Kassie Rong1, Ana Cabrera2,3, Johan Delport2,3, Shannon Schofield3, Fatimah AlMutawa2,3.
Abstract
This study validated the performance of the reverse transcriptase-polymerase chain reaction (rRT-PCR) based Cepheid Xpert® Xpress SARS-CoV-2 assay against the TIB MOLBIOL E-gene/EAV, a standard laboratory rRT-PCR SARS-CoV-2 assay. Upper and lower respiratory tract samples (nasopharyngeal and nasal swabs, bronchoalveolar lavage, and tracheal aspirate) were obtained from patients suspected to have contracted COVID-19. Results from the Xpert® Xpress and standard rRT-PCR assays were compared for positive and negative agreement and analyzed for precision, reproducibility, 95% confidence intervals, and coefficients of variation. The Xpert® Xpress assay demonstrated 100% agreement with the standard lab rRT-PCR for both upper and lower respiratory tract samples. Both the Xpert® Xpress and lab rRT-CPR identified weakly positive (Ct values 35-39) sample replicates with 100% reproducibility and showed 100% precision in identifying triplicates of upper respiratory tract samples. The single-cartridge Xpert® Xpress system has a short turnaround time and can be employed to improve patient management and hospital bed allocation. Further verification of the system is required before implementation and consideration must be paid to its higher cost and impracticality for high-throughput use.Entities:
Year: 2022 PMID: 35286278 PMCID: PMC9036648 DOI: 10.1556/1886.2022.00003
Source DB: PubMed Journal: Eur J Microbiol Immunol (Bp) ISSN: 2062-509X
Acceptance criteria for analytical performance characteristics determined during validation of the Xpert® Xpress assay in comparison with our laboratory rRT-PCR
| Sample Type | Parameter | Acceptance criteria |
| Upper respiratory tract | Accuracy | >90% |
| Precision | CoV <5% | |
| Reproducibility | 95% with CoV <5% | |
| Lower respiratory tract | Accuracy | >95% |
| Reproducibility | 95% with CoV <5% |
Xpert® Xpress assay results on upper respiratory tract specimens compared to expected results from our laboratory rRT-PCR
| Xpert® Xpress | Lab rRT-PCR | |
| Pos | Neg | |
| Pos | 14 | 0 |
| Neg | 0 | 10 |
Xpert® Xpress assay results on lower respiratory tract specimens compared to expected results from our laboratory rRT-PCR
| Sample type | Xpert® Xpress | Lab rRT-PCR | |
| Pos | Neg | ||
| TA | Pos | 15 | 0 |
| Neg | 0 | 5 | |
| BAL | Pos | 5 | 0 |
| Neg | 0 | 5 | |
Note: For two of the samples that tested positive, the E gene target was not detected but the N2 target was detected. One such sample was weakly positive with a Ct of 37.9 while the other returned a Ct of 26.3.