
degradation period, as attained with Polymer A. Yet, the short-lived
conspicuity of Polymer C may render it practical only for procedures
requiring immediate use (i.e., within days of the insertion).It is important to note that even after CT conspicuity is lost,
MR conspicuity was maintained even at 6 months, likely due to macrophage
iron retention.[24] This is also evident
on the gross pathology specimens as a tattooing phenomenon and the
scanning electron microscopy results, apparently independent of polymer
iodine content. Furthermore, regarding retention of iodine, in Polymers
A and B, the iodine-containing Lipiodol bonded, albeit with a decrease
in marker size and reabsorption over time. On the contrary, Polymer
C was unable to hold the Lipiodol, and thus, the oily Lipiodol leaked
out over time from the polymer matrix. This was observed in our study
as a decline in the contrast visualization of Polymer C markers on
CT by week 4 post implantation. The Lipiodol blends with the three
polymers were uniform and stable for months when stored at 4–25
°C. However, when placed in phosphate buffer pH 7.4 at 37 °C,
the PCL–Lipiodol blend deformed, after two weeks, into an oil
droplet, while PLGA rods retained their shape.In addition to
the polymer composition, the payload may also be
hypothetically tailored for specific clinical indications. If a marker
is needed to be visible for a long-term medical situation and specifically
requires visualization on MRI, then the desired polymer may contain
an iron compound. Additionally, iron is beneficial because frank metallic
markers will generally have a large artifact on MRI, but theoretically
the amount of iron can be altered to allow MRI visualization while
minimizing susceptibility artifact. Indeed, iron’s effects
are much less pronounced on T1 gradient echo sequences than on SWI
images. Thus, if needed, radiation planning may be performed based
on MR images using T1 and SWI sequences. However, if the requirement
is for a short-term condition and for CT only, a polymer containing
only Lipiodol may likely be sufficient.The histopathologic
examination showed dissolution of the polymer
marker, with minimal expected inflammatory response, thus adding to
the accumulating data on the safety profile of the polymers. Control
sites did not demonstrate any significant finding beyond minimal inflammatory
changes. Thus, it is likely that the iron deposits seen within macrophages
at the insertion site, although they contribute to the visualization
on MRI even after 24 weeks, are responsible for the mild inflammation.Limitations of this study include the fact that we evaluated predominantly
CT findings, with only one time point for MRI. Although we have previously
documented short-term ultrasonographic conspicuity, long-term US was
not performed in this small-animal model as US conspicuity under these
superficial conditions is not particularly clinically relevant. Further
studies with larger animals may evaluate US conspicuity, as well.
Additionally, the use of healthy male mice in this study should not
affect the subcutaneous degradation profile of the polymeric implants.
Future studies will need to address the rate of degradation in female
breast tissues, as well as in pathologic tissues. MR was evaluated
only at the final time point, but the fact that the contrast signal
remained and early 2-week studies suggest that adequate conspicuity
is achievable for this compound (0.2% iron oxide, as in all polymers
in this study).
Conclusions
In conclusion, we were
able to identify a group of degradable biocompatible
polymers, potentially suitable for clinical use involving CT up to
12 weeks and in MRI for six months. Polymers may be tailored for specific
clinical settings, including modality and length of required visualization.
Additional studies are planned to refine the formulations of the polymers
for optimal balance between visibility and required duration of action
versus degree and rate of degradation. Yet, further evaluation is
required to examine the long-term conspicuity and degradation properties
in larger animal models prior to embarking upon clinical studies.
Authors: Meghan G Lubner; Christopher L Brace; Tim J Ziemlewicz; J Louis Hinshaw; Fred T Lee Journal: Semin Intervent Radiol Date: 2013-03 Impact factor: 1.513