Association of Race With Pulse Oximetry Accuracy in Hospitalized Children.

Erica Andrist1, Mark Nuppnau2, Ryan P Barbaro1, Thomas S Valley2, Michael W Sjoding2.   

Abstract

Entities:  

Mesh:

Substances:

Year:  2022        PMID: 35357460      PMCID: PMC8972025          DOI: 10.1001/jamanetworkopen.2022.4584

Source DB:  PubMed          Journal:  JAMA Netw Open        ISSN: 2574-3805


× No keyword cloud information.

Introduction

Oxygen saturation measured by pulse oximetry (SpO2) is used for triage and management of acute illness in hospitalized children. Adult data indicate that pulse oximetry does not detect hypoxemia as frequently in Black patients as in White patients.[1,2] We aimed to investigate the frequency of occult hypoxemia in children and assess variability by patient race.

Methods

This cross-sectional study conducted at a single center in Ann Arbor, Michigan, analyzed patients aged 17 years or younger admitted between January 1, 2015, and December 31, 2020, who self-identified or were identified by a parent as Black or White. We extracted patient race, age, SpO2, arterial blood gas (ABG) data, respiratory support, vasoactive support, and diagnosis information from electronic health records. The University of Michigan Medical School Institutional Review Board exempted the study from approval and informed consent because it constituted secondary research. This study followed the STROBE reporting guideline. We compared SpO2 readings with arterial oxygen saturation (SaO2) readings directly measured by ABG if the ABG measurement was performed within 10 minutes at the same respiratory support. Exclusion criteria are presented in the eTable in the Supplement. We used the Vasoactive Infusion Score[3] to adjust for cardiovascular severity of illness. We evaluated the mean pulse oximeter bias (SpO2 – SaO2) and frequency of occult hypoxemia (SpO2 ≥92% and SaO2 <88%) by patient race. Statistical analyses were conducted using Stata, version 16 (StataCorp LLC). Significance was set at 2-sided P < .05 (eMethods and eAppendix in the Supplement).

Results

We analyzed 9023 SpO2-SaO2 pairs from 1061 total patients (878 [82.8%] White; 183 [17.2%] Black; 572 [53.9%] male; mean [SD] age, 7 [6] years) (Table). The median time between SpO2 and SaO2 readings was 4 minutes (range, 0 to 9 minutes) for both groups. The overall mean (SD) pulse oximeter bias was 3.5% (5.0%) among White patients and 4.3% (5.0%) among Black patients (P < .001). The frequency of occult hypoxemia was 5.8% (95% CI, 4.6%-7.3%) among White patients and 9.6% (95% CI, 6.3%-14.5%) among Black patients (Figure). After adjusting for age, sex, SpO2, and Vasoactive Infusion Score, the odds ratio for occult hypoxemia among SpO2 measurements for Black patients was 2.16 (95% CI, 1.36-3.44) compared with White patients. At the patient level, 134 of 860 White patients (15.6%; 95% CI, 13.3%-18.2%) and 38 of 180 Black patients (21.1%; 95% CI, 15.7%-27.7%) had occult hypoxemia episodes. After adjusting for the same covariates, the odds ratio for occult hypoxemia was 1.79 (95% CI, 1.07-3.02) for Black patients compared with White patients.
Table.

