Christopher Tiburski1, Ferry Anggoro Ardy Nugroho2, Christoph Langhammer1. 1. Department of Physics, Chalmers University of Technology, 412 96 Göteborg, Sweden. 2. Department of Physics and Astronomy, Vrije Universiteit Amsterdam, De Boelelaan 1081, 1081 HV Amsterdam, The Netherlands.
Abstract
The temperature of nanoparticles is a critical parameter in applications that range from biology, to sensors, to photocatalysis. Yet, accurately determining the absolute temperature of nanoparticles is intrinsically difficult because traditional temperature probes likely deliver inaccurate results due to their large thermal mass compared to the nanoparticles. Here we present a hydrogen nanothermometry method that enables a noninvasive and direct measurement of absolute Pd nanoparticle temperature via the temperature dependence of the first-order phase transformation during Pd hydride formation. We apply it to accurately measure light-absorption-induced Pd nanoparticle heating at different irradiated powers with 1 °C resolution and to unravel the impact of nanoparticle density in an array on the obtained temperature. In a wider perspective, this work reports a noninvasive method for accurate temperature measurements at the nanoscale, which we predict will find application in, for example, nano-optics, nanolithography, and plasmon-mediated catalysis to distinguish thermal from electronic effects.
The temperature of nanoparticles is a critical parameter in applications that range from biology, to sensors, to photocatalysis. Yet, accurately determining the absolute temperature of nanoparticles is intrinsically difficult because traditional temperature probes likely deliver inaccurate results due to their large thermal mass compared to the nanoparticles. Here we present a hydrogen nanothermometry method that enables a noninvasive and direct measurement of absolute Pd nanoparticle temperature via the temperature dependence of the first-order phase transformation during Pd hydride formation. We apply it to accurately measure light-absorption-induced Pd nanoparticle heating at different irradiated powers with 1 °C resolution and to unravel the impact of nanoparticle density in an array on the obtained temperature. In a wider perspective, this work reports a noninvasive method for accurate temperature measurements at the nanoscale, which we predict will find application in, for example, nano-optics, nanolithography, and plasmon-mediated catalysis to distinguish thermal from electronic effects.
Entities:
Keywords:
nanoparticles; nanothermometry; palladium hydride; photothermal; plasmonics; sensing; temperature
Metal nanoparticles
have found
application in a wide range of fields owing to their support of localized
surface plasmon resonances (LSPRs).[1,2] All these applications
have in common that they rely on the irradiation of light to unlock
plasmonic functions utilized, for example, for optical sensing,[3,4] treatment of disease,[5,6] photovoltaic devices,[7,8] optical metamaterials,[9−11] dynamic coloring,[12,13] and active plasmonics[14] and to enhance
catalytic reactions.[15−17] At the same time, it is well known that LSPR excitation
leads to significant light absorption, which via coupling
to lattice phonons increases nanoparticle temperature,[18−20] and that such a temperature increase may have both wanted[21−23] and unwanted[24] consequences for specific
applications. One such area where the potential impact of plasmon-induced
optical heating of nanoparticles is highly debated[25−30] and potentially very significant[31] is
plasmon-mediated catalysis. In this field, several different reaction
enhancement mechanisms have been proposed,[31−34] and it remains a great challenge
to distinguish between photothermal- and hot carrier-induced reaction
enhancement.[35] One reason is that commonly
conducted experiments to disentangle these mechanisms rely on probing
the catalytic activity as a function of photon flux. Such experiments
have been demonstrated to yield ambiguous results since several orders
of magnitude change of the photon flux would be needed for rigorous
analysis,[35] which experimentally is nearly
impossible to achieve. Similarly, actual measurements of light-induced
temperature changes in plasmonic systems is very challenging, especially
in more complex systems and if the temperature probes used are not
placed very carefully.[25] This is the consequence
of a combination of factors that include (i) the high thermal conductivity
of metal nanoparticles that leads to rapid thermal equilibration with
their support,[36] (ii) their small thermal
mass that renders macroscopic temperature probes highly invasive,
and (iii) thermal gradients that readily occur, in particular in three-dimensional
samples.[37] As a consequence, direct and noninvasive measurements of plasmonic nanoparticle temperature in situ are highly desirable to resolve this long-standing
challenge in plasmonic technologies in general[38−43] and in plasmon-mediated catalysis in particular.[25−30] Focusing on catalysis, several attempts in this direction have been
reported. They include anti-Stokes radiation and fluorescence spectroscopy
that either require rather complicated equipment or the use of fluorescent
labels,[44−47] or the application of a specific temperature-sensitive thin film
coating to the sample,[48] which renders
it a destructive measurement method. These and other thermometry methods
for the nano- and microscale can achieve in some cases high temporal
resolution of down to 10–9 s, spatial resolutions
of 10–2 μm, or temperature resolution of 10–5 K.[49] However, although
all of these are impressive achievements, a reasonably simple, noninvasive,
and nondestructive means to directly measure the
temperature of metal nanoparticles with good resolution in general,
and of plasmonic nanoparticles under different levels of illumination
in particular, still does not exist.In response, we here discuss
the concept of optical hydrogen nanothermometry,
which enables direct and noninvasive temperature measurements of Pd
nanoparticles under illumination. We utilize the method to unravel
LSPR-induced nanoparticle heating effects at an absolute scale for
different irradiances of incident visible light and nanoparticle arrays
with different surface coverage, in combination with numerical simulations
and theoretical modeling.
