| Literature DB >> 35356501 |
Jiaxing Huang1,2, Jiawei Wang1,2, Leilei Qin1,2, Bo Zhu1,2, Wei Huang1,2, Ning Hu1,2.
Abstract
Background: Synovial fluid biomarkers have been found to improve the diagnosis of chronic periprosthetic joint infection (PJI); however, no "gold standard" exists yet. Interleukin-4 (IL-4) and polymorphonuclear cell (neutrophil) count in the synovial fluid are crucial in mediating local inflammation during bacterial infections and could be valuable biomarkers for PJI.Entities:
Keywords: chronic periprosthetic joint infection; diagnosis; interleukin 4; polymorphonuclear cell percentage; synovial fluid
Year: 2022 PMID: 35356501 PMCID: PMC8959493 DOI: 10.3389/fsurg.2022.843187
Source DB: PubMed Journal: Front Surg ISSN: 2296-875X
The Musculoskeletal society 2013 criteria for defining periprosthetic joint infection.
|
| |
|---|---|
| Major criteria | 1. A sinus tract communicating with the joint; or 2. Two positive periprosthetic cultures with phenotypical identical organisms |
| Minor criteria | 1. Elevated serum C-reactive protein (CRP) AND erythrocyte sedimentation rate (ESR); or 2. Elevated synovial fluid white blood cell (WBC) count OR ++ change on leukocyte esterase strip; or 3. Elevated synovial fluid polymorphonuclear neutrophil percentage (PMN%); or 4. Positive histological analysis of periprosthetic tissue; or 5. A single positive culture |
Demographic data for the study population.
|
|
| ||
|---|---|---|---|
| Age (year) | 70.58 ± 5.26 | 68.98 ± 6.05 | 0.17 |
|
| |||
| Male | 22 (51.2%) | 27 (48.2%) | 0.77 |
| Female | 21 (48.8%) | 29 (51.8%) | |
|
| |||
| Hip | 17 (39.5%) | 31 (55.4%) | 0.12 |
| Knee | 26 (60.5%) | 25 (44.6%) | |
| BMI (kg/m2) | 22.74 ± 3.91 | 23.21 ± 4.66 | 0.59 |
| Survival time of implantation (month) | 67.91 ± 29.06 | 76.82 ± 21.55 | 0.08 |
Chi-square test,
independent Student's t-test; PJI, periprosthetic joint infection.
Analysis of single markers in patients with chronic PJI and aseptic failure.
|
|
| |||
|---|---|---|---|---|
| ESR (mm/h) | Range | 7.20–120.00 | 2.00–58.00 | |
| Median | 35.00 | 22.00 | 0.001 | |
| P25, P75 | 22.00; 50.00 | 17.50; 32.00 | ||
| Mean ± SD | 42.10 ± 26.23 | 24.38 ± 12.80 | ||
| CRP (mg/L) | Range | 5.80–91.20 | 1.25–28.00 | |
| Median | 21.40 | 6.75 | <0.001 | |
| P25, P75 | 13.20; 33.40 | 3.29; 16.00 | ||
| Mean ± SD | 27.31 ± 20.02 | 9.81 ± 7.75 | ||
| SF-IL4 (pg/mL) | Range | 1.20–14.10 | 0.09–2.88 | |
| Median | 3.30 | 1.10 | <0.001 | |
| P25, P75 | 2.20; 9.7 | 0.55; 1.28 | ||
| Mean ± SD | 5.57 ± 4.37 | 1.00 ± 0.53 | ||
| SF-PMN% | Range | 60.32–96.65 | 23–89.20 | |
| Median | 87.58 | 56.95 | <0.001 | |
| P25, P75 | 83.50, 90.50 | 53.40, 56.95 | ||
| Mean ± SD | 86.15 ± 7.31 | 59.30 ± 15.53 |
SD, standard deviation;
Mann-Whitney U test.
Figure 1ROC curves of the studied markers for chronic PJI diagnosis. (A–D) ROC curves of serum ESR, CRP, SF-IL4, and SF-PMN%. (E) Comparison of these four markers.
AUC, cut-off value, and diagnostic value of each marker for chronic PJI.
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|
| AUC (95% CI) | 0.83 (0.74–0.90) | 0.72 (0.62–0.84) | 0.97 (0.92–0.99) | 0.89 (0.82–0.95) | ||
| Cut-off level | 10 | 30 | 18 | 34 | 1.7 | 75% |
| Sensitivity (95% CI) | 81.4 | 67.4 | 62.79 (46.7–77.0) | 58.14 (42.1–73.0) | 93.02 (80.9–98.5) | 95.35 (84.2–99.4) |
| Specificity (95% CI) | 58.9 | 62.6 | 87.50 (75.9–94.8) | 83.93 (71.7–92.4) | 94.64 (85.1–98.9) | 78.57 (65.6–88.4) |
| PPV | – | – | 79.4 (65.0– 88.9) | 73.5 (59.2–84.2) | 93.0 (81.6–97.6) | 77.4 (67.3–85.0) |
| NPV | – | 75.4 (67.0– 82.10) | 72.3 (64.3 −79.1) | 94.6 (85.6–98.1) | 95.7 (85.0–98.8) | |
| +LR | – | – | 5.02 (2.40– 10.40) | 3.62 (1.9–6.9) | 17.36 (5.8–52.4) | 4.45 (2.7–7.4) |
| –LR | – | – | 0.43 (0.3–0.6) | 0.50 (0.3– 0.7) | 0.07 (0.02–0.2) | 0.06 (0.02–0.2) |
| DOR | – | – | 11.67 | 7.24 | 248.00 | 74.17 |
| Accuracy | 58.6 | 50.5 | 72.7 | 68.7 | 92.9 | 84.8 |
AUC, area under the curve; PPV, positive predictive value; NPV, negative predictive value; +LR, positive likelihood ratio; –LR, negative likelihood ratio; DOR, diagnostic odds ratio.
Diagnostic value of different combinations of markers for chronic PJI diagnosis.
|
| |||
|---|---|---|---|
| Sensitivity | 0.88 | 0.91 | 0.53 |
| Specificity | 0.98 | 0.97 | 0.98 |
| PPV | 0.96 | 0.94 | 0.95 |
| NPV | 0.75 | 0.92 | 0.73 |
| Accuracy | 80.1% | 96% | 78.8% |
PPV, positive predictive value; NPV, negative predictive value.