Characteristics of the Study Sample by Race

CharacteristicHospitalizations, No. (%)aP valueb
WhiteBlackTotal
Hospitalizationsc934 (83.4)186 (16.6)1120 (100)NA
Unique patients878 (82.8)183 (17.2)1061 (100)NA
Sex
Female426 (45.6)90 (48.4)516 (46.1).54
Male508 (54.4)96 (51.6)604 (53.9)
Patient age
<6 mo152 (16.3)33 (17.7)185 (16.5).98
6 to <12 mo83 (8.9)16 (8.6)99 (8.8)
12 to <24 mo75 (8.0)12 (6.5)87 (7.8)
2 to <5 y152 (16.3)29 (15.6)181 (16.2)
5 to <12 y207 (22.2)42 (22.6)249 (22.2)
12 to 17 y265 (28.4)54 (29.0)319 (28.5)
Maximum respiratory support
None35 (3.7)8 (4.3)43 (3.8).002
Supplemental oxygen313 (33.5)36 (19.4)349 (31.2)
Noninvasive positive pressure43 (4.6)11 (5.9)54 (4.8)
Invasive mechanical ventilation543 (58.1)131 (70.4)674 (60.2)
Maximum VIS
0-5551 (59.0)115 (61.8)666 (59.5).51
6-10120 (12.8)18 (9.7)138 (12.3)
11-20111 (11.9)26 (14.0)137 (12.2)
>20152 (16.3)27 (14.5)179 (16.0)
SpO2-SaO2 pairs, No./total No. (%)d7018/9023 (77.8)2005/9023 (22.2)9023/9023 (100)NA
SpO2 values, No./total No. (%)
<88%626/7018 (8.9)248/2005 (12.4)874/9023 (9.7).71
≥88%6392/7018 (91.1)1757/2005 (87.6)8149/9023 (90.3)
SpO2 – SaO2, mean (SD)3.5 (5.0)4.3 (5.0)3.7 (5.0)<.001
SaO2 <88% and SpO2 ≥88%, No./total No. (%)
SpO2 of 88%-91%188/334 (56.3)74/106 (69.8)262/440 (59.6)NA
SpO2 of 92%-96%248/1845 (13.4)127/433 (29.3)375/2278 (16.5)
SpO2 >96%104/4213 (2.5)32/1218 (2.6)136/5431 (2.5)
Occult hypoxemia, No./total No. (%)e352/6058 (5.8)159/1651 (9.6)511/7709 (6.6).03

Abbreviations: NA, not applicable; SaO2, arterial oxygen saturation; SpO2, oxygen saturation measured by pulse oximetry; VIS, Vasoactive Infusion Score.

Some percentages may not sum to 100% owing to rounding.

P values for the patient-level analysis were calculated using the χ2 test or analysis of variance; P values for the SpO2 measurement-level analysis were calculated using mixed models to adjust for clustering within patients.

Sex, age, maximum respiratory support, and VIS are reported for each hospitalization.

The sample included SpO2 and SaO2 directly measured by arterial blood gas analysis. Pairs were measured within 10 minutes at the same level of respiratory support.

Occult hypoxemia was defined as SaO2 less than 88% in the presence of SpO2 of 92% or greater.

Figure.

Racial Differences in the Proportion of Oxygen Saturation Measured by Pulse Oximetry (SpO2)–Arterial Oxygen (SaO2) Pairs Demonstrating Occult Hypoxemia

Vertical lines indicate 95% CIs.

Abbreviations: NA, not applicable; SaO2, arterial oxygen saturation; SpO2, oxygen saturation measured by pulse oximetry; VIS, Vasoactive Infusion Score. Some percentages may not sum to 100% owing to rounding. P values for the patient-level analysis were calculated using the χ2 test or analysis of variance; P values for the SpO2 measurement-level analysis were calculated using mixed models to adjust for clustering within patients. Sex, age, maximum respiratory support, and VIS are reported for each hospitalization. The sample included SpO2 and SaO2 directly measured by arterial blood gas analysis. Pairs were measured within 10 minutes at the same level of respiratory support. Occult hypoxemia was defined as SaO2 less than 88% in the presence of SpO2 of 92% or greater.

Racial Differences in the Proportion of Oxygen Saturation Measured by Pulse Oximetry (SpO2)–Arterial Oxygen (SaO2) Pairs Demonstrating Occult Hypoxemia

Vertical lines indicate 95% CIs.