Results and Discussion
When Pd,
in general, and its nanoparticles, in particular, are
exposed to hydrogen, they transform into a hydride as a consequence
of dissociative chemisorption of hydrogen molecules on Pd surfaces
and their subsequent diffusion into interstitial Pd lattice sites.[50,51] At low hydrogen pressures, the absorbed hydrogen atoms form a solid
solution in the Pd lattice (α-phase) where they are sparsely
distributed and the H–H interactions are weak.[52] At a critical pressure where hydrogen-induced electronic
and strain-field interactions become sizable, the hydride (β-phase)
nucleates.[52] At this point both phases
(α + β) coexist and the system undergoes a first-order
phase transformation, which gives rise to a distinct “plateau”
in a pressure–composition isotherm (Figure a). This phase transformation significantly
changes the electronic structure and volume of the system,[53−55] which enables its tracking using optical methods in both thin films[56,57] and nanoparticles,[58] where LSPR excitations
are important in the latter case. To this end, it has also been demonstrated
experimentally[58,59] and theoretically from first-principles[60] that various descriptors of the LSPR peak, such
as shifts in the spectral position or, as we will use here, the extinction
difference at two rationally selected wavelengths of self-referenced
spectra,[61] are proportional to the atomic
ratio of hydrogen and Pd and therefore can be used to construct optical
isotherms (Figure a).
Figure 1
Hydrogen nanothermometry concept of using a Van’t Hoff plot
to deduce the temperature of Pd nanoparticles. (a) Pressure–composition
isotherms are a common way of characterizing metal–hydrogen
interactions at different temperatures and enable, among others, the
construction of a phase diagram. Specifically, for the Pd–H
case at hand, a pressure–composition isotherm reveals the relation
between external hydrogen (partial) pressure and the amount of hydrogen
atoms occupying the interstitial lattice sites in the Pd host, i.e., the H/Pd ratio. At low hydrogen pressure
the hydrogen atoms form a solid solution in the Pd lattice (α-phase).
At a critical hydrogen pressure, due to now sizable H–H interaction,
the hydride (β-phase) nucleates and both phases coexist, resulting
in a “plateau” in the isotherm that signifies this
first-order phase transformation. As the key point for our present
work, this phase transition will shift to a different pressure at
different temperature (T1, T2, where T2 > T1). (b) Van’t Hoff plot that depicts the phase
transition pressure of the Pd–H system plotted on a semilogarithmic
scale versus the inverse temperature exhibiting a
linear relation. (c) Determination of an unknown temperature (T) of a Pd nanoparticle system
by correlating the obtained phase transition pressure with the premeasured
Van’t Hoff plot.
Hydrogen nanothermometry concept of using a Van’t Hoff plot
to deduce the temperature of Pd nanoparticles. (a) Pressure–composition
isotherms are a common way of characterizing metal–hydrogen
interactions at different temperatures and enable, among others, the
construction of a phase diagram. Specifically, for the Pd–H
case at hand, a pressure–composition isotherm reveals the relation
between external hydrogen (partial) pressure and the amount of hydrogen
atoms occupying the interstitial lattice sites in the Pd host, i.e., the H/Pd ratio. At low hydrogen pressure
the hydrogen atoms form a solid solution in the Pd lattice (α-phase).
