Background: Many procedures to reconstruct osteochondral defects of the elbow radiocapitellar (RC) joint lack versatility or durability or do not directly address the subchondral bone structure and function. Purpose/Hypothesis: To biomechanically characterize the RC joint contact area, force, pressure, and peak pressure before and after reconstruction of osteochondral defects using a novel hybrid reconstructive procedure. It was hypothesized that the procedure would restore the contact characteristics to the intact condition. Study Design: Controlled laboratory study. Methods: A total of 10 cadaveric elbows (mean age 67 ± 2.7 years) were dissected to isolate the humerus and radial head. RC contact area, contact force, mean contact pressure, and peak contact pressure were measured with the elbow at 45° of flexion and neutral forearm rotation at compressive loads of 25, 50, and 75 N. Osteochondral defects 8 and 11 mm in diameter were created at the center of the capitellum; the defects were then reconstructed with a titanium fenestrated threaded implant, countersunk in the subchondral bone, with an acellular dermal matrix allograft sutured in place on top of the implant. Five conditions (intact, 8-mm defect, 8-mm repair, 11-mm defect, and 11-mm repair) were tested and results were compared using repeated-measures analysis of variance. Results: Both 8- and 11-mm defects significantly increased RC mean contact pressure at all compressive loads (P ≤ .008) and significantly increased peak contact pressure at compressive loads of 50 and 75 N (P < .002) compared with the intact condition. Repair of the 8-mm defect significantly decreased RC mean contact pressure at 25- and 50-N loads (P ≤ .009) and significantly decreased peak contact pressure at 50- and 75-N loads (P ≤ .035) compared with the defect condition. Repair of the 11-mm defect decreased mean contact pressure significantly at all compressive loads (P ≤ .001) and peak contact pressure at 50- and 75-N loads (P < .044) compared with the defect condition. Conclusion: RC joint contact pressure was restored to intact conditions while avoiding increased peak contact pressure or edge loading after repairing osteochondral defects related to osteochondrosis with a novel hybrid reconstruction technique. Clinical Relevance: This hybrid procedure that addresses the entire osteochondral unit may provide a new treatment option for osteochondral defects.
Background: Many procedures to reconstruct osteochondral defects of the elbow radiocapitellar (RC) joint lack versatility or durability or do not directly address the subchondral bone structure and function. Purpose/Hypothesis: To biomechanically characterize the RC joint contact area, force, pressure, and peak pressure before and after reconstruction of osteochondral defects using a novel hybrid reconstructive procedure. It was hypothesized that the procedure would restore the contact characteristics to the intact condition. Study Design: Controlled laboratory study. Methods: A total of 10 cadaveric elbows (mean age 67 ± 2.7 years) were dissected to isolate the humerus and radial head. RC contact area, contact force, mean contact pressure, and peak contact pressure were measured with the elbow at 45° of flexion and neutral forearm rotation at compressive loads of 25, 50, and 75 N. Osteochondral defects 8 and 11 mm in diameter were created at the center of the capitellum; the defects were then reconstructed with a titanium fenestrated threaded implant, countersunk in the subchondral bone, with an acellular dermal matrix allograft sutured in place on top of the implant. Five conditions (intact, 8-mm defect, 8-mm repair, 11-mm defect, and 11-mm repair) were tested and results were compared using repeated-measures analysis of variance. Results: Both 8- and 11-mm defects significantly increased RC mean contact pressure at all compressive loads (P ≤ .008) and significantly increased peak contact pressure at compressive loads of 50 and 75 N (P < .002) compared with the intact condition. Repair of the 8-mm defect significantly decreased RC mean contact pressure at 25- and 50-N loads (P ≤ .009) and significantly decreased peak contact pressure at 50- and 75-N loads (P ≤ .035) compared with the defect condition. Repair of the 11-mm defect decreased mean contact pressure significantly at all compressive loads (P ≤ .001) and peak contact pressure at 50- and 75-N loads (P < .044) compared with the defect condition. Conclusion: RC joint contact pressure was restored to intact conditions while avoiding increased peak contact pressure or edge loading after repairing osteochondral defects related to osteochondrosis with a novel hybrid reconstruction technique. Clinical Relevance: This hybrid procedure that addresses the entire osteochondral unit may provide a new treatment option for osteochondral defects.