Discussion

In this study, 21.1% of Black children experienced arterial hypoxemia despite a normal SpO2 reading; occult hypoxemia occurred more frequently than in White children. Among SpO2-Sao2 pairs, the odds of the SpO2 reading failing to detect arterial hypoxemia were more than twice as great if the pair had come from a Black child. Underrecognition and undertreatment of hypoxemia in Black children may contribute to racial disparities in respiratory illness outcomes.[4,5,6] We observed higher mean pulse oximeter bias among all children than has been reported in adults.[2] This study has limitations. Because pediatric ABG values are rarely obtained outside critical care settings, all study patients were likely critically ill. Clinically significant changes could have occurred in the 10-minute window between obtaining SpO2 and SaO2 values. Black children required more respiratory support than White children, potentially confounding the results, although the findings were not different in adjusted analysis. In addition, self-reported race encompasses many phenotypes. Future research may explore more nuanced associations between pulse oximetry accuracy and skin pigmentation and whether pulse oximeter bias is associated with outcomes. Further evaluation of racial differences in pediatric pulse oximetry accuracy in children and investigation of potential clinical outcomes are warranted.
  6 in total

1.  Risk Factors for Respiratory Decompensation Among Healthy Infants With Bronchiolitis.

Authors:  Nina M Dadlez; Nora Esteban-Cruciani; Asama Khan; Lindsey C Douglas; Yi Shi; William N Southern
Journal:  Hosp Pediatr       Date:  2017-09

2.  Racial Bias in Pulse Oximetry Measurement Among Patients About to Undergo Extracorporeal Membrane Oxygenation in 2019-2020: A Retrospective Cohort Study.

Authors:  Valeria S M Valbuena; Ryan P Barbaro; Dru Claar; Thomas S Valley; Robert P Dickson; Steven E Gay; Michael W Sjoding; Theodore J Iwashyna
Journal:  Chest       Date:  2021-09-27       Impact factor: 10.262

3.  Trends in racial disparities for asthma outcomes among children 0 to 17 years, 2001-2010.

Authors:  Lara J Akinbami; Jeanne E Moorman; Alan E Simon; Kenneth C Schoendorf
Journal:  J Allergy Clin Immunol       Date:  2014-08-01       Impact factor: 10.793

4.  Vasoactive-inotropic score as a predictor of morbidity and mortality in infants after cardiopulmonary bypass.

Authors:  Michael G Gaies; James G Gurney; Alberta H Yen; Michelle L Napoli; Robert J Gajarski; Richard G Ohye; John R Charpie; Jennifer C Hirsch
Journal:  Pediatr Crit Care Med       Date:  2010-03       Impact factor: 3.624

5.  Racial Bias in Pulse Oximetry Measurement.

Authors:  Michael W Sjoding; Robert P Dickson; Theodore J Iwashyna; Steven E Gay; Thomas S Valley
Journal:  N Engl J Med       Date:  2020-12-17       Impact factor: 176.079

6.  Rates of Influenza-Associated Hospitalization, Intensive Care Unit Admission, and In-Hospital Death by Race and Ethnicity in the United States From 2009 to 2019.

Authors:  Alissa C O'Halloran; Rachel Holstein; Charisse Cummings; Pam Daily Kirley; Nisha B Alden; Kimberly Yousey-Hindes; Evan J Anderson; Patricia Ryan; Sue Kim; Ruth Lynfield; Chelsea McMullen; Nancy M Bennett; Nancy Spina; Laurie M Billing; Melissa Sutton; William Schaffner; H Keipp Talbot; Andrea Price; Alicia M Fry; Carrie Reed; Shikha Garg
Journal:  JAMA Netw Open       Date:  2021-08-02
  6 in total
  2 in total

Review 1.  Artificial and human intelligence for early identification of neonatal sepsis.

Authors:  Brynne A Sullivan; Sherry L Kausch; Karen D Fairchild
Journal:  Pediatr Res       Date:  2022-09-20       Impact factor: 3.953

2.  Racial bias and reproducibility in pulse oximetry among medical and surgical inpatients in general care in the Veterans Health Administration 2013-19: multicenter, retrospective cohort study.

Authors:  Valeria S M Valbuena; Sarah Seelye; Michael W Sjoding; Thomas S Valley; Robert P Dickson; Steven E Gay; Dru Claar; Hallie C Prescott; Theodore J Iwashyna
Journal:  BMJ       Date:  2022-07-06
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.