At a critical hydrogen pressure, due to now sizable H–H interaction,
the hydride (β-phase) nucleates and both phases coexist, resulting
in a “plateau” in the isotherm that signifies this
first-order phase transformation. As the key point for our present
work, this phase transition will shift to a different pressure at
different temperature (T1, T2, where T2 > T1). (b) Van’t Hoff plot that depicts the phase
transition pressure of the Pd–H system plotted on a semilogarithmic
scale versus the inverse temperature exhibiting a
linear relation. (c) Determination of an unknown temperature (T) of a Pd nanoparticle system
by correlating the obtained phase transition pressure with the premeasured
Van’t Hoff plot.The occurrence of the
α- to β-phase transformation
depends on the temperature of the system and is shifted to higher
pressures when the temperature increases (Figure a). This temperature dependence is widely
used to construct Van’t Hoff plots (Figure b) to extract the hydride formation enthalpy
and entropy change according to the Van’t Hoff equationwhere P is the is
phase transition
pressure, P0 is atmospheric pressure, and are the change in
enthalpy and entropy,
respectively, R is the universal gas constant, and T is temperature.Here, we propose a different consequence
of this intrinsic temperature
dependence of the two-phase coexistence transition, namely, that a
“reversed” Van’t Hoff analysis can be used to
accurately measure the temperature of the hydrogenated system, provided
it has been appropriately calibrated. Specifically, by first generating
a Van’t Hoff plot at well-defined temperatures, it can subsequently
be used to derive the temperature of the system via the phase transition pressure measured at that particular (unknown)
temperature. Projecting this concept onto the case of interest here,
that is, light-induced heating of plasmonic nanoparticles, it means
that it can be used to accurately determine the absolute temperature of Pd nanoparticles under illumination (Figure c). As the key aspect of this
approach, we highlight that the temperature obtained this way corresponds
to the true nanoparticle temperature since it is
only the Pd particles themselves that undergo the
temperature-dependent hydride formation process. Hence, our concept
enables noninvasive, nondestructive direct measurements of nanoparticle
temperature changes induced by light absorption, as we will demonstrate
below.To demonstrate the hydrogen nanothermometry concept,
we nanofabricated
quasi-random arrays of Pd disks with a nominal diameter of 140 nm
and a height of 25 nm on a 9 × 9 mm2 glass substrate
using hole-mask-colloidal lithography,[62] followed by a thermal annealing step to induce particle recrystallization
and ensure a reproducible response upon subsequent hydrogenation (Figure a; see also Methods).[63] The annealed
sample was then placed in a quartz-tube plug-flow reactor inside a
glass pocket holder with optical access[64] and contacted with a spring-loaded thermocouple on its edge to monitor
its global temperature (Figure S1). To
enable measurements of optical isotherms at constant reactor temperature
at different levels of illumination (and thus different local sample
temperatures), the reactor was heated with constant power via a resistive heating coil; that is, no feed-back loop
for temperature control was used. To compensate for this deliberate
lack of active temperature control and ensure a stable reactor temperature
during the experiments, we thermally equilibrated the system for 150
min prior to each measurement to reach a steady-state temperature
value. During a measurement, we increased the hydrogen concentration
in Ar carrier gas at a constant flow rate of 200 mL/min at atmospheric
pressure. Simultaneously, we recorded self-referenced optical spectra
across the 350–900 nm wavelength range, which means that we
used a spectrum initially measured in pure Ar as the reference for
all subsequently measured spectra.[61] To
construct an optical isotherm, we then extracted the difference in
self-referenced optical extinction, , defined
as the average extinction between
450 and 480 nm subtracted by the average extinction between 730 and
760 nm, and plot it versus hydrogen partial pressure
(Figure b; see Figure S2 for raw data and Figure S3a for an explanation of the self-referencing approach).
Figure 2
Constructing
the Van’t Hoff calibration curve. (a) SEM image
of an as-deposited Pd nanodisk quasi-random array sample fabricated
with hole-mask-colloidal lithography. Inset: close-up image of an
annealed Pd particle with thermally stable crystallinity. (b) Optical
isotherms constructed from corresponding measurements at six different
reactor temperatures (38.3, 42.8, 46.3, 51.4, 55.4, and 57.3 °C).
(c) Inverse temperature plotted against the phase transition pressures
extracted from the six isotherms in (b) resulting in a flipped Van’t
Hoff plot. The extracted ΔH and ΔS values are in good agreement with the literature for the
Pd–H system.[65−67] This Van’t Hoff plot serves as the calibration
curve to deduce the temperature of the Pd nanoparticles upon high-power
illumination. The error bars along the x-axis represent
the 95% confidence interval of the fit functions to define the phase
transition pressure (for details see Figure S6). The error bars along the y-axis considers the
deviation from the set temperature during one measurement cycle (cf. Figure S2b).
Constructing
the Van’t Hoff calibration curve. (a) SEM image
of an as-deposited Pd nanodisk quasi-random array sample fabricated
with hole-mask-colloidal lithography. Inset: close-up image of an
annealed Pd particle with thermally stable crystallinity. (b) Optical
isotherms constructed from corresponding measurements at six different
reactor temperatures (38.3, 42.8, 46.3, 51.4, 55.4, and 57.3 °C).
