Abstract
A real-time trace gas detector for benzene is demonstrated. The measurement system takes advantage of modest enrichment through short adsorption periods to reach a ppt-level detection limit with a sampling cycle of 90 s, which includes sample adsorption, desorption, and a spectroscopic measurement. Benzene is collected on Tenax TA sorbent for 30 s and then detected from the enriched samples with photoacoustic spectroscopy. High sensitivity is achieved using cantilever-enhanced photoacoustic spectroscopy and a continuous-wave quantum cascade laser emitting at 14.8 µm wavelength, which corresponds to the absorption wavelength of the strongest benzene infrared band. We reach a detection limit of 150 ppt of benzene, over one sampling cycle. Interference from humidity and other common petrochemicals is evaluated.Entities:
Mesh:
Substances:
Year: 2022 PMID: 35297878 DOI: 10.1364/AO.450407
Source DB: PubMed Journal: Appl Opt ISSN: 1559-128X Impact factor: 1.980