Osteochondral defects, osteochondritis dissecans, or the general term
osteochondrosis may occur in the knee,
elbow,
and ankle.
Osteochondral defects and osteochondrosis involve structural and vascular
compromise of the subchondral bone and progressive damage of the osteochondral
unit via bone hypertrophy or cystic changes.
Etiology includes repetitive microtrauma from abnormal or excessive joint
loading or traumatic injuries such as elbow dislocation.
Questions remain regarding a reliable treatment algorithm, and operative
techniques are still in need of assessment of long-term outcomes and refinement.
A versatile surgical procedure that addresses the mechanical deficiencies
of the bone, preserves marrow communication, and allows a biologic reconstruction
that restores normal surface biomechanics is needed.Humeral capitellar osteochondrosis and related elbow radiocapitellar (RC) joint
osteochondral defects are prevalent in individuals who have played baseball or
performed gymnastics actively since childhood.
Repetitive and excessive compressive force at the RC joint from either
excessive valgus or axial loading have been suggested as the key contributors to
elbow RC defects.
Previous clinical and biomechanical studies have shown that the size and
location of the osteochondral defect affect prognosis in throwing athletes with
capitellar osteochondritis dissecans with larger defects resulting in poorer
clinical outcomes.
Current operative management options include debridement and microfracture
as well as numerous cartilage restoration techniques.
Cartilage restoration procedures include osteochondral autograft transfer,
osteochondral allograft transplantation,
and autologous chondrocyte implantation.
Joint surface restoration and osteochondral defect reconstruction in the
elbow are still in early development, and further studies with long-term clinical
follow-up are necessary to establish the indications and effectiveness of these procedures.
These existing procedures lack versatility or durability or do not directly
address the subchondral bone structure and function.In this study, we introduce a hybrid reconstruction of humeral RC osteochondral
defects utilizing titanium fenestrated threaded implants in the subchondral bone
and human dermal allograft. The fenestrated threaded implant in the subchondral
bone is designed to restore structural stability, preserve marrow communication,
and serve as a fixation platform for osteochondral graft reconstruction. It was
hypothesized that reconstruction of capitellum osteochondral defects would reduce
RC contact pressure toward that of the intact RC joint.
Methods
Specimen Preparation
A total of 10 fresh-frozen cadaveric elbows were acquired for this study
from Science Care. Institutional review board approval was waived by
our institution, as this was a cadaveric basic science study. Elbows
were transected 2 cm below the deltoid insertion and at the midulnar
and midradial shaft and were dissected, removing all skin and
subcutaneous tissue. The mean donor age was 67 ± 2.7 years old (range,
46-73 years) and consisted of 5 males and 5 females. None of the donor
elbows showed any evidence of gross abnormalities.The disarticulated humerus and radius were fixed in PVC pipes with
plaster of Paris and wood screws. The humerus and radius were then
mounted to a custom elbow testing system, with the humeral epicondylar
axis parallel to the floor with the radial head above and directed
toward the capitellum of the humerus (Figure 1). The radius was
positioned in neutral forearm rotation, defined by the radial
tuberosity directed medially in line with the humeral epicondylar
axis.
Figure 1.
Photograph of the testing setup.
Photograph of the testing setup.
Biomechanical Testing
A Tekscan pressure sensor (Model 4000; maximum saturation pressure 10.3
MPa; Tekscan) was placed between the position of contact between the
radial head and capitellum to measure the RC contact force, contact
area, contact pressure, and peak contact pressure at 45° of elbow
flexion based on the humeral and radial shaft. The Tekscan sensor’s
sensitivity was set to 35 and calibrated using a 2-point calibration
protocol with an applied force of 40 N and 80 N using an Instron 4111
load cell (Instron). The average saturation pressure after calibration
was 1565 ± 64 kPa.Specimens were preloaded in compression with 10 N followed by cyclic ramp
loading in compression from 10 to 25 N for 5 cycles, 10 to 50 N for 5
cycles, and 10 to 75 N for 5 cycles using an Instron 3365. Two trials
of 5 cycles at each load were recorded to ensure repeatability
(maximum RC contact force <10% difference).Five conditions were tested: intact, 8-mm defect, 8-mm repair, 11-mm
defect, and 11-mm repair. The defect sizes were chosen to simulate
clinically relevant osteochondral defects in the capitellum. Central
capitellar osteochondral defects located 45° anteriorly to the shaft
of the humerus, centered midway between the medial and lateral edges
of the capitellum, were created using an 8-mm and 11-mm drill by
drilling to a depth of 5 mm.