(c) Inverse temperature plotted against the phase transition pressures
extracted from the six isotherms in (b) resulting in a flipped Van’t
Hoff plot. The extracted ΔH and ΔS values are in good agreement with the literature for the
Pd–H system.[65−67] This Van’t Hoff plot serves as the calibration
curve to deduce the temperature of the Pd nanoparticles upon high-power
illumination. The error bars along the x-axis represent
the 95% confidence interval of the fit functions to define the phase
transition pressure (for details see Figure S6). The error bars along the y-axis considers the
deviation from the set temperature during one measurement cycle (cf. Figure S2b).To now construct a Van’t Hoff plot that serves as
a “calibration
curve” for unknown particle temperature determination via hydrogen nanothermometry, we measured optical isotherms
at six different set reactor temperatures (Figure b and Figure S4), for which we used a low optical power (8.3 mW) halogen lamp to
minimize any light-induced temperature increase. From these isotherms,
we then extracted the phase transition pressure at each temperature[67,68] (Figure S5) and plot them versus the inverse temperature on a semilogarithmic scale (Figure c). As a result, we can extract
values of −34.5 ± 0.3 kJ/(mol H2) and −86.6
± 0.9 J/(K mol H2) for the change in enthalpy, ΔH, and entropy, ΔS, respectively,
which are in good agreement with the literature[65−67] and thus validate
our general procedure. Here we note that the phase transition (Figure b) is not an abrupt
change but rather a gradual transition owing to the fact that the
measurements are done on an array of Pd disks. As each particle has
slightly different shapes, sizes, and grain structure, the phase transition
happens at slightly different pressures, leading to a slanted transition
for the ensemble.[67,69]As the next step of our
analysis, we set out to apply the calibrated
response of our sample to determine the temperature increase induced
by illumination with a high-power continuous wave plasma-arc lamp
at different nominal sample temperatures. Specifically, we again measured
optical isotherms but this time using the light from the plasma arc
lamp for obtaining the self-referenced spectra (Figure S3b). To investigate the dependence of light-induced
heating of the Pd particles on both set reactor temperature and optical
power, we measured isotherms at five different set reactor temperatures
(24.6, 28.9, 34, 38.3, and 42.8 °C) and at five different optical
powers (1, 2, 3, 4, 4.8 W corresponding to 0.88, 1.77, 2.65, 3.54,
and 4.24 W/cm2, respectively, measured outside of the reactor)
obtained by defocusing the collimated beam (see Methods). Here we note that we below chose to discuss our results in terms
of optical power, rather than power density, since the power density
at the sample position is difficult to determine exactly due to the
design of the reactor used and since the illuminated area remains
constant in our experiments. Plotting the corresponding isotherms
measured at five set reactor temperatures at 3 W reveals a distinct
shift of the phase transition pressures compared to the reference
measurement made with the low-power halogen lamp (Figure a). This upward shift indicates
a significant temperature increase induced by illumination, which
we can quantify by determining the intersection with the previously
obtained calibration Van’t Hoff plot, resulting in the absolute
nanoparticle temperature (TH, Figure b). Repeating
this procedure systematically for the five set reactor temperatures
and optical powers reveals a consistent and significantly higher temperature
of the Pd nanoparticles upon illumination (Figure c), ranging from 35.5 °C at 1 W to 77.1
°C at 4.8 W. The data show a clear proportionality with an R2 > 0.98. By subtracting these temperatures
with the corresponding set reactor temperatures, an absolute particle
temperature increase, , of up to 38 °C
is obtained upon high-power
illumination (Figure d). These values agree well with the analytical model for the collective
heating in plasmonic arrays developed by Baffou etal. (Figure S7).[70] Furthermore, we notice a linear dependence of
the absolute temperature increase on the set reactor temperature with
an average R2 of 0.99 (Figure S8). We also note that the overall light absorption
efficiency, Qabs, of the system in the
pristine and hydride state changes by only ∼5%, as revealed
by simulations (Figure S9). This finding
means that the difference in light-induced heating when the system
is in these two different phases is negligible for the overall interpretation
of our results. Furthermore, we note that our approach can also be
used to determine the optical power necessary to reach a specific
particle temperature (see Figure S15 and
corresponding discussion S1).
Figure 3
Determining the temperature increase in Pd nanoparticles
upon illumination.
(a) Representative isotherms measured at 3 W optical power for 24.6,
28.9, 34, 38.3, and 42.8 °C set reactor temperatures. A clear
shift in the phase transition pressure compared to the calibration
measurement with the halogen lamp (red arrow) is observed, indicating
a significantly higher particle temperature. (b) Prerecorded calibration
Van’t Hoff plot (gray line, cf. Figure c) used to determine the absolute
particle temperature upon illumination with higher optical powers
by determining the intersects of the phase transition pressures with
the calibration curve. (c) Extracted absolute particle temperatures
(TH) as a function of optical
power for five different set reactor temperatures and the temperatures
at 0 W (Tt/c). A clear trend of increasing
particle temperature with increasing optical power can be seen for
all reactor temperatures. The error in TH is derived from the 95% confidence interval in the
linear fit to the Van’t Hoff calibration curve, and the error
in the optical power is estimated to be ±0.2 W. (d) Temperature
increase, ΔT, in Pd nanoparticles upon illumination
as a function of optical power for the different set reactor temperatures.
The values are obtained by subtracting the set reactor temperatures
from TH. The error is obtained
from the 95% confidence interval in the linear fit to the Van’t
Hoff calibration curve combined with the standard deviation of the
set reactor temperature measurement.
Determining the temperature increase in Pd nanoparticles
upon illumination.
(a) Representative isotherms measured at 3 W optical power for 24.6,
28.9, 34, 38.3, and 42.8 °C set reactor temperatures. A clear
shift in the phase transition pressure compared to the calibration
measurement with the halogen lamp (red arrow) is observed, indicating
a significantly higher particle temperature. (b) Prerecorded calibration
Van’t Hoff plot (gray line, cf. Figure c) used to determine the absolute
particle temperature upon illumination with higher optical powers
by determining the intersects of the phase transition pressures with
the calibration curve. (c) Extracted absolute particle temperatures
(TH) as a function of optical
power for five different set reactor temperatures and the temperatures
at 0 W (Tt/c). A clear trend of increasing
particle temperature with increasing optical power can be seen for
all reactor temperatures. The error in TH is derived from the 95% confidence interval in the
linear fit to the Van’t Hoff calibration curve, and the error
in the optical power is estimated to be ±0.2 W. (d) Temperature
increase, ΔT, in Pd nanoparticles upon illumination
as a function of optical power for the different set reactor temperatures.
