Whitney York1, M Ryan Smith1, Chin-Chi Liu1.
Abstract
Background: A new, portable bedside coagulation monitor (VCM Vet) has provided a user-friendly, cartridge-based method to perform viscoelastic testing. However, the use of native whole blood limits the time to analyze the sample to minutes. The objective of this study is to assess whether citrated whole blood can be utilized with the cartridge-based system and whether the results are comparable to those of native whole blood. A secondary objective is to assess the viability of citrated whole blood results after up to 4 hours of resting.Entities:
Keywords: VCM Vet; coagulation testing; point of care; resting time; viscoelastic
Year: 2022 PMID: 35330612 PMCID: PMC8940259 DOI: 10.3389/fvets.2022.827350
Source DB: PubMed Journal: Front Vet Sci ISSN: 2297-1769
Results of NWB compared to CWB baseline.
|
|
|
| |
|---|---|---|---|
| CT (min) | 5.97 ± 0.66 | 4.41 ± 1.11 | <0.0001* |
| CFT (min) | 2.90 ± 0.51 | 2.85 ± 1.06 | 0.4139 |
| α-angle (°) | 57.30 ± 4.45 | 59.40 ± 7.76 | 0.1818 |
| A10 (units) | 22.60 ± 3.34 | 22.45 ± 5.31 | 0.3586 |
| A20 (units) | 29.70 ± 3.95 | 27.80 ± 5.52 | 0.1217 |
| MCF (units) | 35.10 ± 3.03 | 31.10 ± 5.42 | 0.0036* |
| LY30 (%) | 99.50 ± 1.27 | 98.45 ± 4.84 | 0.6420 |
| LY45 (%) | 99.20 ± 1.62 | 97.35 ± 4.82 | <0.0001* |
Results are listed as a mean with standard deviation. Asterisks identify values that are significantly different (defined as p <0.05).
Figure 1Selected Bland-Altman plots indicating biases of measured parameters of NWB (control) compared with CWB baseline. Mean difference (bias) is indicated by solid lines with dashed lines indicating 95% agreement interval of +/− standard deviations. Parameters included are (A) CT, (B) CFT, (C) AA, and (D) MCF.
Results of CWB baseline compared to CWB rested for 2 h (Cit-2) and 4 h (Cit-4).
|
|
|
|
|
| |
|---|---|---|---|---|---|
| CT (min) | 4.41 ± 1.11 | 4.68 ± 0.69 | 0.2010 | 4.31 ± 0.61 | 0.6102 |
| CFT (min) | 2.85 ± 1.06 | 3.00 ± 0.83 | 0.4139 | 2.72 ± 0.77 | 0.4139 |
| α-angle (°) | 59.40 ± 7.76 | 58.20 ± 6.10 | 0.1818 | 60.05 ± 5.97 | 0.1818 |
| A10 (units) | 22.45 ± 5.31 | 21.40 ± 4.27 | 0.3586 | 22.40 ± 4.06 | 0.3586 |
| A20 (units) | 27.80 ± 5.52 | 27.50 ± 4.70 | 0.1217 | 28.45 ± 4.37 | 0.1217 |
| MCF (units) | 31.10 ± 5.42 | 30.90 ± 4.52 | 0.9962 | 31.85 ± 4.64 | 0.8416 |
| LY30 (%) | 98.45 ± 4.84 | 98.55 ± 5.61 | 0.6420 | 99.70 ± 1.34 | 0.6420 |
| LY45 (%) | 97.35 ± 4.82 | 96.30 ± 7.89 | 0.4474 | 96.95 ± 5.58 | 0.4474 |
Results are listed as a mean with standard deviation.
Figure 2Selected Bland-Altman plots indicating biases of measured parameters of CWB baseline compared with 4 h rested CWB samples (Cit 4). Mean difference (bias) is indicated by solid lines with dashed lines indicating 95% agreement interval of +/− standard deviations. Parameters included are (A) CT, (B) CFT, (C) AA, and (D) MCF.
Within sample variation (CV, %) listed as a median.
|
|
|
|
| |
|---|---|---|---|---|
| CT (min) | 0 | 68.80 | 6.79 | 6.8 |
| CFT (min) | 0 | 32.93 | 7.90 | N/A |
| α-angle (°) | 0 | 19.46 | 2.36 | 4.7 |
| A10 (units) | 0 | 17.25 | 3.83 | N/A |
| A20 (units) | 0 | 23.14 | 4.49 | N/A |
| MCF (units) | 0 | 27.37 | 3.21 | 10.7 |
| LY30 (%) | 0 | 17.37 | 0.00 | N/A |
| LY45 (%) | 0 | 16.32 | 0.00 | N/A |
Where known, the VCM manufacturer's within sample variation [CV (%)] is also listed. In instances for which the VCM manufacturer's CV is not known, “N/A” is listed instead.