The depths of the defects were assessed with a caliper. For the
repair conditions, 1 complete trial of cyclic loading from 10 to 25 N
for 5 cycles, 10 to 50 N for 5 cycles, and 10 to 75 N for 5 cycles was
performed before measurement to minimize the effect of the graft
viscoelasticity.
Reconstruction of Capitellar Osteochondral Defects
The S-CORE HA Implant (Subchondral Solutions) (Figure 2) is a titanium,
hydroxyapatite-coated, fenestrated, cannulated fracture screw
specifically designed for osteochondral fractures, osteochondral
fracture defect reconstruction, and osteochondritis dissecans. The
implant was secured into the subchondral bone defect, creating a
fixation platform for securing human dermal allografts via suture, as
well as a hybrid reconstruction. Human dermal allografts with an area
of 4 × 6 cm and thickness of 3 mm (Matrix HD; RTI Surgical) were
soaked in 0.9% saline and prepared at room temperature. Circular 8-mm
and 11-mm diameter human dermal allograft discs were punched and
sutured with 3-0 nylon sutures onto the 7-mm and 10-mm diameter
titanium implants, respectively (mean thickness as measured with area
micrometer was 1.82 ± 0.01 mm for 8-mm repairs and 1.81 ± 0.01 mm for
11-mm repairs). The defects were reconstructed with the hybrid
construct, placing the upper portion of the allograft at the level of
the surrounding cartilage.
Figure 2.
(A) Photograph of a 10-mm implant prepared with an 11-mm
dermal allograft. (B) Superior view of dermal allograft.
(C) Photograph of a right elbow following osteochondral
repair of an 11-mm capitellar defect.
(A) Photograph of a 10-mm implant prepared with an 11-mm
dermal allograft. (B) Superior view of dermal allograft.
(C) Photograph of a right elbow following osteochondral
repair of an 11-mm capitellar defect.
Statistical Analysis
The contact force, area, pressure, and peak pressure at the peak force of
the cyclic loading was averaged across the 5 cycles for trial 1 and
then for trial 2. The 2 trials were then averaged together for
analysis. Mean values for the testing conditions were compared using
2-way repeated-measures analysis of variance followed by a post hoc
test with Bonferroni correction for multiple comparisons (IBM SPSS
Statistics 25.0). Statistical significance was defined as
P <.05.A sample size calculation was performed using the difference in contact
pressure between the 11-mm defect and 11-mm repair at 50 N. Based on
the mean and standard deviation of 3 specimens (mean difference: 0.23
kPa - 0.17 kPa = 0.06 kPa; standard deviation of the difference =
0.046 kPa), a total of 10 specimens were determined to be needed for α
= 0.05 and power (1 - β) of 0.80.
RESULTS
There were no statistically significant differences in RC contact area between
any of the testing conditions (P > .502 for all
comparisons) (Figures
3 and 4).
Figure 3.
Representative radiocapitellar contact from the Tekscan images from
1 specimen for each testing condition with calibrated pressure
scale showing saturation pressure of 1550 kPa.
Figure 4.
Radiocapitellar contact area for each testing condition and
compressive load. Data presented as mean with standard error
bars.
Representative radiocapitellar contact from the Tekscan images from
1 specimen for each testing condition with calibrated pressure
scale showing saturation pressure of 1550 kPa.Radiocapitellar contact area for each testing condition and
compressive load. Data presented as mean with standard error
bars.The 8-mm RC defect increased RC mean contact pressure significantly compared
with the intact condition at all compressive loads (25 N, P
= .019; 50 N, P = .039; 75 N, P = .020)
(Figures 3
and 5). Repair of
the 8-mm defect decreased RC mean contact pressure significantly for 25 N
and 50 N compressive loads compared with the defect condition (25 N,
P < .001; 50 N, P = .009) (Figure 5).
Figure 5.