The values are obtained by subtracting the set reactor temperatures
from TH. The error is obtained
from the 95% confidence interval in the linear fit to the Van’t
Hoff calibration curve combined with the standard deviation of the
set reactor temperature measurement.At this point it is also interesting to examine the temperature
resolution of our hydrogen nanothermometry method. To do so, we conducted
a series of isotherm measurements with deliberately small increasing
temperature steps, to see if the method is able to distinguish such
small temperature differences. Figure a shows the resulting isotherms measured at 34.33,
35.58, 36.46, and 36.93 °C, respectively. Clearly, the shift
of the phase transition pressure in each isotherm is discernible,
even for the two highest temperatures, which differ only by ∼0.5
°C. Quantitatively, however, when plotting the phase transition
pressure of the isotherms obtained with the method explained above
(Figure S6), their error confidence intervals
overlap (Figure b).
Looking at the next data points with a temperature interval of ∼0.9
°C, however, such an overlap does not exist. On the basis of
this analysis, we thus conclude that our hydrogen nanothermometer
exhibits a temperature resolution of 1 °C.
Figure 4
Deriving the temperature
resolution of the hydrogen nanothermometry
method. (a) Isotherms measured at four slightly different set reactor
temperatures in a very narrow temperature range. (b) The extracted
phase transition pressure is clearly distinguishable up to a temperature
difference of 0.9 °C between isotherms. At a difference of 0.5
°C the pressure uncertainties (error bars defined by the combination
of the standard errors of the fit functions) intersect and thus define
a temperature resolution of 1 °C for the hydrogen thermometry
in its present form. The error bars along the temperature axis are
defined as the standard deviation of the temperature instability during
one isotherm measurement cycle.
Deriving the temperature
resolution of the hydrogen nanothermometry
method. (a) Isotherms measured at four slightly different set reactor
temperatures in a very narrow temperature range. (b) The extracted
phase transition pressure is clearly distinguishable up to a temperature
difference of 0.9 °C between isotherms. At a difference of 0.5
°C the pressure uncertainties (error bars defined by the combination
of the standard errors of the fit functions) intersect and thus define
a temperature resolution of 1 °C for the hydrogen thermometry
in its present form. The error bars along the temperature axis are
defined as the standard deviation of the temperature instability during
one isotherm measurement cycle.We also note that the final temperature resolution of a measurement
is a trade-off against the speed of the measurement. If a wide temperature
range is to be scanned to determine particle temperature with an accuracy
of 1 °C, the measurement will take several hours, as each temperature
step is on the order of a few minutes. However, by narrowing down
the scanned temperature range or increasing the scanned pressure steps,
the measurement can be sped up at the cost of temperature resolution.
While this is sufficient for most applications, it is still significantly
slower than fast photoluminescence methods that reach measurement
rates of 80 kHz.[71]Having established
the ability of our approach to accurately measure
the absolute temperature of Pd nanoparticles upon illumination with
1 °C resolution, it is now interesting to investigate whether
the direct nature of our measurement also can reveal a temperature difference between the global sample temperature (Tt/c) measured traditionally using a thermocouple
positioned at the side of the sample (Figure S1b) and the real nanoparticle temperature in the sample center measured
by our hydrogen nanothermometer (TH). For this purpose, we plot the difference between Tt/c and TH as a function of irradiated power for four different set reactor
temperatures (Figure a). As the main result, we find that TH is universally higher than Tt/c obtained by the thermocouple. Furthermore, we identify a weak trend
that this difference between TH and Tt/c increases for higher optical
power.
Figure 5
Temperature inhomogeneity on the sample and FDTD-informed conjugated
heat transfer simulations of temperature distribution on the sample.
(a) Temperature difference for a Pd nanoparticle array under illumination
at different optical power and reactor temperatures, as obtained by
comparing the temperature measured by a thermocouple (Tt/c) that is touching the side of the sample and therefore
represents the global sample temperature (including the substrate)
and the temperature extracted by hydrogen nanothermometry (TH) that represents the absolute
Pd particle temperature. A proportionality between increasing temperature
difference and increasing optical power can be identified. The uncertainties
along the y-axis are obtained by using the standard
error of the linear fit function of the Van’t Hoff calibration
curve and the uncertainty along the x-axis is estimated
to be ±0.2 W. (b) Simulated temperature distribution across the
sample and thermocouple mounted inside the reactor pocket (see Figure S18 for details of simulated geometries)
for a set reactor temperature of 25 °C and a gas flow rate of
27 mL/min upon heating with an optical power of 3 W. (c) Extracted
simulated temperature in the center of the sample and at the tip of
the thermocouple plotted for 1, 2, 3, 4, and 4.8 W optical power.
