Q C Nguyen, R M Sly, R L Boeckx, H A Kelly.
Abstract
Determination of theophylline concentrations by Seralyzer in 28 paired serum and plasma samples disclosed almost identical results. Comparison of determinations by Seralyzer and enzyme-multiplied immunoassay in 31 paired serum samples disclosed significant underestimation by Seralyzer when recalibrated whenever results of analysis of a 15-micrograms/mL control exceeded limits of +/- 3 micrograms/mL. Determinations by Seralyzer did not differ significantly from enzyme-multiplied immunoassay with narrower limits of 15 +/- 2 micrograms/mL for Seralyzer calibration for 18 paired samples. Seralyzer results are sufficiently reliable for clinical use if the instrument is recalibrated whenever analysis of the 15-micrograms/mL control falls outside limits of +/- 2 micrograms/mL.Entities:
Mesh:
Substances:
Year: 1986 PMID: 3526988
Source DB: PubMed Journal: Ann Allergy ISSN: 0003-4738