Radiocapitellar mean contact pressure for each testing condition
and each compressive load. Error bars represent standard error.
Statistically significant difference (P <
.05) versus *intact, #8-mm defect, †8-mm
repair, and @11-mm defect.
The 11-mm defect increased mean contact pressure significantly for all
compressive loads compared with intact (25 N, P = .008; 50
N, P = .001; 75 N, P < .001) and
compared with the 8-mm defect condition at 50 N and 75 N compressive loads
(50 N, P = .014; 75 N, P = .009). Repair
of the 11-mm defect decreased mean contact pressure significantly compared
with the 11-mm defect for all compressive loads (25 N, P =
.001; 50 N, P < .001; 75 N, P <
.001) (Figure
5).Radiocapitellar mean contact pressure for each testing condition
and each compressive load. Error bars represent standard error.
Statistically significant difference (P <
.05) versus *intact, #8-mm defect, †8-mm
repair, and @11-mm defect.The 8-mm RC defect significantly increased RC peak contact pressure compared
with the intact condition at 75 N compressive loads (P =
.010) (Figure 6).
Repair of the 8-mm defect significantly decreased peak pressure compared
with the defect condition at 50 and 75 N (50 N, P = .009;
75 N, P = .035). The 11-mm defect increased RC peak contact
pressure significantly at 50 and 75 N compared with intact (50 N,
P = .002; 75 N, P = .001). Repair of
the 11-mm defect significantly decreased peak pressure compared with 11-mm
defect at 50 and 75 N (50 N, P = .006; 75 N,
P = .044).
Figure 6.
Radiocapitellar peak contact pressure for each testing condition
and compressive load. Error bars represent standard error.
Statistically significant difference (P <
.05) versus *intact, #8-mm defect, †8-mm
repair, and @11-mm defect.
Radiocapitellar peak contact pressure for each testing condition
and compressive load. Error bars represent standard error.
Statistically significant difference (P <
.05) versus *intact, #8-mm defect, †8-mm
repair, and @11-mm defect.
Discussion
Capitellum osteochondral defect repair using specialized titanium implants and
human dermal allograft restored RC joint mean and peak contact pressure to
intact levels. There was increased mean RC contact pressure on average
across all loading conditions of 11.7% from intact for 8-mm defects
(P < .039 for all loads) and 28.4% from intact for
11-mm defects (P < .008 for all loads). RC peak contact
pressure increased an average of 25.5% and 46.3% from intact for 8-mm
(P = .010 for 75 N load) and 11-mm defects
(P < .002 for 50 and 75 N loads), respectively.
Both mean and peak contact pressures were restored to intact levels
following repair. There were no significant differences in the contact area
for the defect or repair conditions. Joint incongruity may lend some
explanation to the variation in the values for average contact area. These
findings suggest that osteochondral defect repair of the capitellum with
titanium implants and human dermal allograft can be a feasible option in
treating RC joint osteochondral defects.Literature on surface restoration of the elbow joint is limited, making
validation or comparison of our work difficult. Osteochondral autograft
transplantations, osteochondral allograft transfers, and autologous
chondrocyte implantations have been well studied in the knee.
Kock et al
demonstrated that there were no significant differences in contact
stresses with osteochondral autograft transplantation whether the implant is
bottomed or unbottomed out with the procedure, resulting in 135% of the
intact border contact pressure of human cadaver knees. Harris et al
demonstrated that, with osteochondral autograft transplantation and
synthetic plugs, the peak contact pressure was significantly higher when
placed in a proud position with the osteochondral autograft plug compared
with the synthetic plug in human cadaver knees. In the studies with
implants, there are concerns of increased contact pressure, leading to
long-term complications such as graft subsidence and failure. In the current
study, mean contact pressure after repairing defects was 88.1% to 95.3% of
the intact state, with peak pressure no greater than 119.9% of the intact
state in the elbow RC joint, with no significant differences of peak contact
pressure compared with intact conditions. This would suggest that RC joint
osteochondral defect repair with human dermal allograft was able to restore
mean contact pressure while avoiding an increase in peak contact pressure or
edge loading.Joint congruity after repair is a significant factor, which can be a
limitation. The intact joint has an uninterrupted and continuous cartilage
surface, but a cartilage defect or repaired defect may have an interrupted
surface or step off, which can lead to edge loading. This is akin to the
findings of significantly increased contact pressure with slightly elevated
or sunken osteochondral transplantation plugs during a study performed in
femoral condyles of porcine knees.