Notably, the discrepancy between the thermocouple and sample center
temperature increases for increasing optical power. (d) Difference
between thermocouple and sample center temperatures plotted as a function
of irradiated optical power, revealing a proportionality, in reasonable
agreement with the experiments (cf. Figure a).
Temperature inhomogeneity on the sample and FDTD-informed conjugated
heat transfer simulations of temperature distribution on the sample.
(a) Temperature difference for a Pd nanoparticle array under illumination
at different optical power and reactor temperatures, as obtained by
comparing the temperature measured by a thermocouple (Tt/c) that is touching the side of the sample and therefore
represents the global sample temperature (including the substrate)
and the temperature extracted by hydrogen nanothermometry (TH) that represents the absolute
Pd particle temperature. A proportionality between increasing temperature
difference and increasing optical power can be identified. The uncertainties
along the y-axis are obtained by using the standard
error of the linear fit function of the Van’t Hoff calibration
curve and the uncertainty along the x-axis is estimated
to be ±0.2 W. (b) Simulated temperature distribution across the
sample and thermocouple mounted inside the reactor pocket (see Figure S18 for details of simulated geometries)
for a set reactor temperature of 25 °C and a gas flow rate of
27 mL/min upon heating with an optical power of 3 W. (c) Extracted
simulated temperature in the center of the sample and at the tip of
the thermocouple plotted for 1, 2, 3, 4, and 4.8 W optical power.
Notably, the discrepancy between the thermocouple and sample center
temperature increases for increasing optical power. (d) Difference
between thermocouple and sample center temperatures plotted as a function
of irradiated optical power, revealing a proportionality, in reasonable
agreement with the experiments (cf. Figure a).To put this result into perspective and to prepare grounds for
a discussion of its implications in plasmonic applications, it is
important to elucidate the origin of the observed effect. To this
end, naïvely, one could argue that it is to be expected and
the consequence of the Pd particles being hotter than the substrate
because (i) light is mainly and efficiently absorbed in the particles[72] and (ii) the thermocouple is in contact with
the substrate, rather than with the particles themselves. While this
is a tempting conclusion, calculations by Zhdanov and Kasemo for a
geometrically very similar (quasi-2D) system actually predict that
the temperature difference between a glass substrate and metal nanoparticles
under illumination is almost negligible due to the high thermal conductivity
of the system.[36] These results have also
been confirmed by Baffou etal.[35] Therefore, a more careful analysis of the situation
at hand in our experiments is necessary.To do this, we start
by identifying the geometry of our setup,
as well as the fact that the measurements are carried out in plug-flow,
as two likely factors of importance. Specifically, we note that the
optical measurement of the isotherms used for temperature determination
takes place in the center of a square sample where the light is irradiated
with circular symmetry, whereas the thermocouple is positioned at
the sample edge. This is important because to a first approximation
it is reasonable to assume that the sample is heated homogeneously
from the center since our nanofabrication produces a uniform distribution
of the Pd nanoparticles on the glass substrate at the cm2 level.[62,73] Hence, assuming first that there is no gas
flow along the sample, this homogeneous heating combined with the
square sample geometry will lead to the highest temperatures in the
center of the samples. Consequently, a circular temperature gradient
across the sample can be expected, with the hottest area in the middle
and the coolest areas at the edges. Introducing a gas flow across
the sample will slightly alter this situation due to forced convection
that shifts the hot center downstream, with the amount of shift determined
by the flow rate. This general hypothesis is indeed confirmed by conjugated
heat transfer simulations using COMSOL Multiphysics using the Pd nanodisk
light absorption calculated by the finite-difference time-domain (FDTD)
method (Figure b;
details about the simulations can be found in Figures S16–S19 and the corresponding discussion in
the Supporting Information, SI, Section S2). The temperature gradient
will also contribute to the fact that the phase transition from α-
to β-phase is not abrupt but gradual, as each particle has a
slightly different temperature and hence a slightly different phase
transition pressure.A second effect to consider is the presence
of the thermocouple
itself. As the first aspect, it has a small contact area with the
sample and thus likely a relatively poor thermal contact, which may
contribute to a lower temperature reading. As a second aspect, the
thermocouple has a sizable thermal mass, which in combination with
its placement upstream of the sample in the reactor and a concurrent
cooling effect by the bypassing gas may further reduce the temperature
reading compared to the sample temperature in the center. Again resorting
to simulations, we can indeed confirm the proportionality between
the increasing temperature difference of the thermocouple positioned
at the center of the sample and optical power (Figure c,d), which is in reasonable agreement with
the experiments (cf. Figure a).Having established and characterized
in detail the hydrogen nanothermometry
concept, we now put it to the test by nanofabricating a sample with
higher Pd particle surface coverage compared to the standard sample
(18% vs 11%, respectively; see Figure a). This increase in coverage is also associated
with a decrease in the average interparticle distance from 331 to
250 nm, as determined from the corresponding radial distribution function
(Figure S10).[74] The idea behind this design is to demonstrate that an increased
particle surface coverage leads to a larger temperature increase upon
illumination due to a collective heating effect and that we are able
to accurately measure this increase using hydrogen nanothermometry.
Accordingly, we first compare optical isotherms of the dense and standard
sample acquired at the same conditions and indeed find a significantly
higher phase transition pressure for the surface with higher Pd particle
coverage (Figure b).
Translating these increased phase transition pressures (Figure S11) into a temperature increase induced
by the higher Pd particle coverage, Tdense – Tstandard, we find a value
of approximately 4 °C for 4 W optical power at 25 °C set
reactor temperature, as well as the expected dependence of the temperature
difference on irradiance (Figure c).
Figure 6
Measuring the light-induced temperature increase dependence
on
Pd nanoparticle array density. (a) SEM images of a standard Pd nanoparticle
array with 11% surface coverage (left) and a dense Pd nanoparticle
array with 18% surface coverage (right). (b) Optical isotherms of
the two arrays obtained at identical 2 W optical power and 25 °C
set reactor temperature. The denser array exhibits a higher phase
transition pressure. Note that is
normalized to facilitate direct comparison
between the two isotherms. (c) The measured and calculated (using eq ; details in the SI Section S3) temperature difference between
the standard and the dense sample obtained at 24.6, 28.9, 34, 38.3,
and 42.8 °C set reactor temperature. For all optical powers the
dense sample exhibits higher temperatures. The particles are corrected
for ambient temperature to avoid the influence of slightly varying
room temperature (raw data Figure S13).
The depicted error in temperature difference is derived from the 95%
confidence interval in the linear fit to the Van’t Hoff calibration
curve, and the error in optical power is estimated to be ±0.2
W.
Measuring the light-induced temperature increase dependence
on
Pd nanoparticle array density. (a) SEM images of a standard Pd nanoparticle
array with 11% surface coverage (left) and a dense Pd nanoparticle
array with 18% surface coverage (right). (b) Optical isotherms of
the two arrays obtained at identical 2 W optical power and 25 °C
set reactor temperature. The denser array exhibits a higher phase
transition pressure. Note that is
normalized to facilitate direct comparison
between the two isotherms. (c) The measured and calculated (using eq ; details in the SI Section S3) temperature difference between
the standard and the dense sample obtained at 24.6, 28.9, 34, 38.3,
and 42.8 °C set reactor temperature. For all optical powers the
dense sample exhibits higher temperatures. The particles are corrected
for ambient temperature to avoid the influence of slightly varying
room temperature (raw data Figure S13).
The depicted error in temperature difference is derived from the 95%
confidence interval in the linear fit to the Van’t Hoff calibration
curve, and the error in optical power is estimated to be ±0.2
W.With this value in hand, it is
now interesting to compare it to
the result obtained when using the analytical expression for photoinduced
heating in nanoparticle arrays established by Baffou etal.[70] that suggests
an inverse correlation between interparticle distance and heating
upon illumination aswhere is
the nanoparticle absorption cross section, I is the
irradiance, and are
the thermal conductivity of the substrate
and the medium, respectively, S is the side length
of the array, and p is the interparticle distance.
To obtain of
a Pd nanodisk needed as input to estimate for our system at
hand, we employed FDTD
simulations, which yield an average = 1.32
× 10–14 m2. The remaining required
input parameters used for the calculation,
as well as further details, are summarized in the SI Section S3 and Methods sections.
The accordingly calculated temperature increases upon illumination, , for two samples
with a particle density
difference as in our experiment are plotted in Figure S12. The temperature differences between these two
samples are plotted Figure c, together with the corresponding experimental data points
obtained at five different set reactor temperatures and for four irradiated
powers. Notably, both absolute values and trends with respect to irradiated
power exhibit good agreement. Furthermore, we obtained similar results
using COMSOL simulations for a surface with 18% particle coverage
(Figure S14). Altogether, these results
thus validate the capability of hydrogen nanothermometry to accurately
and directly measure the temperature of Pd nanoparticles upon illumination.
Conclusions
In summary, we have presented a direct, noninvasive optical nanothermometry
concept with 1 °C resolution that employs the temperature-sensitive
first-order hydride formation in Pd nanoparticles as an intrinsic
temperature probe. This use of the Pd nanoparticles themselves as
the temperature reporter enables direct and absolute nanoparticle
temperature measurements and overcomes limitations of traditional
temperature probes that include large thermal mass or convolution
of readout contributions that stem from the particles themselves and
their support/surrounding medium, as well as temperature gradients
on the sample. This was exemplified by identifying a difference between
the temperature of a sample decorated with a quasi-random array of
Pd nanoparticles measured using a thermocouple and the hydrogen nanothermometer,
which arises due to a temperature gradient in the sample and limited
heat transfer between sample and thermocouple. As a further aspect,
applying the hydrogen thermometer, we were able to accurately measure
the temperature difference between two surfaces with different Pd
particle coverages—and thus different collective plasmon heating—at
identical illumination conditions and corroborated these results by
numerical simulations and a theoretical model. Consequently, the hydrogen
nanothermometry developed here provides an alternative solution for
accurate direct temperature measurements at the nanoscale, which we
predict to find application in (plasmon-mediated) catalysis, where
accurate temperature measurements of the catalyst are crucial and
challenging to help distinguish between hot-carrier-mediated and photothermal
reaction pathways. To this end, utilizing plasmonic nanoimaging hydrogen
nanothermometry could be used to acquire thermal images with a spatial
resolution down to 900 nm.[75] Our method
can also be useful in other applications that involve plasmonic nanoparticles
that either are sensitive to temperature change or rely on deliberately
induced temperature variations using the plasmonic effect. We also
predict that the method potentially is capable of near to “real”
time particle temperature determination when utilizing an experimental
setup with low thermal mass (Figure S15). Finally, we also note that hydrogen nanothermometry is not restricted
to Pd alone, since many other metals, such as niobium,[76] yttrium,[77] magnesium,[78] or a wide range of metal alloys[58,79,80] readily form hydrides and are
also plasmonically active.