Similarly, Bobrowitsch et al
performed a biomechanical study on joint contact pressure after
osteochondral graft transplantation in an ovine carpometacarpal joint model,
demonstrating significantly higher contact pressure with the high-implanted
graft. Despite the variation in anatomic congruity of the articular surfaces
in the elbow joint, the trend of contact characteristics in our study was
clear and persistent in all specimens; mean and peak contact pressure were
reduced when comparing the repair conditions with the respective defect
conditions. This could be due to the versatile structure of the dermal
allograft lining the defective surface after repair, which seems to flatten
and adjust to the shape of the articular surface with mechanical compressive
conditioning while its intrinsic viscoelastic property also provides a
cushioning effect for the restored articular surface.Delivery of scaffolds for joint resurfacing or defect reconstruction continues
to be a challenge, especially in cases that involve compromise of the
subchondral bone, osteochondrosis. Bone sequelae include bone mismatch in
osteochondral transplant procedures, intralesional bone hypertrophy after
microfracture, trabecular architectural changes, and cystic changes. As
defects progress in size, edge loading with surrounding osteochondral tissue
occurs; trabecular microcracks, cystic changes, vascular congestion, and
bone edema ensue, leading to pain and further compromise of the subchondral
bone. Therefore, a reconstruction of an osteochondral defect needs a hybrid
reconstruction strategy to restore the subchondral bone’s structural
integrity, permit healing by marrow communication as well as restoring the
articulating surface congruency of the joint.
Limitations
The limitations of this biomechanical study include the removal in our
cadaveric testing of all soft tissue, which can contribute to joint
contact characteristics. It is worth noting that Sabo et al
demonstrated that osteochondral lesions of the capitellum do
not affect elbow kinematics and stability with intact collateral
ligaments. Therefore, further studies could verify this and may focus
more on elucidating joint contact characteristics, possibly with
different flexion angles. A second limitation is the inclusion of a
Tekscan sensor between the articular surface may also affect contact
characteristics but should not affect the comparisons between the
testing conditions. A third limitation is that only unidirectional
loading was applied. Other directions of loading, such as valgus
loading or rotational loading, can occur at the RC joint, which were
not accounted for in this study. A fourth limitation is the inability
to account for healing biology in our cadaveric testing; the results
of this study can represent only the normalization effects at time
zero. The use of human dermal allograft for humeral RC joint
osteochondral defect repair requires further clinical evaluation and
long-term studies.
Conclusion
RC joint contact pressure was restored to intact conditions while avoiding
increased peak contact pressure or edge loading after repairing
osteochondral defects related to osteochondrosis with a novel hybrid
reconstruction using specialized titanium implants and human dermal
allograft. This hybrid reconstruction that addresses the entire
osteochondral unit may provide a new treatment option for osteochondral
defects.
Authors: Evgenij Bobrowitsch; Andrea Lorenz; Johanna Jörg; Ulf G Leichtle; Nikolaus Wülker; Christian Walter Journal: Med Eng Phys Date: 2014-07-25 Impact factor: 2.242
Authors: Teruhisa Mihata; Ryan Quigley; Grant Robicheaux; Michelle H McGarry; Masashi Neo; Thay Q Lee Journal: Am J Sports Med Date: 2013-06-07 Impact factor: 6.202
Authors: Joshua D Harris; Kraig K Solak; Robert A Siston; Alan Litsky; Jason Richards; David C Flanigan Journal: Orthopedics Date: 2011-01-01 Impact factor: 1.390
Authors: Jennifer M Weiss; Kevin G Shea; John C Jacobs; Peter C Cannamela; Ian Becker; Mark Portman; Jeffrey I Kessler Journal: Am J Sports Med Date: 2018-04-03 Impact factor: 6.202
Authors: Niels B Kock; José M H Smolders; Job L C van Susante; Pieter Buma; Albert van Kampen; Nico Verdonschot Journal: Knee Surg Sports Traumatol Arthrosc Date: 2008-05 Impact factor: 4.342