Methods
Nanofabrication
The samples were nanofabricated using
the hole-mask colloidal lithography (HCL) technique.[62] For the isotherm measurements the samples were fabricated
on fused silica substrates and for SEM imaging on an oxidized silicon
wafer. The following instruments were used during the nanofabrication:
spin-coating (Suss, LabSpin6), oxygen ion etching (Plasmatherm, BatchTop
m/95), and electron beam evaporation (Lesker, PVD 225). To define
the mask in the fabrication process, we used a 0.2 wt % 140 nm diameter
sulfate latex bead solution (molecular probes by Life Technologies)
in deionized water. To create a surface with higher particle density,
we added 0.1 mmol/L NaCl to the sulfate latex bead solution to screen
the surface charges of the polystyrene beads and thereby achieve a
higher density array upon their self-assembly.
Experimental Setup
The measurements were executed in
a quartz-tube plug-flow-type reactor (X1, Insplorion AB, Göteborg,
Sweden) enclosed by a heating coil, thermal isolation, and a metal
shield. The sample was mounted in a custom optically transparent pocket
reactor[81] with a spring-loaded thermocouple
touching the side of the sample. The gas flow through the reactor
was controlled by an array of mass flow controllers (Bronkhorst Low-ΔP).
To ensure a constant pressure inside the reactor, an upstream pressure
controller was used (Bronkhorst El-Press P-702CV). The reactor was
also equipped with optical access to allow for transmission measurements
using a fiber-coupled spectrometer (Avantes AvaSpec-1024) and halogen
lamp (Avantes Avalight-HAL-B), high-power halogen light source (250
W halogen lamp in a Newport 67011 QTH housing without optical IR filter),
or a mercury xenon arc light source (Newport 6295NS in a 66921 housing
and a 6123NS liquid optical IR filter). The optical power of the mercury
xenon arc light source was measured with a digital optical power meter
(Thorlabs PM100D) connected to thermal power sensors (Thorlabs S425C-L
and S470C) directly connected to the output of the lamp. H2 (4% ± 2 rel % diluted in Ar carrier gas) and Ar carrier gas
(99.9999% purity) were used. Prior to the first isotherm measurement
the samples were annealed at 500 °C in 4% H2 with
a flow of 200 mL/min for 12 h to induce grain growth and thereby induce
reproducible response upon H2 exposure.[63]
Scanning Electron Microscopy
The
SEM images were recorded
with the in-lens system of a Zeiss Supra 55 VP. The acceleration voltage
was 5 kV, and the working distance was at least 5 mm.
FDTD Simulations
The FDTD simulations were conducted
with Ansys-Lumericals FDTD Solution version 8.26.2717 to calculate
the absorbed power and absorption cross section of a Pd nanodisk.
The simulated geometry was a Pd disk with a 140 nm diameter and a
height of 25 nm placed on a SiO2 substrate. The dielectric
functions for Pd and SiO2 were used out of Ansys-Lumerical’s
database originating from Palik etal.[82] The source was a total-field/scattered
field source with a linearly polarized plane wave.
Heat Distribution
Simulations
To simulate the heat
distribution in the system, COMSOL Multiphysics 5.6 with a conjugated
heat transfer setup was used. The 9 × 9 × 0.5 mm3 fused silica substrate used in the experiment was modeled in a glass
pocket of 25 × 12 × 1 mm3, with a wall thickness
of 1 mm, to mimic the reactor pocket. The inside of the pocket was
filled with air. The thermocouple, touching the side of the sample,
was made of Inconel 600 with a diameter of 0.5 mm. The material constants
were taken from the COMSOL Multiphysics database. The flow rate through
the pocket was 27 mL/min and adjusted to match the temperature in
the experiments at 3 W optical power and 25 °C set reactor temperature
within ±1 °C. The details of the heat distribution simulations
are explained in the Supporting Information (Section S2).
Authors: Daniel Andrén; Lei Shao; Nils Odebo Länk; Srdjan S Aćimović; Peter Johansson; Mikael Käll Journal: ACS Nano Date: 2017-09-20 Impact factor: 15.881
Authors: Nikolai Strohfeldt; Andreas Tittl; Martin Schäferling; Frank Neubrech; Uwe Kreibig; Ronald Griessen; Harald Giessen Journal: Nano Lett Date: 2014-02-27 Impact factor: 11.189